CoStorSol® plus G is intended for the flushing and cold storage of liver, kidney, and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
Device Story
CoStorSol® plus G is a sterile, non-pyrogenic, hypothermic organ storage solution; formulation includes soluble colloids, buffers, sodium/potassium salts, redox stabilizers, and phosphoric compounds. Used by transplant teams during organ procurement; solution is chilled to 2-6°C; organs are flushed to clear blood and stored in solution to depress metabolism and maintain viability. Device is supplied in 1-liter flexible EVA-laminated plastic bags with an optional 0.922g sterile glutathione powder vial. Glutathione is an optional additive previously cleared for use with the predicate; inclusion in the package is for clinical convenience. No change to intended use or clinical methods. Benefits include maintained histological viability of donor organs during transport.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (cytotoxicity, acute systemic toxicity, dermal sensitization, hemolysis, intracutaneous reactivity/irritation per ISO 10993), particulate matter (USP <788>), and shelf-life validation (accelerated and natural aging).
Technological Characteristics
Synthetic medium of organic/inorganic compounds; pH 7.4; osmolality ~320 mOsM. Supplied in 1L flexible EVA-laminated plastic bags. Sterilized via 0.1μm membrane filtration. Biocompatibility per ISO 10993; particulate matter per USP <788>. Shelf life 1 year at 2-25°C.
Indications for Use
Indicated for flushing and cold storage of liver, kidney, and pancreas organs during removal from donor, transport, and preparation for transplantation into a recipient.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2015
Preservation Solutions, Inc. % Neil Burris Principal Consultant Neil Burris and Associates 4250 Grove Street Denver, CO 80211
Re: K151728
> Trade/Device Name: CoStorSol® plus G Regulation Number: 21 CFR§ 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories. Regulatory Class: II Product Code: KDN Dated: June 24, 2015 Received: June 26, 2015
Dear Neil Burris,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151728
Device Name CoStorSol plus G
Indications for Use (Describe)
CoStorSol® plus G is intended for the flushing and cold storage of liver, kidney, and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/>Prescription Use (Part 21 CFR 801 Subpart D)</div> |
|--------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input type="checkbox"/>Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## SUBMITTER INFORMATION
- A. Company Name: Preservation Solutions Inc.
- B. Company Address: 1099 Proctor Drive Elkhorn, Wisconsin 53121
- C. Company Phone: 262 723 6715
- D. Company Facsimile: 262 723 4013
- E. Contact Person: William Wagner Quality Assurance Director
## DEVICE IDENTIFICATION
- A. Device Trade Name: CoStorSol® plus G
- B. Device Common Name: Organ Storage Solution
- C. Classification Name: Isolated kidney perfusion and transport system and accessories
- D. Class II (21 CRF 876.5880)
- E. Device Code: KDN
## IDENTIFICATION OF PREDICATE DEVICES
CoStorSol® cold storage solution cleared to market under:
#### Premarket Notifications K091245, K083453, and K073693.
is an exact predicate to the proposed device CoStorSol® plus G.
#### DEVICE DESCRIPTION
Preservation Solutions, Inc. manufacturers CoStorSol® plus G according to a "recipe" pioneered at the University of Wisconsin by Folkert O. Belzer, MD and James H. Southard, PhD. Indeed, this cold storage solution is often referred to as "Belzer UW" solution. The formulation includes soluble colloids, buffers, sodium and potassium salts, redox stabilizers, and phosphoric compounds.
CoStorSol® plus G is a clear to light yellow, sterile, non-pyrogenic solution for hypothermic flushing and storage of organs. The solution is packaged in 1-liter bags, to which 0.922 g of
{4}------------------------------------------------
sterile glutathione may be added just prior to use. The solution must be chilled to between 2° and 6° C prior to use.
#### INDICATIONS FOR USE
#### INTENDED USE
CoStorSol® plus G is intended for the flushing and cold storage of liver, kidney, and pancreas organs at the time of organ removal from the donor in preparation for storage, transportation and eventual transplantation into a recipient.
## COMPARISON TO PREDICATE DEVICES
CoStorSol® solution received substantial equivalence determinations after submissions filed under premarket notification (510(k)) numbers K091245, K083453, and K073693. The modified device, CoStorSol® plus G, is exactly the same as the predicate device, except that the new device will include a single vial of sterile glutathione powder (0.922 g). The glutathione is included as a convenience should the attending physicians, and/or transplant team, choose to use it as an optional additive to 1 Liter of CoStorSol® at the time of use. The Instructions for Use of CoStorSol® have always included statements advising the optional addition of 0.922 g of glutathione when deemed necessary due to the exact circumstances of the transplant case. Including a vial of glutathione in a package of CoStorSol® plus G thus represents no change to either the intended use or the specific methods of use for the device.
