The intermittent catheter is a single use flexible tubular device that is inserted through the urethra by male, female, and pediatric patients who need to drain urine from the bladder.
Device Story
VaPro Pocket is a single-use, flexible, intermittent urinary catheter; designed for bladder drainage in patients with urinary incontinence. Device features: hydrophilic-coated, phthalate-free PVC tube; two smooth eyelets; protective sleeve; protective introducer tip to shield against distal urethral bacteria; color-coded funnel. Packaging includes a sealed water compartment; water migrates to catheter compartment to activate hydrophilic coating. Device is ready-to-use; no collection bag utilized. Designed for discreet storage and easy access. Operated by patient or caregiver; used in home or clinical settings. Provides non-invasive bladder management; reduces risk of contamination during insertion via protective tip and sleeve.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included package integrity, transportation, and sterility validation (Gamma Irradiation per ANSI/AAMI/ISO 11137-1:2006 & -2:2013). Biocompatibility previously established in K141642; new packaging biocompatibility testing performed with satisfactory results.
Indicated for male, female, and pediatric patients requiring intermittent urinary bladder drainage.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are white and are set against a white background. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2015
Hollister Incorporated Jeanne Lee Sr. Manager, Regulatory Affairs 2000 Hollister Drive Libertyville, IL 60048
Re: K143594
> Trade/Device Name: VaPro™ Pocket Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: GBM Dated: December 17, 2014 Received: December 18, 2014
Dear Jeanne Lee,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -A" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible and well-defined. The overall impression is clean and professional.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
:
510(k) Number (if known)
## K143594
Device Name VaPro™ Pocket 12.18
Indications for Use (Describe)
VaPro™ Pocket 40cm intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.
VaPro™ Pocket 20cm intermittent catheter is a flexible tubular device that is inserted through the urethra by female patients who need to drain urine from the bladder.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Hollister company logo. To the left of the company name is a red geometric shape that resembles a plus sign with additional lines. The company name "Hollister" is written in a serif font.
# 510(k) Summary
| Submitted By: | Jeanne Lee<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 847-996-6350<br>(f) 847-918-3981 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared: | January 6, 2015 |
| Device Name: | Classification Name- Urological catheter and accessories<br>Common/Usual Name- Catheter, Urethral<br>Proprietary Name- VaPro™ Pocket intermittent catheter |
| Device Class/Product Code: | Urological catheter and accessories are Class II devices per<br>876.5130 (Therapeutic Devices). Product code-GBM. |
| Predicate Device: | VaPro intermittent catheter, K141642 |
| Device Description: | The VaPro Pocket intermittent catheter is a single use catheter to<br>be used as a means of managing urinary incontinence by draining<br>urine from the bladder. The catheter has the following elements:<br>∑ hydrophilic-coated PVC catheter (phthalate free)<br>∑ 2 smooth catheter eyelets<br>∑ protective sleeve<br>∑ protective introducer tip as a way to shield the sterile<br>catheter from bacteria in the distal urethra during insertion<br>∑ color coded funnel<br>The packaging contains a sealed water compartment chamber of<br>which the water migrates to the catheter compartment and<br>lubricates the catheter. The outer packaging was designed to be<br>discreet, easy to store, easy to open and facilitate access to the<br>catheter.<br>This product does not utilize a collection bag. |
| Intended Use: | The intermittent catheter is a single use flexible tubular device that<br>is inserted through the urethra by male, female, and pediatric<br>patients who need to drain urine from the bladder. |
{4}------------------------------------------------
Technological Characteristics: The table below summarizes the technological characteristics of the device as compared to the predicate device.
