DEN180055 · Bellco Srl (Now Part of Medtronic, Inc) · QIR · Apr 29, 2020 · Gastroenterology, Urology
Device Facts
Record ID
DEN180055
Device Name
CARPEDIEM System
Applicant
Bellco Srl (Now Part of Medtronic, Inc)
Product Code
QIR · Gastroenterology, Urology
Decision Date
Apr 29, 2020
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 876.5861
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CARPEDIEM System is indicated for use in acute kidney injury or fluid overloaded patients requiring hemodialysis or hemofiltration therapy. It is intended to provide continuous renal replacement therapy (CRRT) to patients weighing between 2.5 and 10 kilograms.
Device Story
CARPEDIEM System provides continuous renal replacement therapy (CRRT) for neonates and infants (2.5-10 kg). Device functions as artificial kidney; removes toxins/excess fluid via hemodialyzer with semipermeable membrane using convection (ultrafiltration) and diffusion (concentration gradient). System includes hemodialysis delivery machine with automated ultrafiltration controller; monitors/controls blood and dialysate pump rates and fluid removal rates. Blood circulates through tubing set connected to hemodialyzer. Operated by clinicians trained in CRRT and complication management. Output allows clinicians to manage fluid balance and solute clearance in critically ill pediatric patients.
Clinical Evidence
Clinical performance testing required to confirm safety, accuracy, precision, and reproducibility of non-clinical data under anticipated conditions of use. Specific clinical metrics or study results are not detailed in the provided document; however, clinical performance testing is mandated as a special control to support the safety and effectiveness of the device.
Technological Characteristics
System comprises hemodialysis delivery machine, hemodialyzer with semipermeable membrane, and blood tubing sets. Features automated ultrafiltration controller, pumps, sensors, and scales. Requires biocompatibility, pyrogenicity, and sterility testing for patient-contacting components. Software-controlled monitoring of flow rates and pressures. Electrical safety and EMC compliance required.
Indications for Use
Indicated for pediatric patients weighing 2.5 to 10 kg with acute kidney injury or fluid overload requiring hemodialysis or hemofiltration therapy.
Regulatory Classification
Identification
A pediatric continuous renal replacement therapy hemodialysis system is a device intended for use as an artificial kidney system for the management of pediatric patients with acute kidney injury and/or fluid overload by performing such therapies as hemodialysis, hemofiltration, hemodiafiltration, and isolated ultrafiltration. Using a hemodialyzer with a semipermeable membrane, the hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via ultrafiltration) and/or diffusion (via a concentration gradient in dialysate). The hemodialysis delivery machine, with an automated ultrafiltration controller, controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. During treatment, a patient's blood is circulated through the blood tubing set connected to the hemodialyzer's blood compartment. Blood access devices and accessories for hemodialysis required for the prescribed treatment are regulated under § 876.5540.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must confirm the safety and the accuracy, precision, and reproducibility of the non-clinical performance data under anticipated conditions of use.
(2) Usability testing must demonstrate that a user can correctly use the hemodialysis delivery device based solely on reading the instructions for use.
(3) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Hemodialysis delivery system performance testing must include:
(A) Fluid flow accuracy testing; and
(B) Functional testing of system components including sensors, pumps, and scales to acceptance criteria.
(ii) Hemodialyzer performance testing must include:
(A) Ultrafiltration;
(B) Blood and dialysate pressure drop;
(C) Clearance rates;
(D) Sieving coefficients;
(E) Mechanical hemolysis;
(F) Structural integrity;
(G) Blood compartment integrity;
(H) Volume of the blood compartment; and
(I) Chemical analysis of the dialyzer membrane.
(iii) Blood tubing set performance testing must include:
(A) Pressure leak testing;
(B) Worst-case endurance testing;
(C) Priming volume assessment;
(D) Tensile testing of joints and materials of all tubing segments;
(E) Pressure transducer leak testing;
(F) Clamp occlusion;
(G) Mechanical hemolysis; and
(H) Kink testing.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and wireless compatibility of the device.
(6) The tissue-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(8) Performance data must validate the reprocessing instructions for the reusable components of the device.
(9) The patient-contacting components of the device must be demonstrated to be non-pyrogenic.
(10) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(11) Device labeling must include:
(i) Hemodialysis delivery system labeling must provide detailed information regarding the safe use of the dialysis machine, including:
(A) Overall description of the device and individual components or accessories labeled for use with the delivery system;
(B) Description of the safety-related components included in the system;
(C) Identification of operational parameters;
(D) Alarms and troubleshooting information;
(E) Cleaning, disinfection, and preventative maintenance procedures; and
(F) A statement that the device is intended for use by operators trained in the administration of continuous renal replacement therapy and in the management of its complications.
