Intermittent Catheter (Not Finalized)

K211436 · Hollister Incorporated · GBM · Jan 27, 2022 · Gastroenterology, Urology

Device Facts

Record IDK211436
Device NameIntermittent Catheter (Not Finalized)
ApplicantHollister Incorporated
Product CodeGBM · Gastroenterology, Urology
Decision DateJan 27, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder.

Device Story

Hydrophilic-coated, single-use intermittent catheter; used for urinary incontinence management. Device inserted through urethra to drain bladder. Available in 7-inch and 16-inch lengths; various French sizes. Features drainage eyelets, color-coded funnel, and optional gripper on 16-inch model. Constructed from phthalate-free, PVC-free thermoplastic elastomer (TPE). Packaged in peelable 4-sided pouch containing catheter and hydration fluid; direct hydration method lubricates coating. Sterile, e-beam irradiated. Used by patients or clinicians in clinical or home settings. Facilitates urine drainage, improving patient quality of life by managing incontinence.

Clinical Evidence

Bench testing only. Performance properties verified against BS EN ISO 20696:2018, including size, lubricity, drainage funnel security, flow rate, kink stability, and tensile force. Biocompatibility testing performed per ISO 10993 series (cytotoxicity, irritation, sensitization, systemic toxicity). Sterilization validated per ISO 11137 and ISO 11737. Packaging and transportation testing confirmed sterile barrier maintenance and performance stability.

Technological Characteristics

Material: Thermoplastic Elastomer (TPE), phthalate/PVC-free. Dimensions: 7" or 16" lengths; 8-16 Fr. Hydrophilic PVP-based coating. Direct hydration system. Sterilization: e-beam irradiation (25-65kGy, SAL 10^-6). Funnel attachment: thermal bond. Packaging: PET/ALU/PE foil laminate.

