Two-way and Three-way Disposable Silicone Foley Catheters are indicated for urethral catheterization for bladder drainage for urological use only. The Two-way and Three-way Disposable Silicone Foley Catheters have an indwell time of no more than 30 days and are indicated for adult use only.
Device Story
Foley balloon catheter for urethral catheterization and bladder drainage; three-way configuration includes irrigation manifold for saline bladder washing. Device consists of silicone/zinc oxide shaft, silicone manifold, and balloon. Input: manual insertion by clinician; balloon inflation via distilled water or saline injection through non-return valve. Output: continuous bladder drainage/irrigation. Features x-ray opaque line for positioning. Used in clinical settings; prescription-only. Benefits: facilitates bladder drainage/irrigation in adults; indwell time up to 30 days.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993), sterilization validation (ISO 11135:2014), shelf-life/package integrity, and functional performance testing per ASTM F623-99.
Indicated for adult male and female patients requiring urethral catheterization for bladder drainage and/or bladder irrigation. Contraindicated for pediatric use.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 30, 2020
Apollon Co., Ltd. % Hvevoung Moon Assistant Manager TSD Life Sciences Co., Ltd. 211, Mallijae-ro, Jung-gu Seoul. 04508 Korea
Re: K201776
Trade/Device Name: Two-way and Three-way Disposable Silicone Foley Catheters Regulation Number: 21 CFR& 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: November 24, 2020 Received: November 24, 2020
Dear Hyeyoung Moon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201776
#### Device Name
Two-way and Three-way Disposable Silicone Foley Catheters
Indications for Use (Describe)
Two-way and Three-way Disposable Silicone Foley Catheters are indicated for urethral catheterization for bladder drainage for urological use only. The Two-way Disposable Silicone Foley Catheters have an indwell time of no more than 30 days and are indicated for adult use only.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <table><tr><td><div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> <polyline fill="none" points="2 7, 6 12, 13 3" stroke="black" stroke-width="2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div></td></tr><tr><td><div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div></td></tr></table> | <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> <polyline fill="none" points="2 7, 6 12, 13 3" stroke="black" stroke-width="2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> <polyline fill="none" points="2 7, 6 12, 13 3" stroke="black" stroke-width="2"></polyline> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | | |
| <div> <span> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" stroke-width="2" width="15"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for a company called APOLLON Co., Ltd. The logo features a stylized letter "A" in a dark blue color, enclosed within two concentric circles. The circles have small, solid black circles attached to them on either side of the letter "A". The company name, "APOLLON Co., Ltd.", is printed in black text below the circular logo.
# 510(k) Summary - K201776
[21 CFR 807.92]
### 1. Submitter Information
| | Company Name | : APOLLON Co., Ltd. |
|----------------------------|--------------------------|--------------------------------------------------------------------------|
| | Company Address | : M-1804/3203, 32, Songdogwahak-ro, Yeonsu-gu, Incheon, Korea |
| | Company Phone | : +82-32-830-4724 |
| | Contact Person | : Donghyuk Shin/ General<br>Manager<br>dhshin@apollonmds.com |
| | Submission Correspondent | : Hyeyoung.moon/RA<br>Hyeyoung.moon@tsdls.co.kr<br>Tel: +82-70-4224-9790 |
| | Date Prepared | : December 28, 2020 |
| <b>2. Device Name</b> | | |
| | Trade Device Name | : Two-way and Three-way Disposable Silicone Foley Catheters |
| | Common Name | : Foley Catheter |
| | Classification Name | : Urological Catheter and Accessories |
| | Classification Number | : 876.5130 |
| | Regulatory Class | : II |
| | Product code(s) | : EZL |
| | Product code Name | : Catheter, Retention Type, Balloon |
| | Advisory Panel | : Gastroenterology/Urology |
| <b>3. Predicate Device</b> | | |
| | 510(k) Number | : K130908: |
| | Device Name | : Disposable Silicone Foley Catheter |
| | Product Code | : EZL |
| | Manufacturer | : Guangdong Baihe Medical Technology Co., Ltd. |
The predicate device is not the subject of a design related recall.
