Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current. Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications. The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the uterus in conductive irrigation fluid as part of a resectoscope system. Specific indications: - transcervical diagnosis and treatment (resection, vaporization, ablation, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis) - lysis of intrauterine septa - endometrial ablation
Device Story
Resection electrodes with HF cable are single-use endoscopic accessories used in urological and gynecological procedures. The device consists of an active tip (loops, bands, rollers, needles, or buttons), PTFE color-coded identification, insulator, guiding tube, and shaft. It connects to an electrosurgical generator via an included HF cable. During surgery, the physician uses the resectoscope system to deliver HF current to the target tissue for cutting, ablation, resection, vaporization, or coagulation in conductive irrigation fluid. The device is operated by a physician in a clinical setting. The output is the physical effect of the electrosurgical current on tissue, which the physician observes directly through the resectoscope to guide treatment. This allows for minimally invasive removal or treatment of intrauterine or urological pathologies, benefiting the patient by reducing the need for open surgery.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (AAMI/ANSI ES 60601-1, IEC 60601-2-2, IEC 60601-2-18), sterilization validation (ISO 11135), and package integrity (ISO 11607-1).
Technological Characteristics
Active tip (loops, bands, rollers, needles, buttons), PTFE insulation, guiding tube, and shaft. HF cable with lantern plugs. 24 Fr shaft diameter; 261.8-336.7mm length. Compatible with 12°-30° telescopes. Sterilized via EtO (ISO 11135).
Indications for Use
Indicated for urological and gynecological endoscopic diagnosis and treatment, including cutting, ablation, resection, vaporization, and coagulation of tissue using HF current. Gynecological use includes transcervical treatment of intrauterine myomas, polyps, synechias, endometrium, lysis of intrauterine septa, and endometrial ablation.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 28, 2017
Olympus Winter & Ibe GmbH % Dolan Mills Sr. Specialist, Regulatory Affairs Gyrus ACMI. Inc. 136 Turnpike Road Southborough, MA 01772
Re: K171965
> Trade/Device Name: Resection Electrodes with HF cable (WA22702S, WA22703S, WA22706S, WA22707S, WA22721S, WA22723S, WA22732S, WA22751S, WA22755S, WA22760S, WA22737S, WA22738S, WA22739S, WA22766S, WA22740S, WA22741S, WA22742S, WA22743S, WA22744S, WA22705S, WA22602S, WA22603S, WA22606S, WA22607S, WA22621S, WA22623S, WA22632S, WA22651S, WA22655S, WA22657S, WA22637S, WA22638S, WA22639S, WA22666S, WA22640S, WA22641S, WA22642S, WA22643S, WA22644S, WA22605S, WA47705S, WA47706S, WA47707S, WA47721S, WA47723S, WA47732S, WA47737S, WA47738S, WA47739S, WA47751S, WA47755S, WA47760S, WA47766S, WA47740S, WA47741S, WA47742S, WA47743S, WA47744S) Regulation Number: 21 CFR§ 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: II Product Code: FAS, FJL, HIH Dated: June 29, 2017
Dear Dolan Mills:
Received: June 30, 2017
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of
Image /page/0/Picture/9 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a symbol that resembles a stylized caduceus, with three figures or forms intertwined, representing health and human services.
{1}------------------------------------------------
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K171965
#### Device Name
Resection Electrodes with HF cable (WA22703S, WA22706S, WA22701S, WA2272SS, WA22732S, WA22732S, WA22751S, WA22751S, WA22760S, WA22737S, WA22739S, WA22766S, WA22740S, WA22741S, WA22742S, WA22743S, WA22744S, WA22705S, WA22602S, WA22603S, WA22606S, WA22621S, WA22621S, WA22635, WA22651S, WA22655S, WA22657S, WA22637S, WA22638S, WA22639S, WA22666S, WA22640S, WA22641S, WA22642S, WA22644S, WA22605S)
#### Indications for Use (Describe)
Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 EF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K171965 Device Name Resection Electrodes with HF cable (WA47705S, WA47707S, WA47721S, WA47723S, WA47732S, WA47737S, WA47738S, WA47739S, WA47751S, WA47755S, WA47760S, WA47740S, WA47740S, WA47741S, WA4742S,
WA47743S, WA47744S)
Indications for Use (Describe)
Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications.
The general indications include transcervical resection, vaporization, cutting and coagulation of tissue in the uterus in conductive irrigation fluid as part of a resectoscope system.
