COMPACTCATH Lubricated Intermittent UrinaryCatheter and OneCath Intermittent Urinary Catheter
K180738 · Compactcath, Inc. · EZD · Jun 14, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180738
Device Name
COMPACTCATH Lubricated Intermittent UrinaryCatheter and OneCath Intermittent Urinary Catheter
Applicant
Compactcath, Inc.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Jun 14, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
CompactCath™ Lubricated Intermittent Urinary Catheter and OneCath™ Intermittent Urinary Catheter are indicated for use in male, female, and pediatric patients (adolescents and transitional adolescents) with chronic urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain.
Device Story
Sterile, single-use intermittent urinary catheters; used for draining urine from bladder in patients with voiding dysfunction. Available in multiple French sizes (8–18 Fr.), lengths (10" and 16"), and tip designs (straight and coudé). Device inserted into urethra to reach bladder; facilitates urine drainage. Used in clinical or home settings; operated by patients or healthcare providers. Benefits include relief of urine retention and management of PVR.
Clinical Evidence
Bench testing only. Functional performance testing conducted per BS EN 1616:1997+A1:1999 and ASTM F623:99. Biocompatibility testing per ISO 10993-1:2009 and FDA guidance. Sterilization validation per ISO 11137-1:2006, ISO 11137-2:2013, and ISO 11135:2014.
Technological Characteristics
Sterile, single-use intermittent urinary catheter. Available in 8-18 Fr sizes, 10" and 16" lengths, straight and coudé tips. Materials evaluated per ISO 10993-1:2009. Functional testing per BS EN 1616:1997+A1:1999 and ASTM F623:99. Sterilization via radiation (ISO 11137) or ethylene oxide (ISO 11135).
Indications for Use
Indicated for male, female, and pediatric (adolescent/transitional adolescent) patients with chronic urine retention or post void residual volume (PVR) due to neurogenic or non-neurogenic voiding dysfunction.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 14, 2018
Compactcath, Inc. % Tehyen Chu Regulatory Affairs Specialist ATOM Health Corporation 12FL, No.122 Songjiang Rd. Taipei City, 104 Taiwan
Re: K180738
> Trade/Device Name: CompactCath™ Lubricated Intermittent Urinary Catheter and OneCath™ Intermittent Urinary Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZD. EZC Dated: May 18, 2018 Received: May 25, 2018
Dear Tehyen Chu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Glenn B. Bell -S" in a large, bold, sans-serif font. The text is black against a white background. The letters are evenly spaced and the text is easy to read.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| 510(k) Number (if known) | K180738 |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | CompactCath <sup>TM</sup> Lubricated Intermittent Urinary Catheter and OneCath <sup>TM</sup> Intermittent Urinary Catheter |
| Indications for Use (Describe) | CompactCath <sup>TM</sup> Lubricated Intermittent Urinary Catheter and OneCath <sup>TM</sup> Intermittent Urinary Catheter are indicated for use in male, female, and pediatric patients (adolescents and transitional adolescents) with chronic urine retention and patients with post void residual volume (PVR) due to neurogenic and non-neurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain. |
| Type of Use (Select one or both, as applicable) | <span> <svg height="12" width="12"> <rect height="12" style="fill:black;" width="12"></rect> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <svg height="12" width="12"> <rect height="12" style="fill:white; stroke:black; stroke-width:1px;" width="12"></rect> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CompactCath. The logo features a stylized blue hand with a circle in the palm, followed by the text "COMPACT" stacked above "CATH". The text is also in blue and has a clean, modern font.
# 510(k) Summary
In accordance with 21CFR 807.92, the following information is provided for the CompactCath™ Lubricated Intermittent Urinary Catheter and OneCath™ Intermittent Urinary Catheter 510(k) Premarket Notification.
#### Company Information
| Submitter | CompactCath, Inc.<br>887 Federation Way<br>Palo Alto, CA 94303<br>Tel: 408-893-9776 |
|-----------|-------------------------------------------------------------------------------------|
| Contact | Naama Stauber Breckler<br>CEO |
#### Device Information
| Trade/Device Name | CompactCathTM Lubricated Intermittent Urinary<br>Catheter and OneCathTM Intermittent Urinary Catheter |
|------------------------|-------------------------------------------------------------------------------------------------------|
| Common Name | Intermittent Urinary Catheter |
| Classification | Class II |
| Regulation Number | 21 CFR 876.5130 |
| Regulation Name | Urological Catheter and Accessories |
| Primary Product Code | EZD |
| Secondary Product Code | EZC |
#### Predicate Devices
- . K160858 CompactCath™ Intermittent Urinary Catheter
- . K140945 CompactCath™ Intermittent Urinary Catheter
#### Indications for Use
CompactCath™ Lubricated Intermittent Urinary Catheter and OneCath™ Intermittent Urinary Catheter are indicated for use in male, female, and pediatric patients (adolescents and transitional adolescents) with chronic urine retention and patients with post void residual volume (PVR) due to neurogenic and nonneurogenic voiding dysfunction. The catheter is inserted into the urethra to reach the bladder allowing urine to drain.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for CompactCath. The logo consists of a stylized hand on the left and the word "COMPACT" stacked on top of the word "CATH" on the right. The hand and the text are in a light blue color.
## Device Description
The CompactCath™ and OneCath™ Intermittent Urinary Catheter are sterile, single use urine drainage catheter for use in draining urine from the bladder in subjects with urine drainage problems.
Both catheters will be offered in multiple French sizes (8 – 18 Fr.), lengths (10" and 16"), two tip designs (straight and coudé).
## Comparison with the predicate device
The CompactCath™ and OneCath™ Intermittent Urinary Catheter have the same technological characteristics as the predicate devices. In addition, the indications for use, clinical application, user operation, method of manufacturing, materials and packaging are all substantially equivalent.
## Performance Testing
The following performance tests were completed and the result met the acceptance criteria.
- Nonclinical functional performance testing was performed in accordance with: .
- 1. BS EN 1616: 1997 + A1:1999 "Sterile urethral catheters for single use" ; and
- 2. ASTM F623:99 "Standard Performance Specification for Foley Catheter"
- . Biocompatibility testing was conducted in accordance with:
- 1. ISO 10993-1:2009 "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process"; and
- 2. FDA Guidance "Use of International Standard ISO 10993 "Biological Evaluation of Medical Devices Part 1: Evaluation of Testing."
- Sterilization validation were conducted according to:
- 1. ISO 11137-1:2006 Sterilization of health care products Radiation Part 1; and
- 2. ISO 11137-2:2013 Sterilization of health care products Radiation Part 2.
- 3. ISO 11135:2014 Sterilization of health-care products -- Ethylene oxide --
- Requirements for the development, validation and routine control of a sterilization process for medical devices
- ●
# Conclusion
Based on the testing results provided in this premarket notification, the proposed device is as safe and effective, has the same intended use, technological characteristics and principals of operation as the predicate device. The minor differences between the two devices do not raise any new questions of safety or effectiveness. Therefore, the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.