Archimedes Pro Automated Peritoneal Dialysis Cycler (3014-00000-1); Archimedes Standard Automated Peritoneal Dialysis Cycler (3014-00000-3); Archimedes Pro Disposable Tubing Set (3014-50000-1); Archimedes Standard Disposable Tubing Set (3014-50000-3); Archimedes Pro Plus Disposable Tubing Set (3014-50000-5)
Archimedes Pro Automated Peritoneal Dialysis Cycler (3014-00000-1); Archimedes Standard Automated Peritoneal Dialysis Cycler (3014-00000-3); Archimedes Pro Disposable Tubing Set (3014-50000-1); Archimedes Standard Disposable Tubing Set (3014-50000-3); Archimedes Pro Plus Disposable Tubing Set (3014-50000-5)
Applicant
Simergent, LLC
Product Code
FKX · Gastroenterology, Urology
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Archimedes APD system is an Automated Peritoneal Dialysis system indicated for acute and chronic peritoneal dialysis for adult patients in clinical and home use. A care partner is not required. The following therapies are supported: Continuous Cyclic Peritoneal Dialysis (CCPD) and Intermittent Peritoneal Dialysis (IPD). Mid-day exchanges are not supported.
Device Story
Automated peritoneal dialysis (APD) system; includes heater unit, control unit, cart, drain containers, and disposable tubing set. Operates via gravity fill/drain controlled by software-actuated pinch valves. Uses AC-powered resistive heater for fluid warming; scales for fluid measurement/volumetric accuracy. Used in clinical or home environments by adult patients without care partner assistance. Healthcare providers program therapy parameters; device executes cycles; monitors fluid volumes; ensures therapy delivery. Benefits include automated, home-based renal replacement therapy.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, human factors validation, and comparison of technological characteristics to predicate devices.
Technological Characteristics
System components: plastic housing, aluminum heater tray, steel cart. Disposable set: plastic tubing and connectors. Sensing/actuation: software-controlled pinch valves, resistive heater, scale-based fluid measurement. Connectivity: not specified. Sterilization: not specified. Software: automated control of dialysis cycles.
Indications for Use
Indicated for acute and chronic peritoneal dialysis in adult patients in clinical and home settings. Supports CCPD and IPD therapies. No care partner required. Mid-day exchanges not supported.
Regulatory Classification
Identification
(1) A peritoneal dialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions, and that consists of a peritoneal access device, an administration set for peritoneal dialysis, a source of dialysate, and, in some cases, a water purification mechanism. After the dialysate is instilled into the patient's peritoneal cavity, it is allowed to dwell there so that undesirable substances from the patient's blood pass through the lining membrane of the peritoneal cavity into this dialysate. These substances are then removed when the dialysate is drained from the patient. The peritoneal dialysis system may regulate and monitor the dialysate temperature, volume, and delivery rate together with the time course of each cycle of filling, dwell time, and draining of the peritoneal cavity or manual controls may be used. This generic device includes the semiautomatic and the automatic peritoneal delivery system.(2) The peritoneal access device is a flexible tube that is implanted through the abdominal wall into the peritoneal cavity and that may have attached cuffs to provide anchoring and a skin seal. The device is either a single use peritioneal catheter, intended to remain in the peritoneal cavity for less than 30 days, or a long term peritoneal catheter. Accessories include stylets and trocars to aid in the insertion of the catheter and an obturator to maintain the patency of the surgical fistula in the abdominal wall between treatments. (3) The disposable administration set for peritoneal dialysis consists of tubing, an optional reservoir bag, and appropriate connectors. It may include a peritoneal dialysate filter to trap and remove contaminating particles. (4) The source of dialysate may be sterile prepackaged dialysate (for semiautomatic peritoneal dialysate delivery systems or “cycler systems”) or dialysate prepared from dialysate concentrate and sterile purified water (for automatic peritoneal dialysate delivery systems or “reverse osmosis” systems). Prepackaged dialysate intended for use with either of the peritoneal dialysate delivery systems is regulated by FDA as a drug.
Special Controls
*Classification.* Class II (special controls). The following accessories are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9: A catheter finger grip that is non-patient contacting and intended for single use with a peritoneal catheter; a continuous ambulatory peritoneal dialysis (CAPD) belt; and a catheter stand that does not include weigh scales.
