K203613 · Shantou Wealy Medical Instrument Co.,Ltd · PNR · Aug 27, 2021 · Gastroenterology, Urology
Device Facts
Record ID
K203613
Device Name
ENFit Reusable Enteral Syringe
Applicant
Shantou Wealy Medical Instrument Co.,Ltd
Product Code
PNR · Gastroenterology, Urology
Decision Date
Aug 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.
Device Story
Piston-style syringe with integral female ENFit connector; sizes 1ml to 60ml. Used for manual dispensing, measuring, and transferring fluids into body orally or enterally. Operates by manual plunger depression; ENFit connector ensures compatibility with enteral access devices to prevent wrong-route administration. Used in non-clinical settings by clinicians or laypersons. Single-patient use; reusable. Benefits include accurate fluid delivery and reduced risk of misconnection via standardized enteral-specific interface.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ISO 80369-3 (small-bore connectors), ISO 7886-1 (manual syringes), and ISO 7886-2 (power-driven syringes). Testing included fluid leakage, sub-atmospheric pressure air leakage, stress cracking, axial load/unscrewing/overriding resistance, plunger force, dead space, and graduated capacity tolerance. Biocompatibility testing per ISO 10993-5, -10, and -11 confirmed no cytotoxicity, irritation, sensitization, or systemic toxicity. Reusability validated via cleaning instructions and use cycle parameters.
Technological Characteristics
Piston-style syringe; barrel with integral female ENFit connector; plunger; ring-type piston. Sizes: 1ml to 60ml. Non-sterile, reusable. Connectivity: Mechanical enteral-specific interface (ISO 80369-3). Materials: Biocompatible (ISO 10993). Standards: ISO 7886-1, ISO 7886-2, ISO 80369-3, ISO 14971.
Indications for Use
Indicated for use as a dispenser, measuring device, and fluid transfer device for oral or enteral fluid delivery. Intended for single-patient, multi-use in non-clinical settings by clinicians and laypersons across all age groups.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Oral/Enteral Syringes with ENFit® connector (12 mL to 60 mL) and Low Dose Tip Oral/Enteral Syringes with ENFit® connector (1 mL to 6 mL) (K183540)
Related Devices
K183540 — Oral/Enteral Syringes with ENFit connector (12 mL to 60 mL), Low Dose Tip Oral/Enteral Syringes with ENFit connector (1 mL to 6 mL) · Neomed, Inc. · Apr 30, 2019
K162188 — Enteral Syringe with ENFit Connector (10 ml to 60 ml), Enteral Syringe with ENFit Low Dose Tip (1 ml to 5 ml) · Vr Medical Technology · Sep 22, 2016
K222155 — O-ring gasket syringe with ENFit connector (single use or reusable) Oral/Enteral syringe with ENFit connector (single use or reusable) · Jiangsu Caina Medical Co.,Ltd · Sep 16, 2022
K211025 — Oral/Enteral Syringe with ENFit connector · Ningbo Tianyi Medical Appliance Co., Ltd. · Jun 30, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
August 27, 2021
Shantou Wealy Medical Instrument Co., Ltd. % Eva Li Consultant Shanghai Sungo Management Consulting Company Limited Room 1309, Dongfang Building, 1500#Century Ave Shanghai, Shanghai 200122 CHINA
Re: K203613
Trade/Device Name: ENFit Reusable Enteral Syringe Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: PNR Dated: July 30, 2021 Received: July 30, 2021
Dear Eva Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general control's provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K203613
Device Name ENFit® Reusable Enteral Syringe
#### Indications for Use (Describe)
The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# 510(K) Summary
# A. Applicant
Shantou Wealy Medical Instrument Co., Ltd. Address: North Jinhuan Road (near of Qishan Mid-school) 515064 Shantou, China Contact Person: Yang Songling Tel: +8615816758600 Fax: 0086-754-82121654
Date Prepared: July 27, 2021
Submission Correspondent Primary contact: Ms. Eva Li Shanghai SUNGO Management Consulting Co., Ltd. Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: eatereva@hotmail.com Secondary contact: Mr. Raymond Luo Room 1309, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com
## B. Device
Trade Name: ENFit® Reusable Enteral Syringe Common Name: Enteral Syringe with Enteral Specific connector Model: 1mlLD, 2.5mlLD, 5ml, 10ml, 20ml, 60ml Product Series Code: WR
Regulatory Information Classification Name: Gastrointestinal tube and accessories Regulator Class: Class II Product code: PNR Regulation Number: 876.5980 Review Panel: Gastroenterology/Urology
## C. Predicate device:
K183540
Oral/Enteral Syringes with ENFit® connector (12 mL to 60 mL) and Low Dose Tip Oral/Enteral Syringes with ENFit® connector (1 mL to 6 mL) NeoMed, Inc.
## D. Intended use of the device:
{4}------------------------------------------------
North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only.
