K990854 · Innerdyne, Inc. · FGE · Aug 2, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K990854
Device Name
INNERDYNE RADIALLY EXPANDING DILATION, RED DEVICE
Applicant
Innerdyne, Inc.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Aug 2, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The InnerDyne, Inc., RED® device is intended for use during minimally invasive surgery to provide temporary dilation access to the abdominal/peritoneal cavity and to the lumen of specified internal organs to establish a port of entry for diagnostic, therapeutic, and operative instruments. Typical procedures for product use include: - Laparoscopic access to the abdominal/peritoneal cavity; - Percutaneous gastrostomy; - Percutaneous enterostomy; - Percutaneous cystostomy; - Percutaneous cholescystostomy; and - Dilation of biliary and urethral strictures.
Device Story
Expandable dilator sheath assembly; includes radially expanding sleeve, inflation syringe, cannula with valve, dilator, and insertion instrument (needle/guidewire). Operation: needle/sleeve inserted into target cavity; needle removed; dilator cannula inserted through sleeve lumen, causing radial expansion of sleeve and surrounding tissue; dilator removed, leaving cannula as sealed working port for diagnostic/therapeutic instruments. Used in minimally invasive surgery; operated by physicians. Benefits: allows initial small-diameter access followed by controlled radial dilation to desired working channel size; reduces trauma compared to non-radial dilation methods.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Medical grade materials; components molded/machined and secured via adhesives, welds, and mechanical interlocks. System includes radially expanding sleeve, inflation syringe, cannula, and dilator. Disposable. No software or electronic components.
Indications for Use
Indicated for patients undergoing minimally invasive surgery requiring temporary dilation access to abdominal/peritoneal/thoracic cavities or hollow organs for instrument passage. Includes laparoscopic access, percutaneous gastrostomy, enterostomy, cystostomy, cholescystostomy, and dilation of biliary/urethral strictures.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
InnerDyne, Inc., R.E.D.® device
InnerDyne, Inc., Step® device
Submission Summary (Full Text)
{0}------------------------------------------------
## SUMMARY OF SAFETY & EFFECTIVENESS INFORMATION
Submitter: InnerDyne, Inc. 5060 West Amelia Earhart Drive Salt Lake City, Utah 84116 (801) 350-3600
Date Prepared: 12 March 1999
- Rick Gaykowski Contact: Corporate Vice President, Regulatory Affairs and Quality Assurance
Classification Name: Dilator (Other) Common/Usual Name: Percutaneous Dilator w/ Sheath Trade/Proprietary Name: InnerDyne , Radially Expanding Dilation "RED® " Device
The RED® device is an expandable dilator sheath assembly, consisting of a radially expanding sleeve, inflation syringe, cannula (with integral valve system), dilator, and insertion instrument (either an access needle or guidewire). The tubular member of the expandable sleeve is configured so as to be radially compressed to reduce the outside diameter of the device prior to insertion. Upon use, the dilator assembly is inserted through the radially expandable sleeve, penetrating the tissue into the target cavity (e.g., abdominal/peritoneal) or hollow organ.
For access to target cavities and hollow organs, the following instructions should be utilized. The needle and radially expanding sleeve should first be inserted into the target location. The needle removed, leaving the radially expanding sleeve in place. The dilator cannula assembly, which consists of a blunt dilation cannula, is then inserted through the lumen of the expandable sleeve which expands radially to accommodate it. This process in turn, radially expands the walls of the surrounding tissue. Following dilation the dilator is removed, leaving the radially expanding dilation sleeve and dilator cannula in place. This technique allows the user to benefit from the unique safety features and controllability of radial dilation by going from a less traumatic initial stick, followed by radial dilation to the desire working channel.
Fill the provided syringe with 2.0 cc of liguid (A 50% concentration is recommended if using contrast). Attach the syringe to the reflux valve and draw a small vacuum to aspirate air. Tilt the syringe so that the air bubble floats away from the reflux valve. Inflate the balloon and remove the syringe. Once the working channel is established, the dilator be removed from the cannula leaving the working cannula in place to provide a sealed port for passage of diagnostic, therapeutic, and/or surgical instruments.
This system configuration allows the user to initially place a small diameter dilator cannula for
{1}------------------------------------------------
K99 0854 pg. 2 of 2
passage of small diagnostic instruments. The dilator cannula can then be removed from the lumen of the radially expanding sleeve while leaving the sleeve inserted through the tissues. A larger diameter dilation cannula can then be inserted through the expanding sleeve to create a larger port for passage of the larger operative instruments.
The device is assembled from medical grade materials under GMP conditions. Components are molded and machined by qualified suppliers. The components are assembled and secured by adhesives, welds, and mechanical interlocks. The RED® device is available in various lengths and working diameters to accommodate the selected size of indicated medical instruments. The subject product may be supplied in both disposable forms.
The subject InnerDyne, Inc., RED® device is substantially equivalent to the predicate InnerDyne, Inc., R.E.D.® &, Step® device versions in basic design, product configuration, composition, utilized materials, function, deployment, warnings and precautions, contraindications, intended use for access to abdominal/thoracic cavities, and for access to hollow body organs.
The subject RED® device is intended for use during minimally invasive surgery for temporary dilation access to the abdominal/peritoneal and thoracic cavities for passage of diagnostic, therapeutic and operative instruments into the abdominal and thoracic cavities, and for percutaneous access to hollow body organs. The device is configured to be used as either a primary or secondary stick.
From the foregoing, we conclude that the subject RED® device is as safe and effective as named predicates and currently marketed competitive devices for the stated indications. -
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its head, representing the department's services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.
AUG - 2 1999
Mr. Rick Gaykowski Corporate Vice President, Regulatory Affairs and Quality Assurance InnerDyne, Inc. 5060 West Amelia Earhart Drive Salt Lake City, UT 84116
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K990854 InnerDyne, Inc., Radially Expanding Dilation, RED®, Device Dated: June 14, 1999 Received: June 16, 1999 Regulatory Class: Il 21 CFR §876.5010/Procode: 78 FGE 21 CFR §876.5090/Procode: 78 FFA 21 CFR §876.5980/Procode: 78 KGC 21 CFR §876.1500/Procode: 78 GCJ
Dear Mr. Gaykowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic cievices), piease contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Page 1 of 1
## 510(k) Number (if known):
Device Name: InnerDyne, Inc., Radially Expanding Dilation, RED®, Device.
Indications for Use: The InnerDyne, Inc., RED® device is intended for use during minimally invasive surgery to provide temporary dilation access to the abdominal/peritoneal cavity and to the lumen of specified internal organs to establish a port of entry for diagnostic, therapeutic, and operative instruments. Typical procedures for product use include:
- Laparoscopic access to the abdominal/peritoneal cavity; .
- Percutaneous gastrostomy; .
- Percutaneous enterostomy; ●
- Percutaneous cystostomy; .
- . Percutaneous cholescystostomy; and
- . Dilation of biliary and urethral strictures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Thmil La. Lagnm
Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.