ARTHROCARE BIPOLAR ELECTROSURGERY SYSTEM

K031029 · Arthrocare Corp. · FAS · Jun 19, 2003 · Gastroenterology, Urology

Device Facts

Record IDK031029
Device NameARTHROCARE BIPOLAR ELECTROSURGERY SYSTEM
ApplicantArthrocare Corp.
Product CodeFAS · Gastroenterology, Urology
Decision DateJun 19, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution. Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization.

Device Story

Bipolar electrosurgical system for endoscopic urological procedures; consists of electrosurgical generator (Controller), disposable bipolar loop electrode, and reusable patient cable. Operates via high-frequency energy in conductive irrigation fluids (saline, Ringer's lactate). Used by surgeons in clinical settings under direct or video-assisted fiberoptic visualization. Enables resection, ablation, excision of urological tissue (prostate, bladder tumors), and hemostasis. System provides controlled energy delivery to target tissue; surgeon monitors procedure via visualization; output affects tissue state directly to achieve surgical goals.

Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and intended use compared to predicate devices.

Technological Characteristics

Bipolar high-frequency electrosurgical system. Components: electrosurgical generator, disposable bipolar loop electrode, reusable patient cable. Operates in conductive irrigation solutions (saline, Ringer's lactate).

Indications for Use

Indicated for patients requiring endoscopic urological surgery, including transurethral prostatectomy (TURP), transurethral incisions in the prostate (TUIP), and bladder wall tumor resection. Used for tissue resection, ablation, excision, and hemostasis.

Regulatory Classification

Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ JUN 1 9 2003 ## 510(k) Summary # ArthroCare Corporation ArthroCare® Bipolar Loop Electrosurgery System ### General Information | Submitter Name/Address: | ArthroCare Corporation<br>680 Vaqueros Avenue<br>Sunnyvale, CA 94085-2936 | |-------------------------|---------------------------------------------------------------------------| | Phone Number: | (408) 736-0224 | | Contact Person: | Valerie Defiesta-Ng<br>Director, Regulatory Affairs | | Date Prepared: | March 31, 2003 | **Device Description** | Trade Name: | ArthroCare Bipolar Loop Electrosurgery<br>System | |----------------------|-------------------------------------------------------------------------------------| | Generic/Common Name: | Electrosurgical Device and Accessories | | Classification Name: | Electrosurgical Cutting and Coagulation<br>Device and Accessories (21 CFR 878.4400) | **Predicate Devices** | ArthroCare Bipolar Loop<br>Electrosurgery System | K955531, K010568, K020832 and K022543 | |--------------------------------------------------|---------------------------------------| | PlasmaKinetic Endourology<br>Generator | K003569 | #### Product Description The ArthroCare Bipolar Loop Electrosurgery System is a bipolar, high frequency electrosurgical system consisting of three components: an electrosurgical generator called the Controller; a disposable, bipolar, single use Loop Electrode; and a reusable Patient Cable. {2}------------------------------------------------ #### Intended Uses The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution. Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization. #### Substantial Equivalence Based on the indications for use statement and technological characteristics to the predicate devices, the ArthroCare Bipolar Loop Electrosurgery System has shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act. #### Summary of Safety and Effectiveness The Bipolar Loop Electrosurgery System, as described in this 510(k), is substantially equivalent to the predicate device with respect to the indications for use and technological characteristics. The modification of the indications for use statement does not raise any new issues of safety or efficacy. There are no changes or modifications being made to the technology, principle of operation, materials, packaging and sterilization parameters that were previously cleared 510(k)s. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing. Public Health Service JUN 1 9 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Valerie Defiesta-Ng Director, Regulatory Affairs ArthroCare Corporation 680 Vaqueros Avenue SUNNYVALE CA 94085-2936 Re: K031029 Trade/Device Name: ArthroCare Bipolar Electrosurgery System, Model TBD Regulation Number: 21 CFR §876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: II Product Codes: 78 FAS and KNS Dated: March 31, 2003 Received: May 13, 2003 Dear Ms. Defiesta-Ng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement ArthroCare® Bipolar Loop Electrosurgery System Device Name 510(k) Number: K 03/029 Indications for Use: The ArthroCare Bipolar Loop Electrosurgery System is a bipolar electrosurgical device intended for use in patients requiring endoscopic surgery for general urological procedures. Urological tissue can be resected using this System, such as the prostate, in procedures including transurethral prostatectomy (TURP) and transurethral incisions in the prostate (TUIP), as well as tumors of the bladder wall. The System has been shown to be effective in tissue resection, ablation, and excision, as well as in hemostasis of blood vessels. It is intended for endoscopic procedures using saline solution, Ringer's lactate, or other conductive solutions as irrigants, under direct or video-assisted fiberoptic visualization. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-the-Counter Use (Per 21 CFR 801.109) David R. Bergman (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devi- X 510(k) Number. ix
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...