MEDLINE PREMIUM LEG BAG, MEDLINE FABRIC BAC LEG BAG, MEDLINE URINARY LEG BAG
K972887 · Medline Industries, Inc. · KNX · Dec 11, 1997 · Gastroenterology, Urology
Device Facts
Record ID
K972887
Device Name
MEDLINE PREMIUM LEG BAG, MEDLINE FABRIC BAC LEG BAG, MEDLINE URINARY LEG BAG
Applicant
Medline Industries, Inc.
Product Code
KNX · Gastroenterology, Urology
Decision Date
Dec 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5250
Device Class
Class 2
Indications for Use
These devices are indicated for use by incontinent persons with other medical conditions that warrant the use of a leg bag and are intended to collect urine. They may be connected to either an indwelling or non-indwelling catheter.
Device Story
Medline Leg Bags are vinyl pouches with sterile fluid paths and adjustable comfort straps. Device attaches to the leg of an incontinent patient to collect urine. Used with either indwelling or non-indwelling catheters. Provides mobility for catheterized patients; typically worn during the day and replaced at night with standard urinary drainage bags. Operated by the patient or caregiver.
Clinical Evidence
No clinical studies performed; bench testing only.
Technological Characteristics
Vinyl pouch construction with sterile fluid paths and adjustable comfort straps. No technological changes from predicate devices.
Indications for Use
Indicated for incontinent persons or those with medical conditions requiring urine collection via leg bag. Compatible with indwelling or non-indwelling catheters.
Regulatory Classification
Identification
A urine collector and accessories is a device intended to collect urine. The device and accessories consist of tubing, a suitable receptacle, connectors, mechanical supports, and may include a means to prevent the backflow of urine or ascent of infection. The two kinds of urine collectors are:(1) A urine collector and accessories intended to be connected to an indwelling catheter, which includes the urinary drainage collection kit and the closed urine drainage system and drainage bag; and (2) A urine collector and accessories not intended to be connected to an indwelling catheter, which includes the corrugated rubber sheath, pediatric urine collector, leg bag for external use, urosheath type incontinence device, and the paste-on device for incontinence.
Special Controls
*Classification* —(1)*Class II (special controls) for a urine collector and accessories intended to be connected to an indwelling catheter.* The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.(2)
*Class I (general controls).* For a urine collector and accessories not intended to be connected to an indwelling catheter, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
K971929 — AMSINO URINARY LEG BAG · Amsino Intl., Inc. · Aug 6, 1997
K963950 — BIODERM EID MALE EXTERNAL INCONTINENCE DEVICE · Bioderm, Inc. · Jan 10, 1997
K183253 — ValPro 2 Plus, VaPro 2 Plus Pocket · Hollister Incorporated · Dec 18, 2018
K080881 — CURE CATHETER CLOSED SYSTEM · Cure Medical, LLC · Jun 11, 2008
K193148 — VaPro Plus Pocket, VaPro Plus · Hollister Incorporated · Dec 13, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Medline. The logo consists of the word "MEDLINE" in a stylized, italicized font. Above and below the text are two parts of a compass rose. The top part of the compass rose is a triangle pointing upwards, and the bottom part is a curved shape pointing downwards.
ﺴﻌ
One Medline Place Mundelein, Illinois 60060.4486
1.847.949.3109 1.800.950.0128 Fax 1.847.949.2643
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
## Corporate Quality Assurance/Regulatory Affairs
SUMMARY OF SAFETY AND EFFECTIVENESS K972887 July 31, 1997 81011 COMPANY INFORMATION: 1. Dynacor Manufacturing DEC I I 1997 A Division of Medline Industries, Inc. One Medline Place Mundelein, IL 60060 Registration #: 1417592 (847) 949-5500 x1131 Phone: Fax: (847) 949-2643 Contact Name: Christine M. Galea Contact Title: Corporate Regulatory Affairs 2. DEVICE NAME: Urine Collection Leg Bag for External Use Premium Leg Bag, Fabric-Bac Leg Bag, Urinary Leg Bag PROPRIETARY NAME: Leg Bag COMMON NAME: I (for devices not intended to be connected to an indwelling catheter) CLASS: II (for devices intended to be connected to an indwelling catheter)
- 3. SUBSTANTIAL EQUIVALENCE: Medline claims substantial equivalence to: Medline's Urine Collector Leg Bag Bard's Leg Bags (K780032)
These new devices are identical to what Medline is currently marketing. The only difference between the devices that are currently on the market and what we intend to market is in the labeling.
- 4. DESCRIPTION:
Medline Leo Bags are vinyl pouches with sterile fluid paths and adjustable comfort straps. This device is attached to the leg of an incontinent person and used to collect urine. This device may be used with an indwelling catheter or an external catheter. A leg bag provides the catheterized person with greater mobility and are usually worn during the day and are replaced at night with a standard urinary drainage bag.
- 5. INTENDED/INDICATIONS FOR USE: These devices are indicated for use by incontinent persons with other medical conditions that warrant the use of a leg bag on the advice of a physician and are intended to collect urine. They may be connected to either an indwelling or non-indwelling catheter.
- 6. TECHNOLOGICAL CHARACTERISTICS: There are no technological characteristics to the new device. The only change Medline is making is in the labeling.
- 7. CLINICAL TESTS: There are no clinical studies that have been performed on these devices, both for the predicate and the new devices.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 11 1997
Ms. Christine M. Galea Corporate Regulatory Affairs Medline Industries. Inc. One Medline Place Mundelein, Illinois 60060-4486 Re: K972887
Urine Collection Leg Bag Dated: November 12, 1997 Received: November 18, 1997 Regulatory class: II and I 21 CFR 8876.5250/Product code: 78 KNX and FAQ
Dear Ms. Galea:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission of affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation of the regulation of the regulation of the "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours.
W.Liau Yu
Lillian Yin. Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## INTENDED USE
510(k) Number (if known): (N/A)
Medline Premium Leg Bag, Fabric-Bac Leg Bag, Urinary Leg Bag Device Name:
Indications for Use:
These devices are indicated for use by incontinent persons with other medical conditions that warrant the use of a leg bag and are intended to collect urine. They may be connected to either an indwelling or non-indwelling catheter.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PA
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use <div style="display:inline;">/</div> |
|------------------------------------------|----|-----------------------------------------------------------|
|------------------------------------------|----|-----------------------------------------------------------|
| (Division Sign-Off) | |
|-----------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices | |
| 510(k) Number | K972887 |
Page 6.0
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.