(ii) Hemodialyzer labeling must include:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Instructions for the preparation of the hemodialyzer, initiation of dialysis, troubleshooting, and discontinuance of dialysis;
(E) Membrane surface area, priming (blood) volume, maximum transmembrane pressure, maximum blood flow and maximum dialysate rate for each model;
(F) Summary of the in vitro performance data; and
(G) A non-pyrogenic statement.
(iii) Blood tubing set labeling must provide detailed information regarding the safe use of the device, including:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Identification of the components in the package;
(E) Total length of the arterial and venous tubing sets;
(F) Outer diameter (OD) of the pump segment;
(G) Priming volume;
(H) Identification of the hemodialysis delivery systems which are compatible with the blood tubing set;
(I) Identification of the largest gauge needle that can be used with the injection port, if applicable; and
(J) Identification of the maximum operating pressures for the transducer protectors.
In combination with the general controls of the FD&C Act, the pediatric continuous renal replacement therapy system is subject to the following special controls:
(1) Clinical performance testing must confirm the safety and the accuracy, precision, and reproducibility of the non-clinical performance data under anticipated conditions of use.
(2) Usability testing must demonstrate that a user can correctly use the hemodialysis delivery device based solely on reading the instructions for use.
(3) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Hemodialysis delivery system performance testing must include:
(A) Fluid flow accuracy testing; and
(B) Functional testing of system components including sensors, pumps and scales to acceptance criteria.
(ii) Hemodialyzer performance testing must include:
(A) Ultrafiltration;
(B) Blood and dialysate pressure drop;
(C) Clearance rates;
(D) Sieving coefficients;
(E) Mechanical hemolysis;
(F) Structural integrity;
(G) Blood compartment integrity;
(H) Volume of the blood compartment; and
(I) Chemical analysis of the dialyzer membrane.
(iii) Blood tubing set performance testing must include:
(A) Pressure leak testing;
(B) Worst-case endurance testing;
(C) Priming volume assessment;
(D) Tensile testing of joints and materials of all tubing segments;
(E) Pressure transducer leak testing;
(F) Clamp occlusion;
(G) Mechanical hemolysis; and
(H) Kink testing.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and wireless compatibility of the device.
(6) The tissue-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(8) Performance data must validate the reprocessing instructions for the reusable components of the device.
(9) The patient-contacting components of the device must be demonstrated to be non-pyrogenic.
(10) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(11) Device labeling must include:
(i) Hemodialysis delivery system labeling must provide detailed information regarding the safe use of the dialysis machine, including:
(A) Overall description of the device and individual components or accessories labeled for use with the delivery system;
(B) Description of the safety-related components included in the system;
(C) Identification of operational parameters;
(D) Alarms and troubleshooting information;
(E) Cleaning, disinfection, and preventative maintenance procedures; and
(F) A statement that the device is intended for use by operators trained in the administration of continuous renal replacement therapy and in the management of its complications
(ii) Hemodialyzer labeling must include:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Instructions for the preparation of the hemodialyzer, initiation of dialysis, troubleshooting, and discontinuance of dialysis;
(E) Membrane surface area, priming (blood) volume, maximum transmembrane pressure, maximum blood flow and maximum dialysate rate for each model;
(F) Summary of the in vitro performance data; and
(G) A non-pyrogenic statement.
(iii) Blood tubing set labeling must provide detailed information regarding the safe use of the device, including:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Identification of the components in the package;
(E) Total length of the arterial and venous tubing sets;
(F) Outer diameter (OD) of the pump segment;
(G) Priming volume;
(H) Identification of the hemodialysis delivery systems which are compatible with the blood tubing set;
(I) Identification of the largest gauge needle that can be used with the injection port, if applicable; and
(J) Identification of the maximum operating pressures for the transducer protectors.
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 29, 2020
Bellco S.r.l.
% Michele Gust
Sr Regulatory Affairs Director
Medtronic, Inc
710 Medtronic Parkway
Minneapolis, MN 55432
Re: DEN180055
Trade/Device Name: CARPEDIEM System
Regulation Number: 21 CFR 876.5861
Regulation Name: Pediatric continuous renal replacement therapy system
Regulatory Class: II
Product Code: QIR
Dated: October 5, 2018
Received: October 9, 2018
Dear Michele Gust:
The Center for Devices and Radiological Health (CDRH) of the Food and Drug Administration (FDA) has completed its review of your De Novo request for classification of the CARPEDIEM System, a prescription device under 21 CFR Part 801.109 with the following indications for use:
The CARPEDIEM System is indicated for use in acute kidney injury or fluid overloaded patients requiring hemodialysis or hemofiltration therapy. It is intended to provide continuous renal replacement therapy (CRRT) to patients weighing between 2.5 and 10 kilograms.
