ENDOVIVE INITIAL PLACEMENT DIRECT PEJ KIT, ENDOVIVE STANDARD PROFILE BALLOON REPLACEMENT, MODELS 6520, 6521, 6220
K020120 · Boston Scientific Corp · KNT · Aug 2, 2002 · Gastroenterology, Urology
Device Facts
Record ID
K020120
Device Name
ENDOVIVE INITIAL PLACEMENT DIRECT PEJ KIT, ENDOVIVE STANDARD PROFILE BALLOON REPLACEMENT, MODELS 6520, 6521, 6220
Applicant
Boston Scientific Corp
Product Code
KNT · Gastroenterology, Urology
Decision Date
Aug 2, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Endovive™ Initial Placement Direct PEJ Kit is indicated for nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated. The Endovive™ Standard Profile Balloon Replacement Tube is indicated for use in percutaneous placement of an enteral feeding tube for feeding and/or administration of medication in conjunction with an established GI stoma tract. Typical uses include the replacement of existing gastrostomy and/or jejunostomy feeding tubes. The Standard Profile Balloon Replacement Tube may also be used for decompression.
Device Story
Endovive™ Initial Placement Direct PEJ Kit and Standard Profile Balloon Replacement Tube facilitate enteral feeding and decompression. Initial placement kit enables direct jejunal access; replacement tube maintains access via established GI stoma. Devices consist of internal bolster and feeding tube. Used in clinical settings by healthcare professionals. Provides direct route for nutrition/medication delivery; bypasses upper GI tract when contraindicated. Enables decompression of GI tract. Benefits include reliable enteral access for patients unable to tolerate oral intake.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Devices consist of an internal bolster and a feeding tube. Materials and design are equivalent to previously cleared Endovive™ gastrostomy and jejunostomy devices. Mechanical device; no energy source or software components.
Indications for Use
Indicated for patients requiring nutritional support, medication administration, or gastric/jejunal decompression when upper GI tract feeding is contraindicated. Used for initial percutaneous endoscopic jejunostomy (PEJ) placement or replacement of existing gastrostomy/jejunostomy tubes in established GI stoma tracts.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
KU20120 10+2
AUG 0 2 2002
#### SECTION 14 510(K) SUMMARY
## FOI RELEASABLE
Pursuant to §513(I)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification "…adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness.
| Date: | January 11, 2002 | | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------|--------|------------|
| Common/Usual Names: | Jejunostomy Feeding Tube | | |
| Trade/Proprietary Names: | - Endovive™ Initial Placement Direct PEJ Kit<br>- Endovive™ Standard Profile Balloon Replacement Tube | | |
| Classification Name &<br>Device Classification: | Class II | | |
| | Name | Number | 21CFR Ref. |
| | Tubes, Feeding | 78 FPD | 876.5980 |
| Device Panel/Branch: | Gastroenterology-Urology (GU)<br>Gastro-Renal (GRDB) | | |
| Owner/Operator: | Boston Scientific Corporation<br>One Boston Scientific Place<br>Natick, MA 01760 | | |
| Contact Person: | Paige Sweeney<br>Regulatory Affairs Specialist | | |
{1}------------------------------------------------
## Description of Devices
The Endovive™ Initial Placement Direct PEJ Kit is used during initial placement for direct feeding. The Endovive™ Standard Profile Balloon Replacement Tube is used for the replacement of direct feeding tubes used for direct feeding.
## Indications for Use
The Endovive™ Initial Placement Direct PEJ Kit is indicated for nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated.
The Endovive™ Standard Profile Balloon Replacement Tube is indicated for use in percutaneous placement of an enteral feeding tube for feeding and/or administration of medication in conjunction with an established GI stoma tract. Typical uses include the replacement of existing gastrostomy and/or jejunostomy feeding tubes when feeding via the upper gastrointestinal tract in contraindicated. The Standard Profile Balloon Replacement Tube may be placed percutaneously as described herein or by using a Stamm surgical procedure (not described). The Standard Profile Balloon Replacement Tube may also be used for decompression.
## Descriptive and Technological Characteristics of Proposed and Predicate Devices
Boston Scientific Corporation believes that the Endovive™ Initial Placement Direct PEJ Kit, and the Endovive™ Standard Profile Balloon Replacement Tube, with the additional direct percutaneous endoscopic jejunostomy indication are substantially equivalent to the currently marketed Endovive™ Initial Placement Gastrostomy Kit, the Endovive™ Balloon Replacement Gastrostomy Tube, and the Endovive™ Percutaneous Endoscopic Jejunostomy Device. The major components of these devices are the internal bolster and the feeding tube. A thorough comparison of the descriptive characteristics between the proposed iejunostomy devices and the predicate devices show equivalence.
# Conclusion
Boston Scientific Corporation has demonstrated that the Endovive™ Initial Placement Direct PEJ Kit, and the Endovive™ Standard Profile Balloon Replacement Tube, are substantially equivalent to the currently marketed Endovive™ Initial Placement Gastrostomy Kit, the Endovive™ Balloon Replacement Gastrostomy Tube, and the Endovive™ Percutaneous Endoscopic Jejunostomy Device.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird in flight, composed of three curved lines. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the bird image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 0 2 2002
Ms. Paige Sweeney Regulatory Affairs Specialist Microvasive Endoscopy Boston Scientific Corporation One Boston Scientific Place NATICK MA 01760-1537
Re: K020120
Trade/Device Name: EndoVive™ Initial Placement Direct PEJ Kit and EndoVive™ Standard Profile Balloon Replacement Tube Regulation Number: 21 CFR \$876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 78 KNT Dated: May 8, 2002 Received: May 9, 2002
Dear Ms. Sweeney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807);
{3}------------------------------------------------
Page 2 - Ms. Paige Sweeney
labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains Iodine swabs, 1% Xylocaine, lubricating jelly, and antibiotic ointment, which are subject to regulation as drugs.
Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html
Sincerely yours,
David A. Leggett
for
Nancy C. Brogdon Director, Division of Reproductive, -Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## SECTION 3 INDICATIONS FOR USE
| 510(k) Number: | To Be Determined K020120 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Names: | - Endovive™ Initial Placement Direct PEJ Kit<br>- Endovive™ Standard Profile Balloon Replacement Tube |
| Indications for Use: | The Endovive™ Initial Placement Direct PEJ Kit is indicated for nutritional support and decompression directly into the jejunum when feeding via the upper gastrointestinal tract is contraindicated.<br><br>The Endovive™ Standard Profile Balloon Replacement Tube is indicated for use in percutaneous placement of an enteral feeding tube for feeding and/or administration of medication in conjunction with an established GI stoma tract. Typical uses include the replacement of existing gastrostomy and/or jejunostomy feeding tubes. The Standard Profile Balloon Replacement Tube may also be used for decompression. |
.
.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| | Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|-----------------------|--------------------------------------------------------|--------------------------|
| Prescription Use | OR | Over-The-Counter Use |
| (Per 21 CFR 801.1091) | | (Optional Format 1-2-96) |
David A. Sagnon
| (Division Sign-Off) | |
|---------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K020120 |
Proprietary and Confidential Information of Boston Scientific Corporation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.