TRANSCOM COLON HYDROTHERAPY MODEL HC-2000

K970482 · Srs Intl. · KPL · Oct 31, 1997 · Gastroenterology, Urology

Device Facts

Record IDK970482
Device NameTRANSCOM COLON HYDROTHERAPY MODEL HC-2000
ApplicantSrs Intl.
Product CodeKPL · Gastroenterology, Urology
Decision DateOct 31, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5220
Device ClassClass 2
AttributesTherapeutic

Intended Use

The indication for use of this device must be restricted to colon cleansing when medically indicated, such as before radiological or endoscopic examination.

Device Story

Transcom Colon Hydrotherapy (Model HC-2000) is a wall-mounted instrument for colonic irrigation. It introduces filtered water at controlled temperatures into the large intestine to cleanse contents. System components include a mixing valve, pressure regulator, water manifold, relief valve, monitoring system (temperature/pressure), viewing assembly, sanitation system, and lighting assembly. Operation is manual, utilizing a syringe-based flow mechanism and gravity drainage. The device features automatic pressure safety, disinfection, and temperature regulation. It includes disposable components for single-use therapy. Used in clinical settings by healthcare providers to prepare patients for medical procedures. Output consists of visual monitoring of fluid discharge via a view tube assembly, allowing the clinician to observe the cleansing process. Benefits include hygienic, painless, and odorless colon preparation.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use.

Technological Characteristics

Wall-mounted colonic irrigator. Materials include brass and stainless steel. Features include thermostatic mixing valve (Grohemix), Norgren pressure regulators and relief valves, and manual flow control. Operates on 110/120 VAC 50/60 Hz. Includes automatic pressure safety, disinfection system, and temperature monitoring. Fluid pathway utilizes disposable components except for the view tube assembly. Dimensions: 1020 x 720 x 270 CM. Weight: 132.3 lbs.

Indications for Use

Indicated for colon cleansing in patients requiring bowel preparation for radiological or endoscopic examination.