Specific indications:
- transcervical diagnosis and treatment (resection, vaporization, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis)
- lysis of intrauterine septa
- endometrial ablation
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K171965
# 510(k) Summary of Safety and Effectiveness September 22, 2017
## 1. General Information
| Manufacturer: | Olympus Winter & Ibe GmbH<br>Kuehnstr. 61<br>22045 Hamburg<br>Germany<br>Establishment Registration Number: 9610773 |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Dolan Mills<br>Sr. Specialist, Regulatory Affairs<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 901.373.0236<br>Fax: 901.373-0220<br>Email: dolan.mills@olympus-osta.com |
| Establishment<br>Registration No | 3003790304 |
## 2. Device Identification
| Common Name: | - Electrode, Electrosurgical, active, urological |
|-------------------------|------------------------------------------------------|
| Regulation Number: | 876.4300 |
| Classification | - Endoscopic electrosurgical unit and accessories |
| Device Class: | II |
| Product Code: | FAS / HIH / FJL |
| Review Panel: | Gastroenterology / Urology / Obstetrics / Gynecology |
| Proprietary/Trade Name: | Resection Electrodes with HF cable |
Model numbers:
WA22702S, WA22703S, WA22705S, WA22706S, WA22707S, WA22721S, WA22723S, WA22732S, WA22737S, WA22738S, WA22739S, WA22751S, WA22755S, WA22760S, WA22602S, WA22603S, WA22605S, WA22606S, WA22607S, WA22621S, WA22623S, WA22632S, WA22637S, WA22638S,
{5}------------------------------------------------
WA22639S, WA22651S, WA22655S, WA22657S, WA22766S, WA22740S, WA22741S, WA22742S, WA22743S, WA22744S, WA22666S, WA22640S, WA22641S, WA22642S, WA22643S, WA22644S
WA47705S, WA47706S, WA47707S, WA47721S, WA47723S, WA47732S, WA47737S, WA47738S, WA47739S, WA47751S, WA47755S, WA47760S, WA47766S, WA47740S, WA47741S, WA47742S, WA47743S, WA47744S
### 3. Predicate Devices
The predicate device was chosen from the following predicate 510(k):
| 510(k)<br>No. | Name | Predicate Model No. | Product code /<br>Reg No. |
|---------------|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------|
| K152092 | Resection Electrodes | WA22302S, WA22503S,<br>WA22306S, WA22507S,<br>WA22521S, WA22523S,<br>WA22332S, WA22351S,<br>WA22355S, WA22557S,<br>WA22537S, WA22538S,<br>WA22539S, WA22566S,<br>WA22540S, WA22541S,<br>WA22542S, WA22543S,<br>WA22544S, WA22305S | FAS / 876.4300<br>FJL / 876.1500<br>HIH / 884.1690 |
These predicates have not been subject to a design-related recall.
No reference devices were used in this submission.
{6}------------------------------------------------
K171965 Page 3 of 6
The Olympus Resection Electrodes with HF cable that are subject to this submission are for application in saline. Depending on the characteristics of electrical current. which is provided by the electrosurgical generator, electrosurgery can be used for coagulation, vaporization and cutting.
The subject HF-Resection Electrodes consist of an active tip, PTFE color code identification, an insulator between the electrode and electrode tube, a guiding tube, telescope clip and arm (shaft). The accompanying HF cables consist of two lantern plugs on the instrument side and one generator plug on the generator side.
The design and dimensions of the electrodes vary to accommodate various procedural conditions. The active tips of the various electrodes may consist of loops, bands, rollers, needles or buttons. The electrodes have a shaft diameter of 24 Fr. range in length from 261.8-336.7mm, and range in compatibility with telescopes with a direction of view of 12° - 30°. The design of the HF cable plugs vary depending on which electrosurgical generator they are compatible with.
All subject Resection Electrodes are single-use electrodes and are delivered sterile. All subject Resection Electrodes are provided with a single-use, sterile cable to connect the electrode to the generator.
### 5. Indications for Use
#### Resection Electrodes with HF cable for use in Urology
Article no. : WA22702S. WA22703S. WA22705S. WA22706S. WA22707S. WA22721S, WA22723S, WA22732S, WA22737S, WA22738S, WA22739S, WA22751S, WA22755S, WA22760S, WA22602S, WA22603S, WA22605S, WA22606S, WA22607S, WA22621S, WA22623S, WA22632S, WA22637S, WA22638S, WA22639S, WA22651S, WA22655S, WA22657S, WA22766S, WA22740S, WA22741S, WA22742S, WA22743S, WA22744S, WA22666S, WA22640S, WA22641S, WA22642S, WA22643S, WA22644S
Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in urological applications: Cutting, ablation, resection, vaporization and coagulation with HF current.