| | Modified Device:<br>VaPro Pocket 20cm<br>intermittent catheter | Modified Device:<br>VaPro Pocket 40cm<br>intermittent catheter | Predicate Device:<br>VaPro intermittent<br>catheter (K141642) |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The intermittent<br>catheter is a single use<br>flexible tubular device<br>that is inserted through<br>the urethra by female<br>patients who need to<br>drain urine from the<br>bladder. | The intermittent catheter is a single use flexible<br>tubular device that is inserted through the urethra<br>by male, female, and pediatric patients who need<br>to drain urine from the bladder. | The intermittent catheter is a single use flexible<br>tubular device that is inserted through the urethra<br>by male, female, and pediatric patients who need<br>to drain urine from the bladder. |
| Condition of Use | Single Use | Single Use | Single Use |
| Prelubricated | Yes-by water vapor<br>hydration | Yes-by water vapor<br>hydration | Yes-by water vapor<br>hydration |
| Ready to use | Yes | Yes | Yes |
| End Design | Funnel | Funnel | Funnel |
| Sterile | Yes | Yes | Yes |
| No touch design | Yes-has protective<br>sleeve | Yes-has protective<br>sleeve | Yes-has protective<br>sleeve |
| Catheter Material | PVC ( phthalate free) | PVC ( phthalate free) | PVC (phthalate free) |
| Lubricant | PVP Based<br>(polyvinylpyrollidone)<br>Coating | PVP Based<br>(polyvinylpyrollidone)<br>Coating | PVP Based<br>(polyvinylpyrollidone)<br>Coating |
| Protective Tip | Yes | Yes | Yes |
| Collection Bag | No | No | No |
The following table outlines the sizes/styles of the VaPro intermittent catheter product line:
| Device | Device Type | Size<br>(Fr) | Length<br>(in) | Type<br>Straight/Nelaton or<br>Curved/Coude |
|--------|-------------|--------------|----------------|---------------------------------------------|
| VaPro™ | Predicate | 6 | 8 | Straight |
| VaPro | Predicate | 8 | 8 | Straight |
| VaPro | Predicate | 10 | 8 | Straight |
| VaPro | Predicate | 12 | 8 | Straight |
| VaPro | Predicate | 14 | 8 | Straight |
| VaPro | Predicate | 6 | 16 | Straight |
| VaPro | Predicate | 8 | 16 | Straight |
| VaPro | Predicate | 10 | 16 | Straight |
| VaPro | Predicate | 12 | 16 | Straight |
| VaPro | Predicate | 14 | 16 | Straight |
| VaPro | Predicate | 16 | 16 | Straight |
{5}------------------------------------------------
| Device | Device Type | Size<br>(Fr) | Length<br>(in) | Type<br>Straight/Nelaton or<br>Curved/Coude |
|--------------|----------------|--------------|----------------|---------------------------------------------|
| VaPro | Predicate | 18 | 16 | Straight |
| VaPro | Predicate | 12 | 16 | Curved |
| VaPro | Predicate | 14 | 16 | Curved |
| VaPro | Predicate | 16 | 16 | Curved |
| VaPro Plus | Predicate | 12 | 8 | Straight |
| VaPro Plus | Predicate | 14 | 8 | Straight |
| VaPro Plus | Predicate | 12 | 16 | Straight |
| VaPro Plus | Predicate | 14 | 16 | Straight |
| VaPro Pocket | Modified, 20cm | 10 | 8 | Straight |
| VaPro Pocket | Modified, 20cm | 12 | 8 | Straight |
| VaPro Pocket | Modified, 20cm | 14 | 8 | Straight |
| VaPro Pocket | Modified, 40cm | 10 | 16 | Straight |
| VaPro Pocket | Modified, 40cm | 12 | 16 | Straight |
| VaPro Pocket | Modified, 40cm | 14 | 16 | Straight |
| VaPro Pocket | Modified, 40cm | 16 | 16 | Straight |
Performance Testing:
There have been no changes to the device, only to the packaging. Therefore, no new biocompatibility testing was performed. For biocompatibility testing on the device in accordance with ISO 10993, please refer to 510(k) K141642.
Biocompatibility testing was performed on the new packaging. This report is available upon request. All results were satisfactory.
Sterility testing was performed using Gamma Irradiation in accordance with Method 1 validation per ANSI/AAMI/ISO 11137-1:2006 & -2:2013. All results were satisfactory.
Package integrity testing was performed to verify the maintenance of the sterile barrier through shelf life. The testing concluded that VaPro Pocket packaging is capable of maintaining a sterile barrier for at least two years.
Transportation testing was performed in order to verify that there is no impact to the safety or efficacy of the catheter performance. The test produced successful results.
Conclusion: Based on the performance testing conducted it is concluded that the modified device, VaPro Pocket, is as safe and effective and performs equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.