Substantial Equivalence Comparison Table for CoStorSol® plus G vs. CoStorSol®
| Device<br>Attribute | CoStorSol® plus G<br>(Proposed Device) | CoStorSol® K091245<br>(Predicate Device) | Equivalent? |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | •CoStorSol® plus G is intended for the<br>flushing and cold storage of liver, kidney,<br>and pancreas organs at the time of organ<br>removal from the donor in preparation for<br>storage, transportation and eventual<br>transplantation into a recipient. | CoStorSol® is intended for the flushing<br>and cold storage of kidney, liver and<br>pancreas organs at the time of organ<br>removal from the donor in preparation<br>for storage, transportation and eventual<br>transplantation into a recipient. | Identical |
| Preparation for<br>Use | Chill to 2° to 6° C without freezing;<br>Required addition to 1L of CoStorSol®:<br>Penicillin G-200,000 units, Insulin-40<br>Units, Dexamethasone-16mg;<br>Optional addition to 1L CoStorSol®:<br>glutathione 0.922g | Chill to 2° to 6° C without freezing;<br>Required addition to 1L of CoStorSol®:<br>Penicillin G-200,000 units, Insulin-40<br>Units, Dexamethasone-16mg;<br>Optional addition to 1L CoStorSol®:<br>glutathione 0.922g | Identical |
| Device<br>Attribute | CoStorSol® plus G<br>(Proposed Device) | CoStorSol® K091245<br>(Predicate Device) | Equivalent? |
| Chemical<br>Composition | A synthetic medium of defined organic,<br>and inorganic, soluble chemical<br>compounds with specified purity.<br>pH = 7.4 nominal<br>osmolality $\approx$ 320 mOsM | A synthetic medium of defined organic,<br>and inorganic, soluble chemical<br>compounds with specified purity.<br>pH = 7.4 nominal<br>osmolality $\approx$ 320 mOsM | Identical |
| Packaging -<br>How Supplied | Filter sterilized using 0.1μm membrane(s)<br>and aseptically filled into biocompatible, 1<br>liter flexible solution bags made of<br>laminated plastic with an ethylene-vinyl-<br>acetate (EVA) layer contacting the<br>solution. The bags have integrated<br>delivery set ports, and are packaged with<br>a single 10mL polyethylene terephthalate<br>(PETG) vial holding 0.922 g sterile non-<br>pyrogenic glutathione.<br>CoStorSol® is non-pyrogenic. Shelf<br>packs contain 6, 1-liter bags and 6 vials<br>of glutathione powder | Filter sterilized using 0.1μm membrane(s)<br>and aseptically filled into biocompatible,<br>1 liter flexible solution bags made of<br>laminated plastic with an ethylene-vinyl-<br>acetate (EVA) layer contacting the<br>solution. The bags have integrated<br>delivery set ports. CoStorSol® is non-<br>pyrogenic. Shelf packs contain 10, 1-<br>liter bags. | Substantially<br>Equivalent -<br>Sterile<br>glutathione, an<br>optional additive<br>to both the<br>subject device<br>and predicate,<br>is provided with<br>1L units of<br>CoStorSol®. |
| Sterility | Sterility Assured via 0.1 μm membrane<br>filtration | Sterility Assured via 0.1 μm membrane<br>filtration | Identical |
| Storage<br>Temperature<br>and Shelf Life | Indoors with temperature controlled at 2°-<br>25°C, without freezing;<br>1year (12 month) shelf life | Indoors with temperature controlled at<br>2°-25°C, without freezing;<br>1 year (12 month) shelf life | Identical |
| Actions and<br>Functions | Use cold solution to flush organ<br>immediately before and/or after removal<br>to clear blood from the vasculature. Store<br>organs cold in aliquot of the same solution<br>to maintain histological viability by<br>depressing metabolism. | Use cold solution to flush organ<br>immediately before and/or after removal<br>to clear blood from the vasculature.<br>Store organs cold in aliquot of the same<br>solution to maintain histological viability<br>by depressing metabolism. | Identical |
| Particulate<br>Matter and<br>Biocompatibility | Particle Counts less than limits for Large<br>Volume Injections per USP <788>;<br>Biocompatible per ISO 10993-1 battery of<br>tests for Externally Communicating Blood<br>Path Indirect Contact for prolonged<br>periods >24 hours. | Particle Counts less than limits for<br>Large Volume Injections per USP<br><788>; Biocompatible per ISO 10993-1<br>battery of tests for Externally<br>Communicating Blood Path Indirect<br>Contact for prolonged periods >24<br>hours. | Identical |
| Interaction with<br>Other Medical<br>Technology | CoStorSol® plus G is not intended for<br>continuous perfusion. Standard<br>transplantation surgical expertise and<br>techniques are required. | CoStorSol® is not intended for<br>continuous perfusion. Standard<br>transplantation surgical expertise and<br>techniques are required. | Identical |
Continued next page
{5}------------------------------------------------
# BIOCOMPATIBILITY, STERILIZATION, PACKAGING, AND BENCH TESTING
Test results have shown CoStorSol® plus G to be a biocompatible solution, supplied in flexible solution bags made of laminated plastic with an ethylene-vinyl-acetate (EVA) layer contacting the solution, which have likewise been tested and shown to be biocompatible. The ISO 10993 series of standards were referenced during the planning and execution of all biocompatibility testing,
{6}------------------------------------------------
including: cytotoxicity, acute systemic toxicity, dermal sensitization, hemolysis, and intracutaneous reactivity/irritation. Particulate matter does not exceed the limits set in USP Section <788>, for large volume injections.
The 1 Liter of CoStorSol®, and the single vial of glutathione, are both supplied sterile and nonpyrogenic in order to assure safety for transplant recipients.
Sterilization processes for CoStorSol® plus G were validated according to the ISO 13408 series applicable to aseptic filling, or the ISO 11137 series of standards application to radiation sterilization, as appropriate.
Preservation Solutions, Inc. has validated shelf life for CoStorSol® plus G using a combination of accelerated aging and natural aging protocols, supporting expiry periods of up to 1 year. Shelf life verification tests show that aged CoStorSol® plus G remains functionally equivalent to newly manufactured solution via tests for biocompatibility, particulate matter, chemical identity, general appearance, weight loss, and package integrity.
#### CONCLUSION
The above statements establish substantial equivalence between the predicate CoStorSol® product, and CoStorSol® plus G.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.