HydroPearl Microspheres

K192684 · MicroVention, Inc. · NOY · Jan 22, 2020 · Gastroenterology, Urology

Device Facts

Record IDK192684
Device NameHydroPearl Microspheres
ApplicantMicroVention, Inc.
Product CodeNOY · Gastroenterology, Urology
Decision DateJan 22, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5550
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HydroPearl® Microspheres are intended for the embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH).

Device Story

HydroPearl Microspheres are pre-formed, compressible, spherical embolic agents made of biocompatible hydrogel (polyethylene glycol diacrylamide and glycerol monomethacrylate). Provided in sterile, pre-filled syringes with phosphate buffered saline. Delivered to treatment sites via catheter under radiographic imaging; mixed with contrast media for radiopacity. Used by physicians in clinical settings for embolization of arteriovenous malformations, hypervascular tumors, and prostatic arteries for BPH. Embolization blocks blood flow to target tissues, providing therapeutic benefit for hypervascular conditions and symptomatic BPH. Device is single-use.

Clinical Evidence

Retrospective analysis of 17 patients (mean age 70) undergoing PAE. Primary endpoint: reduction in International Prostate Symptom Score (IPSS). Mean IPSS improved from 20.7 to 9.1 (p<0.05); mean QoL score improved from 3.7 to 1.2 (p<0.05). Technical success (bilateral embolization) achieved in 100% of patients; clinical success in 94%. One major complication (acute prostatitis) and two minor complications (hematuria, hematospermia) reported. Literature review supports safety/efficacy of microparticle embolic agents for PAE.

Technological Characteristics

Spherical hydrogel microspheres (polyethylene glycol diacrylamide and glycerol monomethacrylate). Diameters 75-1100µm. Supplied in sterile, pre-filled 20ml polycarbonate syringes with phosphate buffered saline. Radiopacity achieved via mixing with contrast media. Steam sterilized. Single-use.

Indications for Use

Indicated for embolization of arteriovenous malformations, hypervascular tumors (including uterine fibroids), and prostatic arteries for symptomatic benign prostatic hyperplasia (BPH).

Regulatory Classification

Identification

A prostatic artery embolization device is an intravascular implant intended to occlude the prostatic arteries to prevent blood flow to the targeted area of the prostate, resulting in a reduction of lower urinary tract symptoms related to benign prostatic hyperplasia. This does not include cyanoacrylates and other embolic agents which act by in situ polymerization or precipitation, or embolization devices used in neurovascular applications (see 21 CFR 882.5950).

