SINGLE USE 3-LUMEN BALLOON CATHETER, MODELS B-230Q-A AND B-230Q-B
Applicant
Aomori Olympus Co., Ltd.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Nov 14, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The subject devices, Single Use 3-Lumen Balloon Catheter B-230Q-A/ B-230Q-B have been designed to be used with an Olympus endoscope to inject contrast media into the biliary or pancreatic tract. They can also be used for retrieval of biliary or pancreatic stones.
Device Story
Single-use 3-lumen balloon catheter designed for use with Olympus endoscopes. Features triple-lumen architecture and multiple-sizing balloon; allows guidewire passage and contrast media injection without guidewire removal; enables balloon inflation to three distinct sizes. Used by physicians in clinical settings during endoscopic procedures. Provides operational assistance by saving time and minimizing inventory requirements. Benefits include improved procedural efficiency and stone retrieval capability.
Clinical Evidence
bench testing only
Technological Characteristics
Triple-lumen catheter with multiple-sizing balloon; single-use; compatible with Olympus endoscopes; includes pre-measured syringes for balloon inflation.
Indications for Use
Indicated for use with Olympus endoscopes for injection of contrast media into the biliary or pancreatic tract and for the retrieval of biliary or pancreatic stones.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
K091495 — SINGLE USE 3-LUMEN EXTRACTION BALLOON V, MODEL V B-V233P-A,B-V233P-B,B-V433P-A,B-V433P-B, BV243Q-A, B-V243Q-B, B-V443Q-A · Olympus Medical Systems Corporation · Sep 3, 2009
K131642 — RSH BILIARY STONE REMOVAL DEVICE · Rsh, LLC · Aug 7, 2013
K970052 — EXTRACTORRX · Boston Scientific Corp · Mar 25, 1997
K253132 — Single use stone retrieval balloons · Zhejiang Chuangxiang Medical Technology Co., Ltd. · Mar 6, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
K033333
NOV 1 4 2003
# SMDA 510(k) SUMMARY
This summary of 510(k) safety and effective information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR Section 807 93.
### A. GENERAL INFORMATION
| <b>1. Applicant:</b><br>Name & Address : Aomori Olympus Optical Co., Ltd<br>248-1 Okkonoki 2-choume Kuroishi-shi<br>Aomori, Japan, 036-0367<br>Registration No. : 9614641 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>2. Initial Importer:</b><br>Name & Address : Olympus America Inc.<br>Two Corporate Center Drive, Melville, NY 11747-315<br>Registration No. : 2429304 |
| <b>3. Submission Correspondent:</b><br>Name & Address : Tina Steffanie-Oak<br>Associate Manager, RA/Clinical Monitor |
| Name & Address : | Tina Steffanie-Oak |
|------------------|----------------------------------------------------|
| | Associate Manager, RA/Clinical Monitor |
| | Olympus America Inc. |
| | Two Corporate Center Drive, Melville, NY 11747-315 |
| Telephone : | 631-844-5477 |
| Facsimile : | 631-844-5416 |
# B. DEVICE IDENTIFICATION
- : Single Use 3-Lumen Balloon Catheter, B-230Q-A / B-230Q-B 1. Trade Name
- 2. Common Name : Balloon catheter
- 3. Classification Name: Endoscope and Accessories
- 4. CFR Number : 876.1500
- 5. Classification Panel : 78
- 6. Product Code : FGE, KOG
#### C. PREDICATE DEVICE INFORMATION
B-5 /B-7 series Balloon Catheter K962925
#### D. DEVICE DESCRIPTION
The subject device(s) is a balloon catheter that has a triple-lumen catheter and a multiple sizing balloon with pre-measured syringes, to be used in accordance with Intended use of the device.
These functions ensure smoother passage of the guidewire and injection of the contrast media without removing the guidewire, and enable to inflate one balloon to 3 different sizes.
{1}------------------------------------------------
These modifications give operational assistance to save time for physicians, or minimize the inventory. The operation is basically identical with the predicate device. We believe that these modifications would not affect safety and effectiveness.
# E. INTENDED USE OF THE DEVICE
The subject devices, Single Use 3-Lumen Balloon Catheter B-230Q-A/ B-230Q-B have been designed to be used with an Olympus endoscope to inject contrast media into the biliary or pancreatic tract. They can also be used for retrieval of biliary or pancreatic stones.
# F. SUMMARY INCLUDING CONCLUSIONS DRAWN FROM NON-CLINICAL TESTS
When compared to the predicate devices, these subject devices, Single Use 3-Lumen Balloon Catheter B-230Q-B / B-230Q-B do not incorporate any significant changes in the intended use, method of operation, materials, or design that could affect safety or effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
NOV 1 4 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Aomori Olympus Corporation Ltd. c/o Ms. Tina Steffanie-Oak Associate Manager, RA/Clinical Monitor Olympus America Inc. Two Corporate Center Drive MELVILLE NY 11747-3157
Re: K033333
Trade/Device Name: Single Use 3-Lumen Balloon Catheter Models B-230O-A and B-230Q-B Regulation Number: 21 CRR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 FGE Dated: October 14, 2003 Received: October 17, 2003
Dear Ms. Steffanie-Oak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx. 1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
510(k) Number(if known): ____________________________________________________________________________________________________________________________________________________ Device Name: B-230Q-A, B230Q-B (Single Use 3-Lumen Balloon Catheter)__
Indications for Use:
These instruments have been designed to be used with Olympus endoscopes to inject contrast media into the biliary or pancreatic tract. They can also be used for retrieval of biliary or pancreatic stones.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <span></span> | OR | Over-The-Counter Use | <span></span> |
|----------------------|---------------|----|----------------------|--------------------------|
| (Per 21 CFR 801.109) | | | | (Optional Format 1-2-96) |
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K033333 |
|---------------|---------|
|---------------|---------|
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.