Fresenius CCPD Cycler Set W/5 Pronged Safe Lock (K822549)
Related Devices
K103220 — VOYAGER PERITONEAL DIALYSIS SYSTEM · Deka Research & Development Corp. · Jul 29, 2011
K124018 — AMIA AUTOMATED PD SYSTEM · Deka Research & Development Corp. · Jun 14, 2013
K243371 — Byonyks X-1 APD Cycler; Byonyks Automated PD Set DS-1 · Byonyks Pvt, Ltd. · May 16, 2025
K212522 — Lilliput APD System · Fresenius Medical Care Renal Therapies Group, LLC · Apr 15, 2022
K123630 — FRESENIUS LIBERTY CYCLER · Fresenius Medical Care North America, Design Cente · Sep 9, 2013
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 2, 2025
Simergent, LLC
Steve Lindo
CEO & Co-Founder
11 NE 11th Street, Suite 229
Oklahoma City, OK 73104
Re: K250523
Trade/Device Name: Archimedes Pro Automated Peritoneal Dialysis Cycler (3014-00000-1); Archimedes Standard Automated Peritoneal Dialysis Cycler (3014-00000-3); Archimedes Pro Disposable Tubing Set (3014-50000-1); Archimedes Standard Disposable Tubing Set (3014-50000-3); Archimedes Pro Plus Disposable Tubing Set (3014-50000-5)
Regulation Number: 21 CFR§ 876.5630
Regulation Name: Peritoneal dialysis system and accessories
Regulatory Class: II
Product Code: FKX
Dated: September 2, 2025
Received: September 2, 2025
Dear Steve Lindo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250523 - Steve Lindo
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K250523 - Steve Lindo
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Maura Rooney -S
Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250523
Device Name
Archimedes Pro Automated Peritoneal Dialysis Cycler (3014-00000-1);
Archimedes Standard Automated Peritoneal Dialysis Cycler (3014-00000-3);
Archimedes Pro Disposable Tubing Set (3014-50000-1);
Archimedes Standard Disposable Tubing Set (3014-50000-3);
Archimedes Pro Plus Disposable Tubing Set (3014-50000-5)
Indications for Use (Describe)
The Archimedes APD system is an Automated Peritoneal Dialysis system indicated for acute and chronic peritoneal dialysis for adult patients in clinical and home use. A care partner is not required. The following therapies are supported: Continuous Cyclic Peritoneal Dialysis (CCPD) and Intermittent Peritoneal Dialysis (IPD). Mid-day exchanges are not supported.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
SIMERGENT
Archimedes Dialysis System
TRADITIONAL 510(k) SUMMARY
# Traditional 510(k) Summary
1. Summary Date: February 7, 2025
2. Applicant Name: Simergent, LLC
11 NE 11th St Suite 229
Oklahoma City, OK 73104
Website: https://www.simergent.com
Owner/Operator Number: Pending
3. Submission Correspondent: On behalf of Simergent, the following person is assigned the responsibility of submission correspondence:
Steve Lindo, CEO & Co-Founder
12920 S. 82nd Ct
Palos Park, IL 60464
Ph: 214-557-5868
4. Authorized Correspondent: On behalf of Simergent, the following consultant is authorized to correspond with the agency regarding this submission:
John F. Ziobro, Principal Consultant
SpectraMedEx, LLC
3215 Golf Drive, #149
Delafield, WI 53018
Ph: 262-719-8922
email: jfz@spectramedex.com
5. Trade Name: Archimedes
6. Model Numbers: Automated Peritoneal Dialysis Cycler
3014-00000-1 Archimedes Pro
3014-00000-3 Archimedes Standard
Archimedes Disposable Tubing Set
3014-50000-1 Archimedes Pro
3014-50000-3 Archimedes Standard
3014-50000-5 Archimedes Pro Plus
7. Suggested Classification Regulation, Class, Product Code, Description & Panel:
| Regulation # | Class | Product Code | Description | Review Panel |
| --- | --- | --- | --- | --- |
| 21 CFR 876.5630 | II | FKX | Peritoneal dialysis system and accessories | Gastroenterology and Urology |
8. Reason for Traditional 510(k): New Submission (No previous submissions)
9. Predicate Device(s): Fresenius PD+ IQCard Cycler and CCPD Cycler Set W/5 Pronged Safe Lock
10. Predicate 510(k) Number: K002892 and K822549
11. Predicate Manufacturer: Fresenius Medical Care
Website: https://fmcna.com/
Owner/Operator Number: 1225714
Simergent - Archimedes Dialysis System
Traditional 510(k) Submission
000_001_003_Traditional 510(k) Summary K250523.docx
Confidential
{5}
S
SIMERGENT
SIMERGENT Archimedes Dialysis System TRADITIONAL 510(k) SUMMARY
# 12. Predicate Classification Regulation, Class, Product Code, Description & Panel
Both predicate devices have the same Classification Regulation, Class, Product Code, Description & Panel.
