UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorganic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing. This System is strictly for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System and is intended for use only with conventional hemodialysis treatments. It is not intended for use with high-flux, CRRT, or any other form of dialysis.
Device Story
Reverse osmosis (RO) water purification system; removes organic/inorganic contaminants and microbial matter from tap water to produce AAMI-compliant water for hemodialysis. Used in clinics/hospitals for multi-patient settings. System includes electronic touch screen interface for monitoring feed/product water pressure, flow, conductivity, and pump pressure. Safety features include automatic diversion to drain if product conductivity exceeds set-points, alarms for low supply pressure, high conductivity, high temperature, and power disturbances. Healthcare providers use output for dialysate preparation and equipment maintenance. Benefits include consistent water quality meeting ANSI/AAMI/ISO 13959:2014 standards, ensuring safe dialysate production for conventional hemodialysis.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and compliance with AAMI/ISO standards for water quality and electrical safety.
Technological Characteristics
Reverse osmosis water purification system; FDA-NSF compliant water contact materials; electronic touch screen interface; wired remote alarm connectivity; no wireless/RF transmission. Electrical components comply with EN/IEC 61000-4-2, 4-3, 4-4, 4-8; UL/CSA recognized pumps. Produces water per ANSI/AAMI/ISO 13959:2014 and ANSI/AAMI/ISO 26722:2014.
Indications for Use
Indicated for multi-patient use in dialysis clinics or hospitals for conventional hemodialysis treatment, dialyzer reprocessing, and dialysis equipment disinfection/rinsing. Not indicated for high-flux, CRRT, or other dialysis forms.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Complete Water System Models MD610, 620, 630 and 640 (K041163)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a human figure with three heads or faces, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 23, 2015
Specialty Water Technologies Alex Clark Compliance Manager 1020 Industrial Drive Orlinda, TN 37141
Re: K151637
> Trade/Device Name: UPT Series Medical Reverse Osmosis Water Treatment System Regulation Number: 21 CFR 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: August 21, 2015 Received: August 24, 2015
Dear Alex Clark,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K151637
Device Name
UPT Series Medical Reverse Osmosis Water Treatment System
Indications for Use (Describe)
UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorganic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing.
This System is strictly for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System and is intended for use only with conventional hemodialysis treatments. It is not intended for use with high-flux, CRRT, or any other form of dialysis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Specialty Water Technologies (SWT). The letters SWT are in a blue, sans-serif font at the top of the image. Below the letters is a graphic of a water droplet falling onto a blue and green surface, creating ripples. The words "SPECIALTY WATER TECHNOLOGIES" are written in a smaller, sans-serif font below the graphic.
# 510(k) Summary
#### 510(k) Number: _______________________________________________________________________________________________________________________________________________________________
| Manufacturer: | Specialty Water Technologies |
|----------------------|--------------------------------------------------------------|
| Address: | 1020 Industrial Drive<br>Orlinda, TN 37141<br>(615) 654-4441 |
| Official Contact: | Alex Clark<br>Compliance Manager |
| Trade Name: | UPT Series Medical Reverse Osmosis Water Treatment<br>System |
| Common Name: | Water Purification System |
| Classification Name: | Water Purification System for Hemodialysis |
| Product Code: | 78 FIP |
| Device Class: | II |
| Classification Reg: | 876.5665 |
Specialty Water Technologies has provided the following information to the US Food and Drug Administration to support substantial equivalence of the UPT Series Medical Reverse Osmosis (RO) Water Purification System to other RO water purification systems currently cleared for sale in the U.S.
#### 1. Device Description
The UPT Series Medical Reverse Osmosis Water Treatment System is an accessory device that is intended for use with hemodialysis applications and is intended to remove orqanic and inorganic substances and microbial contaminates from tap water used to dilute dialysis concentrate to form dialysate. The UPT Series Medical Reverse Osmosis Water Treatment System will produce water as prescribed by the Association for the Advancement of Medical Instrumentation (AAMI) Standard ANSI/AAMI/ISO 13959:2014 Water for Hemodialysis and Related Therapies provided adequate flow rate of the feed (tap) water and compliance with existing drinking water standards and proper pre-treatment.
#### 2. Intended Use
UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorqanic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing.
