K171096 · Fujifilm Medical Systems U.S.A, Inc. · KGE · May 9, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K171096
Device Name
Fujifilm Diathermic Slitter (FlushKnife)
Applicant
Fujifilm Medical Systems U.S.A, Inc.
Product Code
KGE · Gastroenterology, Urology
Decision Date
May 9, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.
Device Story
Single-use electrosurgical instrument for endoscopic tissue removal and hemostasis. Device consists of proximal handle with slider, flexible resin tube, and distal slitter tip. Operator inserts device through endoscope forceps channel; manually extends/retracts slitter via slider. High-frequency (HF) electrical current delivered from external electrosurgical generator to slitter tip to perform tissue cutting/coagulation. Integrated water/fluid supply channel allows delivery of sterile water or fluids to target site via syringe or pump to clear debris/blood. Used in clinical settings by physicians. Benefits include precise tissue management and improved visualization via fluid irrigation.
Clinical Evidence
Bench testing only. Performance testing included fluid supply, temperature rise, dielectric voltage, and durability. All tests met pre-defined acceptance criteria.
Technological Characteristics
Monopolar electrosurgical instrument. Materials: stainless steel, ceramics, fluorine resin, fluorine coating, silver brazing filler metal. Dimensions: 0.5mm slitter diameter, 2.7mm max insertion diameter, 1800mm/2300mm working length. Energy: HF current from external generator. Connectivity: A-cord to generator, fluid port for syringe/pump. Sterilization: Sterile, single-use.
Indications for Use
Indicated for ablation, incision, dissection, avulsion, coagulation, and hemostasis of tissue within the digestive tract using high-frequency current via endoscope.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Fujifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) (K161186)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other. The faces are facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 9, 2017
FUJIFILM Medical Systems U.S.A., Inc. Jeffrey Wan Specialist, Regulatory Affairs 10 High Point Drive Wayne, NJ 07470
Re: K171096
Trade/Device Name: Fujifilm Diathermic Slitter (FlushKnife) Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: KGE Dated: April 11, 2017 Received: April 13, 2017
Dear Jeffrey Wan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known) K171096
Device Name
Fujifilm Diathermic Slitter (FlushKnife)
Indications for Use (Describe)
Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.
Type of Use (Select one or both, as applicable) 区 Prescription Use (Part 21 CFR 801 Subpart D)
□ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Page 1 of 1 FDA
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### XV. 510(K) SUMMARY
### Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc. Endoscopy Division 10 High Point Drive Wayne, NJ 07470 USA FDA Establishment Registration Number: 2431293
# Contact Person:
Jeffrey Wan Specialist, Regulatory Affairs Telephone: (973) 709-2219 Ext. 522219 Facsimile: (973) 633-8818 E-Mail: jeffrey.wan@fujifilm.com
Date Prepared: May 8, 2017
### Identification of the Proposed Device:
| Proprietary/Trade Name: | Fujifilm Diathermic Slitter (FlushKnife) |
|-------------------------|------------------------------------------|
| Common Name: | Electrosurgical Instruments |
| Device Class: | Class II |
| Review Panel: | Gastroenterology/Urology |
#### Classification Information:
Endoscopic electrosurgical unit and accessories 21 C.F.R. § 876.4300 Product Code: KGE
#### Predicate Device:
Fujifilm Diathermic Slitter (FlushKnife) and Diathermic Slitter (ClutchCutter) (K161186)
# Purpose of the Special 510(k) notice:
The Diathermic Slitter (FlushKnife) is a modification to Diathermic Slitter (FlushKnife) (K161186).
#### Intended Use:
Diathermic Slitter (FlushKnife) DK2618J and DK2623J are intended to be used with specified endoscopes to cut tissue using high-frequency current within the digestive tract. The devices are indicated for ablation, incision, dissection, avulsion, coagulation and hemostasis of tissue within the digestive tract.
# Device Description:
Diathermic Slitter (FlushKnife) is sterile and intended for single used electrosurgical instruments that remove tissue and control bleeding by use of high-frequency ("HF") electrical current.
The device is comprised of a proximal handle with slider that is connected to a flexible resin tube. The flexible resin tube covers and insulates the operation wire and slitter (when retracted). The
{4}------------------------------------------------
operation wire controls the mechanical function of and delivers high frequency (HF) electrical current to the slitter. The proximal end of the operation wire is connected to the slider, which allows the operator to manually control the extension and retraction of the sistal end of the operation wire connects to the slitter, which is located at the distal tip of the device.
