K173972 · Fresenius Medical Care Renal Therapies Group, LLC · KDI · May 10, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K173972
Device Name
2008T BlueStar Hemodialysis Machine
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility. Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20 kg and ≤40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤40 kg. The 2008 BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
Device Story
Electromechanical hemodialysis machine for healthcare facilities; controls/monitors dialysate and extracorporeal blood circuits. Pumps blood through dialyzer; semi-permeable membrane uses diffusion/ultrafiltration to remove toxins/excess water. Mixes dialysate concentrates with purified water; heats/degasses/delivers to dialyzer. Balancing chambers control fluid flow. Monitors venous/arterial pressures, air, and blood presence. Optional modules: bibag (bicarbonate), Crit-Line Clip Monitor (CLiC) (non-invasive hematocrit/oxygen saturation/blood volume), BTM, BVM. Clinician/nurse operates device under physician direction; uses real-time data to adjust fluid removal rates to prevent complications like nausea/cramping. Benefits patient by providing effective, controlled dialysis treatment.
Clinical Evidence
Bench testing only. Included biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601-1-2:2014), software verification/validation (unit, integration, system level), mechanical functional verification, environmental/ship testing, and human factors testing. No clinical trial data presented.
Indicated for acute and chronic dialysis therapy in healthcare facilities. Supports Isolated Ultrafiltration, SLED, and low volume hemodialysis for patients 20-40 kg. Contraindicated for patients <20 kg, plasma replacement therapies, or renal therapies using substitution fluid. SLED contraindicated for patients ≤40 kg.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 10, 2018
Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451
Re: K173972
> Trade/Device Name: 2008T BlueStar Hemodialysis Machine Regulation Number: 21 CFR& 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI Dated: April 6, 2018 Received: April 10, 2018
Dear Denise Oppermann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K173972
Device Name 2008T BlueStar Hemodialysis Machine
Indications for Use (Describe) 2008T BlueStar Hemodialysis Machine:
The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20 kg and ≤40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤40 kg. The 2008 BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
### bibag System (Optional):
The bibag system is used with three stream proportioning Hemodialysis Machines equipped with the bibag module such as the 2008T BlueStar Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
### Crit-Line Clip Monitor (CLiC) (Optional):
The Crit-Line Clip Monitor is used with the 2008T BlueStar Hemodialysis Machine to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the monitor provides, the clinician nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three downward-pointing chevrons stacked on top of each other.
### ട. 510(K) SUMMARY
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR §807.92.
### 5.1. Submitter's Information
| Name: | Fresenius Medical Care Renal Therapies Group, LLC |
|-------------------|-------------------------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA 02451-1457 |
| Phone: | (781) 699-4479 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann<br>Senior Director, Regulatory Affairs – Devices |
| Preparation Date: | 4 May 2018 |
### 5.2. Device Name
| Trade Name: | 2008T BlueStar Hemodialysis Machine |
|-----------------------|------------------------------------------------------------------------|
| Common Name: | Dialyzer, High Permeability With or Without<br>Sealed Dialysate System |
| Classification Name: | High Permeability Hemodialysis System |
| Regulatory Class: | Class II per 21 CFR §876.5860 |
| Product Code | KDI |
| Classification Panel: | Gastroenterology/Urology |
### Legally Marketed Predicate Device 5.3.
#### 5.3.1. Primary Predicate - 2008K2 Hemodialysis Machine (K153449)
The 2008K2 Hemodialysis Machine (K153449) is the primary predicate for the 2008T BlueStar Hemodialysis Machine.
#### 5.3.2. Reference Device – 2008T Hemodialysis Machine (K150708)
The 2008T Hemodialysis Machine (K150708) is the reference device for the 2008T BlueStar Hemodialysis Machine.
### Device Description 5.4.
#### 5.4.1. Device Identification
The BlueStar features will be implemented on both New Production Machines and 2008T Machines currently on the market (Table 1). New Production 2008T BlueStar Hemodialysis
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue geometric shape on the left, resembling a stylized downward-pointing arrow composed of three chevrons. To the right of the shape is the text "FRESENIUS" in bold, blue letters, with "MEDICAL CARE" in a smaller font size below it, also in bold, blue letters. The background is white.
