Rubber Utility Catheter

K251864 · C.R. Bard, Inc. · KOD · Feb 23, 2026 · Gastroenterology, Urology

Device Facts

Record IDK251864
Device NameRubber Utility Catheter
ApplicantC.R. Bard, Inc.
Product CodeKOD · Gastroenterology, Urology
Decision DateFeb 23, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The BD Rubber Utility Catheter is indicated for intermittent catheterization of the urethra to drain urine from the bladder by patients who are not capable of voluntary urination.

Device Story

Uncoated, natural rubber latex, single-lumen intermittent urethral catheter; provided sterile for single use. Device inserted through urethra into bladder to facilitate urine drainage. Used in hospitals, nursing homes, rehabilitation facilities, or home environments by healthcare providers or patients. Simple mechanical design; no active components or software. Benefits include relief of urinary retention through controlled drainage.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993 series), functional performance (ISO 20696, ASTM F623-19, ASTM D8389-21), packaging integrity (ASTM F88/F88M-21, ASTM F2096-11), accelerated aging (ASTM F1980-21), and ethylene oxide sterilization validation (ISO 11135).

Technological Characteristics

Natural rubber latex, single-lumen, straight-tip catheter. Dimensions: 8-18 French, 385 mm length. Sterile, single-use. Sterilization: Ethylene oxide (ISO 11135). Compliance with ISO 20696, ASTM F623-19, ASTM D8389-21.

