DEN150001 · Alcresta, Inc. · PLQ · Nov 20, 2015 · Gastroenterology, Urology
Device Facts
Record ID
DEN150001
Device Name
RELIZORB
Applicant
Alcresta, Inc.
Product Code
PLQ · Gastroenterology, Urology
Decision Date
Nov 20, 2015
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 876.5985
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
RELIZORB™ is indicated for use in adults to hydrolyze fats in enteral formula.
Device Story
RELIZORB™ is a point-of-care, single-use, in-line cartridge for enteral feeding circuits; connects between infusion pump and patient feeding tube. Contains polymeric beads with immobilized lipase enzyme. As enteral formula flows through the cartridge, lipase hydrolyzes triglycerides into fatty acids and monoglycerides; mimics natural pancreatic lipase function. Used in clinical or home settings by trained operators or lay users. Output is pre-hydrolyzed formula, facilitating fat and nutrient absorption (e.g., Vitamins A, D, E, LCPUFAs). Benefits patients unable to absorb fats effectively. Does not require external power or electronic processing; purely chemical/mechanical enzymatic action.
Clinical Evidence
Special controls require in vivo testing to demonstrate no adverse tissue response or adverse performance. Non-clinical testing required for mechanical integrity, flow rate/leakage, enzymatic effect, enzyme leaching characterization, nutritional composition impact, and enteral pump alarm compatibility. Human factors testing required to characterize use error risks.
Technological Characteristics
In-line cartridge containing inert polymer with covalently bound enzyme. Components must be biocompatible. Designed to withstand clinical forces, maintain flow rates, and not impede enteral feeding pump alarms. Prescription device.
Indications for Use
Indicated for use in adults to hydrolyze fats in enteral formula.
Regulatory Classification
Identification
An enzyme packed cartridge is an ex vivo prescription device that is used in enzymatic hydrolysis of macronutrients into their essential nutrient forms at the time of delivery. The device consists of an outer casing containing an inert polymer with a covalently bound enzyme through which nutritional formula is directed. The device fits in line with enteral feeding systems.
Special Controls
In combination with the general controls of the FD&C Act, the Enzyme Packed Cartridge is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) The patient contacting components of the device must be demonstrated to be biocompatible.
(2)
*In vivo* testing must be performed and must demonstrate that the device causes neither an adverse tissue response nor adverse performance.(3) Non-clinical testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
(i) Mechanical testing to demonstrate that the device can withstand clinical forces;
(ii) Flow rate and leakage testing to demonstrate that the device does not impede the flow of enteral formula;
(iii) Demonstration of enzymatic effect on intended macronutrient;
(iv) The amount of enzyme that exits the cartridge must be characterized;
(v) Validation that the device does not adversely impact the nutritional composition of enteral formula; and
(vi) Validation that the device does not impede flow alarms on enteral feeding pumps.
(4) Human factors testing must be performed to characterize use error risks.
(5) Performance data must support shelf life by demonstrating package integrity and device functionality over the identified shelf life.
(6) Labeling must include the following:
(i) A detailed summary of
*in vivo* testing pertinent to use of the device, including device-related adverse events;(ii) A detailed summary of compatible formulas that is supported by non-clinical testing, including the expected enzymatic conversion as a percentage;
(iii) Detailed instructions on how to place the device into an enteral feeding circuit;
(iv) A warning regarding the possibility for misconnections; and
(v) Expiration date or shelf life.
(7) Patient labeling must be provided and must include:
(i) Relevant warnings, precautions, adverse effects, and complications;
(ii) A description of the device and how it operates;
(iii) Instructions on how to correctly use the device; and
(iv) The benefits and risks associated with the use of the device.
Submission Summary (Full Text)
{1}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR RELIZORB™
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Enzyme Packed Cartridge: The enzyme packed cartridge is an ex vivo prescription device that is used in enzymatic hydrolysis of macronutrients into their essential nutrient forms at the time of delivery. The device consists of an outer casing containing an inert polymer with a covalently bound enzyme through which nutritional formula is directed. The device fits in line with enteral feeding systems.
