Surgical indication: The Vortek® Single Loop Ureteral Stent is intended for use in ureterostomy or vesical replacement in adult and pediatric (adolescents, children, and infants requiring endourological procedure and /or short-term (less than 30 days) drainage of the upper urinary tract. Endoscopic indication: The Vortek® Single Loop Ureteral Stent is intended for short-term (less than 30 days) drainage of urine from the upper urinary tract over fistulas or ureteral obstacles in adult and pediatric (adolescents, children, and infants) patients.
Device Story
Vortek Single Loop Ureteral Stent is a self-retaining, radiopaque catheter for short-term (<30 days) upper urinary tract drainage. Device inputs: inserted into ureter/bladder via guidewire to bypass obstacles (calculi/stenosis) or maintain drainage post-intervention. Principle: passive urine drainage from kidney to bladder or external collection bag. Used in clinical settings by physicians; kit includes 90cm stent (6-8 Fr), PTFE-coated guidewire, immobilization clamp, and connectors. Output: physical drainage path. Clinical benefit: maintains patency of ureter, prevents obstruction, facilitates urine flow, reduces risk of hydronephrosis. Provider uses visual markings on stent to guide insertion; output confirms successful placement via radiopacity.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1), mechanical performance (flow rate, tensile strength, curl strength, radiopacity, guidewire compatibility), and shelf-life/sterilization validation (EO, SAL 10^-9).
Indicated for adult and pediatric patients (infants, children, adolescents) requiring endourological procedures, ureterostomy, vesical replacement, or short-term (<30 days) upper urinary tract drainage due to fistulas or ureteral obstacles.
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
February 25, 2021
Coloplast Corp. Angela Kilian Head of Regulatory Affairs, North America 1601 West River Road North Minneapolis, MN 55411
Re: K201436
> Trade/Device Name: Vortek® Single Loop Ureteral Stent Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD Dated: February 22, 2021 Received: January 25, 2021
Dear Angela Kilian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201436
### Device Name
Vortek® Single Loop Ureteral Stent
### Indications for Use (Describe)
Surgical indication:
The Vortek® Single Loop Ureteral Stent is intended for use in ureterostomy or vesical replacement in adult and pediatric (adolescents, children, and infants requiring endourological procedure and /or short-term (less than 30 days) drainage of the upper urinary tract.
Endoscopic indication:
The Vortek® Single Loop Ureteral Stent is intended for short-term (less than 30 days) drainage of urine from the upper urinary tract over fistulas or ureteral obstacles in adult and pediatric (adolescents, children, and infants) patients.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY - K201436
#### I. SUBMITTER INFORMATION:
| 510(K) Owner's Name: | Coloplast A/S |
|-------------------------|------------------------------------------------------------|
| Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Office: 612-422-7956<br>Email: usaby@coloplast.com |
| Name of Contact Person: | Angela Kilian<br>Head of Regulatory Affairs, North America |
| Address/Contact: | 1601 West River Road North<br>Minneapolis, MN 55411 |
| Date Prepared: | February 18, 2021 |
#### II. DEVICE CLASSIFICATION:
| Trade or Proprietary Name: | Vortek® Single Loop Ureteral Stent |
|----------------------------|------------------------------------|
| Common or Usual Name: | Ureteral Stent |
| Classification Name: | Stent, Ureteral |
| Classification Number: | 876.4620 |
| Regulatory Class: | II |
| Product Code: | FAD (stent, ureteral) |
| Advisory Panel: | Gastroenterology/Urology |
#### PREDICATE DEVICE III.
Primary predicate: Hydrogel Coated Percuflex® Drainage Catheters (K924608) Secondary predicate: Vortek® Double Loop Ureteral Stents (K180057)
These predicates have not been subject to a design-related recall.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Vortek Single Loop Ureteral Stents are self-retaining catheters used to maintain drainage of the upper urinary tract following an intervention on the ureter/bladder or in the case of an obstacle such as calculi or stenosis. The devices collect urine from the kidney for short-term (<30 day) drainage and can be connected to a urine bag for collection. It is a prescription only device. Vortek® Single Loop Ureteral Stents are supplied as a kit containing:
- . A 90 cm, radiopaque Single Loop Ureteral Stent with diameters from 6 to 8 Fr. The stents come either with an open or closed distal end. The eyelets or holes for drainage are along the entire loop and length of the device for some models or only on the loop for other models. There are markings along the length of the stent to facilitate insertion of the stent. The stents are made of Vortek Material.