# E. Device Description:
The proposed device with Enfit® connector (1ml-60ml) are standard piston style syringes consisting of syringe barrel with integral EnFit syringe tip, syringe plunger and ring type piston. The proposed device are supplied non-sterile, and reusable. The sizes rang from 1ml to 60ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with enteral access devices or accessories having ENFit compliant or compatible male connectors to form a dedicated system that prevents wrong-route administration of fluids.
# F. Comparison with predicate device
| Device | Proposed Device | Predicate Device | Comparison |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Shantou Wealy Medical Instrument Co., Ltd. | NeoMed, Inc. | -- |
| 510(K) number | K203613 | K183540 | -- |
| Product Code | PNR | PNR | Same |
| Regulation Number | CFR 876.5980 | CFR 876.5980 | Same |
| Intend use | The ENFit® Reusable Enteral Syringe is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only. | The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used multiple times in non-clinical settings by users ranging from clinicians to laypersons in all age groups. The device is indicated for single patient use only. | Same |
| Configuration | Barrel with ENFit connector<br>Purple Plunger<br>Piston | Barrel with ENFit connector<br>Purple Plunger<br>Piston | Same |
| Size | Low dose tip<br>ENFit™ syringe<br>1ml, 2.5ml, | Low dose tip ENFit™ 1ml-6ml | Similar* |
Table 1 General Comparison
{5}------------------------------------------------
| | Standard ENFit™<br>syringe | 5ml-<br>60ml | Standard ENFit™<br>syringe | 12ml-60ml | |
|------------------|---------------------------------|--------------|---------------------------------|-----------|------|
| Sterile | No | | No | | Same |
| Reusable | Yes | | Yes | | Same |
| Biocompatibility | No Cytotoxicity | | No Cytotoxicity | | Same |
| | No Irritation and Sensitization | | No Irritation and Sensitization | | |
| | No Acute Toxicity | | No Acute Toxicity | | |
#### *Similar discussion
The size of the proposed device of the low dose tip syringe or standard syringe is similar, all the size of the proposed are pass the acceptance criteria of the related performance standard requirements, so it will not affect the safety and effectiveness of the proposed device.
## G. Non-Clinical Test Conclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The following performance tests are conducted:
- A Finished Device
- Risk management report in accordance with ISO 14971:2007 Medical devices -● Application of risk management to medical devices
- . Summary of the Usability testing for ENfit Reusable Enteral Syringe(human factor and usability validation)
- . Performance testing with ISO 7886-2:2020 "Sterile hypodermic syringes for single use Part 2: Syringes for use with power-driven syringe pumps
- ——Appearance
- -- Pump Force
- -- Short-term Flow Rate Error
- -- Syringe Compliance
- ——Syringe Design critical dimension
- A Reusability
- . Cleaning Instructions Validation and Use Cycle Parameters Study
- > Biocompatibility
- ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity;
- ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization;
- . ISO 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
- A Shelf life
- Package verification test report 0
- . Bench test after real-time aging(ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare application-Part3: Connectors for enteral applications ) ——Fluid Leakage
{6}------------------------------------------------
#### K203613 Page 4 of 5
- North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China
- -- Sub-atmospheric Pressure Air Leakage
- -- Stress Cracking
- -- Resistance to Separation from Axial Load
- -- Resistance to Separation from Unscrewing
- -- Resistance to Overriding
- -- Disconnection by Unscrewing
- . Bench test after real-time aging(ISO 7886-1:2017 Sterile hypodermic syringe for single use-Part 1: Syringe for manual use )
- -- Appearance
- ——Overall Length of Scale to Nominal Capacity Line
- -- Push-button Distance
- -- Force Required to Operate Plunger
- ——Freedom from Air and Liquid Leakage past Piston
- -- Limits for Acidity or Alkalinity
- ——Limits for Extractable Metals
- -- Tolerance on Graducated Capacity
- ——Maximum Dead Space
- Finished Device performance Test
- Critical Dimension verification
- Ink Adhesion
>
- . Bench test (ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare application-Part3: Connectors for enteral applications )
- --Fluid Leakage
- -- Sub-atmospheric Pressure Air Leakage
- -- Stress Cracking
- ——Resistance to Separation from Axial Load
- -- Resistance to Separation from Unscrewing
- -- Resistance to Overriding
- -- Disconnection by Unscrewing
- . Bench test(ISO 7886-1:2017 Sterile hypodermic syringe for single use-Part 1: Syringe for manual use )
- ——Appearance
- ——Overall Length of Scale to Nominal Capacity Line
- ——Push-button Distance
- -- Force Required to Operate Plunger
- ——Freedom from Air and Liquid Leakage past Piston
- -- Limits for Acidity or Alkalinity
- ——Limits for Extractable Metals
- ——Tolerance on Graduated Capacity
- ——Maximum Dead Space
- ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# H. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
# North Jinhuan Road(near Qishan mid-school), 515064 Shantou, China
# I. Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Based on the nonclinical tests performed, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device, Oral/Enteral Syringes with ENFit® connector (12 mL to 60 mL) and Low Dose Tip Oral/Enteral Syringes with ENFit® connector (1 mL to 6 mL) cleared under K183540.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.