NEW REGULATION NUMBER: 21 CFR 876.5985
CLASSIFICATION: II
PRODUCT CODE: PLQ
#### BACKGROUND
DEVICE NAME: RELIZORBTM
SUBMISSION NUMBER: DEN150001
DATE OF DE NOVO: DECEMBER 17, 2014
CONTACT:
ALCRESTA, INC. One Newton Executive Park SUITE 202 NEWTON, MA 02462
#### INDICATIONS FOR USE
RELIZORB™ is indicated for use in adults to hydrolyze fats in enteral formula.
#### LIMITATIONS
The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR §801.109.
RELIZORB™ is for use with enteral feeding only. The "Feed Only" lettering on the RELIZORB™ purple outlet is there to identify that the RELIZORB™ cartridge is intended for connection to enteral feeding lines only.
Medications should not be administered through the RELIZORB™ cartridge. Do
{2}------------------------------------------------
not add medications to the enteral feed line in between the pump and RELIZORB™ (before RELIZORB™). The passage of medications through RELIZORB™ may adversely affect the medications or the ability of RELIZORB™ to hydrolyze fats.
Do not re-use RELIZORB™. RELIZORB™ is a single-use product. Re-use may result in contamination of the product. If re-used, RELIZORB™ may not effectively hydrolyze fats.
Enteral formulas containing insoluble fiber should NOT be used. Insoluble fiber may clog the RELIZORB™ cartridge.
The use of RELIZORB™ along with porcine pancreatic enzyme replacement therapy (PERT) has not been investigated. The appropriate dose and administration of PERT should be evaluated on an individual basis. Patients should continue to follow physician's guidance and PERT product labeling when used in conjunction with RELIZORB™.
The use of RELIZORB™ has not been studied in patients with exocrine pancreatic insufficiency (0) (4) RELIZORB™ has not been evaluated in pediatric populations.
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS, AND CONTRAINDICATIONS.
## DEVICE DESCRIPTION
The RELIZORB™ device is a point-of-care accessory designed to fit in series with currently used enteral feeding circuits. During the submission process, the device was also known as the Enteral Feeding In-Line Cartridge (EFIC). Therefore, the subject device may be referred to as EFIC in some figures within this document. RELIZORB™ is designed to hydrolyze (break down) fats present in enteral formulas from triglycerides into fatty acids and monoglycerides to allow for their absorption and utilization by the body. This breakdown of fats by the RELIZORB™ is intended to mimic the function of the enzyme lipase in patients who do not excrete sufficient levels of pancreatic lipase. The subject device is shown below in Figure 1. The RELIZORB™ is comprised of a cvlindrical. hollow port and a single outlet port connection.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a filtration device with labels pointing to different parts. The labels indicate the inlet, filters, immobilized lipase, and outlet. The device is cylindrical in shape and appears to be made of plastic, with a white filter material visible inside.
Figure 1. Photograph of the RELIZORB™ device.
RELIZORB™ is packed with polymeric beads that have lipase enzyme immobilized on the surface. This lipase enzyme is Generally Regarded as Safe (GRAS) . The chemical action of the lipase enzyme is shown in Figure 2, where triglyceride molecules are broken into constituent monoglycerides and fatty acids. The food contacting substance (FCS) of (b) beads manufactured using the RELIZORBTM are (b) (4) . The lipase enzyme is process. The FCS was an effective notification chemically bound to the FCS and is intended to remain within the cartridge.
Image /page/2/Figure/3 description: This image shows the chemical reaction of a triglyceride being broken down by lipase. On the left side of the image, a triglyceride is shown with a glycerol backbone and three fatty acids attached. The lipase enzyme breaks down the triglyceride into a monoglyceride and two fatty acids, which are shown on the right side of the image.
Figure 2. The lipase enzyme breaks down triglycerides into monoglycerides and fatty acids.
RELIZORB™ is an intermediary between an enteral feeding source (infusion pump) and an implanted feeding tube, as shown in Figure 3. The distal end of compatible infusion pump administration sets (Figure 3A) should have a stepped connector (Christmas tree). This connector plugs into the proximal end of the RELIZORB™ device (Figure 3B). The distal end of the RELIZORB™ (Figure 3C) connects to the enteral funnel of an extension set (Figure 3D). This extension set connects to an enteral feeding tube on the patient, such as a nasogastric or gastrostomy tube.
