K161413 · Applied Medical Technology, Inc. · KNT · Feb 9, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K161413
Device Name
Low Profile Balloon Feeding Device
Applicant
Applied Medical Technology, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Feb 9, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Low Profile Balloon Feeding Device is indicated for use in patients who require long term feeding, are unable to tolerate oral feeding, who are at low risk for aspiration, require gastric decompression and/or medication delivered directly into the stomach through a secured (initial placement) or formed (replacement) stoma. The Low Profile Balloon Feeding Device is intended for all age groups.
Device Story
Low Profile Balloon Feeding Device provides enteral nutrition, medication delivery, and gastric decompression. Device consists of external bolster, feeding catheter, and internal retention balloon. Catheter inserted through stoma; internal balloon secures position. Available in 10Fr to 24Fr diameters and 0.5cm to 10.0cm stoma lengths. Features MR Conditional balloon fill-valve or MR Safe Invisi-Valve. Used in clinical settings for patients requiring long-term access. Healthcare providers manage placement and maintenance. Benefits include reliable gastric access for patients unable to tolerate oral intake.
Clinical Evidence
Bench testing only. Performance testing included stoma pullout, bond strength, fill valve integrity, balloon burst, leak testing, flow rate, and balloon integrity in simulated gastric fluid. Biocompatibility testing performed per ISO 10993-1, -3, -5, -6, -7, -10, -11, and -12 for permanent contact with mucosal membranes.
Technological Characteristics
External bolster, feeding catheter, internal retention balloon. Catheter diameters 10Fr-24Fr; stoma lengths 0.5cm-10.0cm. MR Conditional or MR Safe (with Invisi-Valve). Biocompatibility per ISO 10993 standards. Mechanical retention via balloon.
Indications for Use
Indicated for patients of all ages requiring long-term enteral feeding, gastric decompression, or medication delivery via a secured or formed stoma; patients must be unable to tolerate oral feeding and be at low risk for aspiration.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Kimberly-Clark MIC-KEY SF Low Profile Gastrostomy (K122653)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2017
Applied Medical Technology, Inc. Michelle Scott Regulatory/QA Specialist 8006 Katherine Boulevard Brecksville, OH 44141
Re: K161413
> Trade/Device Name: Low Profile Balloon Feeding Device Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: January 9, 2017 Received: January 10, 2017
Dear Michelle Scott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of
{1}------------------------------------------------
medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161413
Device Name Low Profile Balloon Feeding Device
#### Indications for Use (Describe)
The Low Profile Balloon Feeding Device is indicated for use in patients who require long term feeding, are unable to tolerate oral feeding, who are at low risk for aspiration, require gastric decompression and/or medication delivered directly into the stomach through a secured (initial placement) or formed (replacement) stoma. The Low Profile Balloon Feeding Device is intended for all age groups.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span></span> </div> |
|----------------------------------------------|----------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div> <span></span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## SECTION - 5
## 510(k) Summary
### Low Profile Balloon Feeding Device
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Date Prepared: | February 8, 2017 |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4000<br>Fax: 440-717-4200<br>Contact Person: Michelle Scott - Regulatory/QA Specialist<br>Email: michelle.scott@appliedmedical.net |
| Device<br>Information: | Trade/Device Name: Low Profile Balloon Feeding Device<br>Common Name: Gastrointestinal Feeding Tubes<br>Classification Name: Tubes, Gastrointestinal (And Accessories) (21 CFR 876.5980)<br>Product Code: KNT<br>Regulatory Class: II<br>Review Panel: Gastroenterology and Urology |