Special Controls

In combination with the general controls of the FD&C Act. the prostatic artery embolization device is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) The device must be demonstrated to be biocompatible. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Evaluation of suitability for injection through catheters intended for use in embolization; and (ii) Evaluation of the size distribution of the device. (3) Performance data must support the sterility and pyrogenicity of the device. (4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (5) Clinical data must evaluate post-embolization damage due to non-target embolization under anticipated use conditions. (6) The labeling must include: (i) Specific instructions on safe device preparation and use; (ii) The device shelf life; (iii) Data regarding urinary retention; and (iv) Data regarding post-prostatic artery embolization syndrome.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 22, 2020 MicroVention, Inc. Ms. Analia Staubly Sr. Project Manager, Regulatory Affairs 35 Enterprise Aliso Viejo, California 92656 Re: K192684 Trade/Device Name: HydroPearl Microspheres Regulation Number: 21 CFR 876.5550 Regulation Name: Prostatic Artery Embolization Device Regulatory Class: Class II Product Code: NOY, KRD, NAJ, Dated: September 25, 2019 Received: September 26, 2019 Dear Ms. Staubly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Angel Soler-Garcia, Ph.D. Acting Assistant Director Incontinence and Male Urological Devices Team (THT3B3) DHT3B: Division of Reproductive, Gynecological, and Urological Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K192684 Device Name HYDROPEARL® Microspheres Indications for Use (Describe) The HydroPearl® Microspheres are intended for the embolization of arteriovenous malformations and hypervascular tumors, including uterine fibroids, and for embolization of prostatic arteries (PAE) for symptomatic benign prostatic hyperplasia (BPH). | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-family: sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |----------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-family: sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | CONTINUE ON A SEPARATE PAGE IF NEEDED # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | TRADE NAME: | HydroPearl <sup>TM</sup> Microspheres | |---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | COMMON NAME: | Embolization device | | REGULATION NAMES: | Prostatic artery embolization device<br>Vascular embolization device | | CLASSIFICATION: | II, 21 CFR 876.5550 (NOY)<br>II, 21 CFR 870.3300 (KRD/NAJ) | | PRODUCT CODE: | NOY/KRD/NAJ | | APPLICANT: | MicroVention, Inc.<br>35 Enterprise<br>Aliso Viejo, CA 92656 | | ESTABLISHMENT<br>REGISTRATION NUMBER: | 3013556777 | | CONTACT: | Analia Staubly<br>Sr. Project Manager, Regulatory Affairs<br><a href="mailto:analia.staubly@microvention.com">analia.staubly@microvention.com</a> | | DATE SUMMARY PREPARED: | September 25, 2019 | | PREDICATE DEVICE: | BioSphere Medical S.A.<br>Embosphere Microspheres (DEN160040) | | REFERENCE DEVICE: | HydroPearl <sup>TM</sup> Microspheres (K150870) | | DEVICE DESCRIPTION: | The HydroPearl <sup>TM</sup> Microspheres are a pre-formed,<br>compressible, precisely calibrated, spherical embolic<br>agent consisting of a biocompatible hydrogel. The<br>HydroPearl <sup>TM</sup> Microspheres are offered in a<br>variety of diameters ranging from 75-1100µm and are<br>provided in a sterile syringe pre-filled with microspheres<br>in phosphate buffered saline. The pre- filled syringe is<br>packaged in a sealed sterile dispenser tray. The<br>HydroPearl <sup>TM</sup> Microspheres are delivered to the<br>treatment site through a delivery catheter. | | INDICATIONS FOR USE: | The HydroPearl <sup>TM</sup> Microspheres are intended for the<br>embolization of arteriovenous malformations and<br>hypervascular tumors, including uterine fibroids, and<br>for embolization of prostatic arteries (PAE) for<br>symptomatic benign prostatic hyperplasia (BPH). | {4}------------------------------------------------ # TECHNOLOGICAL CHARACTERISTICS AND PRODUCT FEATURE COMPARISON: | Characteristics | Predicate Device:<br>Embosphere Microspheres<br>(DEN160040) | Reference Device:<br>HydroPearl Microspheres<br>(K150870) | Subject Device:<br>HydroPearl<br>Microspheres | |--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------| | Intended Use | Indicated for use in embolization<br>of arteriovenous malformations,<br>hypervascular tumors, including<br>symptomatic uterine fibroids and<br>prostatic arteries for<br>symptomatic benign prostatic<br>hyperplasia (BPH). | Intended for the<br>embolization of<br>arteriovenous<br>malformations and<br>hypervascular tumors,<br>including uterine fibroids. | Same as<br>Predicate Device | | TECHNICAL | | | | | Microsphere<br>Material | Acrylic polymer impregnated<br>with porcine delivered gelatin. | Polyethylene glycol<br>diacrylamide and glycerol<br>monomethacrylate | Same as<br>Reference<br>Device | | Microsphere<br>Diameter | Size Range: 40 – 1200 $\mu$ m<br><br>Labeled size range:<br>40 - 120 $\mu$ m<br>100 - 300 $\mu$ m<br>300 - 500 $\mu$ m<br>500 - 700 $\mu$ m<br>700 - 900 $\mu$ m<br>900 - 1200 $\mu$ m | Size Range: 75 – 1100 $\mu$ m<br><br>Labeled size:<br>75 $\mu$ m<br>200 $\mu$ m<br>400 $\mu$ m<br>600 $\mu$ m<br>800 $\mu$ m<br>1100 $\mu$ m | Same as<br>Reference<br>Device | | Microsphere<br>Container | Contained in a sterile, 20 ml<br>polycarbonate syringe. | Same as Predicate Device | Same as<br>Predicate Device | | Microsphere<br>volume per<br>syringe | 1.0 or 2.0 ml, in physiological<br>saline. | 2.0 ml in phosphate<br>buffered saline (PBS) | Same as<br>Reference<br>Device | | Radiopacity<br>Method | Mixed with contrast media prior<br>to injection. | Same as Predicate Device | Same as<br>Predicate Device | | Delivery Method | Via catheter under radiographic<br>imaging. | Same as Predicate Device | Same as<br>Predicate Device | | STERILIZATION AND PACKAGING | | | | | Sterilization<br>Method. | Steam sterilization. | Same as Predicate Device | Same as<br>Predicate Device | | Packaging | Pre-filled