The BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.
Device Story
Percutaneous drainage catheter set used for abscess, cyst, gall bladder, nephrostomy, and urinary fluid drainage. Device utilizes Seldinger technique for access. Catheter features soft, biocompatible TPU material, radiopaque markers, and hydrophilic coating; distal end includes pigtail or closed-loop configuration with drainage holes. Operated by clinicians in clinical settings. Output is fluid drainage; healthcare provider monitors drainage to manage patient fluid accumulation. Benefits include minimally invasive fluid removal.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including sterilization validation, shelf-life, biocompatibility (ISO 10993-1), and mechanical performance testing (EN 1617, ISO 20697, ASTM F640, EN 1618, ISO 594-2, ISO 80369-7, ISO 11070, ASTM F1828, ISO 7864).
Technological Characteristics
Material: TPU. Distal configuration: Pigtail/closed-loop. Features: Hydrophilic coating, radiopaque markers, shaft depth printing. Sizes: 8F-14F. Sterilization: Ethylene Oxide. Standards: ISO 11135, ISO 10993-7, ISO 11737, USP <85>, ASTM F1140, ASTM F1980, ASTM F1929, EN 1617, ISO 20697, ASTM F640, EN 1618, ISO 594-2, ISO 80369-7, ISO 11070, ASTM F1828, ISO 7864.
Indications for Use
Indicated for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and other fluids in adults; pediatric use determined by clinician.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
BIOTEQ Drainage Catheter Set (One Step Type) (K200103)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern. To the right of the seal, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" below. The logo is simple and professional, reflecting the FDA's role in regulating food and drugs.
April 13, 2022
Bioteque Corporation Stella Hsu RA Specialist 5F-6, No. 23, Sec. 1, Chang'an E. Rd., Zhongshan Dist. Taipei City, 104 Taiwan
Re: K210419 Trade/Device Name: BIOTEO Drainage Catheter Set (Seldinger Type) Model Name: BT-PDS-series Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: II Product Code: FGE, LJE, GBO Dated: March 15, 2022 Received: March 16, 2022
Dear Stella Hsu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Je Hi An. Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210419
### Device Name
BIOTEQ Drainage Catheter Set (Seldinger Type) Model Name: BT-PDS-series
Indications for Use (Describe)
The BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------|
| <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k), K210419 Summary
# 510(k) Summary
- Traditional 5.1 Type of Submission: 5.2 Date of Summary: March 31, 2022 BIOTEQUE CORPORATION 5.3 Submitter: Address: 5F-6, No. 23, Sec. 1, Chang'an E. Rd., Zhongshan Dist. Taipei City 10441, Taiwan Phone: +886-2-2571-0269 Fax: +886-2-2536-1967 Contact: Stella Hsu
#### 5.4 Identificationof the Device:
| Proprietary/T de name: | BIOTEQ Drainage Catheter Set (Seldinger Type)<br>Model Name: BT-PDS-series |
|------------------------------|----------------------------------------------------------------------------|
| Classification Product Code: | FGE |
| Subsequent Product Code: | LJE, GBO |
| Regulation Number: | 876.5010 |
| Regulation Description: | Biliary catheter and accessories; |
| Review Panel: | Gastroenterology/Urology; General & Plastic Surgery |
| Device Class: | II |
| Basis for the Submission: | New Device |
#### ર્સ્ડ Identificationof the Predicate Device:
| Predicate Device Name: | BIOTEQ Drainage Catheter Set (One Step Type) | |
|------------------------------|----------------------------------------------|--|
| | Model Name: BT-PD1-series | |
| Manufacturer | BIOTEQUE CORPORATION | |
| Classification Product Code: | FGE | |
| Subsequent Product Code: | LJE, GBO | |
| Regulation<br>number | 876.5010 | |
| Device Class: | II | |
| 510(k) Number: | K200103 | |
{4}------------------------------------------------
Traditional 510(k), K210419 Summary
#### Device Descrintion ર્સ્વ
The BIOTEO Drainage Catheter Set (Seldinger Type), BT-PDS-series percutaneous drainage catheter with hydrophilic coating, are percutaneous drainage catheters used for drainage of abscess and fluid collections. The catheter is made from a soft, biocompatible plastic, a material that is radiopaque for X-rays. The distal end of catheter contains a pigtail or close loop and drainage holes.
The operator can use different drainage sets according to the type of accumulated fluid and place of accumulation. These sets are classified according to the catheter size (8F~14F pertaining to various fluid viscosity) and according to the position of the accumulation. The operator should use Seldinger Technique to provide access.
#### 5.7 Indications for Use
The BIOTEQ Drainage Catheter Set (Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids.