- A Polytetrafluoroethylene (PTFE) coated radiopaque guidewire (Seldinger) with a fixed core and flexible end.
- A clamp for immobilizing the guidewire inserted in the stent during the procedure.
- . Luer connector to connect the Single Loop Stent to the latex connector.
- . A latex connector, for connection to a urine bag conical connector.
#### V. INDICATIONS FOR USE
## Surgical indication
The Vortek® Single Loop Ureteral Stent is intended for use in ureterostomy or vesical replacement in adult and pediatric (adolescents, children, and infants) patients requiring endourological procedure and /or short-term (less than 30 days) drainage of the upper urinary tract.
## Endoscopic indication
The Vortek® Single Loop Ureteral Stent is intended for short-term (less than 30 days) drainage of urine from the upper urinary tract over fistulas or ureteral obstacles in adult and pediatric (adolescents, children, and infants) patients.
The subject and predicate device have the same intended use.
{5}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
| Device & Predicate<br>Device(s): | K201436 | K924608 | K180057 |
|----------------------------------|----------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------|
| Materials | Vortek Material | copolymer | Vortek Material |
| Configuration | Image: One loop | Image: One loop | Image: Two loops |
| Indwelling duration | Less than 30 days | Up to 90 days | Up to 180 days |
| Size (Fr) | 6 to 8 Fr | 6 to 8 Fr | 4.8 to 8.0 Fr |
| Length (cm) | 90 cm | 80 cm | 12- 30 cm |
| Eyes | Eyelets on straight part and<br>loop or only on the loop | Eyelet on loop and<br>straight part | Eyelets on straight part<br>and loop or only on the<br>loop |
| Coating | No | Hydrogel | No |
| Sterile | Yes | Yes | Yes |
| Shelf Life | 5 years | Unknown | 5 years |
| Guidewire | PTFE-coated stainless steel | Unknown | PTFE-coated stainless<br>steel or hydro-coated<br>Nitinol guidewire |
As evidenced by the above table, the subject and predicate devices have different technological characteristics. However, the differences in technological characteristics do not raise different questions of safety or effectiveness.
#### SUMMARY OF TESTING PERFORMED VII.
The sponsor provided the following performance testing to support substantial equivalence:
# Biocompatibility Testing
Biocompatibility testing was conducted based upon ISO 10993-1 (2009): Biological evaluation of medical devices – Part 1: "Evaluation and testing within a risk management process" and FDA Guidance for Use of International Standard ISO 10993-1, "Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process" - Guidance for Industry and Food and Drug Administration Staff - June 16, 2016. The biocompatibility of the Vortek Single Loop Ureteral Stent included:
{6}------------------------------------------------
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- . Acute Systemic toxicity
- Material-mediated pyrogenicity
- Subchronic toxicity
- Genotoxicity
- Implantation ●
- Chemical characterization
- Toxicological Risk evaluation
# Mechanical Testing
The following mechanical testing was completed using the FDA guidance document "Guidance for the content of premarket notifications for ureteral stents" February 1993.
- Flow Rate
- Elongation/Yield and tensile strength ●
- Curl (Loop) Strength ●
- Shelf Life/Expiration date
- Visual and dimensional testing ●
- Guidewire compatibility
- Radiopacity
- Magnetic resonance compatibility ●
## Sterilization
The Vortek Single Loop Ureteral Stents and accessories are sterilized using ethylene oxide in a validated cycle demonstrating a sterility assurance level (SAL) of 10-9.
# Shelf Life
The Vortek Single Loop Ureteral Stent was subjected to package integrity testing and performance verification testing to support the proposed shelf life.
The protocol and results of the provided performance testing to support substantial equivalence are acceptable.
#### VIII. CONCLUSION
The subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.