MENTOR UROLOGY TRIPLE RIB OCCLUSION WRAP FOR MALE INCONTINENCE
Applicant
Mentor Urology, Inc.
Product Code
FHA · Gastroenterology, Urology
Decision Date
Apr 2, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5160
Device Class
Class 1
Attributes
Therapeutic
Intended Use
The Mentor Urology Triple Rib Occlusion Wrap for Male Incontinence is indicated for any male, who has urinary incontinence, either on a temporary or long term basis. It is also indicated for those patients who have an adverse reaction to natural latex rubber.
Device Story
Two-piece compression device for male urinary incontinence; consists of silicone gel-filled penile compression wrap and silicone ring. Provides temporary urinary continence via urethral occlusion. Designed as reusable, soft-sided alternative to rigid clamps. Used by patients with moderate to heavy incontinence. Patient applies device to penis; requires clean, dry skin. Benefits include non-rigid comfort and latex-free material for sensitive users.
Clinical Evidence
Bench testing only. Biological safety testing performed on all components per Tripartite and ISO 10993 standards; all results satisfactory. Physical testing and design verification performed to confirm device adequacy for intended use.
Technological Characteristics
Materials: Silicone gel and silicone. Form factor: Two-piece compression wrap with triple rib design. Energy source: None (mechanical compression). Sterilization: Not specified.
Indications for Use
Indicated for males with temporary or long-term urinary incontinence, including those with natural latex rubber allergies. Contraindicated for patients with broken or poor skin quality or adverse reactions to device components.
Regulatory Classification
Identification
A urological clamp for males is a device used to close the urethra of a male to control urinary incontinence or to hold anesthetic or radiography contrast media in the urethra temporarily. It is an external clamp.
K971992 — UROMED ALTERNATIVE BLADDER CONTROL CONTINENCE DEVICE(ABCPN-0026) · Uromed Corp. · Aug 1, 1997
K964580 — CAP-AID (IMPROVED) · Nebel, Inc. · Jan 10, 1997
K963950 — BIODERM EID MALE EXTERNAL INCONTINENCE DEVICE · Bioderm, Inc. · Jan 10, 1997
K962900 — SELF RETAINING MALE EXTERNAL URINARY CATHETER · Hollister, Inc. · Sep 16, 1996
K963265 — PROSYS SAMEC LA · Convatec, A Division of E.R. Squibb & Sons · Oct 25, 1996
Submission Summary (Full Text)
{0}
APR 01 '97 01:40PM MENTOR UROLOGY P.2
APR - 2 1997 K970308
p1072
# SUMMARY OF SAFETY AND EFFECTIVENESS
## MENTOR UROLOGY TRIPLE RIB OCCLUSION WRAP FOR
## MALE INCONTINENCE
## (510K SUMMARY)
### 1. Submitter's Data
Mentor Urology, Inc.
1615 West River Road North
Minneapolis, MN 55411
Date Prepared: 23 January 1997
### 2. Device Name, Classification
Name: Mentor Urology Triple Rib Occlusion Wrap for Male Incontinence
Trade Name: To be determined
FDA Classification: Class I per 21CFR876.5250, Urological Clamp for Males, Code 78FHA.
### 3. Identification of Substantially Equivalent Device
Bard Cunningham Incontinence Clamp, #K884426
### 4. Device Description
The Mentor Urology Triple Rib Occlusion Wrap for Male Incontinence is a two piece compression device for males with urinary incontinence. The device consists of a silicone gel filled penile compression wrap for occlusion of the urethra and a silicone ring. It is intended for users who would prefer a reusable soft sided device instead of a rigid clamping device. The device is intended for users with moderate to heavy urinary incontinence to provide a temporary means of urinary continence.
### 5. Indications for Use
The Mentor Urology Triple Rib Occlusion Wrap for Male Incontinence is indicated for any male, who has urinary incontinence, either on a temporary or long term basis. It is also indicated for those patients who have an adverse reaction to natural latex rubber.
153
{1}
APR 01 '97 01:40PM MENTOR UROLOGY
K970308 P.3
p2072
6. Contraindications for Use
Clinical contraindications for use of the Mentor Urology Triple Rib Occlusion Wrap for Male Incontinence include:
Use on patients with broken or poor skin quality.
Use on patients with adverse reactions to any component of the system.
7. Warnings, Precautions
Patient skin must be clean and dry before use.
The correct size must be used. Physician advise may be helpful.
8. Biological Safety Testing
Mentor Urology has performed biological safety testing on all components used in this device. These tests meet or exceed Tripartite and ISO 10993 standards. All tests were satisfactory.
9. Physical Testing, Design Verification
Mentor Urology has performed various physical tests on this device and its components to show that the device and its design are adequate for the use intended.
140
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.