(HD) Machines will be available in four (4) configurations. Clinics will be given the option to upgrade the 4 current 2008T Hemodialysis Machine configurations with the new BlueStar software and hardware via Upgrade Kits.
| Product<br>Code | Product Code Description | BlueStar<br>Features<br>Implementation |
|-----------------|----------------------------------------|----------------------------------------|
| 190713 | 2008T Hemodialysis System with CDX | |
| 190766 | 2008T Hemodialysis System with BIBAG | Field Upgrade |
| 190858 | 2008T Hemodialysis System w/o CDX | (Standard or<br>Premium) |
| 190895 | 2008T GEN2 BIBAG without CDX | |
| 191124 | 2008T HD SYS. CDX BLUESTAR | |
| 191126 | 2008T HD SYS. CDX W/BIBAG BLUESTAR | New Production |
| 191128 | 2008T HD SYS. W/O CDX BLUESTAR | |
| 191130 | 2008T HD SYS. W/O CDX W/BIBAG BLUESTAR | |
Table 1: BlueStar Hemodialysis Machines and Field Upgrade Machines
Both Standard and Premium Upgrade Kits will be offered to clinics to upgrade current 2008T Hemodialysis Machines to BlueStar software and hardware.
### 5.4.2. Device Characteristics
The 2008T BlueStar Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
### 5.4.3. Environment of Use
The 2008T BlueStar Hemodialysis Machine is to be used in healthcare facilities.
### 5.4.4. Brief Written Description of the Device
The 2008T BlueStar Hemodialysis Machine provides hemodialysis treatment by controlling and monitoring both the dialysate circuit and the extracorporeal blood circuit. The machine pumps blood from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzer contains a semi-permeable membrane that uses diffusion to transfer toxins and ultrafiltration to transport excess water from the blood into the dialysate circuit. In this separate dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, and delivered to the dialyzer. Balancing chambers control the incoming flow and outgoing flow of the dialysate fluid during ultrafiltration. During hemodialysis, the extracorporeal blood circuit is monitored for venous and arterial blood pressures as well as for the presence of air and blood. The 2008T BlueStar Hemodialysis Machine accommodates the following accessory devices and options:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
### Accessories
- Diasafe®plus Filter (K070049) ●
- Patient Card (subject to review in this 510(k)) ●
- Patient Card Reader (subject to review in this 510(k)) ●
- Bloodlines: 6.35mm and 8mm (K962081. K000451. K001107. K022536. K070049 . and K120823)
- Dialyzers: Any commercially-available dialyzer equipped with ISO 8637 Standard dialysis connectors
## Options
- bibag® – K162716 (stand-alone disposable) and K121341 (bibag disposable cleared with 2008T HD Machine)
- CDX (Clinical Data Exchange) K093902 (CDX cleared with 2008T HD Machine) ●
- CLiC (Crit-Line in a Clip Monitor) K121599 (Stand-alone CLiC) and K131908 ● (CLiC with 2008T HD Machine)
- BTM (Blood Temperature Monitor) K941460 (Stand-alone BTM) and K080964 ● (BTM with 2008T HD Machine)
- BVM (Blood Volume Monitor) – K982926 (Stand-alone BVM) and K994267 (BVM with 2008K HD Machine)
- Single Needle System K080964 (Single Needle with 2008T HD Machine) .
#### 5.4.5. Materials of Use
The 2008T BlueStar Hemodialysis Machine's hydraulic system is composed of the following indirect, prolonged contact, externally communicating materials:
- Plastic/Rubber:
- PAEK (Polyaryletherketone) —
- PEI (Polyetherimide) —
- PESU (Polyethersulfone) -
- PSU (Polysulfone) -
- PET (Polyethylene terephthalate) —
- PUR (Polyurethane) -
- -PET (Polyethylene terephthalate)
- PP (Polypropylene) -
- PPO (Polyphenylene oxide) —
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons, stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
- PPS (Polyphenylene Sulfide) -
- -PPSU (Polyphenylsulfone)
- PTFE (Polytetrafluoroethylene) -
- PVDF (Polyvinylidene fluoride) -
- EPDM (Ethylene Propylene Diene Monomer Rubber) —
- -Silicone
- Metals ●
- Stainless Steel -
- Tantalum Tungsten -
- -Tungsten
- Glass
- Borosilicate Glass -
The hydraulic lines of the machines are in contact with the dialysate circuit. The dialysate circuit has prolonged, indirect blood contact. Modifications have been made to components of the hydraulic system. Biocompatibility testing was conducted to verify the changes.