{4}------------------------------------------------
Image /page/3/Figure/0 description: The image shows a final setup of a medical device with labels A, B, C, and D. The device includes a pump on the left side and a connection to the patient on the right side, labeled as 'Downstream'. A Y-connector extension set is present, and a note indicates that medications or other solutions may be added at point D.
# Figure 3. Illustration showing a proposed enteral feeding circuit with RELIZORB™!
## SUMMARY OF NON-CLINICAL/BENCH STUDIES
Non-clinical/bench studies conducted on the RELIZORB™ to demonstrate a reasonable assurance of safety and effectiveness of the device are summarized in the sections below.
## BIOCOMPATIBILITY/MATERIALS
| The RELIZORBTM cartridge is constructed using a | (b) (4) housing, | (b) (4) |
|-------------------------------------------------------------------------------|-------------------------------|----------------------|
| inlet connector, | (b) (4) outlet connector, and | (b) (4) filters. The |
| food contacting materials are the acrylic copolymer beads, lipase enzyme, and | | |
| (b) (4) stabilizer. | | |
All materials within the RELIZORB™ have indirect contact with the patient's gastrointestinal tract. The RELIZORB™ is categorized as a permanent (>30 days) contact device that is externally communicating. The following biocompatibility testing was conducted on the final RELIZORB™ device in Table 1.
| Test | Purpose | Results |
|----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Cytotoxicity – MEM Elution<br>Test (ISO 10993-5: 2009<br>(b) (4) | To assess the biological activity of L-929 mouse fibroblast cells after exposure to extracts prepared from the final RELIZORB™ device. | Non-cytotoxic |
| Irritation – Intracutaneous<br>reactivity (ISO 10993-10:<br>2010) | To determine if extracts from the final RELIZORB™ device produce an irritation reaction when injected intracutaneously in the rabbit model. | Non-irritating |
| Sensitization – Guinea Pig<br>Maximization Test (ISO<br>10993-10: 2010) | To determine the potential for sensitization of extracts prepared from the final RELIZORB™ device using the guinea pig animal model. | Non-sensitizing |
| Acute systemic toxicity (ISO<br>10993-11: 2010) | To determine the potential for acute systemic toxicity of extracts prepared from the final RELIZORB™ device by injections into mice. | Not systemically toxic |
| Genotoxicity – In Vitro Gene<br>Mutations in Bacteria (Ames<br>Assay, ISO 10993-3: 2003<br>(b) (4) | To determine the potential mutagenic activity of extracts prepared from the final RELIZORB™ device by measuring reversion rates in bacteria. | Non-mutagenic |
| Genotoxicity – In Vitro<br>Mouse Lymphoma (ISO<br>10993-3: 2003 (b) (4) | To determine if extracts prepared from the final RELIZORB™ device induce forward mutations at the thymidine kinase locus | Non-mutagenic |
{5}------------------------------------------------
| Test | Purpose | Results |
|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Genotoxicity – In Vivo<br>Mouse Micronucleus Assay<br>(ISO 10993-3:2003 <span style="background-color:gray">(b) (4)</span> ) | To determine the potential for extracts prepared from the<br>final RELIZORBTM device induce micronuclei<br>formation in immature polychromatic erythrocytes in the<br>bone marrow of mice. | Non-mutagenic |
# Table 1. RELIZORB™ Biocompatibility Testing.
The sponsor justified that for non-polar extractions, sesame oil would be degraded by the lipase enzyme. Because the sponsor did leachability testing on the lipase enzyme from the beads, the sponsor used a surrogate bead for non-polar extractions. This approach was deemed acceptable by FDA.
## PERFORMANCE TESTING - BENCH
The sponsor conducted bench tests to demonstrate mechanical integrity, package integrity, enzyme activity, and that RELIZORB™ can be integrated into existing enteral feeding circuits. Table 2 summarizes each of the bench tests.
