ELITONE Urge Urinary Incontinence Device

K223884 · Elidah, Inc. · QAJ · Feb 24, 2023 · Gastroenterology, Urology

Device Facts

Record IDK223884
Device NameELITONE Urge Urinary Incontinence Device
ApplicantElidah, Inc.
Product CodeQAJ · Gastroenterology, Urology
Decision DateFeb 24, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5330
Device ClassClass 2
AttributesTherapeutic

Indications for Use

ELITONE URGE is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues.

Device Story

ELITONE Urge is a wearable, battery-powered muscle stimulator for treating urge urinary incontinence in women. The device consists of a control unit (55mm x 45mm x 10mm) and a disposable, flexible cutaneous electrode applied to the perineal region. The electrode uses mildly adhesive, electrically conductive hydrogel to transfer stimulation to pelvic floor muscles. The user operates the device via two pushbuttons to adjust intensity (0-35 levels, 0-15mA RMS) and control operation (start/pause/stop). The device delivers a symmetric, amplitude-modulated, 10 Hz sinusoidal waveform in 20-minute sessions, recommended 4-5 times per week. The 10 Hz frequency is intended to calm detrusor muscle activity. The device is used at home by the patient. The healthcare provider prescribes the device; the patient self-administers treatment. The output helps reduce urgency symptoms and urge leaks, potentially improving patient quality of life.

Clinical Evidence

Clinical study was a single-arm, prospective study (n=34) of women with urge urinary incontinence. Subjects self-administered treatment for 6 weeks. Primary endpoint: percent change in urge leaks per day. Results: 76% of participants were responders (≥50% reduction in urge leaks). Secondary endpoints: 76% responder rate for urgency episodes, 91% for I-QoL, and 32% for pad use. No adverse events resulting in injury or requiring medical treatment were reported. Supporting evidence included a clinical literature review on perineal-applied EMS for female urinary incontinence.

Technological Characteristics

Non-implanted electrical muscle stimulator. Materials: ABS (controller housing), polyethylene, hydrogel, PET (electrode). Energy: Battery-powered. Waveform: Symmetric, amplitude-modulated, 10 Hz sinusoidal. Dimensions: 55mm x 45mm x 10mm. Connectivity: Standalone. Sterilization: Non-sterile. Software: Embedded, moderate level of concern. Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, ISO 10993-1.

Indications for Use

Indicated for women with urge urinary incontinence. Designed to treat urge urinary incontinence by applying electrical stimulation to pelvic floor muscles and surrounding tissues.

Regulatory Classification

Identification

A transcutaneous electrical continence device consists of cutaneous electrodes that are used to apply external stimulation to reduce urinary incontinence.

