K171158 · Olympus Medical Systems Corp. · KNS · Nov 17, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K171158
Device Name
Single Use Electrosurgical knife
Applicant
Olympus Medical Systems Corp.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Nov 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These instruments have been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current. These instruments are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.
Device Story
Single-use electrosurgical knife designed for Endoscopic Submucosal Dissection (ESD); integrates saline feeding function with high-frequency tissue cutting. Device used with compatible endoscopes and electrosurgical units; operated by physicians during endoscopic procedures. Saline induction lifts mucosal lesions; high-frequency current enables tissue cutting under direct visualization. Benefits include combined functionality for lifting and cutting, facilitating ESD procedures. Device is sterile, single-use, and disposable.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated using ex-vivo porcine digestive tract models (esophagus, stomach, colon). Tests compared cutting efficacy, mucosal lift size, and thermal damage against predicate devices. Additional testing included knife operation, dimensional verification, durability, shelf-life (ASTM F1980-16), sterilization (ISO 11135:2014), biocompatibility (ISO 10993), and electromagnetic/electrical safety.
Technological Characteristics
Endoscopic electrosurgical knife with integrated saline irrigation channel. Materials validated per ISO 10993. Sterilization via ethylene oxide (ISO 11135). Electrical safety per AAMI/ANSI ES60601-1, IEC 60601-2-2, and IEC 60601-2-18. Shelf-life validated per ASTM F1980-16. Device is a standalone accessory for use with external electrosurgical generators and endoscopes.
Indications for Use
Indicated for patients requiring endoscopic mucosal lesion lifting and tissue cutting within the digestive tract via high-frequency electrosurgery.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 17, 2017
Olympus Medical Systems Corp. % Daphney Germain-Kolawole Senior Project Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway P.O. Box 610 Center Vallev. PA 18034-0610
Re: K171158
Trade/Device Name: Single Use Electrosurgical knife Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KNS Dated: October 20, 2017 Received: October 23, 2017
Dear Daphney Germain-Kolawole:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/6 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clear and easy to read, with a simple and professional appearance.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K171158
Device Name
Single Use Electrosurgical Knife
Indications for Use (Describe)
These instruments have been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current.
These instruments are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Underneath the word is a thin, horizontal, yellow line. Below the line is the company's slogan, "Your Vision, Our Future," written in a smaller font.
# 510(k) Summary
April 19, 2017
### I. General Information OLYMPUS MEDICAL SYSTEMS CORP. = Applicant: 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No: 8010047 ■ Daphney Germain-Kolawole Official Correspondent: Senior Project Manager, Regulatory Affairs Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 484-896-5691 FAX: 484-896-7128 Email: daphney.germain-kolawole@olympus.com = Manufacturer: Aomori Olympus Co., Ltd. 248-1 Okkonoki 2-chome Kuroishi-shi, Aomori, Japan 036-0357 Establishment Registration No.: 9614641 II. Device Identification ■ Device Trade Name: Single Use Electrosurgical knife ■ Common Name: Single Use Electrosurgical knife . Regulation Number: 876.4300 ■ Regulation Name: Endoscopic electrosurgical unit and accessories ■ Regulatory Class: II ■ Classification Panel: Gastroenterology and urology
- Product Code: KNS
{4}------------------------------------------------
| Single Use Electrosurgical knife | | |
|-----------------------------------------------------------------|----------------------------------------|---------|
| Primary Predicate Device1 | | |
| Device name | Applicant | 510(k)# |
| SINGLE USE<br>ELECTROSURGICAL KNIFE<br>SERIES | OLYMPUS MEDICAL<br>SYSTEMS CORPORATION | K092309 |
| Additional Predicate Device2 | | |
| Device name | Applicant | 510(k)# |
| ERBE WaterJet Model<br>ERBEJET®2 System with<br>HybridAPC Probe | ERBE USA, INC. | K143306 |
| Additional Predicate Device3 | | |
| Device name | Applicant | 510(k)# |
| ERBE HYBRID KNIFE | ERBE USA, INC. | K083608 |
# III. Predicate Device and Reference Devices Information
#### Device Description IV.
The subject device is an endoscopic electrosurgical knife incorporating the saline feeding function for Endoscopic Submucosal Dissection (ESD). The indications for use of the subject device is a combination of predicate devices 1 and 2.
#### V. Indications for Use
These instruments have been designed to be used with Olympus endoscopes and electrosurgical units to cut tissue within the digestive tract using high-frequency current. These instruments are indicated for the induction of sterile normal saline into the submucosa to lift mucosal lesions using direct visualization through an endoscope.
# VI. Comparison of Technological Characteristics
Compared to the primary predicate device (KD-620LR, K092309), the subject device has a different knife shape. Additionally, the subject device incorporates the saline feeding function for ESD procedures which is an equivalent function available in both predicate device 2 (HybridAPC Probe, K143306) and predicate device 3 (ERBE HYBRID KNIFE, K083608).
## VII. Summary of non-clinical testing
- Performance testing was conducted to demonstrate the essential performance of the subject device and confirmed that the subject device works as intended with the compatible devices. Additionally, the results of the tests below were evaluated as substantially equivalent to the cited predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a thin, yellow line. Underneath the line, the words "Your Vision, Our Future" are written in a smaller, blue font.
K171158
### <Test A>
A test to evaluate the performance to cut the digestive tract mucosa and compare the result with that of the predicate device. The subject device and the primary predicate device 1 (KD-620LR, K092309) were compared for the equivalency in cutting the ex-vivo porcine digestive tract mucosa (esophagus, stomach, and colon).
### <Test B>
A test to evaluate the lifting of the digestive tract mucosa and compare the size of mucosal lift with that of the predicate device. The subject device and the predicate device 3 (Hybridknife, K083608) were compared for the equivalency in lifting of the ex-vivo porcine digestive tract mucosa (esophagus, stomach, and colon).
### <Test C>
A test composed of the test items "cut the mucosa" and "lift the mucosa by induction of saline to the submucosal layer" to compare the degree of thermal damage on the ex-vivo porcine digestive tract (esophagus, stomach, and colon) wall caused by the subject device and that caused by the predicate device 3 (Hybridknife, K083608). The histological study was carried out to evaluate the degree of thermal damage.
<Other performance testing>
- > Knife operation with the compatible endoscopes
- >> Dimension of each part of the knife
- > General durability
- Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verification tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment.
- Shelf-life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16, the standard guide for accelerated aging of sterile barrier systems for medical devices. Three-year aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test.
- Sterilization validation was carried out with Methods Half-cycle approach in accordance with ISO 11135:2014.
- Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process' issued on June 16, 2016.
- Electromagnetic compatibility, electric safety, and thermal safety had been confirmed.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Olympus. The word "OLYMPUS" is written in large, bold, blue letters. Below the word is a yellow line. Below the line, the words "Your Vision, Our Future" are written in a smaller font.
K171158 Page 4 of 4
The following standards have been applied to the Single Use Electrosurgical knife.
- > ISO 11135
- > ASTM F1980-16
- > ISO 10993-1
- > AAMI/ANSI/ISO 10993-5
- > AAMI/ANSI/ISO 10993-10
- > ISO 10993-11
- > AAMI/ANSI ES60601-1
- > IEC 60601-1-2
- > IEC 60601-1-6
- > IEC 60601-2-2
- > IEC 60601-2-18
- > ISO 14971
### VIII. Conclusion
Compared to the predicate devices, the proposed design modifications to the Single Use Electrosurgical Knife were verified and validated and do not raise any significant issues with safety and effectiveness. Therefore, the subject device is substantially equivalent to the cited predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.