K023090 · Opticon Medical, Inc. · EZL · Mar 19, 2003 · Gastroenterology, Urology
Device Facts
Record ID
K023090
Device Name
OPTION-VF URINARY CATHETER, FV14218
Applicant
Opticon Medical, Inc.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Mar 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OPTION-vf is intended to provide drainage of the urinary bladder. The OPTION-yf is indicated for use only for urinary bladder drainage in female patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated.
Device Story
Indwelling urinary catheter for female patients; provides controlled bladder drainage. Device consists of flexible silicone shaft, retention balloon, adjustable retainer ring, and discharge bulb with integral patient-operated valve. Inserted through urethra by healthcare professional; balloon inflated with 5-10 cc sterile water. Patient squeezes discharge bulb to open valve for voiding; release closes valve. Eliminates need for external collection bags/tubing; allows normal bladder cycling and improved patient mobility/comfort. Used in clinical or home settings; operated by patient. Benefits include reduced self-consciousness, pain, and sleep disruption compared to continuous drainage systems.
Clinical Evidence
Multi-center randomized clinical trial comparing OPTION-vf to standard Foley catheter in female patients requiring short-term catheterization. Results showed no significant difference in incidence of significant bacteriuria or mean residual bladder volume. Adverse event rates were similar and mild. Patient questionnaires indicated significant benefits regarding mobility, comfort, self-consciousness, and ability to perform daily activities (showering/dressing) due to elimination of collection bags.
Technological Characteristics
Biocompatible silicone elastomers; flexible shaft with two proximal eyelets; retention balloon; adjustable retainer ring; self-sealing microvalve port; discharge bulb with integral valve. Conforms to ASTM F 623-99 (Standard Performance Specification for Foley Catheter). Sterile, single-use, disposable.
Indications for Use
Indicated for urinary bladder drainage in female patients requiring short-term (≤14 days) management for acute conditions, who are capable of self-operating the device, and for whom normal bladder cycling is not contraindicated.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Opticon Female Urinary Catheter for Continuous Drainage (K961697)
{0}------------------------------------------------
K023C090
Page 1 of 3
# 510(k) SUMMARY
# OPTICON MEDICAL, INC.
# OPTION-vf™ Urinary Catheter
#### Submitter's Name and Contact Information
Opticon Medical, Inc. 7001 Post Road, Suite 100 Dublin, OH 43016
Primary Contact: Glenn D. Brunner, President Phone Number: (614) 366-2000 Fax Number: (614) 336-2059
Date Prepared: September 13, 2002
#### Device Name
Trade / Proprietary Name: OPTION-vfTM Urinary Catheter Common / Usual Name: urinary catheter or Foley catheter Classification Name: catheter, retention type, balloon (product code EZL; 21 CFR 876.5130)
### Predicate Devices
- K961697, Medical Device International, Inc. Opticon Female Urinary Catheter for Continuous Drainage ("Opticon Direct").
- K760093, C.R. Bard, Inc. Bardex® Silicone Foley Catheter ("Bardex"). ●
#### Intended Use
The OPTION-vf is intended to provide drainage of the urinary bladder.
The OPTION-yf is indicated for use only for urinary bladder drainage in female patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated.
{1}------------------------------------------------
# Technological Characteristics and Substantial Equivalence
## Device Description:
The OPTION-vf is an indwelling catheter that provides controlled drainage of the urinary bladder. It is a sterile, single-use, disposable device that is to be prescribed by a physician and inserted and removed by an appropriate health care professional. The device is composed of biocompatible silicone elastomers, and consists of a flexible shaft with two opposing eyelets in the proximal tip for urine entry, one internal lumen for urine drainage and a second lumen for balloon inflation; a retention balloon; an adjustable retainer ring; a self-sealing balloon inflation microvalve port; and a urine discharge bulb with integral valve.
The device is inserted through the urethra to position the proximal tip and balloon into the urinary bladder. Once the catheter is properly inserted, the teardrop-shaped balloon is inflated with 5 to 10 cc of sterile water through the microvalve port via a standard syringe and a blunt cannula. The retainer ring is then adjusted to the urinary meatus, thus gently anchoring the device in place. Removal of the OPTION-vf follows aspiration of the fluid from the balloon.
