The MD 420-490 series reverse osmosis system is intended to purify water for hemodialysis when used as a component of a complete water purification system.
Device Story
MD 420-490 series is a multi-patient reverse osmosis (RO) water purification system for hemodialysis. It replaces the MD 400 A-F series. The device utilizes a Programmable Logic Controller (PLC) and touch-screen interface to monitor water room parameters, including pre-treatment, storage tank levels, line/loop pressures, temperatures, and flow rates. It calculates and displays membrane percentage rejection and recovery. Data is logged locally or transmitted via modem to a remote location for monitoring services. The system uses SCH 80 PVC and 316L stainless steel piping. It is operated by clinic staff to ensure water quality for dialysis. The PLC-based automation improves maintenance, reduces noise, and provides centralized data management compared to the mechanical-only predecessor.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Reverse osmosis water purification system. Materials: SCH 80 PVC and 316L stainless steel piping. Control: Programmable Logic Controller (PLC) with touch-screen interface. Connectivity: Modem for remote data transmission. Monitoring: Conductivity controller (Myron L), pressure, temperature, and flow sensors. Automated calculation of membrane rejection and recovery percentages.
Indications for Use
Indicated for patients requiring hemodialysis; intended to purify water for use as a component of a complete water purification system.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Multi-Patient Reverse Osmosis (RO) Water Treatment System MD 400 Series (K944385)
Submission Summary (Full Text)
{0}------------------------------------------------
JAN 2 4 2000
Image /page/0/Picture/1 description: The image shows the word "isopure" in a bold, sans-serif font. Below the word, there is a graphic of a series of wave-like shapes. The word and the graphic are both in black, contrasting with the white background. The overall design is simple and eye-catching.
## 1993200
Ph. (502) 267-(800)-280- PURF Ph. 8x. (502) 267 11700 Commonwealth Drive, I ouisville. Kv. 40290
## 510 (K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is K993200.
- 1) Submitter's Identification: Isopure Corporation 11700 Commonwealth Drive Suite 605 Louisville, Kentucky 40299
Date Summary Prepared: September 20, 1999
- 2) Name of The Device: MD 420/440/460/470/480/490 Multi- Patient Reverse Osmosis (RO) Machine.
- 3) Predicate Device Information: Multi-Patient Reverse Osmosis (RO) Water Treatment System MD 400 Series.
- 4) Device Description: The MD 420-490 Series is a Reverse Osmosis Machine used in purifying water.
- 5) Intended Use: The MD 420-490 Series is intended to purify water for Hemodialysis when used as a component of a complete water purification system.
- Comparison to Predicated Device: () The MD 420-490 series features an advanced electronic package including a PLC with software and validation procedures and a Touch-screen. The new advanced system has the capability to monitor both the pre-treatment and posttreatment of the water room. The data can be printed for a daily log or modem to a remote location.
Manufacturers of Advanced Water Purification and Reclamation Technologies
{1}------------------------------------------------
- Discussion of Non-Clinical Test Performed for Determination of Substantial 7) Equivalence are as follows: Based on the information submitted, there are no changes in the safety of effective issucs.
- Discussion of Clinical Test Performed: 8) N/A
- 9) Conclusions:
As was true for the original registered HD400A-F Reverse Osmosis (RO) nachine device, the intended use of the modified RO machine device, i.e. the MD 420-490 series, is intended to purify water for Hemodialysis when used as a component of a complete water purification system.
{2}------------------------------------------------
## EXHIBIT A:
## SECTION 1: DEVICE MODIFICATION DESCRIPTION
1.1 Intended Use: As was true for the original registered HD400A--F Reverse Osmosis (RO) Machine device (510(k) application #K944385), the intended use of the modified RO machine device, i.e., the MD420-490 series, is intended to purify water for Hemodialysis when used as a component of a complete water purification system. The device name change from MD 400 A-F to MD 420-490 was to facilitate recognition of a medical device (MD) 400 series as multi-patent and 20-90 as the total machine membranes 420 (2 membranes) 440 (4 membranes) etc. The text of the original 510(k), though of limited explicit information characteristic of its time of submission, is included for reference (See Exhibit E: Section 10).
1.2 Rationale: The rationale for modifying the original registered device is straightforward. First, the original design, although very useful at the time, now proves to be very nosey and cluttered with pipes, valves, etc. This made the machine very hard to operate and maintenance proved very costly. Second, the horizontal position of the Goulds pump resulted in complete pump failures costing IPS large sums of money in pump failures resulting in the sale of the company in 1998 to Isopure Corp. Third, with the FDA now requiring a complete registration of the entire water system, this machine with the use of a PLC, will record data for all aspects of the water system tying the pre-treatment and storage tank back to the RO for one central location for all of the data. This information is also modemable to a remote location. Fourth, the machine is now simpler, quieter, and easier to maintain, run, and service compared to the MD 400 A-F.
1.3 Background: It was, and remains, the opinion of the applicant that the change from a mechanical device entrained changes attendant to a automated device, were subject to documentation for the models of the original registered device series (MD400A-F, now MD 420--490). The PLC change and entrained changes are listed later in this exhibit (See Exhibit A: Section 1.9, "Table of Differences"). Marketing of these devices continues. In addition, long before changing to the new PLC, MD 400D-- E models were constructed and sold using the same original circuitry to meet the growing purified water volume demands of the dialysis market. The MD400 D-- E devices, now the MD 420-490 also underwent alteration and are equipped with PLC circuitry. Again, it was felt that the PLC could offer a much broader control of the entire system per the FDA requirements. The original conductivity control a Myron L conductivity controller used to monitor the conductivity of the water and also control the " Diversion to Drain" of the machine remains the same as on the original registration. The PLC can monitor all aspects of the water room including the RO, pre-treatment, storage tank, line and loop pressures, temperatures, and flows. The system will display membrane % rejection and % recovery. All of this data is sent back to the RO where it is displayed. The data also has the capability to be modemed to a remote location. Isopure would also like to offer a "Remote Monitoring Service" where we would call up the machine from our location, review the data of the water room and fax or e-mail the results back to the clinic for their reports. The hydraulic design of the machine remains unchanged for the original 510(k) application, however, we now use SCH 80 PVC pipe and 316L Stainless Steel piping in place of the tubing originally used of the HD- MD series.
lsopure
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
24 2000 JAN
Mr. Kevin Gillespie President and CEO Isopure Corporation 11700 Commonwealth Drive Louisville, KY 40299
Re: K993200 MD 420/440/460/470/480/490 Reverse Osmosis Machine Dated: October 24, 1999 Received: October 28, 1999 Regulatory Class: Il 21 CFR §876.5665/Procode: 78 FIP
Dear Mr. Gillespie:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
:
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regult in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K993200
Device Name: MD 420/440/460/470/480/490 Multi-patient Reverse Osmosis (RO) Machine
Indications For Use:
The MD 420-490 series reverse osmosis system is intended to purify water for hemodialysis when used as a component of a complete water purification system.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign-Off
(Division Sign Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices K993200 510(k) Number _
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.