Special Controls

In combination with the general controls of the FD&C Act, the transcutaneous electrical continence device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must characterize the electrical stimulation, including the following: Waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density. (2) The patient-contacting materials must be demonstrated to be biocompatible. (3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, thermal safety, and mechanical safety of the device. (4) Software verification, validation, and hazard analysis must be performed. (5) Labeling must include the following: (i) Instructions for use, including specific instructions regarding the proper placement of electrodes; (ii) A summary of electrical stimulation parameters; and (iii) Cleaning instructions and reuse information.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. February 24, 2023 Elidah, Inc. Gloria Kolb CEO 810 Main St., Suite C Monroe, CT 06468 Re: K223884 Trade/Device Name: ELITONE Urge Urinary Incontinence Device Regulation Number: 21 CFR§ 876.5330 Regulation Name: Transcutaneous Electrical Continence Device Regulatory Class: II Product Code: OAJ Dated: December 27, 2022 Received: December 27, 2022 Dear Gloria Kolb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Negeen Haghighi -S for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223884 Device Name ELITONE Urge Urinary Incontinence Device #### Indications for Use (Describe) ELITONE Urge is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------|-----------------------------------------------------------| |------------------------------------------------------------|-----------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Sponsor: | Elidah, Inc.<br>810 Main St, Suite C.<br>Monroe, CT 06468<br>Contact: Gloria Kolb<br>Phone: 978-435-4324<br>Date Prepared: 2/20/23 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Information: | Trade Name: ELITONE Urge Urinary Incontinence Device<br>Common Name: Non-implanted electrical continence device<br>Classification Name: Transcutaneous electrical continence device<br>Classification Product Code: QAJ<br>Regulation Number: 876.5330<br>Device Classification: Class II | | Predicate Device: | ELITONE (K183585), manufactured by Elidah, Inc.<br>The predicate device has not been subject to a design-related recall. | # Device Description: The ELITONE Urge Urinary Incontinence Device (hereafter ELITONE URGE or ELITONE UUI) provides electrical muscle stimulation (commonly called EMS or E-stim) to treat urge urinary incontinence. Stimulation is delivered to the pelvic floor muscles and surrounding structures through a disposable cutaneous electrode applied to the perineal region. One end of the thin, flexible electrode is positioned proximate the pubis and the other end is positioned proximate the ischial tuberosity. Mildly adhesive and electrically conductive hydrogel portions of each electrode to the perineal tissues and allow comfortable transfer of electrical stimulation through the pelvic floor. The stimulation is controlled by a battery powered, wearable control unit. The housing of the control unit is approximately 55mm x 45mm x 10mm. It includes two pushbuttons that allow the user to increment the voltage (i.e., intensity), and to start, pause and stop device operation. The stimulation utilizes a symmetric, amplitude-modulated, sinusoidal waveform. The output intensity operates over a range of levels (0-35) that approximately correspond to a 0-15mA RMS current. The stimulation frequency is 10 Hz, a frequency typically used to calm detrusor muscle activity, decreasing symptoms of urgency type incontinence. Each treatment session lasts 20 minutes, after which the device turns off automatically. Women are recommended to use the device 4-5 times per week. # Intended Use: ELITONE URGE is a non-implanted muscle stimulator designed to treat urge urinary incontinence in women. It applies stimulation to the pelvic floor muscles and surrounding tissues. #### Comparison to the Predicate Device: The table below provides comparisons of the proposed ELITONE UUI device to the predicate ELITONE device (hereafter ELITONE SUI for clarity). The difference between ELITONE UUI and the predicate is the {4}------------------------------------------------ composition of the output signal. Each stimulation cycle of the predicate's output combines a 50Hz signal (understood to treat stress urinary incontinence) and a 10Hz signal (understood to treat urge urinary incontinence), whereas the proposed ELITONE UUI output is solely 10Hz. This difference is achieved through a modification to the embedded software. Product labeling has been changed to reflect the new Indication for Use. All other elements of the two products are identical. Differences in technological characteristics do not raise different questions of safety and effectiveness. | Technological<br>Characteristic | ELITONE UUI Device<br>(New Device) | ELITONE SUI Device<br>(K183585) | Comparison and Impact<br>on Safety/Performance | |---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | Indication for Use | non-implanted muscle<br>stimulator designed to treat<br>urge urinary incontinence in<br>women. It applies stimulation<br>to the pelvic floor muscles<br>and surrounding tissues. | non-implanted muscle<br>stimulator designed to treat<br>stress urinary incontinence<br>in women. It applies<br>stimulation to the pelvic floor<br>muscles and surrounding<br>tissues. | Substantially Equivalent.<br>Both have an intended<br>use for the treatment of<br>urinary incontinence in<br>women. | | Therapeutic modality | Electrical muscle stimulation | Electrical muscle stimulation | Identical | | Rx or OTX | Rx and OTC | Rx and OTC | Identical | | Targeted tissue | Pelvic floor muscles and<br>surrounding structures | Pelvic floor muscles and<br>surrounding structures | Identical | | Anatomic site of<br>stimulation application | Perineal region | Perineal region | Identical | | Device Materials | Controller Housing - ABS<br>Electrode - Polyethylene,<br>Hydrogel, PET | Controller Housing - ABS<br>Electrode - Polyethylene,<br>Hydrogel, PET | Identical | | Sterility Status | Non-sterile | Non-sterile | Identical | | Shelf-Life | 2-Year shelf life on GelPads | 2-Year shelf life on GelPads | Identical | | Number of output<br>modes | 1 | 1 | Identical | | Number of output<br>channels | 1 | 1 | Identical | | Controls | 2 buttons provide all<br>functionality | 2 buttons provide all<br>functionality | Identical | | Compliance with<br>voluntary standards | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60606-1-2 | IEC 60601-1<br>IEC 60601-2-10<br>IEC 60606-1-2 | Identical | | Waveform | Modulated, symmetric<br>(250µs/phase). | Modulated, symmetric<br>(250µs/phase). | Identical | | Stimulation Frequency | 10 Hz<br>Amplitude modulated at<br>2000Hz | 50 Hz, 10 Hz<br>Amplitude modulated at<br>2000Hz | Substantially equivalent:<br>Both utilize frequencies<br>known to treat UUI (10-15<br>Hz). | | Time (Cycle<br>Duration?) | 6s at 10 Hz<br>6s no stimulation | 4s at 50 Hz<br>2s at 10 Hz<br>6s no stimulation | Substantially equivalent:<br>Both utilize contract/relax<br>sequencing consistent<br>with known treatments for<br>incontinence. | | Max output voltage<br>(500Ω) | Vrms = 8V<br>Vpp = 43V | Vrms = 15V at 50Hz, 8V at<br>10Hz<br>Vpp = 43V | Identical to 10Hz portion<br>of predicate | | Technological<br>Characteristic | ELITONE UUI Device<br>(New Device) | ELITONE SUI Device<br>(K183585) | Comparison and Impact<br>on Safety/Performance | | Max output current<br>(500Ω) | Ipp — 97mA<br>Irms -11.4mA | Ipp — 97mA<br>Irms -28.2mA at 50Hz,<br>11.4mA at 10 Hz | Identical to 10Hz portion<br>of predicate | | Max current density | RMS: 0.4 mA/cm²<br>Peak: 3.5 mA/cm² | RMS: 1.0 mA/cm² at 50Hz,<br>0.4mA/cm² at 10Hz<br>Peak: 3.5 mA/cm² | Identical to 10Hz portion<br>of predicate | | Maximum Power<br>Density (W/cm²) | 0.0032 W/cm² | 0.015 W/cm² at 50Hz,<br>0.0032 W/cm² at 10Hz | Identical to 10Hz portion<br>of predicate | | Maximum Phase<br>Charge (µC @<br>500Ohms) | 9.09 µC | 9.09 µC | Identical | {5}------------------------------------------------ # Testing Summary: The following performance testing was provided in support a substantial equivalence determination. - Software Verification and Validation Testing - The software for the device was considered to be a "moderate" level of concern. Verification and validation testing was conducted in accordance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued May 11, 2005. - . Electrical Safety and Electromagnetic Compatibility (EMC) Testing - Testing was completed by external labs per IEC 60601-1 and IEC 60601-2-10 standards for safety and IEC 60601-1-2 standard for EMC. - Other Performance Testing - Performance of the electrode component was assessed by measuring its impedance, its adhesiveness and its current dispersion. Comparisons were made to legally marketed EMS electrodes. - . Biocompatibility Assessment – A biocompatibility evaluation was completed in accordance with the ISO 10993-1 standard. The electrode is considered tissue contacting for a duration of less than 24 hours. The materials, manufacturing processes and use application are the same as those of legally marketed electrodes. Accordingly, no new biocompatibility testing was needed to demonstrate safety and efficacy. - . Clinical Literature Evaluation – A literature review was conducted to demonstrate the safety and efficacy of perineal-applied EMS in the treatment of female urinary incontinence. Articles were reviewed to assess their relevance to the performance and safety of the device. - Clinical Testing – Clinical testing was conducted on women with urge urinary incontinence in the US. The study design was a single-arm study with the patient serving as own control (i.e., comparison to baseline). Subjects self-administered treatment for 6-weeks. The pre-specified primary efficacy endpoint was the percent change in urge leaks per day, determined by comparing 7-day baseline data with data from the 6th week of treatment. Responders were defined as achieving ≥50% reduction in urge leaks. Pre-specified secondary efficacy endpoints included: changes in urgency episodes (i.e., combined number of urge leaks and urgency events that did not result in a leak) with responders defined as achieving ≥50% reduction, changes in pad use with responders defined as achieving ≥50% reduction, and changes in I-QoL score with responders defined as achieving ≥2.5 points change. Subjects with urge incontinence symptoms predominant to stress incontinence symptoms (if present) were randomly assigned to treat with one of two {6}------------------------------------------------ devices, one of which was the subject ELITONE UUI device. 15 subjects were included in analysis of the ELITONE UUI device. Clinical testing was later expanded (under a second IRB-approved protocol) to include a cohort of women with urge urinary incontinence without stress incontinence symptoms. These subjects self-administered treatment for 6-weeks with the ELITONE UUI device according to the same protocol as the first cohort. Pre-specified primary and secondary endpoints aligned with those from the first cohort of subjects. 19 subjects were included in the analysis of this second cohort. The protocol detailing the expanded data collection specifies intent to pool data from the two testing periods. Accordingly, 34 subjects were included in the final analysis. For the primary efficacy endpoint, 76% of participants responded to the treatment. Responder rates for the secondary endpoints were: 76% for urgency episodes, 91% for I-QoL, and 32% for pad use. The primary safety endpoint was defined as adverse events that resulted in an injury or required medical treatment. None were reported. The self-administered, at home treatment limited the opportunity for direct oversight by study personnel, making it relatively easy for subjects to enroll and then subsequently opt-out of participation before initiating treatment. In the first cohort, 26 subjects were shipped the ELITONE UUI device (i.e., enrollment), 15 returned the study device demonstrating at least one treatment session (i.e., treatment), and 15 were found to have completed the study protocol (i.e., analysis). Enrollment, treatment, and analysis values for the second cohort were 30, 23, and 19 respectively. # Risk Management Summary: ELITONE UUI has been designed according to Elidah's internal procedures with traceability between the design inputs, design outputs, verification and validation activities. ELITONE UUI has been evaluated for risks according to Elidah's internal procedures based on ISO 14971. The risks associated with ELITONE UUI were individually and collectively assessed, and the risk/benefit analysis was acceptable. # Conclusion: The non-clinical data demonstrate that ELITONE UUI is as safe as the predicate. The software verification and validation demonstrate that ELITONE UUI performs as intended in the specified use conditions. The clinical literature evaluation supports the use of transcutaneous electrical stimulation as an effective treatment of urge urinary incontinence in women. The clinical study data demonstrates that ELITONE UUI is as safe and effective as the predicate device. Collectively, performance testing supports a Substantial Equivalence determination.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...