K193269 · Neotract, Inc. · PEW · Dec 20, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K193269
Device Name
UroLift System (UL400)
Applicant
Neotract, Inc.
Product Code
PEW · Gastroenterology, Urology
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5530
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.
Device Story
UroLift System (UL400) is an implantable transprostatic tissue retractor system used to treat BPH-related urinary obstruction. The system consists of a single-use delivery device pre-loaded with a UroLift implant. The physician inserts the delivery device through the penile orifice into the urethra; the device displaces the urethra toward the prostatic capsule and deploys the implant transversely through the prostatic tissue. Multiple implants (typically 2-6) are placed to retract the obstruction, securing the inner urethral tissue to the outer capsular tissue. This maintains an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). The procedure is performed in a clinical setting. The device modification in this submission involves increasing the contraindication threshold for prostate volume from >80 cc to >100 cc, supported by clinical literature review.
Clinical Evidence
Clinical literature review of sponsored and independent studies comparing patients with prostate volumes >80 cc to those <80 cc. Results indicate equivalent symptom response, quality of life, uroflowmetry, adverse events, and catheterization rates. No new clinical testing was performed for this submission; reliance on predicate data and literature.
Technological Characteristics
Implantable transprostatic tissue retractor system. Biocompatibility per ISO 10993-1. Single-use delivery device with pre-loaded implant. Mechanical deployment mechanism. No software or electronic components.
Indications for Use
Indicated for men 45+ with urinary outflow obstruction secondary to BPH (lateral/median lobe hyperplasia). Contraindicated for prostate volume >100 cc, UTI, urethral conditions preventing insertion, incompetent sphincter, or gross hematuria.
Regulatory Classification
Identification
An implantable transprostatic tissue retractor system is a prescription use device that consists of a delivery device and implant. The delivery device is inserted transurethrally and deploys the implant through the prostate. It is designed to increase prostatic urethral patency by providing prostate lobe tissue retraction while preserving the potential for future prostate procedures and is intended for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia in men.
Special Controls
In combination with the general controls of the FD&C Act, the UroLift® System is subject to the following special controls:
- (1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
- (2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
- (3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity and device functionality over the requested shelf life.
- (4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
- (A) Deployment testing must be conducted
- (B) Mechanical strength must be conducted
- (C) Resistance-to-degradation testing must be conducted
- (5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment
- (6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of
- use. The following performance characteristics must be tested:
- (A) Deployment testing must be conducted
- (B) Implant migration must be conducted
- (7) Labeling must bear all information required for safe and effective use of the device, and must include:
- (A) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device
- (B) Information on the patient population for which the device has been demonstrated to be effective
- (C) A detailed summary of the device technical parameters
- (D) Information on how the device operates and the typical course of treatment
- (E) An expiration date/shelf life
- (F) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.
*Classification.* Class II (special controls). The special controls for this device are:(1) The elements of the device that may contact the patient must be demonstrated to be biocompatible.
(2) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(3) Performance data must support shelf life by demonstrating continued sterility of the device (of the patient-contacting components), package integrity, and device functionality over the requested shelf life.
(4) Non-clinical testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Mechanical strength must be conducted.
(iii) Resistance-to-degradation testing must be conducted.
(5) Non-clinical testing must evaluate the compatibility of the device in a magnetic resonance environment.
(6) In vivo testing must demonstrate safe and effective use, assess the impact of the implants on the ability to perform subsequent treatments, document the adverse event profile associated with clinical use, and demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Deployment testing must be conducted.
(ii) Implant migration must be conducted.
(7) Labeling must bear all information required for safe and effective use of the device, and must include:
(i) Specific instructions, warnings, cautions, limitations, and the clinical training needed for the safe use of the device.
(ii) Information on the patient population for which the device has been demonstrated to be effective.
(iii) A detailed summary of the device technical parameters.
(iv) Information on how the device operates and the typical course of treatment.
(v) An expiration date/shelf life.
(vi) A detailed summary of the device- and procedure-related complications or adverse events pertinent to use of the device.