#### Essential Performance Characteristics 5.4.6.
The essential performance characteristics of the 2008T BlueStar Hemodialysis Machine are listed in Table 2. A comparison of its essential performance characteristics with those of the 2008K2 Hemodialysis Machine (K153449) and 2008T Hemodialysis Machine (K150708) is provided in Section 12.
| Feature | Specification | |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Blood Flow Rates | Blood line | Blood flow rate |
| | 8 mm | 20–600 mL/min* |
| | 6.35 (displayed as 6.4) mm | 20–465 mL/min |
| | 4.8 mm | 10–274 mL/min |
| | 2.6 mm | 6–86 mL/min |
| | *Not available with the Low Volume feature enabled<br>Accuracy: ± 10% tested at -200 mmHg | |
| Dialysate Flow Rates | Dialysate flow rates are selectable on the Home screen in the<br>following mL/min increments:<br>(0)/100 ‡/150‡/200‡/300†/400/500/600/700/800 | |
### Table 2 2008T BlueStar Hemodialysis Machine Essential Performance Characteristics
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the Fresenius Medical Care logo. The logo consists of a blue emblem on the left and the text "FRESENIUS MEDICAL CARE" on the right. The emblem is a stylized design with three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
## 2008T BlueStar Hemodialysis Machine Traditional 510(k)
| Feature | Specification | | | | | | | | | | | | | | | | | | | | | | | | |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| | † Sustained Low Efficiency Dialysis (SLED) | | | | | | | | | | | | | | | | | | | | | | | | |
| | ‡ Flow rate requires that the Allow Slow Flow option be<br>selected in Service mode. | | | | | | | | | | | | | | | | | | | | | | | | |
| | The dialysate flow rates (Qd) for both 1.5x or 2.0x dialysate<br>flow (Auto Flow), based on the Blood Pump rate (Qb). | | | | | | | | | | | | | | | | | | | | | | | | |
| | Qb w/1.5x Qd Qb w/2.0x Qd Qd 0 – 165* 0 – 150* 300 166 – 215* 151 – 215* 400 216 – 315* 216 – 265* 500 315 and below** 265 and below** 500 316 – 415 266 – 315 600 416 – 480 316 – 365 700 481 and above 366 and above 800 | | | | | | | | | | | | | | | | | | | | | | | | |
| | Note: All flow rates are approximate. Dialysate flow will not<br>adjust unless the blood pump is adjusted at least 15–<br>20 mL/min.<br>* If Auto Flow Minimum of 300 Qd is set in Service mode<br>** If Auto Flow Minimum of 500 Qd is set in Service mode | | | | | | | | | | | | | | | | | | | | | | | | |
| Net Fluid Removal | 0–4000 mL/hr | | | | | | | | | | | | | | | | | | | | | | | | |
| | Dialysate flow rate Accuracy (on total vol. removed) 100 mL/min ± (1% UF rate + 18 mL/hr) 500 mL/min ± (1% UF rate + 30 mL/hr) 800 mL/min ± (1% UF rate + 48 mL/hr) | | | | | | | | | | | | | | | | | | | | | | | | |
| Dialysis Time | Dialysis type Time Dialysis 0-9:59 hours* SLED Fixed at 12 hours | | | | | | | | | | | | | | | | | | | | | | | | |
| | *Time can be adjusted manually<br>Accuracy: ± 1 second per hour | | | | | | | | | | | | | | | | | | | | | | | | |
| Dialysis Fluid Composition | Volumetric, selectable:<br>Acid adjustment range: 130–155 mEq/L Na+<br>Bicarbonate adjustment range: 20-40 mEq/L Bicarbonate<br>(post-reaction, after mixing with the acid and purified water), | | | | | | | | | | | | | | | | | | | | | | | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
| Feature | Specification |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Monitoring conductivity average accuracy: ± 1.5% | |
| Dialysis Fluid Temperature | Range 35°C–39°C with alarm limit window automatically<br>adjusted to 2°C above and below set point. Alarm window will<br>not adjust to below 34°C (or 30°C during BTM recirculation<br>measurement) or above 41 °C. |
| Accuracy: ± 0.3°C | |
| Heparin Delivery Rate | 0 – 9.9 mL/hr |
| Accuracy: ± 5% | |
### ર.ડ. Intended Use
The intended use of the 2008T BlueStar Hemodialysis Machine is identical to the primary and reference devices' intended use.
The 2008T Hemodialysis Machine is intended for acute and chronic dialysis therapy.
### 5.6. Indications for Use
2008T BlueStar Hemodialysis Machine: The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥20 kg and ≤40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤40 kg. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
## bibag System (Optional):
The bibag system is used with three stream proportioning Hemodialysis Machines equipped with the bibag module such as the 2008T BlueStar Hemodialysis Machine and is intended for use in bicarbonate hemodialysis for acute and chronic renal failure. The bibag is intended for extracorporeal bicarbonate hemodialysis according to a physician's prescription.
