← Product Code [QTD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/QTD) · K232544

# Apollo ESG NXT System, Apollo REVISE NXT System (K232544)

_Apollo Endosurgery, Inc. · QTD · Sep 18, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/QTD/K232544

## Device Facts

- **Applicant:** Apollo Endosurgery, Inc.
- **Product Code:** [QTD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/QTD.md)
- **Decision Date:** Sep 18, 2023
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5983
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Apollo ESG NXT System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss by reducing stomach volume through endoscopic sleeve gastroplasty in adult patients with obesity with BMI between 30-50 kg/m2 who have not been able to lose weight, or maintain weight loss, through more conservative measures. The Apollo REVISE NXT System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss in adult patients with between 30-50 kg/m2 by enabling transoral outlet reduction as a revision to a previous bariatric procedure.

## Device Story

Endoscopic suturing system for gastric remodeling; used by gastroenterologists or surgeons in clinical settings. System includes needle driver assembly, anchor exchange device, tissue helix, suture cinch, and suture-anchor assembly. Physician attaches needle driver to single-channel endoscope; uses tissue helix to acquire gastric tissue; actuates needle driver handle to place anchor-sutures and locking cinches. 'MaxFlex' pull-string feature allows additional control of endoscope retroflexion. Procedure reduces stomach volume (ESG) or revises bariatric outlet (TORe) to facilitate weight loss. Devices are sterile, single-use, and compatible with 1.5T/3T MRI.

## Clinical Evidence

No new clinical data presented. Substantial equivalence relies on clinical data from the predicate De Novo (DEN210045). The procedures, implants, and training requirements remain identical to the predicate, providing reasonable assurance of equivalent weight loss outcomes and adverse event profiles.

## Technological Characteristics

Materials: Thermoplastic, silicone, stainless steel. Sensing/Actuation: Mechanical needle driver assembly with pull-string retroflexion control. Form factor: Single-use, compatible with 8.8-9.8 mm diameter single-channel endoscopes. Sterilization: Ethylene Oxide (EO). Standards: Biocompatibility per ISO 10993; Packaging per ASTM F2096-11, F1980-21, D4169-22, F88/F88M-21.

## Regulatory Identification

This endoscopic device uses suturing to approximate gastric tissue to restrict the volume of the stomach for the intended purpose of weight loss.

## Predicate Devices

- Apollo ESG Sx System, Apollo Revise Sx System ([DEN210045](/device/DEN210045.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 18, 2023

Boston Scientific Inc. James Shene Senior Regulatory Affairs Specialist 1120 South Capital of Texas Hwy Suite 300 Austin, Texas 78746

Re: K232544

Trade/Device Name: Apollo ESG NXT System, Apollo REVISE NXT System Regulation Number: 21 CFR 876.5983 Regulation Name: Endoscopic suturing device for altering gastric anatomy for weight loss Regulatory Class: Class II Product Code: QTD Dated: August 18, 2023 Received: August 22, 2023

Dear James Shene:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Glenn B. Bell -S

Glenn B. Bell Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232544

Device Name Apollo ESG NXT System

#### Indications for Use (Describe)

The Apollo ESG NXT System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss by reducing stomach volume through endoscopic sleeve gastroplasty in adult patients with obesity with BMI between 30-50 kg/m2 who have not been able to lose weight, or maintain weight loss, through more conservative measures.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                  |                                                                                                |                                                                                      |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <table border="0"><tr><td><div> <span> <b> </b>        Prescription Use (Part 21 CFR 801 Subpart D)       </span> </div></td></tr><tr><td><div> <span>        Over-The-Counter Use (21 CFR 801 Subpart C)       </span> </div></td></tr></table> | <div> <span> <b> </b>        Prescription Use (Part 21 CFR 801 Subpart D)       </span> </div> | <div> <span>        Over-The-Counter Use (21 CFR 801 Subpart C)       </span> </div> |
| <div> <span> <b> </b>        Prescription Use (Part 21 CFR 801 Subpart D)       </span> </div>                                                                                                                                                   |                                                                                                |                                                                                      |
| <div> <span>        Over-The-Counter Use (21 CFR 801 Subpart C)       </span> </div>                                                                                                                                                             |                                                                                                |                                                                                      |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K232544

Device Name Apollo Revise NXT System

#### Indications for Use (Describe)

The Apollo REVISE NXT System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss in adult patients with between 30-50 kg/m2 by enabling transoral outlet reduction as a revision to a previous bariatric procedure.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)   </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)   </span>  |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Boston Scientific. The words "Boston Scientific" are written in a serif font. The color of the text is blue. 

Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, MA 01752 (508) 382-4000 www.bostonscientific.com

# 510(k) Summary

| Owner's Name & Address: | Boston Scientific Coporation<br>300 Boston Scientific Way<br>Marlborough, MA 01752                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:         | James Shene<br>Senior Regulatory Affairs Specialist<br>(512) 852-3793<br>james.shene@apolloendo.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Date:                   | September 14, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Trade Name:             | Apollo ESG NXTTM System, Apollo REVISE NXTTM System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common Name:            | Endoscopic suturing device for altering gastric anatomy for<br>weight loss                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Product Code:           | QTD                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification:         | Class II (21 CFR 876.5983)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Classification Name     | Endoscopic suturing device for altering gastric anatomy for<br>weight loss                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Devices:      | DEN210045-Apollo ESG Sx System, Apollo Revise Sx System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Description      | These devices are used to perform gastric remodelling through<br>the placement of anchor-sutures and locking cinch device. The<br>Apollo ESG NXT system is for intended for endoscopic sleeve<br>gastroplasty (ESG) while the Apollo REVISE NXT device is<br>intended for transoral outlet reduction, as a revision to a previous<br>bariatric procedure. The two device systems differ in terms of the<br>number of anchor-sutures and cinches needed to perform ESG<br>and TORe procedures.<br>Both systems are comprised of a Needle Driver Assembly and<br>Anchor Exchange Device, and accessories such as the Tissue<br>Helix, Suture Cinch and Suture-Anchor Assembly devices. All<br>devices are sterile packaged and designed for single use and are<br>manufactured from various thermoplastic, silicone, stainless steel<br>and other medical grade materials. |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Boston Scientific. The logo is in blue and features the words "Boston" on the top line and "Scientific" on the bottom line. The font is a serif font and the letters are all capitalized.

bostonscientific.com

## Apollo ESG NXT and Apollo REVISE NXT

The Apollo ESG NXT and Apollo REVISE NXT needle driver assembies are designed to be compatible with single channel endoscopes. The endcap assembly of the needle driver is mounted onto the endoscope using polyester medical tape. The external catheter sheath has two working channels through which the Anchor Exchange and other accessories can operate, independent of the endoscope channel. The needle driver assembly also incorporates a pull string feature that, when pulled, allows the physician to further control the retroflexion capabilities of the endoscope.

The handle of the Needle Driver Assembly is squeezed to actuate the needle body and exchange the proprietary Suture-Anchor Assembly with the Anchor Exchange to perform stitching operations. A previously cleared Tissue Helix is provided with Apollo ESG NXT while the NXT Tissue Helix Pro is provided with Apollo REVISE NXT. Both instruments are used for manipulating tissue into the suturing window of the needle driver assembly. These devices also include a matching number of anchor-sutures and cinches to perform the gastroplasty or revision procedure.

## NXT Tissue Helix Pro

The NXT Tissue Helix Pro is used to acquire tissue by rotating the device's handle to to gather tissue onto the exposed exposed helix coil. The acquired tissue is then pulled into proximity of the needle body to complete the stitching operation. Tissue is released by rotating the handle in the opposite direction.

Indications for Use: The APOLLO ESG System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss by reducing stomach volume through endoscopic sleeve gastroplasty in adult patients with obesity with BMI 30 -50 kg/m² who have not been able to lose weight, or maintain weight loss, through more conservative measures.

> The APOLLO REVISE System is intended to be used by trained gastroenterologists or surgeons that perform bariatric procedures to facilitate weight loss in adult patients with obesity with BMI 30 - 50 kg/m² by enabling transoral outlet reduction as a revision to a previous bariatric procedure.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Boston Scientific. The logo is in blue and features the words "Boston" on the top line and "Scientific" on the bottom line. The font is a serif typeface, and the overall design is clean and professional.

Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, MA 01752 (508) 382-4000 www.bostonscientific.com

| Technological Characteristics: | Modifications to design and materials as compared to the predicate device cleared via DEN210045 are as follows:<br>Apollo ESG NXT and Apollo Revise NXT <ul><li>"MaxFlex" Feature: A new feature, integrated into the needle driver assembly, that incorporates the use of a "pull string" to provide physicians with the ability to further control the retroflexion capabilities of their endoscope. This feature does not need to be used by the physician.</li><li>Endoscope Attachment Method : The method and materials used to attach the needle driver to the endoscope were updated.</li><li>Design/Material Modifications: New materials and design modifications were implemented in order to improve product durability and the manufacturability of the device.</li></ul> |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                | NXT Tissue Helix Pro<br>The design of this tissue helix has been updated to lower catheter stiffness and improve maneuverability. Additionally, the handle interface was updated to allow the device to be controlled entirely by the physician and the helix tip was updated to an electro-cut design.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Non-Clinical Performance Data: | Appropriate product testing was performed on the subject device to evaluate conformance to product specifications and equivalence to the predicate devices. Verification and validation for the proposed system of devices was conducted in accordance with written protocols, previously applied to the predicate device. All devices were evaluated to their individual functional and reliability requirements, as well as system compatibility.                                                                                                                                                                                                                                                                                                                                    |
|                                | Performance Testing<br>Bench testing included suture drag testing, tensile testing, torque load testing, endoscope compatibility, sterility, reliability and bond strength. Additionally, the Tissue Helix Pro device was tested for tissue acquisition reliability.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                | Functional Testing<br>An ex vivo model was utilized, using the same test methods as the predicate to evaluate the device function and intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                | Packaging Integrity<br>Packaging Integrity was confirmed by repeating testing in accordance with ASTM F2096-11, ASTM F1980-21, ASTM D4169-22 and ASTM F88/F88M-21.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                | Acceptance criteria were met in each of the studies listed above. The results confirmed equivalency between the subject and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

