MODEIFICATION TO: SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEM
Applicant
Cordis Corp.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Jul 7, 2003
Decision
SESU
Submission Type
Special
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is intended for use in the palliation of malignant neoplasms in the biliary tree.
Device Story
The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is a self-expanding stent delivery system used for the palliation of malignant biliary neoplasms. The device consists of a nickel-titanium (Nitinol) alloy stent with tantalum micromarkers, mounted on a 7 French delivery system. It is designed for use by clinicians in a clinical setting. The stent is deployed via a proximal handle mechanism to maintain patency in the biliary tree. The device is compatible with 0.035" guidewires and is available in various diameters (12, 14 mm) and lengths (20-80 mm). The system facilitates the palliative treatment of biliary obstructions caused by malignant tumors, helping to improve patient outcomes by restoring bile flow.
Clinical Evidence
No clinical data. Substantial equivalence was confirmed through pre-clinical testing.
Technological Characteristics
Stent material: Nickel-Titanium (Nitinol) alloy with tantalum micromarkers. Delivery system: 7 French profile, 80/120 cm usable length, 0.035" guidewire compatible. Deployment: Proximal handle. Biocompatible materials.
Indications for Use
Indicated for the palliation of malignant neoplasms in the biliary tree.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Cordis S.M.A.R.T.™ CONTROL™ Nitinol Stent Transhepatic Biliary System (K021898)
S.M.A.R.T.™ Nitinol Stent Transhepatic Biliary System (K994068)
Bard LUMINEXX™ 6 Fr Biliary Stent and Delivery System (K020682)
Related Devices
K042969 — MODIFICATION TO SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEM · Cordis Corp. · Nov 8, 2004
K020052 — SMART NITINOL STENT TRANSHEPATIC BILIARY SYSTEM · Cordis Corp. · Feb 11, 2002
K030409 — SMART NITINOL STENT ENDOSCOPIC BILIARY SYSTEM · Cordis Corp. · Mar 10, 2003
K032457 — MODIFICATION TO SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEM · Cordis Corp. · Sep 4, 2003
K021898 — SMART CONTROL NITINOL STENT TRANSHEPATIC BILIARY SYSTEM · Cordis Corp. · Aug 14, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows handwritten text on a white background. The text at the top reads "K0317777", which appears to be a reference or identification number. Below this, the text "page 1 of 2" indicates that this is the first page of a two-page document.
## 510(k) Summary of Safety and Effectiveness
| General<br>Provisions | The name of the device is: | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | Proprietary Name | Common or Usual Name |
| | SMART™ Control™ Nitinol Stent<br>Transhepatic Biliary System | Biliary Stent |
| Name of<br>Predicate<br>Devices | The device is substantially equivalent to:<br>Cordis S.M.A.R.T.™ CONTROL™ Nitinol Stent Transhepatic Biliary System (510(k) # K021898 – August 14, 2002) - Cordis Corporation. S.M.A.R.T.™ Nitinol Stent Transhepatic Biliary System (510(k) # K994068 - December 23, 1999 – Cordis Corporation. Bard LUMINEXX™ 6 Fr Biliary Stent and Delivery System (510(k) # K020682 - April 02, 2002) - C.R. Bard Inc. | |
| Classification | Class II. | |
| Performance<br>Standards | Performance standards have not been established by the FDA under section<br>514 of the Food, Drug and Cosmetic Act. | |
| Indications<br>for Use | The SMART™ Control™ Nitinol Stent Transhepatic Biliary System is<br>intended for use in the palliation of malignant neoplasms in the biliary tree. | |
| Device<br>Description | The device description of the proposed SMART™ Control™ Nitinol Stent<br>Transhepatic Biliary System is as follows. 7 French stent delivery system profile; Stent material – Nickel Titanium alloy and tantalum micromarkers; Expanded stent diameters 12, and 14 mm; Stent lengths: 20, 30, 40, 60, and 80mm; Stent delivery system usable length 80 and 120 cm; Guidewire lumen 0.035" compatible; and Proximal Deployment handle | |
| Biocompatibility | All materials used in the SMART™ Control™ Nitinol Stent<br>Transhepatic Biliary System are biocompatible. | |
{1}------------------------------------------------
K031777
page 2 of 2
The SMARTM ControlTM Nitinol Stent Transhepatic Biliary System is Summary of substantially equivalent to the predicate devices. The equivalence was Substantial confirmed through pre-clinical testing. Equivalence
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white logo. The logo consists of a circular border with text around the top half. Inside the circle is a symbol that resembles three curved lines, stacked on top of each other, and angled to the right. The symbol is reminiscent of a stylized bird in flight.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 0 7 2003
Mr. Sam Mirza Manager Regulatory Affairs Cordis Corporation 14201 N.W. 60th Avenue MIAMI LAKES FL 33014
Re: K031777
Trade/Device Name: S.M.A.R.T.™ Control ™ Nitinol Stent Transhepatic Biliary System Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: 78 FGE Dated: June 9, 2003 Received: June 10, 2003
Dear Mr. Mirza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established.
{3}------------------------------------------------
Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
David C. Schultz, M.E.
Daniel G. Schultz, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page__________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): _K031777
Device Name: Cordis S.M.A.R.T.™ Control™ Nitinol Stent Transhepatic Biliary System
FDA's Statement of the Indications For Use for device:
The Cordis S.M.A.R.T.™ Control™ Nitinol Stent Transhepatic Biliary System is indicated for use in the palliation of malignant neoplasms in the biliary tree.
Prescription Use
(Per 21 CFR 801.109)
OR
Over-the-Counter Use
Nancy C. Hogdon
Division Sign-Off
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.