The device connects to a HF electrosurgical power supply unit by an active cord ("A-Cord") connector. HF electrical current generated by a HF electrosurgical power supply unit flows to the slitter from the HF electrosurgical power supply unit via the A-cord, the A-cord connector, and the operation wire.
The distal tip of the device is inserted through the forceps channel of the specified endoscope. Once inserted, the operator can extend the slitter from the tip of the endoscope using the slitter is extended to the target site of a patient. Cleavage, resection, incision, ablation, hemostasis, coagulation, or excision of tissue is achieved by delivering HF current to the target tissue through the slitter.
The device is provided sterile for single-use only.
The device can supply sterile water or fluids to its distal end target site via a water/fluid supply channel. The port for the water/fluid supply channel is located on the handle at the proximal end of the device. The sterile water or fluid also aids in the removal of debris, blood and extraneous material from the distal end of the device or a target site.
To supply sterile water or fluid to the distal end target site, a syringe filled with water or fluid is connected to the water/fluid inlet at the proximal handle.
Besides a syringe, alternate delivery methods can also be used to supply sterile water or fluid to the distal end target site of the device.
# Technological Characteristics:
A table comparing the technological characteristics between FlushKnife and the predicate devices is shown below.
| | Predicate Device | Proposed Device | |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Device Name | Fujifilm Diathermic Slitter<br>(FlushKnife) | Fujifilm Diathermic Slitter<br>(FlushKnife) | |
| Device Models | DIATHERMIC SLITTER, MODEL<br>DK2618J -N10-, DK2618J -N15-,<br>DK2618J -N20-, DK2618J -N25-,<br>DK2618J -N30-, DK2618J -B15-,<br>DK2618J -B20-, DK2618J -B25-,<br>DK2618J -B30-, DK2623J -N15-,<br>DK2623J -N20-, DK2623J -B15-,<br>DK2623J -B20- | Same as predicate device | |
| 510(k) Number | K161186 | Pending | |
| | Intended Use | These instruments have been<br>designed to be used with specified<br>endoscopes to cut tissue using<br>high-frequency current within the<br>digestive tract. Both the ball tip<br>type and the needle type<br>instruments are indicated for<br>ablation, incision, dissection,<br>avulsion, cauterization,<br>coagulation and hemostasis of<br>tissue within the digestive tract. | Same as predicate device |
| | | | |
| Technological<br>Characteristics | Slitter Length | 1.0/1.5/2.0/2.5/3.0mm (for<br>DK2618 series)<br>1.5/2.0mm (for DK2623 series) | Same as predicate device |
| | Slitter Diameter | 0.5mm | Same as predicate device |
| | Slitter Shape | Needle Type<br>(With Ball Tip:-BXX-)<br>(Without Ball Tip:-NXX-) | Same as predicate device |
| | Maximum Diameter of<br>Insertion Portion | 2.7mm | Same as predicate device |
| | Working Length | 1800mm/2300mm | Same as predicate device |
| | Method of Operation | Manually (handle slider) | Same as predicate device |
| Energy used or delivered | | energy delivered from a<br>electrosurgical generator | Same as predicate device |
| Monopolar/Bipolar | | Monopolar | Same as predicate device |
| Sterilization | | Yes (Single Use Device) | Same as predicate device |
| Combination equipment | | Endoscope<br>Electrosurgical generator<br>A Cord<br>Syringe | Endoscope<br>Electrosurgical generator<br>A Cord<br>Syringe<br>Pump |
| Compatible fluids | | Sterile water | Sterile water and other sterile<br>fluids |
| Material of Insertion Portion | | Stainless steel<br>Ceramics<br>Fluorine resin<br>Fluorine coating<br>Silver brazing filler metal | Same as predicate device |
# Summary of differences between modified FlushKnife from its predicate device
{5}------------------------------------------------
# Performance Data
Fujifilm conducted the following performance testing on the proposed device FlushKnife to ensure that the modified device performs equivalently to the predicate device:
- Fluid supply ●
- Temperature rise .
- Dielectric voltage ●
- Durability .
In all cases, the device met the pre-defined acceptance criteria for the test.
{6}------------------------------------------------
# Substantial Equivalence
The company's FlushKnife has the same intended use as the previously cleared predicate FlushKnife (K161186). In addition, the proposed device FlushKnife has the same intended use, indications for use, technological characteristics, and principles of operation as its predicate. Although there is a minor difference between the proposed device and its predicate device, namely the addition of compatible sterile fluids and alternate delivery systems, this difference does not raise new or additional questions of safety or effectiveness of the proposed device. Thus, the proposed device FlushKnife is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.