predicate devices, and that no new issues of safety or efficacy were raised.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Boston Scientific. The logo is in blue and features the words "Boston Scientific" stacked on top of each other. The font is a serif typeface, and the overall design is clean and professional.

v.bostonscientific.com

Biocompatibility: Biocompatibility testing and toxicological assessments were performed on subject OverStitch devices in accordance with their risk category requirements, as defined in ISO 10933-1. Testing included cytotoxicity, irritation and skin sensitization, systemic toxicity, and material mediated pyrogenicity.

Clinical Performance Data: The DEN210045 clearance associated with the predicate devices created a special controls. One control was for clinical performance, specifically that the device performs as intended under anticipated conditions of use evaluating weight loss and adverse events. Another control was to ensure physicians were trained on the use of the systems to remodel gastric anatomy. These devices, which use the De Novo clearance devices as predicates, leverage the clinical data presented in that De Novo. The ESG and outlet revisions procedures performed with these devices are the same (same implant placements and the same implants), and training requirements are the same. There is reasonable assurance that the weight loss and adverse events will the equivalent to the predicate.

Basis of Substantial Equivalence: Based on a comparison of indications for use and technological characteristics, the proposed devices have demonstrated substantial equivalence to their predicates.

|                            | Predicate Apollo ESG Sx and Apollo Revise<br>Sx (DEN210045)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Apollo ESG NXT and<br>Apollo Revise NXT                                                      |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Indications for Use        | The APOLLO ESG System is intended to be<br>used by trained gastroenterologists or surgeons<br>that perform bariatric procedures to facilitate<br>weight loss by reducing stomach volume<br>through endoscopic sleeve gastroplasty in adult<br>patients with obesity with BMI 30 -50 kg/m2 who<br>have not been able to lose weight, or maintain<br>weight loss, through more conservative<br>measures.<br>The APOLLO REVISE System is intended to be<br>used by trained gastroenterologists or surgeons<br>that perform bariatric procedures to facilitate<br>weight loss in adult patients with obesity with<br>BMI 30 - 50 kg/m2 by enabling transoral outlet<br>reduction as a revision to a previous bariatric<br>procedure. | Same, with the revision of the<br>product name (APOLLO ESG<br>NXT and APOLLO REVISE<br>NXT). |
| Endoscope<br>Compatibility | Compatible with single channel endoscopes<br>having an outer diameter range of 8.8 to 9.8 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                         |
| Working Length             | Maximum working length of 110 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                         |
| Sterilization Method       | EO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                         |
| Shelf Life                 | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1 year                                                                                       |
| Usage                      | Single-use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                         |

#### Table 1: Summary Table of Equivalence

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the logo for Boston Scientific. The logo is in blue and features the words "Boston Scientific" stacked on top of each other. The font is a serif font, and the overall design is simple and professional.

Boston Scientific Corporation
300 Boston Scientific Way
Marlborough, MA 01752 (508) 382-4000 www.bostonscientific.com

|                                                 | Predicate Apollo ESG Sx and Apollo Revise<br>Sx (DEN210045)                                                                                                                                                                                                          | Apollo ESG NXT and<br>Apollo Revise NXT |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Functional<br>Durability                        | Validated to perform 8 stitches for each of 8<br>sutures.                                                                                                                                                                                                            | Same                                    |
| Biocompatibility                                | Tested per ISO 19993.                                                                                                                                                                                                                                                | Same                                    |
| MR Compatibility                                | Safe with 1.5 and 3 T MR scanners with spatial<br>field gradient of 2500 Gauss/cm (extrapolated or<br>less) and SAR of 2.0 W/kg for 15 minutes of<br>continuous scanning                                                                                             | Same                                    |
| Labeling<br>(Special Control<br>from DEN210045) | Labeling must include a summary of clinical<br>performance testing with the device, including a<br>discussion of adverse event and clinical benefit<br>reported as percent total body weight loss; and<br>shelf-life.                                                | Same                                    |
| Training<br>(Special Control<br>from DEN210045) | Training must be provided so that upon<br>completion of the training program, the user can<br>use the device correctly to approximate tissue to<br>alter the gastric anatomy for the purpose of<br>weight loss with minimal implact to the safety of<br>the patient. | Same                                    |

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/QTD/K232544](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/QTD/K232544)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
