UPT Series Medical RO Water Treatment System

K250514 · Specialty Water Technologies, Inc. · FIP · Mar 17, 2025 · Gastroenterology, Urology

Device Facts

Record IDK250514
Device NameUPT Series Medical RO Water Treatment System
ApplicantSpecialty Water Technologies, Inc.
Product CodeFIP · Gastroenterology, Urology
Decision DateMar 17, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5665
Device ClassClass 2

Intended Use

UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorqanic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing. This System is for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System.

Device Story

System removes organic/inorganic contaminants from tap water for hemodialysis applications. Components include pretreatment (sediment filters, carbon filters, softeners), reverse osmosis (membrane separation), and post-treatment/distribution (storage tanks, final filters, pumps, UV disinfection). Used in clinical/hospital settings by staff to supply purified water for dialysate preparation, dialyzer reprocessing, and equipment maintenance. System ensures water meets quality standards for hemodialysis; benefits patient by providing purified water for safe treatment.

Clinical Evidence

No clinical data. Bench testing not required for this labeling update; device remains compliant with existing specifications and performance criteria established by the primary predicate.

Technological Characteristics

System includes pretreatment, reverse osmosis membrane separation, and post-treatment distribution. Components: sediment filtration, carbon removal, water softeners, storage tanks, final filters, delivery pumps, and optional UV disinfection. Operates via membrane separation to remove 95-99% of total dissolved solids and bacteria. Standalone system for clinical/hospital use.

Indications for Use

Indicated for multi-patient use in dialysis clinics or hospitals to purify tap water for hemodialysis dialysate preparation, dialyzer reprocessing, and dialysis equipment disinfection/rinsing.

Regulatory Classification

Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services-USA seal. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below the words "U.S. FOOD & DRUG". March 17, 2025 Specialty Water Technologies, Inc. % Mark Spreeman Regulatory, Quality, and Compliance Consultant DuVal & Associates Suite 1820, Medical Arts Building 825 Nicollet Mall Minneapolis, MN 55402 Re: K250514 Trade/Device Name: UPT Series Medical RO Water Treatment System Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water Purification System For Hemodialysis Regulatory Class: II Product Code: FIP Dated: February 21, 2025 Received: February 21, 2025 Dear Mark Spreeman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ 3 For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Maura Rooney -S Maura Rooney Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use Submission Number (if known) ### K250514 Device Name UPT Series Medical RO Water Treatment System ### Indications for Use (Describe) UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorqanic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing. This System is for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Specialty Water Technologies. The logo features a stylized water droplet in shades of blue, with the company name "SPECIALTY WATER" in bold, dark blue letters below it. Underneath the company name, the word "TECHNOLOGIES" is written in a smaller, lighter font. ## 510(k) Summary | 510(k) Owner: | Specialty Water Technologies, Inc.<br>104 Flex Ave.<br>Portland, TN 37149<br>Contact: Alex Clark<br>615-654-4441 | |----------------|------------------------------------------------------------------------------------------------------------------| | Date prepared: | February 20, 2025 | | Device Name: | Trade Name: UPT Series Medical RO Water Treatment System | | | Common Name: Subsystem, Water Purification | | | Classification Name: Water purification system for hemodialysis | | | Regulation: 21 CFR §876.5665 | | | Regulatory Classification: 2 | Product Code: FIP Predicate Devices: Primary: UPT Series Medical Reverse Osmosis Water Treatment System (151637) Additional: Isopure Complete Water Purification System (K041163) ### Device Description: The UPT Series Medical Reverse Osmosis Water Treatment System and its components consisting of; pretreatment, reverse osmosis, post-treatment, and the product water distribution components, are designed to remove microbiological, organic, and inorganic contaminants from the tap water to supply dialysis machines for the preparation of dialysate solutions for hemodialysis treatments. Pretreatment components can include a tap water boosting system, blending valve, sediment filtration, carbon removal filters, water softeners, and all the necessary interconnecting plumbing. The purpose of this part of the system is to ensure that properly conditioned water is supplied to the reverse osmosis machine to ensure its safe and trouble-free operation. After the tap water has been pre-treated, it then enters the R.O. (reverse osmosis), where total dissolved solids are removed to pertinent water quality standards. This is accomplished by utilizing a membrane separation process, whereby the incoming water is separated into a product stream, and a concentrate stream. The molecular weight cut-off determines what and how many contaminants are passed through into the product stream. R.O.s used for this application typically remove 95-99% of all total dissolved solids and bacteria. {5}------------------------------------------------ The post treatment and product water distribution part of the system is in place to store, provide additional purification if needed, and deliver the purified water to wherever needed. These components can include items such as a storage tank, final filters (for bacteria and endotoxins), and delivery pumps and controls. Some systems can also utilize an ultraviolet light for additional disinfection properties if needed. ### Indications for Use UPT Series Medical Reverse Osmosis Water Treatment System is intended to be used to remove organic and inorganic contaminants from a tap water supply to dilute a dialysate concentrate for hemodialysis treatments, as well as for use for dialyzer reprocessing and dialysis equipment disinfecting and rinsing. This System is for multi patient use, such as for a dialysis clinic or hospital serving multiple patients with a Central Water Purification System. ### Substantial Equivalence Comparison with the Predicate Devices Indications for Use are identical to the primary predicate device with the exception of excluding limitations for certain hemodialysis applications that existed in the predicate. These limitations were inadvertently incorporated into the Indications for Use and have now been corrected. No other changes have been made to the Indications for Use. This is a minor change with no impact on the use of the device, and so does not constitute a new intended use. The technological characteristics between the subject device and both predicates are identical in Regulation Number, Product Code, Pre-Treatment Options, Operational Features, Post-Treatment, and Performance in providing water purification systems for hemodialysis. Hence, there are no changes between the subject device and predicates that would raise different questions of safety or effectiveness. As set out in section 513(i) of the FD&C Act, a determination of substantial equivalence requires: 1) the same intended use, and 2) either the same technological characteristics or differences in technological characteristics that do not raise different questions of safety or effectiveness. Both of these conditions have been met and support a determination of substantial equivalence. ### Non-Clinical Performance Data This modification is to the current UPT Series Medical Reverse Osmosis Water Treatment System (primary predicate), and is specific to clarifying the Indications for Use. No verification or validation performance testing was required for this labeling update and the system remains compliant to its current specifications and requirements. ### Conclusion Inclusive of this labeling update, the UPT Series Medical Reverse Osmosis Water Treatment System continues to meet all of its established acceptance criteria that were based upon performance of the primary predicate device. This demonstrates that the device function is as safe, as effective, and performs as well as the primary predicate device. The information provided supports substantial equivalence of the UPT Series Medical Reverse Osmosis Water Treatment System to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...