K072756 · Promedic, Inc. · KNT · Dec 19, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K072756
Device Name
NON-IV ADAPTER FOR INFANT NG TUBE
Applicant
Promedic, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Dec 19, 2007
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
A non-IV connector for use with Infant Feeding tubes that are intended to be placed into the stomach to permit the introduction of fluids as directed by the physician. Intended for nasogastric or orogastric placement. Limited to < 30 day placement. Not intended for transpyloric placement.
Device Story
Infant feeding tube kit consisting of small-diameter PVC tube (5, 6, 8 French) with integral female non-IV connector, two distal eyelets, shaft markings, and radiopaque line; includes non-sterile syringe with matching non-IV male connector. Used in hospital or home settings by clinicians or caregivers to deliver fluids into the stomach. Modification from standard slip-fit luer to non-IV connector prevents accidental IV connection. Device provides enteral access; healthcare providers use markings and radiopaque line to verify placement; output is fluid delivery to stomach. Benefits include reduced risk of misconnection errors.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Materials: PVC tubing and connector. Dimensions: 5, 6, 8 French diameters; 14.5", 35", 41" lengths. Features: Two distal eyelets, shaft markings, integral radiopaque line. Connector: Non-IV slip-fit female. Sterilization: Sterile (tube), non-sterile (syringe). Connectivity: None.
Indications for Use
Indicated for infants and pediatric patients requiring nasogastric or orogastric feeding tube placement for fluid administration. Limited to < 30 days duration. Contraindicated for transpyloric placement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
## Non-Confidential Summary of Safety and Effectiveness
| | Page 1 of 2 |
|--|-------------|
| | 26-Sept-07 |
| ProMedic, Inc. | Tel (239) 498-2155 |
|-------------------------------|--------------------|
| 3460 Pointe Creek Court #102. | Fax (239) 236-0354 |
| Bonita Springs, FL 34134 | |
| Official Contact: | Paul Dryden |
|-------------------|-------------|
|-------------------|-------------|
| | DEC 1 9 2007 |
|--|--------------|
|--|--------------|
| Proprietary or Trade Name: | Infant Feeding Tube with Non-IV fittings |
|----------------------------|-------------------------------------------|
| Common/Usual Name: | Feeding Tube or NG /OG Tube |
| Classification Name: | Tubes, Gastrointestinal (and Accessories) |
| Predicate Devices: | ProMedic - Infant Feeding Tubes - K052903 |
#### Device Description:
The proposed modification to Infant Feeding Tube is to change the fittings from a standard slip fit luer to a non-IV (luer) size connector. The feeding tubes are a small diameter tube of various diameters, 5, 6, and 8 French, and lengths, 14.5", 35" and 41". It has an integral female luer fitting. There are 2 eyelets near the tip of the tube. It has marking along the shaft of the tubing and an integral radiopaque line. It is provided sterile. We will offer a non-sterile syringe with the appropriate mating, non-IV connector as part of a kit.
### Indications for Use:
A non-IV connector for use with Infant Feeding tubes that are intended to be placed into the stornach to permit the introduction of fluids as directed by the physician. Intended for nasogastric placement. Limited to < 30 day placement. Not intended for transpyloric placement.
### Environment of Use:
Hospital or environment where placement of a feeding tube is required.
| Attribute | Proposed device with<br>modification | Predicate ProMedic Infant<br>Feeding Tube - K052903 |
|---------------------------------|--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Indications<br>General | To be placed into the stomach to<br>permit the introduction of fluids<br>as directed by the physician. | To be placed into the stomach<br>to permit the introduction of<br>fluids as directed by the<br>physician. |
| Type of placement | Nasogastric or orogastric<br>Not for transpyloric placement. | Nasogastric or orogastric<br>Not for transpyloric<br>placement. |
| Length of placement | < 30 days | < 30 days |
| Intended for single patient use | Yes | Yes |
| Prescription | Yes | Yes |
### Comparison to Predicate Devices:
{1}------------------------------------------------
4072056
may 2012
# Non-Confidential Summary of Safety and Effectiveness
Page 2 of 2 26-Sept-07
| Attribute | Proposed device with modification | Predicate ProMedic Infant<br>Feeding Tube – K052903 |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Intended population | Infants / pediatrics | Infants / pediatrics |
| Intended Environment of Use | Hospital, home or environments<br>where placement of a Feeding<br>tube is required. | Hospital, home or environments<br>where placement of a Feeding tube is<br>required. |
| Design Features | | |
| Provided in various diameters<br>from 4 - 12 Fr | 5, 6, 8 Fr | 5, 6, 8 Fr |
| Connector | Non-IV slip fit female<br>connection<br>Must be used with an<br>appropriate syringe with non-<br>IV male mating connector | Standard slip fit female luer |
| Two (2) eyelet holes near tip | Yes | Yes |
| Radiopaque line entire length<br>of tubing | Yes | Yes |
| Markings along the length of<br>the tubing | Yes | Yes |
| Materials | | |
| Tubing and Connector - PVC | Yes | Yes |
| Packaging | | |
| Sterile | Yes<br>Syringe offered non-sterile | Yes |
| Performance | | |
| None under Section 514 | Yes | Yes |
# Differences between Other Legally Marketed Predicate Devices
There are no significant differences between the intended device and the predicate - ProMedic Infant Feeding Tube - K052903.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 9 2007
Mr. Paul Dryden President ProMedic, Inc. 3460 Pointe Creek Court, #102 BONITA SPRINGS FL 34134
Re: K072756
> Trade/Device Name: Feeding tube adapter Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: September 26, 2007 Received: September 27, 2007
Dear Mr. Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2 - Mr. Paul Dryden
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh.dsmamain.html.
Sincerely yours,
Herbert Lemon for
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# 1072756
# Indications for Use Statement
Page 1 of 1
(To be assigned) 510(k) Number:
Device Name:
Feeding tube adapter
Indications for Use:
A non-IV connector for use with Infant Feeding tubes that are intended to be placed into the stomach to permit the introduction of fluids as directed by the physician. Intended for nasogastric or orogastric placement. Limited to < 30 day placement. Not intended for transpyloric placement.
Prescription Use XX (Part 21 CFR 801 Subpart D) 01
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert Leiner
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.