K162254 · Cedic S.R.L. · PIO · May 4, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K162254
Device Name
Cedic Enteral ENFit Transition Connectors
Applicant
Cedic S.R.L.
Product Code
PIO · Gastroenterology, Urology
Decision Date
May 4, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The Cedic F 00126 ENFit to Reverse Luer Transition Connector is intended to connect an enteral female ENFit giving set to the "g" port of the AbbVie PEG-J enteral male Luer catheter. The Cedic F 00128 Luer to ENFit Transition Connector is intended to connect the AbbVie Duopa Medication Cassette Reservoir delivery system to an enteral male ENFit catheter.
Device Story
Transition connectors facilitate physical interconnection between legacy enteral feeding devices (Luer) and new-generation ISO 80369-3 (ENFit) compliant devices. Device consists of two specific configurations: F 00126 (ENFit to Reverse Luer) and F 00128 (Luer to ENFit). Used in clinical settings where legacy AbbVie PEG-J catheters or Duopa Medication Cassette Reservoirs must interface with ENFit-compliant giving sets. Operated by clinicians; connectors provide a fluid path for enteral nutrition or medication. Not intended for direct patient contact, though accidental contact may occur. Benefits include enabling continued use of legacy equipment while transitioning to standardized, safer ENFit connections, reducing risk of misconnection with non-enteral devices.
Clinical Evidence
Bench testing only. Performance evaluated via enteral connector misconnection assessment, human factors, fluid leakage, stress cracking, resistance to axial load, resistance to unscrewing, and resistance to overriding. No clinical data provided.
Technological Characteristics
Injection-molded polypropylene. ENFit ends comply with ISO 80369-3; Luer ends comply with ISO 594-2. Single-use, non-sterile. Biocompatibility established per ISO 10993. No software or energy source.
Indications for Use
Indicated for patients requiring enteral feeding or medication delivery using AbbVie PEG-J enteral male Luer catheters or AbbVie Duopa Medication Cassette Reservoir systems, facilitating connection between legacy Luer and ISO 80369-3 ENFit connectors.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol with three figures facing right, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 4, 2017
Cedic S.r.1. % Roger Gray VP. Ouality And Regulatory Donawa Lifescience Consulting S.r.1. Piazza Albania, 10 Rome, 00153 Italy
Re: K162254
> Trade/Device Name: Cedic Enteral ENFit Transition Connectors Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PIO Dated: March 29, 2017 Received: April 5, 2017
Dear Roger Gray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Image /page/1/Picture/8 description: The image shows the name "Benjamin R. Fisher -S" in a large, bold font. The text is black and appears to be centered on a white background. The letters are clearly legible and well-defined.
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K162254
Device Name
Cedic Enteral ENFit Transition Connectors
Indications for Use (Describe)
The Cedic F 00126 ENFit to Reverse Luer Transition Connect an enteral female ENFit giving set to the "g" port of the AbbVie PEG-J enteral male Luer catheter.
The Cedic F 00128 Luer to ENFit Transition Connect the AbbVie Duopa Medication Cassette Reservoir delivery system to an enteral male ENFit catheter.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div><span style="font-size: 12pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the word "Cedic" in a stylized, cursive font. The word is white and has a subtle shadow effect, giving it a three-dimensional appearance. Above the "i" in "Cedic" is a small, stylized "e" symbol, which appears to be part of the logo. The background is a gradient of blue and teal, with the darker shade at the top and the lighter shade at the bottom.