Indications for Use

Indicated for intermittent urethral catheterization to drain urine from the bladder in adult and pediatric patients (2 years or older) unable to urinate voluntarily.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 23, 2026 C. R. Bard, Inc. Cameron Boyd Regulatory Affairs Specialist 8195 Industrial Boulevard Covington, Georgia 30014 Re: K251864 Trade/Device Name: Rubber Utility Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: II Product Code: KOD Dated: January 30, 2026 Received: February 2, 2026 Dear Cameron Boyd: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251864 - Cameron Boyd Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K251864 - Cameron Boyd Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, JESSICA K. NGUYEN -S Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K251864 Device Name Rubber Utility Catheter Indications for Use (Describe) The BD Rubber Utility Catheter is indicated for intermittent catheterization of the urethra to drain urine from the bladder by patients who are not capable of voluntary urination. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} BD C. R. Bard, Inc. 8195 Industrial Boulevard Covington, Georgia 30014 USA K251864 bd.com # 510(k) Summary K251864 In accordance with 21 CFR 807.92, the following information is provided for the BD Rubber Utility Catheter Traditional 510(k) Premarket Notification. This submission was prepared in accordance with 21 CF 807.90 Subpart E and the FDA's electronic submission template eSTAR. ## 510(k) Owner/Submitter Information C. R. Bard, Inc. 8195 Industrial Boulevard Covington, GA 30014 Phone: 1-844-823-5433 Contact Person: Cameron Boyd, Regulatory Affairs Specialist Contact Email: Cameron.Boyd@BD.com Date Prepared: 20 February 2026 ## Subject Device | Trade Name: | Rubber Utility Catheter | | --- | --- | | Common Name(s): | Red Rubber Catheter, Urological Catheter | | Recommended Classification Regulation: | 21 CFR 876.5130 | | Classification: | II | | Product Code: | KOD | | Device Classification Name: | Catheter, Urological | | Review Panel: | Gastroenterology/Urology | | Basis for Submission: | Modified Device | ## Predicate Device | Trade Name: | Dover Red Rubber Robinson Catheter | | --- | --- | | Predicate 510(k) Number: | K041243 | | Common Name(s): | Urinary Drainage Catheter | | Recommended Classification Regulation: | 21 CFR 876.5130 | | Classification: | II | | Product Code: | KOD | | Device Classification Name: | Catheter, Urological | | Review Panel: | Gastroenterology/Urology | Additionally, a reference device was selected within the 510(k) to provide a clinically, commercially, and legally marketed comparator as an additional benchmark to ensure the subject device does not raise different questions of safety or effectiveness. Advancing the world of health Page 1 of 4 {5} BD C. R. Bard, Inc. 8195 Industrial Boulevard Covington, Georgia 30014 USA K251864 bd.com # Reference Device | Trade Name: | Intermittent Nelaton Catheter for Single Use | | --- | --- | | Predicate 510(k) Number: | K222677 | | Common Name: | Urethral Catheter | | Recommended Classification Regulation: | 21 CFR 876.5130 | | Classification: | II | | Product Code: | EZD | | Device Classification Name: | Catheter, Straight | | Review Panel: | Gastroenterology/Urology | # Device Description The BD Rubber Utility Catheter is an uncoated, latex, intermittent urethral catheter that is provided sterile and intended for single use. It is intended for the drainage of urine from the bladder via intermittent urethral catheterization. The device is intended for use in adult and pediatric patients 2 years of age or older. Selection of the appropriate catheter size should be determined by the patient's healthcare provider based on clinical judgement. # Indications for Use The BD Rubber Utility Catheter is indicated for intermittent catheterization of the urethra to drain urine from the bladder by patients who are not capable of voluntary urination. # Non-Clinical Data The following data is provided in support of the substantial equivalence determination and the safety and effectiveness claims of the subject device based on FDA recognized consensus standards and others at the time of submission. # Biocompatibility: The biocompatibility evaluation was conducted in accordance with FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and applicable sections of - ISO 10993-1 – Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process - ISO 10993-5 – Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - ISO 10993-6 – Biological evaluation of medical devices - Part 6: Tests for local effects after implantation - ISO 10993-10 – Biological evaluation of medical devices - Part 10: Tests for skin sensitization - ISO 10993-11 – Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - ISO 10993-12 – Biological evaluation of medical devices - Part 12: Sample preparation and reference materials - ISO 10993-23 – Biological evaluation of medical devices - Part 23: Tests for irritation Advancing the world of health Page 2 of 4 {6} BD C. R. Bard, Inc. 8195 Industrial Boulevard Covington, Georgia 30014 USA K251864 Page 3 of 4 bd.com Non-Clinical Functional Performance Testing: Non-clinical performance testing of the subject device was performed in accordance with applicable sections of - ISO 20696 – Sterile urethral catheters for single use - ASTM F623-19 (2023) – Standard Performance Specification for Foley Catheter - ASTM D8389-21 – Standard performance specification for urinary intermittent catheters Non-clinical performance testing of the subject device’s primary packaging was performed in accordance with applicable sections of - ASTM F88/F88M-21 – Standard Test Method for Seal Strength of Flexible Barrier Materials - ASTM F2096-11 – Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test) To demonstrate device and packaging functionality over time, accelerated aging was performed to simulate aging in accordance with - ASTM F1980-21 – Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices Sterilization validation was performed according to - ISO 11135 Second Edition 2014-07-15 – Sterilization of health-care products - Ethylene oxide - Requirements for the development validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2018)] All tests met the requirements of the pre-determined acceptance criteria. ## Technological Characteristic Comparison The subject device is substantially equivalent to the predicate device cited above. The differences between the subject device and the predicate device are minor, and all associated risks have been appropriately mitigated through testing. These differences in technological characteristics do not raise different questions of safety or effectiveness. Table 1 below provides a comparison of the technological characteristics of the subject device and predicate device. ## Conclusion The BD Rubber Utility Catheter has the same intended use and similar technological characteristics as the predicate device. The subject device is substantially equivalent to the legally marketed predicate device as demonstrated by the substantial equivalence comparisons and performance data, and the subject device does not raise different questions of safety and effectiveness for its intended use. Advancing the world of health Page 3 of 4 {7} BD C. R. Bard, Inc. 8195 Industrial Boulevard Covington, Georgia 30014 USA K251864 bd.com Table 1. Technological Characteristic Comparison | Characteristic | Subject Device | Predicate Device (K041243) | | --- | --- | --- | | Regulation Number | 21 CFR 876.5130 | 21 CFR 876.5130 | | Regulation Description | Urological Catheter and Accessories | Urological Catheter and Accessories | | Product Code | KOD | KOD | | Classification | II | II | | Intended Use | Urine drainage from the bladder via intermittent urethral catheterization | Urine drainage from the bladder via intermittent urethral catheterization | | Indications for Use | The BD Rubber Utility Catheter is indicated for intermittent catheterization of the urethra to drain urine from the bladder by patients who are not capable of voluntary urination. | The proposed device is intended for use in the drainage of urine from the urinary bladder. The product is intended for intermittent catheterization on patients who are not capable of voluntary urination. This product is not designed for use as an indwelling catheter. | | Target Population | Adult and Pediatric | Not stated | | Sterile State | Provided sterile | Provided sterile | | Condition of Use | Single Use | Single Use | | Principle of Operation | Device is passed to the urinary bladder via the urethra. | Device is passed to the urinary bladder via the urethra. | | Where Used | Hospital, nursing home, rehabilitation facilities, restrooms, or home use environment | Hospital, nursing home, rehabilitation facilities, restrooms, or home use environment | | Catheter Design | Single lumen catheter with eyelets to funnel urine | Single lumen catheter with eyelets to funnel urine | | Catheter Material | Natural Rubber Latex | Natural Rubber Latex | | Catheter Coating | None | None | | Tip Configuration | Straight tip | Straight tip | | Diameter French Sizes | 8-18 | 12 inch length: 8-10 16 inch length: 12-30 | | Catheter Length | Effective Length: 385 mm | Overall Lengths: 12 inches (305 mm) or 16 inches (406 mm) | | Sterilization Method | Ethylene Oxide | Ethylene Oxide | Advancing the world of health Page 4 of 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...