| Test | Purpose | Acceptance Criteria | Results |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Torque strength | To determine the torque<br>necessary to separate the<br>small bore connectors from<br>the cartridge body at both<br>the distal and proximal ends<br>of RELIZORB™ | The sponsor estimated the<br>clinical force that<br>RELIZORB™ may<br>encounter and built in a<br>three-fold safety factor.<br>The sponsor defined a<br>passing result as a torque<br>at separation higher <span>(b) (4)</span> | RELIZORB™ met the acceptance<br>criteria for this test. |
| Tensile strength | To determine the force<br>required to separate the<br>device small bore<br>connectors from the<br>cartridge using a linear<br>tensile force. | RELIZORB™ shall have<br>a linear tensile<br><span>(b) (4)</span><br>break. | RELIZORB™ met the established<br>acceptance criteria. |
| Air leakage test | To establish that the<br>RELIZORB™'s material<br>bonds would not fail or leak<br>when challenged with<br>pressurized air. | RELIZORB™ material<br>bonds shall not leak when<br>challenged with <span>(b) (4)</span><br>compressed air. | RELIZORB™ did not leak when<br>pressurized with air. The device<br>met the acceptance criteria. |
| Filter integrity | To ensure that the<br>FCS/enzyme beads are<br>retained within the<br>cartridge. The sponsor<br>connected RELIZORB™ to<br>an enteral feeding circuit<br>and subjected the device to<br>the maximum pump flow<br>rate for five minutes. The<br>sponsor conducted this test<br>in both a forward and | RELIZORB™ should not<br>allow for beads <span>(b) (4)</span><br>to pass the filter and leave<br>the cartridge. | RELIZORB™ allowed for five<br>particles to leave the cartridge in<br>three repetitions in the forward<br>flow direction. RELIZORB™<br>allowed one particle to exit the<br>cartridge in the reverse flow<br>direction. Out of these six<br>particles, only one had a diameter<br><span>(b) (4)</span><br>The sponsor investigated the |
| Test | Purpose | Acceptance Criteria | Results |
| | reverse flow orientation<br>The sponsor collected any<br>potential escaped particles<br>during this test using a (b)(4)<br>filter. | | source of this particle using<br>spectroscopy and concluded that it<br>was a contamination from their<br>testing laboratory. The Agency<br>reviewed the sponsor's data and<br>agrees with their justification. |
| | | | Because the sponsor did observe<br>some particles exiting the<br>cartridge, FDA considered clinical<br>or toxicological adverse events<br>resulting from escaped particles. A<br>clinical assessment determined that<br>so few escaped beads would likely<br>pass in a patient's stool.<br>Additionally, the sponsor<br>submitted acceptable<br>biocompatibility data and the FCS-<br>enzyme complex are acceptable<br>for use in foods. Based on these<br>assessments, this is an acceptable<br>result. |
| Fat hydrolysis | Using a simulated enteral<br>circuit, this test is to<br>determine the amount of<br>free fatty acids (FFA) that<br>RELIZORB™ produces by<br>enzymatic hydrolysis. The<br>sponsor used a<br>commercially available<br>colorimetric assay to<br>determine the amount of<br>free fatty acids produced by<br>RELIZORB™ after<br>simulated use. | RELIZORB™ should<br>produce (b)(4)FFA per<br>serving. | The sponsor submitted bench<br>testing data demonstrating that<br>RELIZORB™ breaks down ≈90%<br>of fats in most enteral formulas.<br>The sponsor added these results to<br>their labeling. This is an<br>acceptable result. |
| Unconjugated<br>lipase analysis | The sponsor evaluated the<br>potential leaching of the<br>lipase enzyme from the<br>beads. The sponsor<br>extracted the beads for (b)(4)<br>at room<br>temperature. | There were no formal<br>acceptance criteria for<br>this test. The sponsor<br>measured the lipase<br>concentration using a | Using several lots, the sponsor<br>observed (b)(4) leaching by the<br>BCA assay and (b)(4) leaching by<br>absorbance. Based on the<br>biocompatibility testing and the<br>GRAS status of lipase, there are no<br>safety concerns with this small<br>amount of enzyme released. |
| Assessment of<br>impact to other<br>nutrients | Using simulated use<br>conditions, the sponsor<br>flowed enteral formula<br>through their device. The<br>sponsor analyzed | There were no formal<br>acceptance criteria for<br>this bench test. The<br>sponsor conducted a<br>nutritional analysis of | After exposure to RELIZORBTM,<br>there was no meaningful<br>difference for any vitamins or<br>minerals. This is an acceptable<br>result. |