Indications for Use

Indicated for male, female, and pediatric patients requiring bladder drainage via urethral catheterization.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. January 27, 2022 Hollister Incorporated Michelle Schiltz-Taing Regulatory Affairs Manager 2000 Hollister Drive Libertyville, IL 60048 Re: K211436 > Trade/Device Name: Intermittent Catheter (not finalized) Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: December 29, 2021 Received: December 30, 2021 Dear Michelle Schiltz-Taing: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211436 #### Device Name Intermittent Catheter (not finalized) Indications for Use (Describe) The intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. **Type of Use (Select one or both, as applicable)** > Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Applicant: | Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60048 | |------------------------|------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Michelle Schiltz-Taing<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 224-864-0431 | | Date Prepared: | January 25, 2022 | | Trade Name: | Intermittent Catheter (not finalized | | Common Name: | Catheter, Urethral | | Product Code/Class: | GBM/Class II | | Classification Name: | Urological catheter and accessories | | Classification Number: | 21 CFR 876.5130 | #### Predicate Device: Onli Intermittent Catheter by Hollister Incorporated (K163179) The predicate device has not been subject to a design-related recall. ## Indications for Use: This intermittent catheter is a flexible tubular device that is inserted through the urethra by male, female and pediatric patients who need to drain urine from the bladder. ## Device Description: The Intermittent Catheter (final name to be determined) is a hydrophilic-coated, single use catheter with drainage eyelets that is used to manage urinary incontinent Catheter is inserted into the urethra to drain urine from the bladder. The Intermittent Catheter is available in 7 in. and 16 in. lengths. The 7 in. catheter is available in sizes 8, 10, 12, 14 Fr and the 16 in. catheter is available in sizes 8, 10, 12, 14, 16 Fr. The Intermittent Catheter is provided sterile, using e-beam Irradiation sterilization. The device is made from Thermo-plastic Elastomer (TPE), and is phthalate free, PVC free. This directly hydrated catheter is packaged in a peel-able 4 sided pouch which was designed to be easy to open and to facilitate access to the catheter. The primary pack contains both the catheter assembly and the hydration fluid in one compartment. This means that the coated catheter surface is lubricated by direct contact with the hydration fluid There is a gripper on the 16 inch catheter to aid with the insertion process, if desired. {4}------------------------------------------------ # Comparison of Technological Characteristics: The table below summarizes the technological characteristics of the Intermittent Catheter as compared to the predicate and reference devices. | | Predicate Onli<br>K163179 | Intermittent Catheter | Same or Different/Rational for no<br>impact to efficacy or safety | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | The intermittent catheter is a flexible tubular device that is<br>inserted through the urethra by male, female and pediatric<br>patients who need to drain urine from the bladder. | | No change | | Condition of Use | | Single Use | No change | | Device Material | Polyvinyl Chloride (PVC);<br>phthalate free funnel and<br>catheter<br>Thermo-plastic Elastomer<br>(TPE); phthalate free,<br>PVC free gripper | Thermo-plastic Elastomer<br>(TPE); phthalate free,<br>PVC free funnel, catheter,<br>and gripper | Testing shows differences do not<br>affect safety and effectiveness<br>compared to the predicate device. | | Length/Fr Sizes | 7 inches: Fr 8, 10, 12, 14<br>16 inches: Fr 8, 10, 12, 14, 16 | | No change | | Gripper on 16 in.<br>Catheter | Yes | | No change | | Attachment of<br>Funnel to<br>Catheter Tube | Solvent Bonded | Insert molded funnel<br>(created by a thermal<br>bond) | Although the attachment of the<br>funnel to the catheter tube is<br>different, the differences do not<br>affect safety and effectiveness<br>compared to the predicate device. | | End of Catheter<br>Design | Rounded tip | | No change | | End Design | Color-coded funnel | | No change | | Number of<br>Eyelets | 2 | | No change | | Catheter Color | Opaque | Clear | The appearance of the catheter does<br>not impact functionality. | | Coating | PVP Based Coating<br>(polyvinylpyrollidone) | | The Hydrophilic coating on the<br>predicate and the Intermittent Catheter<br>is PVP based. Testing shows<br>differences do not affect safety and<br>effectiveness compared to the<br>predicate device. | | Fluid | DI water | Hydration fluid | Although the hydration fluid is<br>different between the predicate and the<br>Intermittent Catheter, testing shows<br>differences do not affect safety and<br>effectiveness compared to the<br>predicate device. | | Hydration<br>Method | Vapor Hydration | Direct Hydration | Although a different method of<br>hydration is utilized between the<br>predicate device and the Intermittent<br>Catheter, substantial equivalency of<br>catheter lubricity is demonstrated<br>via testing. | | Packaging<br>Material | Oriented<br>Polyamide/Aluminum/Polyethyle<br>ne (OPA/ALU/PE) foil<br>(foil laminate) | Polyethlyene<br>Terephthalate/Aluminum/<br>Polyethylene<br>(PET/ALU/PE) foil<br>(foil laminate) | Testing shows differences do not<br>affect safety and effectiveness<br>compared to the predicate device. | | Sterilization<br>Method | Gamma Irradiation<br>Dose 25-40kGy<br>SAL 10-6 | e-beam Irradiation<br>Dose 25-65kGy<br>SAL 10-6 | The difference in sterilization method<br>does not affect the safety and<br>effectiveness compared to the<br>predicate device. | | Expiration<br>Dating | 2 years | | No change | | Storage<br>Conditions | 15-30°C / 59-86°F | | No change | {5}------------------------------------------------ ## Brief Description of Non-Clinical Testing: The physical performance properties of the Intermittent Catheter met all applicable requirements of BS EN ISO 20696:2018, Sterile urethral catheters for single use. Testing was conducted to support size designation, show equivalence of lubricity, and determination of the catheter, the security of fit of the drainage funnel, flow rate through catheter, kink stability and peak tensile force of urethral catheter. Biocompatibility testing met the requirements of the following standards: - ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and testing within ● a risk management process - ISO 10993-3:2014, Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity - ISO 10993-5:2009, Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity ● - ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin ● sensitization - ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity - ISO 10993-12:2021, Biological evaluation of medical devices - Part 12: Sample preparation and reference materials - ISO 10993-17:2002, Biological evaluation of medical devices Part 17: Establishment of allowable ● limits for leachable substances - ISO 10993-18:2020, Biological evaluation of medical devices Part 18: Chemical characterization ● of medical device materials within a risk management process {6}------------------------------------------------ The following biological endpoints were addressed: cytotoxicity, intracutaneous irritation, vaginal irritation, sensitization, acute systemic toxicity, and subacute systemic toxicity. Sterilization met all requirements of the following FDA-recognized standards: - ISO 11137-1:2006. Sterilization of health care products Radiation Part 1: Requirements for . development, validation and routine control of a sterilization process for medical devices. - ISO 11137-2:2013, Sterilization of health care products Radiation Part 2: Establishing the ● sterilization dose - ANSI/AAMI/ISO 11737-1:2018, Sterilization of health care products Microbiological ● methods - Part 1: Determination of a population of microorganisms on products - ANSI/AAMI/ISO 11737-2:2019, Sterilization of medical devices Microbiological methods -● Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process Packaging integrity testing was conducted to verify the maintenance of the sterile barrier through shelf life. Transportation testing was conducted in order to verify that there is no impact to the device safety or efficacy of the catheter performance due to the hazards associated with the transportation environment. #### Conclusion: It is concluded that the information supplied in this submission has demonstrated that the Intermittent Catheter is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...