BALLARD MEDICAL PRODUCTS FACILETOME BOW STRING PAPILLOTOME
K990376 · Ballard Medical Products · KNS · May 3, 1999 · Gastroenterology, Urology
Device Facts
Record ID
K990376
Device Name
BALLARD MEDICAL PRODUCTS FACILETOME BOW STRING PAPILLOTOME
Applicant
Ballard Medical Products
Product Code
KNS · Gastroenterology, Urology
Decision Date
May 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ballard Medical Products® Faciletome™ Bow String Papillotome(s) is intended to be used for patients requiring transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi.
Device Story
The Faciletome™ Bow String Papillotome is an electrosurgical device used during ERCP procedures. It consists of a sheath housing a stainless steel mono-filament or stranded wire. Operated by a physician under direct vision of a duodenoscope (with or without fluoroscopy), the device is inserted into the biliary tree. The handle allows the operator to apply tension to the distal end, forming a 'bow' shape. When monopolar surgical power is applied, the wire incises the Papilla of Vater or Sphincter of Oddi. The sheath insulates the non-cutting portion of the wire. The device is available in single, double, and triple lumen designs with various wire lengths and precurved tip configurations. It facilitates biliary access and tissue incision, potentially benefiting patients by enabling therapeutic biliary interventions.
Clinical Evidence
Bench testing only. In vitro tests performed: high-frequency leakage current, capacitive coupling between active wire and guidewire, frequency and voltage signal verification (800V p-p @ 1MHz), and dielectric withstand testing.
Technological Characteristics
Electrosurgical device; stainless steel mono-filament or stranded cutting wire; single/multiple lumen sheath; one-piece handle system; precurved tip options. Energy source: monopolar electrosurgical power. Connectivity: compatible with 0.035 inch guidewires and duodenoscope biopsy channels. Optional Touhy Borst attachment for guidewire lumens.
Indications for Use
Indicated for patients requiring transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Wiltek Medical, Inc. papillotomes (K894861) (marketed as Accuratome™, Accuratome™, and Double Lumen™ Papillotome)
Submission Summary (Full Text)
{0}------------------------------------------------
5/3/99
Image /page/0/Picture/1 description: The image shows the logo for "POLLARD Medical Products". The word "POLLARD" is in large, bold, sans-serif font, with a registered trademark symbol to the right of the word. Below "POLLARD" is the phrase "Medical Products" in a smaller, bold, sans-serif font. There are horizontal lines above and below the phrase "Medical Products".
12050 Lone Peak Park Draper, Utah 84020 (801) 572-6800 FAX (801) 572-6999
K 990376
pg 1 of 2
Appendix F Attachment 1.0 Summary Statement
l
#### 510(k) Premarket Notification Summary Statement
Ballard Medical Products® Faciletome™ Bow String Papillotome
Friday, February 05, 1999
#### Submitter Information per 807.92(a)(1):
E. Martin Chamberlain Ballard Medical Products 12050 Lone Peak Parkway Draper, UT., 84020 Tel. (801) 572 - 6800 Fax. (801 572 - 6869
## Proprietary Name per 807.92(a)(2):
The Ballard Medical Products® Faciletome™ Bow String Papillotome
Common Name per 807.92(a)(2):
Biliary Bow String Papillotome or Sphincterotome
#### Classification per 807.92(a)(2):
The Ballard Medical Products® Faciletome™ Bow String Papillotome has been classified a Class II device through the Gastrointestinal -- Urology Panel per 21 CFR Part 876.4300. Classification name is: Unit, Electrosurgical, Endoscopic (With or Without Accessories). Product Code 78KNS.
Legally marketed equivalent(s) per 807.92(a)(3):
The Ballard Medical Products® Faciletome™ Bow String Papillotome device is substantially equivalent to the Wiltek Medical, Inc. papillotomes; the subject of 510(k) #K894861, marketed under the trade names of Accuratome TM, Accuratome™, and Double Lumen™ Papillotome.