{4}------------------------------------------------
This System is strictly for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System and is intended for use only with conventional hemodialysis treatments. It is not intended for use with high-flux, CRRT, or any other form of dialysis.
| | Specifications | Predicate Devices | |
|-------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|
| Specifications | UPT Series Medical Reverse<br>Osmosis Water Treatment System | Matrix Series Digital Reverse<br>Osmosis | Complete Water System Models<br>MD610, 620, 630 and 640 |
| Manufacturer | Specialty Water Technologies | Better Water, Inc | Isopure Corporation |
| 510(k) Number | | K051620 | K041163 |
| Decision Date | | 11/18/2005 | 10/7/2004 |
| Classification | 876.5665, Class II | 876.5665, Class II | 876.5665, Class II |
| Product Code | 78 FIP | 78 FIP | 78 FIP |
| Water Contact<br>Materials | FDA-NSF Compliant | FDA-NSF Compliant | FDA-NSF Compliant |
| Pre-Treatment<br>Options | Based upon site specific feed<br>water chemical analysis | Based upon site specific feed water<br>chemical analysis | Based upon site specific feed water<br>chemical analysis |
| | Feed water Pre-Treatment may<br>include: | Feed water Pre-Treatment may<br>include: | Feed water Pre-Treatment may<br>include: |
| | Carbon Filtration | Carbon Filtration | Carbon Filtration |
| | Softeners | Softeners | Softeners |
| | Temperature Control | Temperature Control | Temperature Control |
| | Pressure Compensation | Pressure Compensation | Pressure Compensation |
| | Special Filtration | Special Filtration | Special Filtration |
| | Ultra-Violet (UV) | Ultra-Violet (UV) | Ultra-Violet (UV) |
| Operational<br>Features | Electronic Touch Screen Interface | Electronic Touch Screen Interface | Electronic Touch Screen Interface |
| | All components mounded on an<br>open frame for accessibility | All components mounted on an<br>open frame for accessibility | All components mounted on an open<br>frame for accessibility |
| | <b>Electronic Monitoring of:</b> | <b>Electronic Monitoring of:</b> | <b>Electronic Monitoring of:</b> |
| | Feed Water Pressure / Flow | Feed Water Pressure / Flow | Feed Water Pressure |
| | Product Water Pressure / Flow | Product Water Pressure / Flow | Product Water Pressure |
| | Membrane Feed Pressure | Membrane Feed Pressure | Membrane Feed Pressure |
| | Pump Pressure | Pump Pressure | Pump Pressure |
| | Reject Water Pressure / Flow | Reject Water Pressure / Flow | Reject Water Pressure / Flow |
| | Product Water Conductivity | Product Water Conductivity | Product Water Conductivity |
| | Feed Water Conductivity | Feed Water Conductivity | Feed Water Conductivity |
| Safety Features | 100% Diversion to drain above<br>product conductivity set-point | 100% Diversion to drain above<br>product conductivity set-point | 100% Diversion to drain above<br>product conductivity set-point |
| | Low Supply Pressure Alarm | Low Supply Pressure Alarm | Low Supply Pressure Alarm |
| | High Product Water Conductivity<br>Alarm | High Product Water Conductivity<br>Alarm | High Product Water Conductivity Shut<br>Down |
| | High Supply Water Temperature<br>Shut Down | High Supply Water Temperature<br>Shut Down | High Supply Water Temperature Shut<br>Down |
| | Power Disturbance Alarm | Power Disturbance Alarm | Power Disturbance Alarm |
| | Remote Alarm (Nurses Station) | Remote Alarm (Nurses Station) | Remote Alarm (Nurses Station) |
| | May include 0.5 micron or better<br>ultra-filtration | May include 0.5 micron or better<br>ultra-filtration | May include 0.5 micron or better<br>ultra-filtration |
| Post-Treatment<br>Performance | Delivery of AAMI Standard Water | Delivery of AAMI Standard Water | Delivery of AAMI Standard Water |
#### 3. Substantial Equivalence to Predicate Device
{5}------------------------------------------------
#### 4. Safety and Effectiveness
The intended use and technological characteristics of the UPT Series Medical Reverse Osmosis Water Treatment System are similar or equivilant to the predicate devices indicated. Any differences between the device and the predicate devices have no significant influence on the safety or effectiveness of the product.
#### 5. Clinical Performance Data
Not required for determination of substantial equivilance for this type and class of device.
#### 6. Electromagnetic Compatibility & Safety
The water system does not contain any wifi, wireless or other transmitting or receiving devices. Any electrical connections between components are wired to include any remote alarms.
The electrical control components purchased and used in the construction of the components used for this water system comply with:
EN/IEC 61000-4-2 Electrostatic Discharges EN/IEC 61000-4-3 Radiated Fields and Transients / Burst EN/IEC 61000-4-4 Electrical Fast Transients / Burst EN/IEC 61000-4-8 Maqnetic Field UL Approved / Recognized for safety
All pumps used are UL recognized (UR) and CSA Approved for safety
## 7. Conclusion Drawn from Clinical and Nonclinical Test Data
As compared with the predicate devices, K051620 and K041663, the UPT Series Medical Reverse Osmosis Water Treatment System will produce water that will meet the minimum water quality requirements as specified by Association for the Advancement of Medical Instrumentation (AAMI) standard ANSI/AAMI/ISO 13959:2014, Water for Hemodialysis and Related Therapies; and ANSI/AAMI/ISO 26722:2014, Water Treatment Equipment for Hemodialysis Applications and Related Therapies, when used as directed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.