| Predicate<br>Device: | Primary Predicate: Kimberly-Clark MIC-KEY SF Low Profile Gastrostomy (K122653)<br>Reference Device: Kimberly-Clark MIC-KEY Low Profile Gastrostomy Tube (K043114)<br>Reference Device: Low Profile Balloon Feeding Device (K945618) |
| Intended Use: | The Low Profile Balloon Feeding Device is indicated for use in patients who<br>require long term feeding, are unable to tolerate oral feeding, who are at low risk<br>for aspiration, require gastric decompression and/or medication delivered<br>directly into the stomach through a secured (initial placement) or formed<br>(replacement) stoma. The Low Profile Balloon Feeding Device is intended for<br>all age groups. |
| Device<br>Description: | The Low Profile Balloon Feeding Device is used to provide nutrition,<br>medication, and decompression access into the stomach through a secured<br>(initial placement) or formed (replacement) stoma. |
| Technological<br>Characteristics: | The Low Profile Balloon Feeding Device consists of an external bolster, feeding<br>catheter, and internal retention balloon, similar to the predicate devices. The<br>feeding catheter is inserted into the stomach through a stoma and is held in place<br>with an internal retention balloon. The device can have an MR Conditional<br>balloon fill-valve or will be MR Safe when assembled with the Invisi-Valve.<br>The devices are provided in a number of sizes to accommodate different stoma |
| | diameters and lengths and can have either a blunt or tapered tip balloon. The<br>catheter tubes range in diameter from 10Fr to 24Fr, which is the same as the<br>primary predicate. The stoma lengths are available in a wider range of options,<br>from 0.5cm to 10.0 cm. |
| Biocompatibility<br>Testing: | The Low Profile Balloon Feeding Device has been tested for biocompatibility<br>based on the applicable sections of the following standards:<br>ISO 10993-1: 2009 Biological Evaluation of Medical Devices - Part 1:<br>Evaluation and testing within a risk management process ISO 10993-3: 2003 Biological Evaluation of Medical Devices - Part 3:<br>Tests for genotoxicity, carcinogenicity, and reproductive toxicity. ISO 10993-5: 2009 Biological Evaluation of Medical Devices - Part 5:<br>Tests for in vitro cytotoxicity. ISO 10993-6: 2007 Biological Evaluation of Medical Devices - Part 6:<br>Test for local effects after implantation. ISO 10993-7: 2008 Biological Evaluation of Medical Devices - Part 7:<br>Ethylene Oxide Sterilization Residuals. ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part<br>10: Tests for irritation and skin sensitization. ISO 10993-11:2006 Biological Evaluation of Medical Devices – Part<br>11: Tests for systemic toxicity ISO 10993-12:2007 Biological Evaluation of Medical Devices – Part<br>12: Sample preparation and reference materials An independent risk assessment was completed on the patient contacting<br>materials and it was determined that the Low Profile Balloon Feeding Device<br>met the acceptance criteria for permanent contact (greater than 30 days) with<br>mucosal membrane and breached/compromised surfaces. |
| Performance<br>Testing: | AMT conducted various performance tests on the components contained within<br>the Low Profile Balloon Feeding Device. Testing found that all components and<br>materials met or exceeded design specifications established by AMT.<br>Bench tests have been carried out on samples of the Low Profile Balloon<br>Feeding Device. The tests carried out included:<br>Stoma Pullout Minimum Overmold Bond Strength Fill Valve Blow Out Fill Valve Pullout Fill Valve Cycle Test Balloon Assembly Bond Peel/Tear Strength Balloon Burst Assembled Balloon Size Interlock Pullout Leak Test |
| Balloon Concentricity Ratio Tubing Tensile Testing Flow Rate Balloon Volume Maintenance Balloon Integrity Balloon Integrity in Simulated Gastric Fluid Testing | |
| Conclusion: | The Low Profile Balloon Feeding Device is substantially equivalent to the predicate device cleared under K122653 in intended use, design, biocompatibility, performance, and principles of operation. |
Applied Medical Technology, Inc. – 510(k) Submission Low Profile Balloon Feeding Device Section 5 – Page 5.1
{4}------------------------------------------------
## SECTION - 5
Applied Medical Technology, Inc. – 510(k) Submission Low Profile Balloon Feeding Device Section 5 – Page 5.2
{5}------------------------------------------------
# SECTION - 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.