The patient-operated discharge valve within the OPTION-vf is housed within the discharge bulb. When the bulb is squeezed, the valve opens allowing urine to pass. When the bulb is released, the valve closes, providing the patient with direct control over urinary drainage. The OPTION-vf thus allows the patient to mimic normal voiding and permits patient mobility and comfort by eliminating the need for urine collection bags and tubing.
## Substantial Equivalence Comparison:
The OPTION-vf and the predicate devices are all indwelling urinary catheters, which are used in the same anatomic location for the same purpose. The OPTION-vf is similar in design, materials of construction, physical dimensions, methods of insertion, maintenance, and removal, etc. to the predicate catheters. It also follows equivalent principles of operation, with the exception that the predicate catheters continuously drain into a urine collection bag, whereas the OPTION-yf has an integral valve that allows the user to control bladder drainage.
The OPTION-vf is substantially equivalent to the other legally marketed catheters referenced above. The OPTION-yf and the predicate devices are all indwelling urinary catheters used to drain the urinary bladder, thus the OPTION-vf raises no new issues of safety or effectiveness.
### Device Performance
The company has completed both the engineering testing typically required for clearance of a standard Foley catheter, and sufficient laboratory and clinical performance testing to demonstrate the comparative safety and effectiveness of the OPTION-vf.
### Engineering Testing:
Engineering verification testing confirms that results obtained from bench testing according to the recognized Foley standard; ASTM F 623-99, Standard Performance Specification for Foley Catheter, as well as additional verification testing demonstrate conformance of the OPTION-vf device with clinical and design requirements. The device conforms to all
{2}------------------------------------------------
KU23090
page 3 of 3
applicable elements of the ASTM Foley catheter standard, and demonstrates proper performance in all valve-related testing.
#### Clinical Testing:
In a multi-center randomized clinical trial, female patients requiring short-term urinary catheterization were treated with either a standard Foley indwelling catheter or the experimental Opticon-vf™ catheter.
There was no significant difference between groups with respect to the incidence of significant bacteriuria following catheter removal. The mean residual bladder volume at the time of catheter removal was also comparable between groups.
A review of adverse events by treatment group showed that adverse event rates were similar between the groups. In general, adverse events associated with the Opticon catheter were of mild severity and were resolved quickly either with appropriate patient counseling or catheter removal.
Results from the patient questionnaire revealed significant benefits of the Opticon catheter as compared to the Foley catheter with respect to mobility, pain/discomfort, self-consciousness, sleep disruption, and ability to shower or get dressed. These benefits follow directly from the design of the Opticon catheter in which bulky urine collection bags and connecting tubes are eliminated, as is the need to periodically empty the urine collection bag. Moreover, patients are able to maintain a more normal urinary function, with normal bladder cycling as compared to a continuously contracted bladder with constant drainage, and with the ability to discretely drain the bladder when desired.
In all instances, the OPTION-vf functioned to provide drainage of the urinary bladder and the device performance was as expected.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 9 2003
Mr. Glenn D. Brunner President Opticon Medical, Inc. 7001 Post Road Suite 100 DUBLIN OH 43016
Re: K023090
Trade/Device Name: OPTION-vf™ Urinary Catheter Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: 78 EZL Dated: December 23, 2002 Received: December 24, 2002
Dear Mr. Brunner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Statement of Indication for Use
| 510(k) Number: | K023090 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | OPTION-vf <sup>TM</sup> Urinary Catheter |
| Indication for use: | The OPTION-vf is indicated for use only for urinary bladder drainage in female patients: 1) who have acute conditions that require short-term (14 days or less) urinary management; 2) who are capable of operating the device in accordance with its instructions for use; and 3) for whom normal bladder cycling is not contraindicated. |
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <div> OR </div> | Over-The-Counter Use |
|------------------|---------------------------------------------|----------------------|
|------------------|---------------------------------------------|----------------------|
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K023090 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.