### 4. Device Description:
This product is a Foley Balloon Catheter. It is a 'prescription only' device. It consists of a Manifold part and Shaft parts. The Manifold Part is equipped with a non-return valve for inflating the balloon and a band to fix the non-return valve and identify the catheter's size by its color. The Shaft parts have an x-ray opaque line for checking the position of the catheter by X-ray and a balloon to place the catheter in the bladder. After the catheter is inserted, distilled water or saline is injected to expand the balloon and fix the catheter in place. The shaft also contains drainage eyes for urine drainage, and a rounded tip to aid insertion of the catheter. This Foley Balloon Catheter will be provided in Two and Three-Way configurations. The Three-Way configuration contains an 'Irrigation Manifold' for washing the bladder with saline solution. The shaft and
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for a company named APOLLON Co., Ltd. The logo features a stylized letter "A" in a dark blue color, enclosed within two concentric circles. The circles have small, circular extensions on either side, resembling electrical contacts. The company name "APOLLON Co., Ltd." is printed in a simple font below the circular logo.
## 510(k) Summary - K201776
[21 CFR 807.92]
balloon of the catheter are made of silicone and zinc oxide. The tip and the manifold are made of silicone only.
The proposed device is supplied in French sizes ranging from 12 to 24. It is available in 337(±15) mm and 345(±15) mm lengths with various balloon sizes. The devices for adults only.
### 5. Indication for use
Two-way and Three-way Disposable Silicone Foley Catheters are indicated for urethral catheterization for bladder drainage for urological use only. The Two-way and Three-way Disposable Silicone Foley Catheters have an indwell time of no more than 30 days and are indicated for adult use only.
### 6. Comparison of Technological Characteristics with Predicate Device
| Features | Subject Device | Predicated Device |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Two-way and Three-way Disposable<br>Silicone Foley Catheters | Disposable Silicone Foley Catheter |
| 510(k)<br>Number | K201776 | K130908 |
| Manufacturer | APOLLON Co., Ltd. | Guangdong Baihe Medical Technology Co., Ltd. |
| Indication for<br>use | Two-way and Three-way Disposable Silicone<br>Foley Catheters are indicated for urethral<br>catheterization for bladder drainage for<br>urological use only. The Two-way and Three-<br>way Disposable Silicone Foley Catheters have<br>an indwell time of no more than 30 days and are<br>indicated for adult use only. | Two-way Disposable Silicone Foley Catheter:<br>Urethral catheterization for bladder drainage for<br>urological use only; the indwell time of the<br>proposed device is no more than 30 days.<br>Three-way Disposable Silicone Foley Catheter:<br>Urethral catheterization for bladder drainage and<br>bladder irrigation for urological use only; the<br>indwell time of the proposed device is no more<br>than 30 days. |
| Population | Male and female (adult only) | Pediatric, male and female |
| Lumen | Two way and three way | Two way and three way |
| Indwell Time | No more than 30 days | No more than 30 days |
| Single Use | Yes | Yes |
| Size Range | Two way: 12,14,16,18,20,22,24Fr<br>Three way: 14,16,18,20,22,24Fr | Two way: 6,8,10,12,14,16,18,20,22,24 and 26Fr<br>Three way: 16,18,20,22,24 and 26Fr |
| Length | 337( $\pm$ 15) and 345( $\pm$ 15) mm | 310mm and 400mm |
| Balloon | Yes | Same |
| Balloon size | 5cc and 30cc | 1.5cc,3cc,5cc,10cc,15cc,20cc,30cc |
| Sterile | Yes | Same |
| Main Shaft<br>Material | Silicone and Zinc Oxide | Silicone |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for Apollon Co., Ltd. The logo features a stylized letter "A" in blue, enclosed within two concentric circles. The circles have small extensions with dots on either side, resembling an electrical circuit. Below the circular logo, the text "APOLLON Co.,Ltd." is printed in a simple, sans-serif font.
## 510(k) Summary - K201776
[21 CFR 807.92]
| Eyes in Tip | Yes | Same |
|---------------------|--------------|------|
| Tip Shape | Rounded | Same |
| Performance<br>test | ASTM F623-99 | Same |
The subject and predicate device have the same intended by the above table, the subject and predicate device have different technological characteristics. The differences in technological characteristics do not raise different questions of safety or effectiveness.
## 7. Summary of Testing Performed
The sponsor provided the following performance testing to support substantial equivalence:
- . Biocompatibility per ISO 10993 - cytotoxicity, sensitization, irritation or intracutaneous reactivity, material-mediated pyrogenicity, acute systemic toxicity, subacute toxicity, and implantation
- Sterilization validation per ISO 11135:2014
- . Shelf-life - transportation simulation, package integrity and functional testing
- . Performance/Functional testing per ASTM F623
The protocol and results of the provided performance testing is acceptable.
## 8. Conclusion
The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.