FDA concludes that this device should be classified into Class II. This order, therefore, classifies the CARPEDIEM System, and substantially equivalent devices of this generic type, into Class II under the generic name Pediatric Continuous Renal Replacement Therapy System.
FDA identifies this generic type of device as:
**Pediatric continuous renal replacement therapy system.** A pediatric continuous renal replacement therapy hemodialysis system is a device intended for use as an artificial kidney system for the management of pediatric patients with acute kidney injury and/or fluid overload by performing such therapies as hemodialysis, hemofiltration, hemodiafiltration, and isolated ultrafiltration. Using a hemodialyzer with a semipermeable membrane, the hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via ultrafiltration) and/or diffusion (via a concentration gradient in dialysate). The hemodialysis delivery machine, with an automated ultrafiltration controller, controls and monitors the parameters related to this processing, including the
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
DEN180055 - Michele Gust
Page 2
rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. During treatment, a patient’s blood is circulated through the blood tubing set connected to the hemodialyzer’s blood compartment. Blood access devices and accessories for hemodialysis required for the prescribed treatment are regulated under 21 CFR § 876.5540.
Section 513(f)(2) of the Food, Drug and Cosmetic Act (the FD&C Act) was amended by section 607 of the Food and Drug Administration Safety and Innovation Act (FDASIA) on July 9, 2012. This law provides two options for De Novo classification. First, any person who receives a "not substantially equivalent" (NSE) determination in response to a 510(k) for a device that has not been previously classified under the Act may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act. On December 13, 2016, the 21st Century Cures Act removed a requirement that a De Novo request be submitted within 30 days of receiving an NSE determination. Alternatively, any person who determines that there is no legally marketed device upon which to base a determination of substantial equivalence may request FDA to make a risk-based classification of the device under section 513(a)(1) of the Act without first submitting a 510(k). FDA shall, within 120 days of receiving such a request, classify the device. This classification shall be the initial classification of the device. Within 30 days after the issuance of an order classifying the device, FDA must publish a notice in the Federal Register announcing the classification.
On October 9, 2018, FDA received your De Novo requesting classification of the CARPEDIEM System. The request was submitted under section 513(f)(2) of the FD&C Act. In order to classify the CARPEDIEM System into class I or II, it is necessary that the proposed class have sufficient regulatory controls to provide reasonable assurance of the safety and effectiveness of the device for its intended use. After review of the information submitted in the De Novo request, FDA has determined that, for the previously stated indications for use, the CARPEDIEM System can be classified in class II with the establishment of special controls for class II. FDA believes that class II (special) controls provide reasonable assurance of the safety and effectiveness of the device type. The identified risks and mitigation measures associated with the device type are summarized in the following table:
| Identified Risks to Health | Mitigation Measures |
| --- | --- |
| Adverse tissue reaction | Biocompatibility evaluation
Pyrogenicity testing
Non-clinical performance testing |
| Death | Labeling
Clinical performance testing
Usability testing |
| Infection | Labeling
Reprocessing validation
Pyrogenicity testing
Shelf life testing
Usability testing |
| Inadequate or incomplete treatment | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing |
{3}
DEN180055 - Michele Gust
Page 3
| | Usability testing |
| --- | --- |
| Clearance of essential blood substances or medications | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing
Usability testing |
| Blood loss or blood cell destruction | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing |
| Thermal injury | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing
Usability testing |
| Blood leak into the dialysis fluid | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing |
| Fluid imbalance | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing |
| Air embolism | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing
Usability testing |
| Fluid pump(s) reversal resulting in air infusion via the arterial bloodline | Non-clinical performance testing
Clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Shelf-life testing
Usability testing |
| Electrical shock | Electrical safety testing |
| Electromagnetic interference with other devices/equipment | Electromagnetic compatibility (EMC) testing |
In combination with the general controls of the FD&C Act, the pediatric continuous renal replacement therapy system is subject to the following special controls:
{4}
DEN180055 - Michele Gust
Page 4
(1) Clinical performance testing must confirm the safety and the accuracy, precision, and reproducibility of the non-clinical performance data under anticipated conditions of use.
(2) Usability testing must demonstrate that a user can correctly use the hemodialysis delivery device based solely on reading the instructions for use.