| Test | Purpose | Acceptance Criteria | Results |
| | nutritional components<br>such as vitamins and<br>minerals to ensure that<br>RELIZORB™ does not<br>adversely affect other<br>nutrients in enteral formula. | vitamins and minerals. | |
| Flow rate | The sponsor conducted this<br>test to ensure that<br>RELIZORB™ does not<br>restrict the flow of formula<br>using an enteral feeding<br>pump. | The presence of<br>RELIZORB™ in an<br>enteral feeding circuit<br>should not affect the flow<br>rate of formula. | There were no statistical<br>differences between the flow rate<br>with or without RELIZORB™ in<br>the enteral feeding line. This result<br>is acceptable. |
| Liquid leakage<br>test | After simulated feeding, the<br>sponsor inspected material<br>joints to determine if the<br>device leaked during<br>priming or flow rate testing. | RELIZORB™ should not<br>leak under normal and<br>worst-case conditions. | The sponsor used two formulas<br>and two different enteral feeding<br>pumps. In all cases, the sponsor<br>did not observe any leaking from<br>any material joints. |
| Pump alarm<br>verification | After performing flow rate<br>testing, the sponsor kinked<br>the tubing while the pump<br>was in run mode to verify<br>that the flow error alarm<br>works both before and after<br>RELIZORB™. | RELIZORB™ shall not<br>cause pump alarm failure. | The sponsor verified that the flow<br>alarm sounds if the tubing<br>becomes occluded before or after<br>the RELIZORB™. This is an<br>acceptable result. |
{6}------------------------------------------------
{7}------------------------------------------------
Table 2. RELIZORB™ Bench Testing.
In accordance with the recommendations outlined in the guidance document "Safety Considerations to Mitigate the Risks of Miscomnections with Small-bore Connectors Intended for Enteral Applications," mechanical misconnection testing was conducted and a warning is included in the labeling to reduce the risk of misconnection.
## SHELF LIFE/STERILITY
RELIZORB™ is provided non-sterile in a foil package. RELIZORBTM is a single use only device that is disposed of after each use. The identified shelf life for RELIZORBTM is six months.
In support of the six-month shelf life, the sponsor completed several tests using real-time aged devices after simulated shipping. The shelf life tests have identical acceptance criteria to those identified in the Performance Testing - Bench section.
- Fat hydrolysis
- · Tensile strength
- · Filter integrity
- · Flow rate
- · Package integrity
- o Visual inspection
{8}------------------------------------------------
o Peel strength (ASTM F88)
o Bubble leak test (ASTM F2096)
The sponsor reported fat hydrolysis conversions that were in agreement with their baseline data:
| | TwoCal HN | Peptamen |
|------------|-----------|----------|
| t=0 | 51% | 85% |
| t=1 month | 44% | 89% |
| t=3 months | 49% | 84% |
| t=6 months | 47% | 92% |
Table 3. Fat hydrolysis results (% conversion of fats).
Because there was no meaningful difference between the baseline and aged product, the Agency determined that the fat hydrolysis results were acceptable. The tensile strength, filter integrity, and flow rate were not compromised after aging. After simulated shipping, the clean barrier for the RELIZORB™ was not compromised. The six-month shelf life is acceptable.
## PERFORMANCE TESTING - ANIMAL
The sponsor completed four animal tests using exocrine pancreatic insufficient (0) ( (b) (4) pig models.
The four animal studies are summarized below.
## 1. Chronic Porcine Study
## Protocol
The purpose of this study was to determine if the absorption of long-chain polyunsaturated fatty acids (LCPUFAs) is enhanced when formulas were prehydrolyzed with soluble microbial lipase.
Image /page/7/Picture/12 description: The image is mostly obscured by a gray block. The visible text says "hydrolyzed with soluble microbial lipase."
The pigs were monitored for a reduction of total and PUFA fecal fats, and change in coefficient of fat absorption (% CFA). The pigs were also monitored for an increase in arachidonic acid (AA) and docosahexaenoic acid (DHA) in plasma and tissues such as the liver, retina, heart, and fat. This result is consistent with increased absorption of LCPUFAs. After the study endpoint, the sponsor performed necropsy of study animals to harvest specified tissues, such as the liver, retina. heart, fat, red blood cells, hippocampus, cerebellum, and brain cortex.