#### Comparison of Technological Characteristics with Predicate Device 5.8
Equivalence, same and difference between the subject and predicate devices are cited as below.
| Item | Subject device | Predicate device | Substantial equivalence<br>determination |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Proprietary Name | BIOTEQ Drainage Catheter Set<br>(Seldinger Type) | BIOTEQ Drainage Catheter Set<br>(One Step Type) | Substantial equivalence<br>determination |
| 510(k) No. | K210419 | K200103 | |
| Intended Use | The BIOTEQ Drainage Catheter Set<br>(Seldinger Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders nephrostomy, urinary and other fluids. | The BIOTEQ Drainage Catheter Set<br>(One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids. | Equivalent<br>Both the devices are used for percutaneous drainage in Gastroenterology and urology. |
| Type of Use | Prescription Use | Prescription Use | Same |
| Intended User | Adults and recommending clinician should choose an appropriate catheter size for pediatric use. | Adults and recommending clinician should choose an appropriate catheter size for pediatric use. | Same |
| Item | Subject device | Predicate device | Substantial equivalence determination |
| Proprietary Name | BIOTEQ Drainage Catheter Set<br>(Seldinger Type) | BIOTEQ Drainage Catheter Set<br>(One Step Type) | Substantial equivalence<br>determination |
| 510(k) No. | K210419 | K200103 | |
| Catheter Shaft<br>Material | TPU | TPU | Same |
| Distal Configuration | String Locking Pigtail,<br>Non-String Locking Pigtail | String Locking Pigtail,<br>Non-String Locking Pigtail | Same |
| Distal Shape | Pigtail, Closed-Pigtail | Pigtail, Closed-Pigtail, Mini-Pigtail,<br>Mini-closed Pigtail, J shape | Equivalent<br>Both the devices have<br>pigtail and closed shape. |
| Distal Hydrophilic<br>Coating | Yes | Yes | Same |
| Shaft Depth Printing<br>Markers | Yes | Yes | Same |
| Proximal Hub<br>Assembly | Hub (for String Lock Pigtail),<br>F.L.L. Adapter | Hub (for String Lock Pigtail),<br>F.L.L. Adapter | Same |
| Size | 8, 10, 12, 14 Fr | 5 Fr (Non-String Lock), 6, 7, 8, 10,<br>12, 14, 16 Fr | Equivalent<br>The size range of subject<br>device is equivalent to<br>the predicate device. |
| Useable Length | 40, 45, 50 cm | 20, 25, 30, 35, 40, 45, 50 cm | Equivalent<br>The length range of<br>subject device is<br>equivalent to the<br>predicate device. |
| Included Insert<br>Accessory | • Metal Stiffening Cannula<br>• Flexible (plastic) Stiffening<br>Cannula<br>• Wire cap<br>• Suture Wire<br>• Curve Straightener | • Trocar Needle<br>• Trocar Stylet<br>• Flexible (plastic) Stiffening<br>Cannula<br>• Wire cap<br>• Suture Wire | Equivalent<br>The accessories of<br>subject device equivalent<br>to the predicate device. |
| Item | Subject device | Predicate device | Substantial equivalence determination |
| Proprietary Name | BIOTEQ Drainage Catheter Set<br>(Seldinger Type) | BIOTEQ Drainage Catheter Set<br>(One Step Type) | |
| 510(k) No. | K210419 | K200103 | |
| | • Radiopaque band | • Curve Straightener<br>• Radiopaque band | |
| Packaging | Tyvek/Mylar (PET/LDPE) pouch | Tyvek/Mylar (PET/LDPE) pouch | <i>Same</i> |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | <i>Same</i> |
{5}------------------------------------------------
### Traditional 510(k), K210419 Summary
{6}------------------------------------------------
Traditional 510(k), K210419 Summary
#### રું. જેવી સાથે તે જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં લોકોનો મુખ્ય વ્યવસાય ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુખ્યત્વે ખેતી, ખેતમજૂરી તેમ જ પશુપાલન છે. આ ગામનાં મુ Similarity and Difference
The BIOTEQ Drainage Catheter Set (Seldinger Type) has been compared with "BIOTEQ Drainage Catheter Set (One Step Type)". The subject device has the same intended use, and similar principle of operation and technological characteristics as the predicate device. The subject device has undergone safety and performance tests, and the results complied with the test requests. Although there are some different specifications between these devices, the performance test has been completed to demonstrate that the differences between these parameters do not raise different types of safety and effectiveness questions. The subject device is substantially equivalent to the predicate device in intended use, design and performance claims.
# 5.10 Performance Data - Non-clinical Testing
The following performance data were provided in support of the substantial equivalence determination.
- Sterilization Validation
The ethylene oxide (EO) sterilization and related validation testing were conducted in accordance and complied with ISO 11135, ISO 10993-7, ISO 11737-1, ISO 11737-2, USP <85>, and ASTM F1140/F1140M.
- . Shelf-life
The shelf-life testing were conducted in accordance and complied with ASTM F1980, ASTM F1929, and ASTM F1140/F1140M.
- . Biocompatibility
The biocompatibility evaluation was conducted in accordance and complied with the FDA
{7}------------------------------------------------
Traditional 510(k), K210419 Summary
Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process''' and International Standard ISO 10993-1:2018. Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process as recognized by FDA. The endpoint testing included the following tests:
- -Cytotoxicity
- -Sensitization
- -Irritation
- Acute Systemic Toxicity -
- Pyrogenicity -
- Subchronic Systemic Toxicity -
- Genotoxicity -
- -Implantation
- Performance
The performance testing were conducted in accordance and complied with EN 1617, ISO 20697, ASTM F640, EN 1618, ISO 594-2, ISO 80369-7, ISO 11070, ASTM F1828 and ISO 7864.
All the test results demonstrate BIOTEQ Drainage Catheter Set (Seldinger Type) meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate device.
# 5.11 Performance Data - Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
# 5.12 Conclusion
The data demonstrate the subject device is as safe and effective as the primary predicate device. The data support the safety and performance of the subject device, and demonstrate that the subject device should perform as intended in the specified use conditions. The data demonstrates that the BIOTEQ Drainage Catheter Set (Seldinger Type) performs comparably to the predicate device that is currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.