The Dornier MINNOW Ureteral catheter is indicated for access and catheterization of the urinary tract, including the following applications: Delivery of contrast media Drainage of fluids from the urinary tract Delivery of irrigation fluids to the urinary tract Navigation of a tortuous ureter Access, advancement, or exchange of wire quides The target population is for adults only (at least 22 years old).
Device Story
Sterile, single-lumen, thermoplastic elastomer (Pebax) ureteral catheter; used by physicians in clinical settings for retrograde access to upper urinary tract. Device facilitates drainage, irrigation, or contrast media delivery; compatible with guidewires up to 0.038" diameter. Includes Touhy-Borst adapter for fluid aspiration/injection; removable to allow scope exchange. Radiopaque material allows visualization under fluoroscopy. Benefits include facilitated urinary tract access and fluid management during endoscopic procedures.
Clinical Evidence
Bench testing only. Performance verified via testing for sterility, packaging, biocompatibility, radiopacity, effective working length, shaft ID, tensile strength, and stiffness. All results met acceptance criteria and demonstrated equivalence to the predicate.
Technological Characteristics
Single-lumen catheter constructed of medical-grade thermoplastic elastomer (Pebax). Radiopaque via BaSO4 compounding. Available in 5Fr and 6Fr sizes, 70cm length. Features open tip and Touhy-Borst adapter. Compatible with up to 0.038" guidewires. Biocompatibility per USP Class VI.
Indications for Use
Indicated for adults (≥22 years) requiring urinary tract access/catheterization for contrast media delivery, fluid drainage, irrigation, navigation of tortuous ureters, or wire guide exchange.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 18, 2019
Dornier MedTech America Inc John Hoffer Vice President Quality, Regulatory, Clinical 1155 Roberts Blvd, Suite 100 Kennesaw, GA 30144
Re: K191187 Trade/Device Name: Dornier MINNOW Ureteral Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EYB Dated: July 17, 2019 Received: July 19, 2019
Dear John Hoffer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
{1}------------------------------------------------
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Acting Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known)
# K191187
Device Name
### Dornier MINNOW Ureteral Catheter
Indications for Use (Describe)
The Dornier MINNOW Ureteral catheter is indicated for access and catheterization of the urinary tract, including the following applications: Delivery of contrast media Drainage of fluids from the urinary tract Delivery of irrigation fluids to the urinary tract Navigation of a tortuous ureter Access, advancement, or exchange of wire quides
The target population is for adults only (at least 22 years old).
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801
Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## 510(k) SUMMARY
# Dornier MINNOW Ureteral Catheter
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Dornier MedTech America, Inc. 1155 Roberts Blvd., Suite 100 Kennesaw, GA 30144
Date Prepared: 4/30/19
Contact Person: John Hoffer Phone: 770-514- 6163 Name/Address of Sponsor and Name of Device
Dornier MedTech America, Inc. 1155 Roberts Blvd. Kennesaw, GA 30144
| Device Name - | Dornier MINNOW Ureteral Catheter<br>5F x 70cm Open Tip Ureteral Catheter - MIOTO570R<br>6F x 70cm Open Tip Ureteral Catheter - MIOTO670L |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name - | Ureteral Catheter |
| Classification Number - | 876.1530 |
| Classification Name - | Urological catheter and accessories |
| Product Code - | EYB |
| Product Code Name - | Catheter, ureteral, gastro-urology [GU] |
| Regulatory Class - | II |
Predicate Device Cook Open End Ureteral Catheter (K171662).
Purpose of the 510(k) Notice
The purpose of this submission is to obtain marketing clearance for the Dornier MINNOW Ureteral Catheter product line.
Intended Use/Indications for Use
The Dornier MINNOW Ureteral catheter is indicated for access and catheterization of the urinary tract, including the following applications:
Delivery of contrast media
Drainage of fluids from the urinary tract Delivery of irrigation fluids to the urinary tract Navigation of a tortuous ureter
Access, advancement, or exchange of wire guides
The target population is for adults only (at least 22 years old).
{4}------------------------------------------------
# Device Description
The MINNOW Ureteral Catheter is a sterile single lumen medical grade thermoplastic elastomer (Pebax®) catheter. It is designed to assist in access to the upper urinary tract using standard endoscopic technique for drainage and delivery of gels or fluids. The device is offered without side holes and is available in 5 and 6 Fr. Sizes, which are common to the industry. The catheter is indicated for use by physicians for facilitating access to the urinary tract through a retrograde route and may be used in conjunction with a quidewire
or for the injection of gels or fluids into the urinary tract. The catheter tip is radiused and has a .041" ID that allows for a smooth transition and passage when placed over a guidewire of up to 0.038" inches in diameter that is pre-positioned through the urological tract. The MINNOW Ureteral Catheters are packaged with a standard Touhy-Borst adapter that allows for injection or aspiration of fluids, but removable to allow for scope exchange without having to remove the catheter. The MINNOW Ureteral Catheters are available in 5 and 6 French (Fr) diameter, with a catheter length of 70 centimeters (cm). Dimensional details are below.
| Catheter Size | OD | ID | Tip ID | Length |
|---------------|-----------------|-----------------|-----------------|--------------|
| 5FR | .065" +/- .002" | .041" +/-.002" | .041" +/- .001" | 70cm +/- 2cm |
| 6FR | .078" +/- .002" | .049" +/- .002" | .041" +/- .001" | 70cm +/- 2cm |
The MINNOW Ureteral Catheters are constructed of a medical grade thermoplastic elastomer (Pebax®). This material has been USP Class VI tested and has passed all applicable biocompatibility testing. The base material is compounded with BaSO4 to render the MINNOW Ureteral Catheter radiopaque under x-ray fluoroscopy. All material colorants used are compliant with FDA standards.
## Technological Characteristics
The basic characteristics of the Dornier MINNOW Ureteral Catheter are substantially equivalent to the predicate device. They are both made from medical grade material and are a single lumen design with an open tip. Both are radiopaque and can be visualized under fluoroscopy. They are able to be passed over a 0.038 guidewire during clinical use.
## Performance Data
The MINNOW Ureteral Catheter was subjected to the following tests to assure design and performance under the specified testing parameters:
- Sterility Packaging Biocompatibility Radiopacity Effective Working Length Catheter Shaft ID Catheter Tensile Strength Catheter Stiffness
All testing was found to be acceptable and substantially equivalent to those of the predicate device.
## Substantial Equivalence
A comparison of design characteristics has been performed and demonstrates that the proposed MINNOW Ureteral Catheter is substantially equivalent to the predicate device in terms of intended use, technological characteristics, type of materials and performance characteristics. Therefore, the proposed MINNOW Ureteral Catheter is as safe, as effective, and performs as well as the predicate device.
# Conclusion
Based on the data and information comparing the MINNOW Ureteral Catheter and the predicate device we conclude they are substantially equivalent as they have the same intended use, basic design, principle of operation, technology, materials, and performance to the predicate. Any minor differences between the subject and predicate devices do not raise any concerns regarding the overall safety or effectiveness. Thus, the MINNOW Ureteral Catheter is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.