Crit-Line Clip Monitor (CLiC) (Optional):
The Crit-Line Clip Monitor is used with the 2008T BlueStar Hemodialysis Machine to noninvasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC device measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the clinician. Based on the data that the
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image contains the logo for Fresenius Medical Care. On the left side of the logo is a blue symbol that consists of three downward-pointing chevrons stacked on top of each other. To the right of the symbol, the words "FRESENIUS" and "MEDICAL CARE" are written in blue, with "FRESENIUS" on top of "MEDICAL CARE."
monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common complications of dialysis which include nausea, cramping and vomiting
### 5.7. Comparison of Technological Characteristics with the Predicate Device
A comparison of the technological characteristics of the 2008T BlueStar Hemodialysis Machine with the 2008K2 Hemodialysis Machine (primary predicate) and the 2008T Hemodialysis Machine (reference device) is provided in Table 3.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a blue symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right, also in blue. The symbol is made up of three chevron-like shapes stacked on top of each other, pointing downwards.
| Parameter | Modified Device – 2008T<br>BlueStar Hemodialysis<br>Machine | Primary Predicate Device –<br>2008K2 Hemodialysis<br>Machine (K153449) | Reference Device<br>– 2008T Hemodialysis<br>Machine (K150708) | Equivalence Discussion |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Intended Use | The device is intended for<br>acute and chronic dialysis<br>therapy. | The device is intended for<br>acute and chronic dialysis<br>therapy. | The device is intended for<br>acute and chronic dialysis<br>therapy. | Same |
| Indications for Use | 2008T BlueStar Hemodialysis<br>Machine: The 2008T BlueStar<br>Hemodialysis Machine is<br>indicated for acute and<br>chronic dialysis therapy in a<br>healthcare facility. | The 2008K2 Hemodialysis<br>Machine is indicated for acute<br>and chronic hemodialysis<br>therapy in a healthcare<br>facility. | N/A. Substantial Equivalence<br>discussion is limited to the<br>modified and primary<br>predicate devices. | Same |
| | Additional therapy options<br>for patients receiving<br>hemodialysis include:<br>Isolated Ultrafiltration,<br>Sustained Low Efficiency<br>Dialysis (SLED), and low<br>volume hemodialysis<br>(patientsweighing≥ 20 kg<br>and≤ 40 kg). This machine<br>accommodates the use of both<br>low flux and high flux<br>dialyzers. The SLED therapy<br>option is not to be used for<br>patients weighing ≤ 40 kg.<br>The 2008T BlueStar<br>Hemodialysis Machine is not<br>to be used for plasma<br>replacement therapies, for<br>patients weighing less than<br>20 kg, or for renal therapies<br>using substitution fluid | Additional therapy options<br>for patients receiving<br>hemodialysis include:<br>Isolated Ultrafiltration,<br>Sustained Low Efficiency<br>Dialysis (SLED), and low<br>volume hemodialysis<br>(patients weighing ≥ 20 kg<br>and ≤ 40 kg). This machine<br>accommodates the use of both<br>low flux and high flux<br>dialyzers. The SLED therapy<br>option is not to be used for<br>patients weighing ≤ 40 kg.<br>The 2008K2 Hemodialysis<br>Machine is not to be used for<br>plasma replacement therapies,<br>for patients weighing less<br>than 20 kg, or for renal<br>therapies using substitution fluid | | Same |
| Parameter | Modified Device – 2008T<br>BlueStar Hemodialysis<br>Machine | Primary Predicate Device –<br>2008K2 Hemodialysis<br>Machine (K153449) | Reference Device<br>– 2008T Hemodialysis<br>Machine (K150708) | Equivalence Discussion |
| bibag System (Optional):<br>The bibag system is used with<br>three stream proportioning<br>Hemodialysis Machines<br>equipped with the bibag<br>module such as the 2008T<br>BlueStar Hemodialysis<br>Machine and is intended for<br>use in bicarbonate<br>hemodialysis for acute and<br>chronic renal failure. The<br>bibag is intended for<br>extracorporeal bicarbonate<br>hemodialysis according to a<br>physician's prescription. | Bibag functionality is not<br>available on the 2008K2<br>Hemodialysis Machine. | | Substantially Equivalent<br>The Indications for Use<br>language for the optional<br>bibag system was<br>leveraged from the<br>currently marketed<br>(unmodified) 2008T<br>Hemodialysis Machine.<br>There were no<br>modifications made to<br>the optional bibag<br>system with the 2008T<br>BlueStar Hemodialysis<br>Machine. | |