K162345 — UroLift System UL500, UroLift Delivery Handle, UroLift Implant Cartiridge · Neotract, Inc. · Sep 21, 2016
K172359 — UroLift System (UL500) · Neotract, Inc. · Aug 18, 2017
K173087 — UroLift System (UL400 and UL500) · Neotract, Inc. · Dec 28, 2017
K200441 — UroLift Advanced Tissue Control (ATC) System · Neotract, Inc. · Jun 5, 2020
K190377 — UroLift System UL400 · Neotract, Inc. · Mar 21, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
December 20, 2019
Neotract, Inc. Brian Gall Regulatory Affairs Manager 4155 Hopyard Rd. Pleasanton, CA 94588
Re: K193269
Trade/Device Name: UroLift System (UL400) Regulation Number: 21 CFR 876.5530 Regulation Name: Implantable Transprostatic Tissue Retractor System Regulatory Class: Class II Product Code: PEW Dated: November 25, 2019 Received: November 26, 2019
Dear Brian Gall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193269
Device Name UroLift System (UL400)
Indications for Use (Describe)
The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 03 510(k) SUMMARY
#### COMPANY INFORMATION
NeoTract, Inc. 4155 Hopyard Rd. Pleasanton, CA 94588 Registration Number: 3015181082
#### SUBMISSION CORRESPONDENT
Brian Gall Requlatory Affairs Manager, Interventional Urology NeoTract, Inc. 4155 Hopvard Rd. Pleasanton, CA 94588
Telephone - 925.329.6547 E-mail - brian.gall@teleflex.com
### DATE PREPARED
25 November 2019
#### DEVICE INFORMATION
| Trade Name: | NeoTract® UroLift® System (UL400) |
|-----------------------|----------------------------------------------------|
| Common Name: | Implantable transprostatic tissue retractor system |
| Classification Name: | Implantable transprostatic tissue retractor system |
| Product Code: | PEW |
| Regulation Number: | 876.5530 |
| Classification: | II |
| Classification Panel: | Gastroenterology/Urology |
### DEVICE DESCRIPTION
The UroLift System is designed to access the prostatic urethra and deliver one UroLift Implant through a lobe of the prostate. The UroLift System is inserted into the urethra through the penile orifice and used to displace the urethra toward the prostatic capsule. The UroLift Implant is then deployed transversely through the prostatic tissue. Multiple implants are deployed in the UroLift System procedure. The implants secure the retracted position of the urethra, thereby maintaining an expanded urethral lumen, reducing fluid obstruction and improving lower urinary tract symptoms (LUTS). This is accomplished by holding the approximated position of the inner (urethral) tissue and the outer (capsular) tissue of the prostate with the UroLift Implant. The procedure typically requires 2-6 implants to retract the obstruction. The UL400 (most recently cleared in K173087), consists of two main components, the UroLift Delivery Device (single use), and the UroLift Implants (one implant per delivery device). Each Delivery Device comes pre-loaded with one UroLift Implant.
Image /page/3/Picture/14 description: The image features the logos of NeoTract and Teleflex, two companies in the interventional urology field. The NeoTract logo is on the left, while the Teleflex logo is on the right. Below the logos, the words "INTERVENTIONAL UROLOGY" are written in a smaller font, indicating the area of specialization for these companies.
{4}------------------------------------------------
## INDICATIONS FOR USE
The UroLift System is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH), including lateral and median lobe hyperplasia, in men 45 years of age or older.
### CONTRAINDICATIONS
The UroLift System should not be used if the patient has:
- Prostate volume of >100 cc
- A urinary tract infection
- Urethra conditions that may prevent insertion of delivery system into bladder
- · Urinary incontinence due to incompetent sphincter
- · Current gross hematuria
# PREDICATE DEVICE
The predicate device is the UroLift System by NeoTract (K173087).
| Trade Name: | NeoTract UroLift System (UL400) |
|-----------------------|----------------------------------------------------|
| Common Name: | Implantable transprostatic tissue retractor system |
| Classification Name: | Implantable transprostatic tissue retractor system |
| Product Code: | PEW |
| Regulation Number: | 876.5530 |
| Classification: | II |
| Classification Panel: | Gastroenterology/Urology |
### COMPARISON WITH THE PREDICATE DEVICE
The UroLift System (UL400) described in this submission is substantially equivalent to the previously cleared generations of the device. The UL400 was previously cleared in K173087. This submission concerns the modification of a contraindication of the device and does not impact the device itself. The modification is to change one contraindication from "The UroLift System is contraindicated for men with Prostate volume of >80 cc" to "The UroLift System is contraindicated for men with Prostate volume of >100 cc". This is based on a clinical literature review. The indications for use and remaining contraindications do not change as a result of this submission. Minor device modifications which were determined to not require a pre-market submission based on Deciding When to Submit a 510(k) for a Change to an Existing Device Guidance for Industry and Food and Drug Administration Staff are included in this submission.
### PERFORMANCE TESTING
The modification of the contraindication does not impact the performance testing of the existing UroLift System. As such, the performance testing data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication. The minor device modifications discussed were tested using test methods equivalent to the predicate device.
The basis for increasing the maximum prostate volume indicated in the contraindication for the UroLift System is based on clinical review of both sponsored and independent clinical studies that included men with prostate volumes greater than 80cc. These studies show that the symptom response, quality of life, uroflowmetry, adverse events,
Image /page/4/Picture/19 description: The image features the logos of two companies, "neotract" and "Teleflex", along with the text "INTERVENTIONAL UROLOGY". The "neotract" logo is on the left, with the company name in blue and black. The "Teleflex" logo is on the right, with the company name in blue. The text "INTERVENTIONAL UROLOGY" is centered below the two logos.
{5}------------------------------------------------
and catheterization rates are equivalent to the outcomes of patients with prostate volumes less than 80cc.
# BIOCOMPATIBILITY TESTING
The UroLift System has been tested for biocompatibility and passed the relevant tests according to ISO 10993-1: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk manaqement process. The modification of the contraindication does not impact the biocompatibility of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication.
# STERILIZATION AND SHELF-LIFE TESTING
The modification of the contraindication does not impact the sterilization or shelf life of the existing UroLift System. As such, the data provided with the predicate device (K173087) is applicable for the proposed device with a modified contraindication.
# CONCLUSION
The data provided demonstrated the NeoTract UroLift System with a modified contraindication is as safe and effective, has the same intended use, technological characteristics and principles of operation as the predicate device. Therefore, the NeoTract UroLift System is substantially equivalent to the predicate devices.
Image /page/5/Picture/9 description: The image features the logos of two companies, "NeoTract" and "Teleflex", side by side. The "NeoTract" logo is on the left, while the "Teleflex" logo is on the right, both in a dark blue color. Below the logos, the text "INTERVENTIONAL UROLOGY" is displayed in a smaller font size, indicating the field of expertise or focus of these companies.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.