The Waters Medical Systems (Waters Instruments, Inc.) RM3 Renal Preservation System is intended to be used to maintain kidneys for transplant.
Device Story
RM3 Renal Preservation System is a transportable, self-contained unit for hypothermic pulsatile perfusion of kidneys during transport and storage. It monitors, displays, and trends perfusion parameters including perfusate flow rate, temperature, systolic/mean/diastolic pressures, and renal resistance. The system features user-configurable audio/visual alarms for parameter limits and includes an integrated printer for data and waveforms. Operated by clinical staff, the device automates monitoring and calculation tasks previously performed manually on legacy systems. By maintaining physiologic conditions via controlled hypothermic perfusion, the device supports kidney viability for transplantation.
Clinical Evidence
Bench testing only. Performance testing demonstrated that the RM3 control unit meets or exceeds the performance of the predicate MOX-100 system for all technologically different features.
Technological Characteristics
Transportable, self-contained control unit for hypothermic pulsatile perfusion. Features automated monitoring, trending, and calculation of renal resistance, flow, and pressure. Includes audio/visual alarm system and integrated printer. Operates via pulsatile hypothermic perfusion methodology.
Indications for Use
Indicated for the maintenance of kidneys intended for transplant.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
Predicate Devices
MOX-100 Renal Preservation System TM2 Transport Module
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/18 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of flowing lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 9 1997
Re: K971571
Mr. Dave Schollman General Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Waters Instruments, Inc. 2411 Seventh Street, N.W. P.O. Box 6117 Rochester, Minnesota 55903-6117
RM3 Renal Preservation System-control unit ---Dated: April 28, 1997 Received: April 30, 1997 Regulatory class: II 21 CFR §876.5880/Product code: 78 KDN
Dear Mr. Schollman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(K) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/odth/dsmamain.html".
Sincerely vours
H.L.Liau Yu
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{1}------------------------------------------------
1. Submitter Information
## -- 510[k] SUMMARY --
JUL 2 9 1997
| Waters Medical Systems, a division of Waters Instruments Inc.<br>2411 Seventh Street N.W.<br>Rochester, MN 55903-6117<br>Estab't. Registration Number: 2123774 | Contact: | Dave Schollman, General Manager |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------|
| | Phone: | 800-426-9877 tel<br>507-252-3700 fax |
| | Summary Preparation Date: April 28, 1997 | |
2. Device Names
| Proprietary Name: | RM3 Renal Preservation System--contol unit |
|----------------------|-------------------------------------------------------------------------------------|
| Common/Usual Name: | renal preservation system--control unit |
| Classification Name: | Isolated kidney perfusion and transport system and accessories<br>(21 CFR 876.5880) |
| Classification: | Class II, Gastroenterology/Urology Panel |
3. Predicate devices
MOX-100 Renal Preservation System TM2 Transport Module, a pre-amendment device.
### 4. Device Description
The RM3 Renal Preservation System is a lightweight, transportable and self-contained renal preservation system, designed to support static monitoring and transportation of kidneys. The RM3 system control unit provides controlled kidney perfusion of hypothermic physiologic solutions, and monitors, displays, trends, and saves important perfusion parameters, including; perfusate flow temperature, rate and pressure, and resistance. The RM3 system control unit can be configured to signal an audio and visual alarm for user-selected limits, and print user-selected data and waveforms.
### 5. Intended use
The RM3 Renal Preservation System is intended to be used to maintain kidneys for transplant.
### 6. Comparison of technological characteristics
The RM3 and the MOX-100 TM2 control units are identical in intended use and methodology. Both systems utilize pulsatile hypothermic perfusion to maintain kidneys for transplant. They are substantially similar in the means by which the pulsatile flow through the kidneys is accomplished. and both allow the user to maintain and monitor nearly the same important perfusion parameters within similar operating ranges. The control units are different in the RM3 automatically monitors, displays, and trends the perfusate pump rate, systolic, mean and diastolic pressures, perfusate temperature, and real-time arterial flow through each kidney, and calculates the renal resistance of each kidney. In the MOX-100 TM2 units the system operator must visually monitor these parameters by reading analog meters and must manually trend the data and calculate renal resistance. Additionally, control of perfusion pump power and rate and coolant circulation pump power different. Test results demonstrate that the RM3 control unit performance meets or exceeds the MOX-100 performance for each of the technologically different features.
April 28, 1997
{2}------------------------------------------------
– ﺮﺩ ﺳ
# ATTACHMENT 1: INTENDED USE STATEMENT
The Waters Medical Systems (Waters Instruments, Inc.) RM3 Renal Preservation System is
intended to be used to maintain kidneys for transplant.
Raler R. Ritting/
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number 971
Prescription Use
(Per 21 CFR 801.109)
Over-the-Counter Use_
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.