#### Description of the device 807.92(a)(4):
The bow string papillotome is one of the most commonly used E.R.C.P. (Endoscopic Retrograde Colangio-Pancreatography) devices and has been established in the industry for several years. It has evolved into several variations on one general theme, facilitating entrance into the biliary tree. It performs this by housing a stainless steel mono-filament or stranded wire that is to be the active wire when monopolar, surgical power is introduced into the device handle connection. A sheath runs the length of the wire and allows only that portion of the wire that is needed to cut be exposed and insulates the remainder from the patient. The bow geometry and wire tension aid in cutting the tissue at the precise location. The wire is attached to a handle, that when manipulated, allows the operator to apply tension forming a " bow" at the distal end.
The Ballard Medical Products® Faciletome™ Bow String Papillotome is placed under direct vision of a duodenscope, with or without fluoroscopic aid. The single lumen, double lumen and triple lumen designs have a variety of cutting wire lengths as well as a variety of precurved tip configurations in the precurved model type.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "POLLARD Medical Products". The word "POLLARD" is in large, bold, sans-serif font, with a registered trademark symbol to the right of the "D". Below "POLLARD" is the phrase "Medical Products" in a smaller, sans-serif font. There are horizontal lines above and below the text.
12050 Lone Peak Parkway Draper, Utah 84020 (801) 572-6800 FAX (801) 572-6999
#### Intended Use per 807.92(a)(5):
The Ballard Medical Products® Faciletome™ Bow String Papillotome(s) is intended to be used for patients requiring transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi.
## Technological Characteristics (equivalence to predicate devices) per 807.92(a)(6):
The Ballard Medical Products® Faciletome™ Bow String Papillotome(s) are substantially equivalent to the Wiltek Medical, Inc. papillotomes; K894861, marketed under the trade names of Accuratome3m, Accuratome™, and Double Lumen™ Papillotome.
The general design characteristics and functionality are similar in that they are all one piece construction which include: single or multiple lumen shafts, a one piece handle system, multiple cutting wire lengths and construction with precurved tips in multiple lengths.
The above devices may be advanced with or without employing a 0.035 inch guidewire through the biopsy channel of a duodenoscope with optional use of fluoroscopy.
In all the devices, current is applied through the cutting wire to incise the Papilla of Vater or Sphincter of Oddi. A Touhy Borst may be attached to the guidewire lumens of the multiple lumen devices to assist in placement, and minimize contrast leakage.
## Determination of Substantial Equivalence (non-clinical data) per 807.92(b)(1):
The following in vitto tests were performed on the proposed Ballard Medical Products® FaciletomeTM Bow String Papillotome:
- A) High Frequency Leakage Current
- B) Capacitive Coupling between device active wire and recommended guidewire(s).
- C) Verification of Frequency and Voltage Signal (800V p-p @ 1MHz)
- D) Dielectric Withstand Test
#### Conclusions from non-clinical data per 807.92(b)(3):
Based on the indications for use, technological characteristics, and performance testing, the Ballard Medical Products® Faciletome™ Bow String Papillotome has been shown to be safe and effective for its intended use.
990376
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the seal's perimeter in a circular fashion. The seal is presented in black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
3 1030 MAY
Mr. E. Martin Chamberlain V.P. Regulatory Affairs Ballard Medical Products 12050 Lone Peak Parkway Draper, UT 84020
Re: K990376 Ballard Medical Products®Faciletome™ Bow String Papillotome Dated: February 5, 1999 Received: February 8, 1999 Regulatory Class: II 21 CFR 876.4300/Procode: 78KNS
Dear Mr. Chamberlain:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
APPENDIX A ATTACHMENT 3.0 INTENDED USE STATEMENT
# Ballard Medical Products® Faciletome™ Bow String Papillotome
## INTENDED USE STATEMENT
The Ballard Medical Products® Faciletome™ Bow String Papillotome(s) is intended to be used for patients requiring transendoscopic cannulation and sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi.
Prescription
t
Over-the-counter.
(Division Sign-
Division of Re ctive, Abdominal, ENT,
and Radiologi vices
510(k) Numbe. K990376
Image /page/3/Picture/10 description: The image shows the logo for "POLLARD Medical Products". The word "POLLARD" is in a bold, sans-serif font, with a registered trademark symbol to the right of the word. Below "POLLARD" is the phrase "Medical Products" in a smaller, sans-serif font.
12050 Lone Peak Parkv Draper, Utah 84020 (801) 572-6800 FAX (801) 572-6999
1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.