{9}------------------------------------------------
## Results
There were no adverse clinical effects or pathologic macroscopic findings in the gut or liver after seven days of administration. Both the RO and CV pre-hydrolyzed formulas increased the absorption of LCPUFA as reflected by reduced total stool fat, fecal AA, and DHA. The sponsor included a summary showing the %CFA for the control and treated groups. The sponsor's data supports their conclusion that the enriched diets enhanced fat absorption with an improvement in %CFA of 20-30% in comparison to controls.
| WEEK | CFA% (MEAN ± SD) | | |
|-----------|------------------|-----------|-----------|
| | CONTROL | CV LIPASE | RO LIPASE |
| CONTROL | 66.5±7.8 | 51.7±14.3 | 76.9±1.9 |
| TREATMENT | 67±5.8 | 86.6±4.3* | 87.1±3.5* |
| P value | | 0.002 | 0.003 |
Shown is a mean (± SD) CFA values during Control and Treatment weeks for Control RO organe. Difference was consider sinnificant tocol 05 for company on helpers of and RO groups during treatment week, n=3/arm
Image /page/8/Figure/4 description: The image shows the title of a table. The title is "Table 4. Mean of 24 hour %CFA during 3 consecutive days (Study 1)". The table will likely contain data related to the mean percentage of CFA over a 24-hour period for three consecutive days.
## 2. Chronic Porcine Study
## Protocol
The sponsor tested the safety and effectiveness of continuous feeding of the RO lipase enzyme in the b)(4) porcine model over six weeks. The sponsor used the same lipase bound beads that are packed into the final device in this study.
During the study, the sponsor measured total and PUFA fecal fats, %CFA, and AA and DHA levels in both plasma and tissue. At study termination, the sponsor completed a gross postmortem examination of tissue and a blinded GLP histopathology examination.
## Results
The sponsor reported food intake and body weight of all pigs. The formula intake The spensor reported root make and body wergit of were its (b) (4) control pigs grew 2-4 kg per week. The sponsor observed a 38% and 53% reduction in omega-3 and omega-6 fecal LCPUFA in (0) pigs treated with the RO enzyme. Similarly, the sponsor reported a 66% and 50% respective reduction in fecal AA and DHA levels for (b)(0) pigs treated with RO enzyme. These data suggest that the inability for (0) (4) pigs to absorb fat was reversed by feeding with pre-hydrolyzed formula treated with the lipase enzyme. Additionally, the pigs fed a pre-hydrolyzed diet had a more normalized blood lipid profile, improved consumption of LCPUFA. and improved Vitamins A and E absorption.
The sponsor harvested the cranial alimentary. GI tract, and the liver to send the samples for a blinded, GLP histopathology examination. The sponsor included analyses on the tongue, esophagus, stomach, small intestines, large intestines, liver,
{10}------------------------------------------------
and liver fat. The pathologist found no safety signals relative to the consumption of the lipase enzyme. The pathologist had the following observations:
- Increased periportal fat in the liver (all study animals). .
- Inflammatory lesions in the liver, stomach, and tongue of (b)(4) ● animals, but not healthy control animals.
- . An increase in distribution and severity of inflammatory lesions in EPI animals compared to healthy control animals.
The sponsor's findings seemed related to the pigs' 1010 status and not the 10/4/ treatment group. Some of the findings showed milder and less frequent issues in the treatment (4) animals in comparison to (1) controls. The observations noted in the (b)(4) treatment group were not related to the treatment itself and were of low incidence. The histopathological examination followed GLP regulations and was complete. Because the pathologist's findings were also peer reviewed, this study supported the accuracy of the sponsor's conclusions.
# 3. 12 Day Efficacy study with Gastrostomy Tube Feeding
## Protocol
The sponsor conducted a simulated use study of RELIZORB™. The sponsor did nightly gastrostomy tube feeding over 12 days in the 014) porcine model.
During the last three days of the study, the sponsor collected two 24 hour stool and urine samples. On the last day after overnight fasting, the sponsor collected blood samples for protein and fat profiles, as well as DHA and eicosapentaenoic acid (EPA) measurements.
## Results
The sponsor did not report any adverse events during this study. While the food intake between (8) (4) control and [8] treatment pigs was similar, the sponsor reported a statistically significant decrease in the stool weight for the treatment group (p=0.014). The sponsor reported a statistically significant increase in %CFA, as shown in Figure 4.