syringe, placed in a<br>polycarbonate tray and sealed<br>with a peel away Tyvek lid,<br>placed inside bleached sulfate<br>carton box. | Same as Predicate Device | Same as<br>Predicate Device | | Method of<br>Supply | Sterile and single use. | Same as Predicate Device | Same as<br>Predicate Device | {5}------------------------------------------------ # VERIFICATION AND TEST SUMMARY: Non-clinical testing was performed on the HydroPearl Microspheres, reference device, to provide reasonable assurance that the proposed device has been designed and tested to assure conformance to special requirements for its intended use. A risk assessment has been performed and demonstrates that the expanded indication did not require additional bench, biocompatibility, packaging or sterilization. All existing testing in the previous HydroPearl Microsphere premarket submission remains applicable. Comparative mechanical testing was conducted to compare the performance of the HydroPearl Microspheres to that of legally marketed devices for the same indication. The HydroPearl Microspheres performed equivalently to the Embosphere Microspheres, predicate device, in comparative mechanical testing. Pre-clinical animal testing was also performed in animal models representing the renal and prostatic vasculature. The HydroPearl performed similarly to the Embosphere Microspheres and had favorable embolization results in both studies. ## CLINICAL INFORMATION: Existing real-world data from a single, US center was analyzed retrospectively for the purpose of expanding the existing indication for the HydroPearl Microspheres to include embolization of prostatic artery embolization (PAE) for symptomatic benign prostatic hyperplasia (BPH). The retrospective analysis consisted of 17 patients that underwent prostatic artery embolization (PAE) using HydroPearl Microspheres for the treatment of symptomatic benign prostatic hyperplasia (BPH). The primary objective was to collect data of the procedure safety and efficacy, which included age, prostate volume baseline International Prostate Symptom Score + Quality of Life score (IPSS+QoL), unilateral embolization, size of particles used, complications and follow-up IPSS+OoL. The primary endpoint was reduction in IPSS score during the follow-up period. ### Demographics and Baseline Characteristics: Twenty-three patients were identified. 17 patients returned consent forms and their data was included in this study. The mean age of the patients included was 70 (range; 55-85) and their mean prostate volume was 63 ml (range; 27-140). One of the patients relied on clean intermittent catheterization (CIC) for voiding and the rest were able to void without a catheter. Fourteen of seventeen patients were on BPH medications at the time of PAE. All 17 patients underwent bilateral prostatic artery embolization was performed with either 75 um HydroPearl (n=2), 200 um HydroPearl (n=3) or 75 um followed by 200 um HydroPearl (n=12). ### Primary Endpoint Results: Mean follow up period was 8.6 months (range; 1-12). Mean baseline International Prostate Symptom Score was 20.7 (range; 4-30) at baseline and 9.1 (range; 2-18) at latest follow-up (p<0.05). Mean Quality of Life Score was 3.7 (range; 1-6) at baseline and 1.2 (range; 0-3) at follow-up (p<0.05). The single patient who relied on CIC for voiding catheterized 5 times a day before PAE and 0 or 1 time a day after PAE. Seven of the 14 patients on BPH medications at {6}------------------------------------------------ baseline no longer took these medications at their latest follow-up and two additional patients were able to reduce their daily medications. #### Technical and Clinical Success Results : Technical success, defined as bilateral prostate artery embolization, was achieved in all 17 patients. Clinical success was achieved in 16/17 patients (94%). The single patient that was considered a clinical failure underwent PAE with the goal of no longer requiring BPH medication after the procedure because of the side-effects that he experienced. His LUTS were well controlled on the medication prior to PAE (IPSS of 4). Post-PAE, he was able to cease taking medication, but his symptoms did mildly worsen (IPSS of 10). #### Safety Results: One major complication was reported that involved hospitalization of one subject 2 days after PAE to treat acute prostatitis. The patient was discharged after a 4-day admission without any residual effects. Additional minor complications that were reported included hematuria (n=1) and hematospermia (n=1). These two adverse events were likely the result of mild non-target embolization and are expected complications of PAE. #### Study Conclusion: PAE performed with HydroPearl Microspheres resulted in clinical success in nearly all the patients included in this analysis. The single exception was a patient who already had his LUTS well controlled on alpha-blocker medication. but who desired to no longer take this medication, which was accomplished after PAE. Additionally, half of the patients who presented at baseline on BPH medication, no longer required them after PAE. In addition to the retrospective analysis, MicroVention Inc. conducted a comprehensive review of relevant clinical literature to assess whether any issues of safety or effectiveness were raised using microparticle embolic devices for prostatic artery embolization (PAE) for symptomatic benign hyperplasia (BPH). The literature review concludes that PAE is a safe, minimally invasive, and efficacious procedure with low rate of adverse events, particularly those related to the embolic agent. ### CONCLUSION: Based on the indications for use, technological characteristics, safety and performance testing, and clinical evidence, the HydroPearl Microspheres have been shown to support safe effective use in humans for prostatic artery embolization for symptomatic benign hyperplasia (BPH) and are substantially equivalent to Embosphere Microspheres The HydroPearl Microspheres demonstrates compliance with the Special Controls under 21 CFR 876.5550 for the expanded indication for use and 21 CFR 870.3300.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...