| Crit-Line Clip Monitor<br>(CLiC) (Optional):<br>The Crit-Line Clip Monitor is<br>used with the 2008T BlueStar<br>Hemodialysis Machine to<br>non-invasively measure<br>hematocrit, oxygen saturation<br>and percent change in blood<br>volume. The CLiC device<br>measures hematocrit, percent<br>change in blood volume and<br>oxygen saturation in real time<br>for application in the<br>treatment of dialysis patients<br>with the intended purpose of<br>providing a more effective<br>treatment for both the dialysis | Crit-Line Clip Monitor<br>(CLiC) functionality is not<br>available on the 2008K2<br>Hemodialysis Machine. | | Substantially Equivalent<br>The Indications for Use<br>language for the optional<br>CLiC functionality was<br>leveraged from the<br>currently marketed<br>(unmodified) 2008T<br>Hemodialysis Machine.<br>There were no<br>modifications made to<br>the optional (CLiC)<br>functionality with the<br>2008T BlueStar<br>Hemodialysis Machine. | |
| MEDICAL CARE | | | | |
| Parameter | Modified Device – 2008T<br>BlueStar Hemodialysis<br>Machine | Primary Predicate Device –<br>2008K2 Hemodialysis<br>Machine (K153449) | Reference Device<br>– 2008T Hemodialysis<br>Machine (K150708) | Equivalence Discussion |
| | patient and the clinician.<br>Based on the data that the<br>monitor provides, the<br>clinician/nurse, under<br>physician direction,<br>intervenes (i.e., increases or<br>decreases the rate at which<br>fluid is removed from the<br>blood) in order to remove the<br>maximum amount of fluid<br>from the dialysis patient<br>without the patient<br>experiencing the common<br>complications of dialysis<br>which include nausea,<br>cramping and vomiting. | | | |
| Dimensions | | N/A. Substantial Equivalence<br>discussion is limited to the<br>modified and reference<br>devices. | | |
| Floor Space | Approximately 54 cm wide<br>by 63 cm deep | | Approximately 54 cm wide<br>by 63 cm deep | Same as the reference<br>(unmodified) 2008T |
| Height | 149 cm | | 149 cm | device. |
| Operating<br>Temperature and<br>Humidity | $60°F – 100°F (15.5°C –38°C)$<br>Relative Humidity 10% to<br>90%, non-condensing | $60°F - 100°F (15.5°C –38°C)$<br>Relative Humidity 10% to<br>90%, non-condensing | | |
| Electrical | | | | |
| Power Supply – Main | Single phase AC 117 V ±<br>10% 60 Hz ± 3 Hz; must be<br>connected to a circuit which<br>is equipped with a hospital<br>grade receptacle and is<br>protected by circuit breaker<br>and ground fault interrupter<br>(GFI). | Single phase AC 117 V ±<br>10% 60 Hz ± 3 Hz; must be<br>connected to a circuit which<br>is equipped with a hospital<br>grade receptacle and is<br>protected by circuit breaker<br>and ground fault interrupter<br>(GFI). | Single phase AC 117 V ±<br>10% 60 Hz ± 3 Hz; must be<br>connected to a circuit which<br>is equipped with a hospital<br>grade receptacle and is<br>protected by circuit breaker<br>and ground fault interrupter<br>(GFI). | Same |
| MEDICAL CARE | | Traditional 510(k) | | |
| Parameter | Modified Device – 2008T<br>BlueStar Hemodialysis<br>Machine | Primary Predicate Device –<br>2008K2 Hemodialysis<br>Machine (K153449) | Reference Device<br>– 2008T Hemodialysis<br>Machine (K150708) | Equivalence Discussion |
| | chassis to ground must be<br>< 0.2 ohm. | chassis to ground must be<br>< 0.2 ohm. | chassis to ground must be<br>< 0.2 ohm. | |
| Power Consumption | Does not exceed 12.6 amps | Does not exceed 12.5 amps | Does not exceed 12.5 amps | Substantially Equivalent |
| Protection Against<br>Electric Shock | Type: Safety class I Degree:<br>Type B<br>Type CF: Only BPM Blood<br>Pressure Cuff | Type: Safety class I Degree:<br>Type B<br>Type CF: Only BPM Blood<br>Pressure Cuff | Type: Safety class I Degree:<br>Type B<br>Type CF: Only BPM Blood<br>Pressure Cuff | Same |
| Materials | | | | |
| Patient-Contacting<br>Materials | All direct and indirect patient-<br>contacting materials have<br>been evaluated for<br>Biocompatibility | All direct and indirect patient-<br>contacting materials have<br>been evaluated for<br>Biocompatibility | All direct and indirect patient-<br>contacting materials have<br>been evaluated for<br>Biocompatibility | Same |
| Water | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.