{11}------------------------------------------------
Image /page/10/Figure/0 description: This bar graph compares the %CFA of PepAF and PepAF+EFIC. PepAF has a %CFA of approximately 42, while PepAF+EFIC has a %CFA of approximately 60. The error bars for PepAF+EFIC are larger than the error bars for PepAF. The y-axis ranges from 0 to 100.
Figure 4. Improved fat absorption expressed as %CFA for EFIC/RELIZORBIM (p=0.036, Study 3).
## 4. 24 Hour Pharmacodynamics Study
#### Protocol
The sponsor conducted a 24 hour, randomized, cross-over study in 11 pigs to assess the pharmacodynamics of gastrostomy tube feeding using a prototype of the RELIZORBTM device through a gastrostomy tube. While the sponsor used a device prototype for the treatment arm, this prototype was nearly identical in form and function to the finished device.
The sponsor collected baseline samples and at several time points after tube feeding began.
#### Results
Due to the short study duration, food consumption and body weight were not endpoints. The sponsor reported a statistically significant improvement in fat absorption in the treatment arm in comparison to the control. The plasma omega-3 fat (DHA and EPA) concentrations over 24 hours are shown in Figure 5. The sponsor reported a statistically significant (p <0.05) increase in plasma omega-3 fats when using RELIZORB™ in comparison to formula that was not prehydrolyzed with RELIZORB™
{12}------------------------------------------------
Image /page/11/Figure/0 description: The image contains two line graphs comparing the levels of DHA and EPA in different conditions over time. The left graph shows DHA levels (ng/mL) over 24 hours for three conditions: RELIZORB, No-RELIZORB, and HEALTHY. The right graph shows EPA levels (ng/mL) over the same 24-hour period for the same three conditions. Both graphs show that the RELIZORB and HEALTHY conditions have a higher peak in DHA and EPA levels around the 7-hour mark compared to the No-RELIZORB condition.
Figure 5. Plasma DHA and EPA concentration over 24h period after single administration of hydrolyzed formula (500 mL) using RELIZORB™ in a porcine model of exocrine pancreatic insufficiency(b) (4)
The sponsor completed a proof of concept (Study 1), safety (Study 2), and a simulated use study (Study 3) using the same enzyme as the finished device. These studies demonstrated safety using the enzyme and simulated use conditions. The sponsor demonstrated improvements in fat absorption and Vitamins D and E absorption in these studies. The sponsor also observed increased concentrations of LCPUFAs in several organs after the studies were completed. The Agency expects these results would also be seen in humans using pre-hydrolyzed formula due to the nutritional benefits observed from fat hydrolysis in the literature.
## HUMAN FACTORS TESTING
The sponsor conducted a human factors summative protocol in the submission. This study evaluated both the clinical care setting with trained operators and the home care setting with lay users. The sponsor conducted a subjective assessment on the performance of the critical tasks. The sponsor noted clinician errors after they neglected to change the device after one feeding session versus institutional policies that change feeding sets after 24 hours. This resulted in an update to the directions for use with a precaution that the device is for single use only. Based on the low risk of injury to the patient, the Agency determined that the sponsor's human factors activities are adequate.
## LABELING
The sponsor provided labeling that includes the physician's instructions for use, patient guide, and package labels for the RELIZORBTM
The labeling is acceptable and meets the requirements of 21 CFR 8801.109 for prescription devices. The patient guide also follows the principles identified in the FDA Guidance Document Medical Device Patient Labeling.
{13}------------------------------------------------
## RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of Enzyme Packed Cartridges and the measures necessary to mitigate these risks.
| Identified Risk | Mitigation Measure |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Adverse tissue reaction | Biocompatibility testing Non-clinical testing <i>In vivo</i> testing Labeling |
| Mechanical failure Deprivation of care Device clogging Filter becomes dislodged and releases beads into enteral formula | Non-clinical testing Shelf life testing Labeling |
| Reduced enzymatic effect | Non-clinical testing <i>In vivo</i> testing Shelf life testing Labeling |
| Use error | Human factors testing Labeling |
| Infection | Shelf life testing Labeling |
## SPECIAL CONTROLS:
In combination with the general controls of the FD&C Act, the Enzyme Packed Cartridge is subject to the following special controls:
- 1. The patient contacting components of the device must be demonstrated to be biocompatible.
