Malecot Nephrostomy Catheter/Stent Set

K180029 · Cook Incorporated · FAD · Sep 21, 2018 · Gastroenterology, Urology

Device Facts

Record IDK180029
Device NameMalecot Nephrostomy Catheter/Stent Set
ApplicantCook Incorporated
Product CodeFAD · Gastroenterology, Urology
Decision DateSep 21, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4620
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Malecot Nephrostomy Catheter/Stent Set is intended for use as a nephrostomy drainage catheter and ureteral stent. The catheter is placed percutaneously through an existing nephrostomy tract.

Device Story

Device is a combined nephrostomy drainage catheter and ureteral stent; used for percutaneous drainage and stenting via existing nephrostomy tract. Components include radiopaque polyurethane catheter/stent with distal/proximal malecot wing design, flexible stylets, connecting tube, and retention disc. Operated by clinicians in clinical settings. Device facilitates internal/external drainage; provides mechanical support for ureteral patency. Benefits include effective fluid drainage and maintenance of ureteral access. Output is physical drainage of fluids; clinician monitors device position and drainage efficacy.

Clinical Evidence

Bench testing only. Testing included biocompatibility, tensile strength (shaft, hub-to-shaft, malecot, connection), retention strength, flow rate, liquid leakage, radiopacity, component compatibility, MR imaging safety, and shelf-life testing (accelerated aging). All predetermined acceptance criteria were met.

Technological Characteristics

Constructed from radiopaque polyurethane. Features malecot wing design at distal/proximal ends. Stent outer diameter 7 Fr, length 20 cm; catheter outer diameter 12-24 Fr, lengths 25.5-27.4 cm. Supplied sterile for single-use. No electronic components or software.

Indications for Use

Indicated for use as a nephrostomy drainage catheter and ureteral stent in patients requiring percutaneous placement through an existing nephrostomy tract.

Regulatory Classification

Identification

A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 21, 2018 Cook Incorporated Minjin Choi Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404 Re: K180029 > Trade/Device Name: Malecot Nephrostomy Catheter/Stent Set Regulation Number: 21 CFR§ 876.4620 Regulation Name: Ureteral Stent Regulatory Class: II Product Code: FAD, LJE Dated: August 24, 2018 Received: August 27, 2018 Dear Minjin Choi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {2}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Glenn B. Bell -S for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) #### K180029 Device Name Malecot Nephrostomy Catheter/Stent Set #### Indications for Use (Describe) The Malecot Nephrostomy Catheter/Stent Set is intended for use as a nephrostomy drainage catheter and ureteral stent. The catheter is placed percutaneously through an existing nephrostomy tract. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) V Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {4}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is in white, sans-serif font, and is placed at the top of the logo. Below "COOK" is the word "MEDICAL" in white, sans-serif font, set against a dark red background. 510(k) Summary COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM # Malecot Nephrostomy Catheter/Stent Set 21 CFR §807.92 Date Prepared: December 29, 2017 # Submitted By: | Submission: | Traditional 510(k) Premarket Notification | |--------------------|-------------------------------------------| | Applicant: | Cook Incorporated | | Contact: | Minjin Choi | | | Andrew Breidenbach | | Applicant Address: | Cook Incorporated | | | 750 Daniels Way | | | Bloomington, IN 47404 | | Contact Phone: | (812) 339-2235 x104901 | | Contact Fax: | (812) 332-0281 | #### Device Information: | Trade Name: | Malecot Nephrostomy Catheter/Stent Set | |------------------------------------|----------------------------------------| | Common Name: | Stent, Ureteral | | Classification Regulation: | 21 CFR§876.4620, Ureteral Stent | | Product Code: | FAD, LJE | | Device Class/Classification Panel: | Class II, Gastroenterology/Urology | # Predicate Devices: - Primary predicate device: 트 Expel Nephroureteral Drainage Stent with Twist-Loc Hub System (K141344) - י Secondary predicate device: Universa Malecot Drainage Catheter Exchange Set (K140085) {5}------------------------------------------------ Image /page/4/Picture/15 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The word "MEDICAL" is smaller than the word "COOK". ### Device Description: The Malecot Nephrostomy Catheter/Stent Set consists of a catheter/stent, two flexible stylets, a connecting tube, and a retention disc with pull tie. The Malecot catheter/stent is a single device consisting of an externally draining catheter in line with an internally draining ureteral stent. It is constructed from radiopaque polyurethane and has a malecot wing design located on the distal end of the catheter and proximal end of the stent. The stent (distal) portion of the device is available in an outer diameter of 7 Fr with a length of 20 cm. The catheter (proximal) portion of the device is available in an outer diameter between 12 Fr to 24 Fr with lengths of 25.5, 27.0, or 27.4 cm. The set will be supplied sterile and is intended for one-time use. The set is packaged in a peel-open pouch with a three-year shelf life. #### Indications for Use: The Malecot Nephrostomy Catheter/Stent Catheter Set is intended for use as a nephrostomy drainage catheter and ureteral stent. The catheter is placed percutaneously through an existing nephrostomy tract. # Comparison to Predicate Devices: The Malecot Nephrostomy Catheter/Stent Set and the primary predicate device, Expel Nephroureteral Drainage Stent with Twist-Loc Hub System (K141344), are substantially equivalent in that these devices have similar intended uses, methods of operation, and designs. The subject device and secondary predicate device, Universa Percutaneous Drainage Catheter Set (K140085) are also similar in intended uses, methods of operation, and dimensions. The modifications from the predicate devices include: - . Indications for Use - Catheter/Stent Size - 트 Catheter/Stent Retention Configuration - 트 Catheter/Stent Materials - . Stylet {6}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, uppercase letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, uppercase letters on a red background. The word "MEDICAL" is smaller than the word "COOK". Differences between the characteristics of the subject device and the predicate devices are supported by testing. #### Performance Data: The following testing was performed in order to demonstrate that the subject device, Malecot Nephrostomy Catheter/Stent Set, met applicable design requirements. - Biocompatibility - Tensile Strength ● - Catheter and Stent Shaft O - Hub to Shaft Bond of Catheter and Flexible Stylet O - Malecot o - Connection of Catheter to Connecting Tube O - Retention Strength - Catheter to Connecting Tube and Retention Disc O - Flow Rate and Liquid Leakage - 0 Curved and Straight Flow Rate - Radiopacity - Component Compatibility ● - Magnetic Resonance (MR) Imaging Safety - Shelf Life following Accelerated Aging to Three-year Real-time Equivalency ● #### Conclusion: All predetermined acceptance criteria for the testing were met. Therefore, the results of these tests support a conclusion that the Malecot Nephrostomy Catheter/Stent will perform as intended and support a determination of substantial equivalence to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...