(3) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Hemodialysis delivery system performance testing must include:
(A) Fluid flow accuracy testing; and
(B) Functional testing of system components including sensors, pumps and scales to acceptance criteria.
(ii) Hemodialyzer performance testing must include:
(A) Ultrafiltration;
(B) Blood and dialysate pressure drop;
(C) Clearance rates;
(D) Sieving coefficients;
(E) Mechanical hemolysis;
(F) Structural integrity;
(G) Blood compartment integrity;
(H) Volume of the blood compartment; and
(I) Chemical analysis of the dialyzer membrane.
(iii) Blood tubing set performance testing must include:
(A) Pressure leak testing;
(B) Worst-case endurance testing;
(C) Priming volume assessment;
(D) Tensile testing of joints and materials of all tubing segments;
(E) Pressure transducer leak testing;
(F) Clamp occlusion;
(G) Mechanical hemolysis; and
(H) Kink testing.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and wireless compatibility of the device.
(6) The tissue-contacting components of the device must be demonstrated to be biocompatible.
(7) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(8) Performance data must validate the reprocessing instructions for the reusable components of the device.
(9) The patient-contacting components of the device must be demonstrated to be non-pyrogenic.
(10) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(11) Device labeling must include:
(i) Hemodialysis delivery system labeling must provide detailed information regarding the safe use of the dialysis machine, including:
(A) Overall description of the device and individual components or accessories labeled for use with the delivery system;
(B) Description of the safety-related components included in the system;
{5}
DEN180055 - Michele Gust
Page 5
(C) Identification of operational parameters;
(D) Alarms and troubleshooting information;
(E) Cleaning, disinfection, and preventative maintenance procedures; and
(F) A statement that the device is intended for use by operators trained in the administration of continuous renal replacement therapy and in the management of its complications
(ii) Hemodialyzer labeling must include:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Instructions for the preparation of the hemodialyzer, initiation of dialysis, troubleshooting, and discontinuance of dialysis;
(E) Membrane surface area, priming (blood) volume, maximum transmembrane pressure, maximum blood flow and maximum dialysate rate for each model;
(F) Summary of the in vitro performance data; and
(G) A non-pyrogenic statement.
(iii) Blood tubing set labeling must provide detailed information regarding the safe use of the device, including:
(A) Description of compatibility;
(B) Shelf life;
(C) Storage conditions;
(D) Identification of the components in the package;
(E) Total length of the arterial and venous tubing sets;
(F) Outer diameter (OD) of the pump segment;
(G) Priming volume;
(H) Identification of the hemodialysis delivery systems which are compatible with the blood tubing set;
(I) Identification of the largest gauge needle that can be used with the injection port, if applicable; and
(J) Identification of the maximum operating pressures for the transducer protectors.
In addition, this is a prescription device and must comply with 21 CFR 801.109.
Although this letter refers to your product as a device, please be aware that some granted products may instead be combination products. If you have questions on whether your product is a combination product, contact CDRHProductJurisdiction@fda.hhs.gov.
Section 510(m) of the FD&C Act provides that FDA may exempt a class II device from the premarket notification requirements under section 510(k) of the FD&C Act, if FDA determines that premarket notification is not necessary to provide reasonable assurance of the safety and effectiveness of the device type. FDA has determined premarket notification is necessary to provide reasonable assurance of the safety and effectiveness of the device type and, therefore, the device is not exempt from the premarket notification requirements of the FD&C Act. Thus, persons who intend to market this device type must submit a premarket notification containing information on the pediatric continuous renal replacement therapy system they intend to market prior to marketing the device.
{6}
DEN180055 - Michele Gust
Page 6
Please be advised that FDA's decision to grant this De Novo request does not mean that FDA has made a determination that your device complies with other requirements of the FD&C Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the FD&C Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and if applicable, the electronic product radiation control provisions (Sections 531-542 of the FD&C Act); 21 CFR 1000-1050.
A notice announcing this classification order will be published in the Federal Register. A copy of this order and supporting documentation are on file in the Dockets Management Branch (HFA-305), Food and Drug Administration, 5630 Fishers Lane, Room 1061, Rockville, MD 20852 and are available for inspection between 9 a.m. and 4 p.m., Monday through Friday.
As a result of this order, you may immediately market your device as described in the De Novo request, subject to the general control provisions of the FD&C Act and the special controls identified in this order.
For comprehensive regulatory information about medical devices and radiation-emitting products, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
If you have any questions concerning the contents of the letter, please contact Dr. Carolyn Neuland, at 301-796-6523
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D.
Director
OHT3: Office of GastroRenal, ObGyn,
General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.