| Regulation # | Class | Product Code | Description | Review Panel |
| --- | --- | --- | --- | --- |
| 21 CFR 876.5630 | II | FKX | Peritoneal dialysis system and accessories | Gastroenterology and Urology |
# 13. Proposed Indication for Use:
The Archimedes APD device is an Automated Peritoneal Dialysis system indicated for acute and chronic peritoneal dialysis for adult patients in clinical and home use. A care partner is not required. The following therapies are supported: Continuous Cyclic Peritoneal Dialysis (CCPD) and Intermittent Peritoneal Dialysis (IPD). Mid-day exchanges are not supported.
# 14. Guidance Documents/Special Control/Regulations
The recognized consensus standards under 21 CFR 876.5630 which are applicable to the proposed device are as follows:
- 9-149 IEC 60601-2-39 Edition 3.0 2018-04
The following FDA guidance documents were applied to the design and software verification of the proposed device:
- Content of Premarket Submissions for Device Software Functions (June 14, 2023)
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions (September 27, 2023)
- Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016)
- Design Considerations for Devices Intended for Home Use (November 24, 2014)
The following FDA guidance document was applied to the design and biocompatibility verification of the proposed device:
- Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (September 8, 2023)
The following FDA guidance document was applied to the design and sterility verification of the proposed device:
- Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (January 8, 2024)
- Guidance for Industry Pyrogen and Endotoxins Testing: Questions and Answers (June 2012)
# 15. Device Description:
The proposed system is an automated peritoneal dialysis (APD) cycler which consists of a heater unit, control unit, cart, drain containers, and disposable tubing set.
# 16. Materials:
The cycler consists of the following materials:
- Plastic housing
- Aluminum heater tray
- Steel cart
The disposable tubing set consists of the following materials:
- Plastic tubing
- Plastic connectors
# 17. Human Factors Validation Testing:
The Archimedes cycler was validated for safe and effective use in accordance with FDA guidance Applying Human Factors and Usability Engineering to Medical Devices (February 3, 2016).
Simergent - Archimedes Dialysis System
Traditional 510(k) Submission
000_001_003_Traditional 510(k) Summary K250523.docx
Confidential
{6}
S
SIMERGENT
SIMERGENT Archimedes Dialysis System TRADITIONAL 510(k) SUMMARY
18. Comparison of Technological Characteristics with the Predicate Device:
The Archimedes cycler has similar technical characteristics as those of the predicate device. Basic principle of operation (gravity fill/drain via software-controlled valves), use of vertically-oriented mobile cart/stand, fluid measurement mechanism (scales), fluid heating mechanism (AC-powered resistive heater), and volumetric accuracy for fill and drain are all similar to the predicate device.
The Archimedes disposable tubing set has similar technical characteristics as those of the predicate device. Basic principle of operation (gravity fill/drain fluid control via pinch valves), use of materials, connection methods, interactions with third party devices, and sterility are all similar to the predicate device.
19. Comparison Summary / Conclusions
Simergent believes the proposed Simergent Archimedes device under review and the predicate device Fresenius PD+ IQCard Cycler cleared under K002892 are substantially equivalent. Furthermore, both devices have the same/equivalent technological characteristics, physical characteristics and safety standards. The differences that exist between the devices do not affect the relative safety and/or effectiveness.
Additionally, Simergent believes the proposed Simergent Archimedes device (disposable tubing set) under review and the predicate device Fresenius CCPD Cycler Set W/5 Pronged Safe Lock cleared under K822549 are substantially equivalent. Furthermore, both devices have the same/equivalent technological characteristics, physical characteristics and safety standards. The differences that exist between the devices do not affect the relative safety and/or effectiveness.
Simergent - Archimedes Dialysis System
Traditional 510(k) Submission
000_001_003_Traditional 510(k) Summary K250523.docx
Confidential
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.