Regulatory Classification

Identification

A colonic irrigation system is a device intended to instill water into the colon through a nozzle inserted into the rectum to cleanse (evacuate) the contents of the lower colon. The system is designed to allow evacuation of the contents of the colon during the administration of the colonic irrigation. The device consists of a container for fluid connected to the nozzle via tubing and includes a system which enables the pressure, temperature, or flow of water through the nozzle to be controlled. The device may include a console-type toilet and necessary fittings to allow the device to be connected to water and sewer pipes. The device may use electrical power to heat the water. The device does not include the enema kit (§ 876.5210).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} 10f6 41 # SRS INTERNATIONAL CORPORATION Suite 1000 • 1625 K Street, NW • Washington, DC 20006-1604 Telephone (202) 223-0157/0298 • Telecopy (202) 835-8970 OCT 31 1997 ## PREMARKET NOTIFICATION [510(K)] SUMMARY ### 1. Name/Address/Phone/Fax, Contact Person, Date: Submitter's Name: SRS International Corporation Address: 1625 K St., NW, Suite 1000 Washington, DC 20006 Telephone: 202-223-1057 Fax: 202-835-8970 Contact Person: Michael G. Farrow, Ph.D. Date of Summary: February 8, 1997 ### 2. Name of Device, Proprietary Name, Common Name, Classification Name of Device: Transcom Colon Hydrotherapy Proprietary Name: Transcom Colon Hydrotherapy Common Name: Colonic Irrigation System Classification: Irrigator, Colonic (Gastro/Urology) 78 KPL ### 3. Legally Marketed Predicate Devices of Substantial Equivalence Specialty Health Models A&B Clearwater PPC-101 Doltolo Research Corp. Toxygen ### 4. Description of Device This device is not a kit. This device is an instrument for hydrotherapy of the colon. It introduces filtered water at a comfortable temperature into the large intestine. It is an automatic system for filling and emptying water into and out of the colon and thus cleansing the colon of its contents when medically indicated such as before radiological or endoscopic examination. It is hygienic, painless and odorless. The instrument has an automatic pressure safety system; automatic disinfection; a system for collection of samples for analysis; remote and frontal panel control; a counter for the number of liters of water used; a three-way faucet; flow regulator; thermostatic mixer and an automatic temperature safety feature. It contains disposable components for use only once in the therapy ### 5. Intended Use of Device The indication for use of this device must be restricted to colon cleansing when medically indicated, such as before radiological or endoscopic examination. {1} 2016 412 # 6. Summary of Technological Characteristics Compared to Predicates Summaries between Transcom and three predicates manufactured by Specialty Health, Clearwater, and Dotolo Research, appear below. Differences in technological characteristics between Transcom and the predicate devices include the following. Between Transcom and Specialty Health 1. Specialty Health uses a gravity flow system, Transcom a syringe. Between Transcom and Clearwater 1. A sink and faucet are needed for the Clearwater unit; Transcom is plumbed. 2. Clearwater has no speculum collection; Transcom does. 3. Clearwater's unit is portable; Transcom is wall mounted 4. Clearwater uses a pump; Transcom uses a syringe Between Transcom and Doltolo 1. Dotolo uses a gravity flow system, Transcom a syringe Details in the technological characteristics for Transcom and the predicates are itemized on the following pages. {2} 31 | Parameter | TRANSCOM | SPECIALTY HEALTH Hydro-San | CLEARWATER PPC-101 | DOTLO RESEARCH TOXYGEN | | --- | --- | --- | --- | --- | | Water Source | Household/Commercial | Household/Commercial | Household/Commercial | Household/Commercial | | Water Flow (Pressure) | Yes | Yes | Yes | Yes | | Water Flow Control Valve | Yes | Yes | Yes | Yes | | Drainage System (Gravity) | Yes | Yes | Yes | Yes | | Closed System | | | | | | Water | Yes | Yes | Yes | Yes | | Drainage | Yes | Yes | Yes | Yes | | Mixing Valve | | | | | | Grohemix | Yes | Yes | Yes | Yes | | Powers Hydroguard | N/A | | Yes | | | Fittings | | | | | | Brass | Yes | Yes | Yes | Yes | | Stainless | Yes | Yes | Yes | Yes | | Cabinet Composition | | | | | | Stainless Steel | Yes | N/A | Yes | N/A | | Sheet Metal (Painted) | N/A | Yes | N/A | Yes | | Sink & Faucet | N/A | N/A | Yes | N/A | | Cabinet Design | Wall-Mount | Wall-Mount | Portable (on-wheels) | Wall-Mount | | Weight | 132.3 lbs. | Approx 40 lbs. | 70 lbs. | Approx 40 lbs. | | Physical Dimensions | 1020 x 720 x 270 CM | 20" w x 7" d x 20" h | 19" w x 16" d x 41" h | 20" w x 7" d x 20" h | | Pumps | None | None | Sanitation System | None | {3} 414 | Parameter | TRANSCOM | SPECIALTY HEALTH Hydro-San | CLEARWATER PPC-101 | DOTLO RESEARCH TOXYGEN | | --- | --- | --- | --- | --- | | Mode of Operation | Continuous Water Flow Gravity Flow Drainage Manual Operation | Continuous Water Flow Gravity Flow Drainage Manual Operation | Continuous Water Flow Gravity Flow Drainage Manual Operation | Continuous Water Flow Gravity Flow Drainage Manual Operation | | Major Separate System Components | Mixing Valve Pressure Regulator Water Manifold System Relief Valve System Monitoring System (Temp. Pressure) Viewing Assembly Sanitation System Lighting Assembly Drainage Assembly None | Mixing Valve Pressure Regulator Water Manifold System Relief Valve System Monitoring System (Temp. Pressure) Viewing Assembly Sanitation System Lighting Assembly Drainage Assembly None | Mixing Valve Pressure Regulator Water Manifold System Relief Valve System Monitoring System (Temp. Pressure) Viewing Assembly Sanitation System Lighting Assembly Drainage Assembly Sink & Faucet | Mixing Valve Pressure Regulator Water Manifold System Relief Valve System Monitoring System (Temp. Pressure) Viewing Assembly Sanitation System Lighting Assembly Drainage Assembly None | | Intended Usage | Colonic Irrigator | Colonic Irrigator | Colonic Irrigator | Colonic Irrigator | | Monitoring Systems Water Temperature Pressure View Tube Assembly (Fluid Discharge Path) | Yes Yes Yes | Yes Yes Yes | Yes Yes Yes | Yes Yes Yes | | Fluid Pathway | Disinfected, and disposable except the view tube assembly which must be cleaned & sterilized prior to each procedure | Disinfected, and disposable except the view tube assembly which must be cleaned & sterilized prior to each procedure | Disinfected, and disposable except the view tube assembly which must be cleaned & sterilized prior to each procedure | Disinfected, and disposable except the view tube assembly which must be cleaned & sterilized prior to each procedure | | Specimum Collection | Yes | Yes | No | Yes | {4} 410 | Parameter | TRANSCOM | SPECIALTY HEALTH Hydro-San | CLEARWATER PPC-101 | DOTLO RESEARCH TOXYGEN | | --- | --- | --- | --- | --- | | Regulators (Water Pressure) | Norgren Yes | Norgren Yes | Norgren Yes | Norgren Yes | | Relief Valve (Water Pressure) | Norgren Yes | Norgren Yes | Norgren Yes | Norgren Yes | | Over Pressure Relief System | Yes | Yes | Yes | Yes | | Systems Check Valves | Yes | Yes | Yes | Yes | | Sanitation System Pump | Yes Syringe | Yes Gravity Flow System | Yes Yes | Yes Gravity Flow System | | Gauges Water Temperature Water Pressure | Yes Yes | Yes Yes | Yes Yes | Yes Yes | | Electrical Requirements | 110/120 VAC 50/60 Hz service, power outlet to be grounded and polarized and GFI | 110/120 VAC 50/60 Hz service, power outlet to be grounded and polarized and GFI | 110/120 VAC 50/60 Hz service, power outlet to be grounded and polarized and GFI | 110/120 VAC 50/60 Hz service, power outlet to be grounded and polarized and GFI | | View Tube Assembly Back Lighting | Yes | Yes | Yes | Yes | {5} 6 of 6 416 ## 7. Non-Clinical Performance Data Substantial equivalence is not based on an assessment of non-clinical performance data. ## 8. Clinical Performance Data Substantial equivalence is not based on an assessment of clinical performance data. ## 9. Conclusions from Non-Clinical and Clinical Not applicable {6} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 31 TRANSCQM Transcendencias Comerciales, S.L. c/o Michael G. Farrow, Ph.D. SRS International Corporation 1625 K Street, N.W. Washington, D.C. 20006-1604 Re: K970482 TRANSCOM Colon Hydrotherapy Model HC-2000 Dated: September 10, 1997 Received: September 10, 1997 Regulatory class: II 21 CFR §876.5220/Product code: 78 KPL Dear Dr. Farrow: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Enclosure {7} Page 1 of 1 510(k) Number (if known): K 970482 Device Name: TRANSCOM COLON HYDROTHERAPY MODEL HC-2000 Indications For Use: The indication for use of this device must be restricted to colon cleansing when medically indicated, such as before radiological or endoscopic examination. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K970482 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...