#### Resection Electrodes with HF cable for use in Gynecology
Article no. : WA47705S, WA47706S, WA47707S, WA47721S, WA47723S, WA47732S, WA47737S, WA47738S, WA47739S, WA47751S, WA47755S, WA47760S, WA47766S, WA47740S, WA47741S, WA47742S, WA47743S, WA47744S
{7}------------------------------------------------
Electrodes with HF-cable are part of a resectoscope system for endoscopic diagnosis and treatment in gynecological applications.
The general indications include transcervical resection, vaporization, ablation, cutting and coagulation of tissue in the uterus in conductive irrigation fluid as part of a resectoscope system.
# Specific indications:
- transcervical diagnosis and treatment (resection, vaporization, ablation, biopsy, cutting and coagulation) of intrauterine myomas, intrauterine polyps, synechias and endometrium (TCRis)
- lysis of intrauterine septa -
- endometrial ablation -
Although the subject and predicate devices have different indications for use statements, the intended use is the same. The subject device is indicated for gynecologic applications while the predicate device is indicated for urological indications. However, when the decision-making criteria specified in the FDA guidance document, "General/Specific Intended Use," (issued November 4, 1998) are applied, the subject device indications are determined to fall within the scope of the intended use of the predicate device. The specific gynecology indications of the subject device do not represent a different intended use since there is an extensive knowledge base regarding the use of bipolar electrosurgery for gynecological applications, including the disease states listed in the specific gynecologic indications. Therefore, the intended use comparison supports substantial equivalence.
# 6. Comparison of Technological characteristics
At a high level. the subject and predicate devices are based on the same technological principle with the same elements:
- I Resection electrodes consisting of an active (distal) tip, PTFE color code identification at the distal and proximal ends, an insulator between the electrode and electrode tube, a stabilizing (guiding) tube, and arm (shaft)
- . HF cables consisting of two lantern plugs on the instrument side and one generator plug on the generator side
- I Used in combination with a resectoscope system
- 트 Like the predicate electrodes, the subject device resection electrode series features loops, bands, needles, rollers, and a button as active tip shapes
- . Resection electrodes utilizing ablation or for resection in saline (dependent on model)
- I Respectively identical or similar outer dimensions
{8}------------------------------------------------
K171965 Page 5 of 6
- Design changes of the electrodes and cables are minor and do not negatively impact safety or effectiveness of the subject devices
- I The same or similar materials in patient contact are used in predicate and subject device and have all been successfully tested for biocompatibility.
As stated above, the subject and predicate devices have similar design characteristics and performance specifications, with the exception of the addition of the HF cables. These minor differences, however, do not raise different questions of safety or effectiveness. As demonstrated in the non-clinical testing (e.g., electrical safety, sterilization validation, and package integrity), the different technological characteristics do not affect the safety and effectiveness of the subject devices.
## 7. Performance Data
The subject device electrodes, including all patient-contacting materials, are identical in design and manufacturing to the predicate device (K152092). Therefore, the current submission relies on performance testing previously reviewed and deemed acceptable in K152092.
### 8. Electrical safety and electromagnetic compatibility (EMC)
Electrical Safety was tested according to
| AAMI/ANSI ES 60601-<br>1:2005 + A1:2012, C1:2009<br>and A2:2010 | Medical Electrical Equipment - Part 1.1 General<br>requirements for safety and essential performance. |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/ANSI/IEC 60601-2-2<br>2009 | Medical Electrical Equipment - Part 2-2: Particular<br>Requirements for the Basic Safety And Essential<br>Performance of High Frequency Surgical Equipment<br>and High Frequency Surgical Accessories |
| IEC 60601-2-18:2009 | Medical electrical equipment - Part 2-18: Medical<br>Electrical Equipment - Part 2-18: Particular<br>requirements for the basic safety and essential<br>performance of endoscopic equipment. |
Electromagnetic compatibility (EMC) testing according to IEC 60601-1-2 and IEC 60601-2-2 is not applicable to the Resection Electrodes. They do not contain any electrical components that can be influenced by electromagnetic emission as well as electrostatic discharge. From an electromagnetic compatibility point of view, it is a metal bar that cannot be influenced at any time.
{9}------------------------------------------------
## 9. Sterilization and Shelf Life
Sterilization is performed according to ISO 11135 and packaging conforms with AAMI ANSI ISO 11607-1:2006. The EtO sterilization cycle has been validated.
A sterility assurance level (SAL) of 106 was reached during validation and will be used for routine sterilization in compliance with regulations in force for sterile medical devices.
The EtO residuals are within the limits after tunnel degassing time.
Shelf Life testing, including package integrity testing in accordance with ISO 11607-1:2006, was conducted to support a shelf life of 3 years for the resection electrodes.
### 10. Conclusion
The performance data support the safety and effectiveness of the subject device and demonstrate that the subject device is substantially equivalent to the predicate device.
In conclusion, the Resection Electrodes with HF cable are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.