- 2. In vivo testing must be performed and must demonstrate that the device causes neither an adverse tissue response nor adverse performance.
- 3. Non-clinical testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be demonstrated:
- (A). Mechanical testing to demonstrate that the device can withstand clinical forces.
- Flow rate and leakage testing to demonstrate that the device does not impede the (B). flow of enteral formula.
- Demonstration of enzymatic effect on intended macronutrient. (C).
- The amount of enzyme that exits the cartridge must be characterized. (D).
- Validation that the device does not adversely impact the nutritional composition (E). of enteral formula.
{14}------------------------------------------------
- (F). Validation that the device does not impede flow alarms on enteral feeding pumps.
- 4. Human factors testing must be performed to characterize use error risks.
- 5. Performance data must support shelf life by demonstrating package integrity and device functionality over the identified shelf life.
- 6. Labeling must include the following:
- (A). A detailed summary of in vivo testing pertinent to use of the device, including device-related adverse events.
- (B). A detailed summary of compatible formulas that is supported by non-clinical testing, including the expected enzymatic conversion as a percentage.
- (C). Detailed instructions on how to place the device into an enteral feeding circuit.
- A warning regarding the possibility for misconnections. (D).
- (E). Expiration date or shelf life.
- 7. Patient labeling must be provided and must include:
- (A). Relevant warnings, precautions, adverse effects, and complications.
- A description of the device and how it operates. (B).
- Instructions on how to correctly use the device. (C).
- The benefits and risks associated with the use of the device. (D).
## BENEFIT/RISK DETERMINATION
The observed risks of the device are based on the animal studies described above. The sponsor did not include any clinical data. The sponsor presented safety data from four different porcine studies. The sponsor did not observe any adverse events in their animal studies. Since the sponsor's small bore connector is based on a legacy design, there is a risk for misconnection with devices from other healthcare applications.
The observed probable benefits of the device are also based on nonclinical laboratory and animal study data as described above. The sponsor demonstrated in laboratory testing that the device was able to hydrolyze >90% of fat in most enteral formulas. The four preclinical animal studies appeared to provide benefit to pigs displaying clinical signs of exocrine pancreatic insufficiency (b) (4) when fed a diet pre-hydrolyzed by the lipase enzyme manually or through the RELIZORB™ device. In their animal studies, the sponsor demonstrated that treating (9) (4) pigs using RELIZORB™ led to increases in total fat absorption, improved update of omega-3 fatty acids in plasma, and improved Vitamin D and E levels. FDA expects these benefits would also be seen using pre-hydrolyzed formula due to the nutritional benefits observed from fat hydrolysis in the literature.
Additional factors to be considered in determining probable risks and benefits for the RELIZORB™ include:
- The cumulative data from the animal studies were robust. Thev were designed logically, . systematically, and built on each other. The first two studies did not use the device. The
{15}------------------------------------------------
pigs in the first study were fed soluble lipase enzyme, while the second study used the beads bound with lipase in a mesh bag. The final two studies used a device prototype that was functionally equivalent to the final device.
- . Study 2 had a safety endpoint and the histopathology was conducted at a GLP pathology lab under blinded conditions. Study results were peer-reviewed prior to release of final study report.
- Study 4 was a randomized, blinded, cross-over study.
- Patients can have abdominal cramps, steatorrhea (loose, greasy, foul-smelling . voluminous stools), and malnutrition with weight loss. The condition is currently treated with oral administration of pancreatic enzyme replacement therapy tablets (PERTs).
- . Human factors testing and clear labeling help mitigate risks.
- The device is beneficial because no enteral formulas currently contain pre-hydrolyzed fat. ●
In conclusion, given the available information above, the data support that the probable benefits for the RELIZORB™ device to hydrolyze fats in enteral formula outweigh the probable risks. The device provides probable benefits and the risks can be mitigated by the use of general and the identified special controls.
## CONCLUSION
The De Novo request for the RELIZORB™ is granted and the device is classified under the following:
Product Code: PLQ Device Type: Enzyme Packed Cartridge Class: II Regulation: 21 CFR 876.5985
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.