# 510(k) Summary
| Device Name: | Cedic F 00126 ENFit to Reverse Luer Transition Connector, and<br>Cedic F 00128 Luer to ENFit Transition Connector |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission: | Traditional |
| Date of Submission: | 8 August 2016 |
| Manufacturer: | Cedic S.r.l.<br>Biomedical Division<br>Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy |
| 510(k) Owner and Submitter: | Cedic S.r.l.<br>Biomedical Division<br>Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy |
| Phone:<br>Fax:<br>Contact: | +39 02 5530 0174<br>+39 02 5530 1487<br>Ms Tiziana Melis<br>Regulatory Affairs Manager |
| Email: | tiziana.melis@cedicbio.com |
| FDA Registration Number: | 3003593728 |
| Owner/Operator Number: | 9063446 |
| 510(k) Application Correspondent: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy |
| Phone:<br>Fax:<br>Email: | +39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com |
| FDA Product Code: | PIO |
| FDA Regulation Number: | 876.5980 |
| FDA Classification Name: | Enteral Specific Transition Connectors |
| Classification Panel: | Gastroenterology and Urology |
| Common Name: | Transition Connectors for Enteral Applications |
| FDA Classification: | Class II |
| FDA Identification: | Facilitates enteral specific connections between ENFit<br>connectors and non 80369-3 compliant enteral connectors. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image features the word "Cedic" in a stylized, cursive font. The word is white and has a subtle shadow effect, giving it a three-dimensional appearance. Above the "ic" portion of the word, there is a small, circular design element that resembles the letter "e". The background is a gradient of teal, with the color becoming lighter towards the top of the image.
## Device Description
The introduction of new connectors to devices and accessories for enteral feeding applications, in order to help avoid misconnections with devices intended for other clinical applications, has resulted in the short-term need to connectors that will allow devices with existing ('legacy') end connectors to be connected with newer devices having end connectors meeting the relevant requirements of the ISO 80369 series of standards.
Cedic Enteral ENFit Transition Connectors are designed to facilitate enteral specific connections between ISO 80369-3 compliant connectors and non ISO 80369-1 compliant legacy enteral connectors.
This submission describes the Cedic ENFit to Reverse Luer Transition Connector and the Cedic Luer to ENFit Transition Connector. These Cedic ENFit Transition Connectors are intended for prescription use only.
The two transition connectors introduced by means of this submission are intended only for use with specific products commercialized in the US by Abbvie Inc., North Chicago, IL 60064-1802, USA, these being the AbbVie PEG-J enteral male Luer catheter (cleared under K142816) and the AbbVie Duopa Medication Cassette Reservoir (details available in DMF 26089) respectively.
#### Indications for Use
The Cedic F 00126 ENFit to Reverse Luer Transition Connector is intended to connect an enteral female ENFit giving set to the "g" port of the AbbVie PEG-J enteral male Luer catheter.
The Cedic F 00128 Luer to ENFit Transition Connector is intended to connect the AbbVie Duopa Medication Cassette Reservoir delivery system to an enteral male ENFit catheter.
#### Principle of operation, mechanism of action, and interaction with the patient
The Cedic Enteral ENFit Transition Connectors operate by providing a means of interconnecting incompatible enteral feeding device end fittings together, so that patient connection can take place when 'new generation' end fittings in accordance with ISO 80369-3 need to be connected to previous 'legacy' designs of end fitting.
The Cedic Enteral ENFit Transition Connectors are not intended to come into contact with the patient, but accidental contact may occur. The Cedic Enteral ENFit Transition Connectors have a central fluid path through which fluids flow into the patient during use.
Cedic ENFit Transition Connectors are not intended for a specific patient population: they can be used for any patients who need to use the Abbvie products to which they will be connected..
#### Device Specification
The ENFit (PGLock) ends of the connectors are designed in compliance with FDA-recognized standard ISO 80369-3 "Small-bore connectors for liquids and gases in healthcare applications — Part 3: Connectors for enteral applications", and tested in accordance with FDA-recognized standard ISO 80369-20 " Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methods". The Luer ends of the connectors are designed in compliance with ISO 594-2.
#### Manufacture
The Cedic ENFit Transition Connectors are manufactured by injection molding from polypropylene.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "Cedic" in white font on a blue background. The font is stylized and cursive. The letter "e" has a small circle above it, which is also in white.
## Performance Data
In relation to performance data for enteral transition connectors, the subject transition connectors follow the guidance provided in FDA publication 'Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications: Guidance for Industry and Food and Drug Administration Staff', February 11, 2015.
Bench tests have been carried out on samples of the Cedic Transition Connectors. The tests carried out include:
- Enteral connector misconnection assessment ●
- Human factors ●
- Fluid leakage
- Stress cracking ●
- Resistance to separation from axial load ●
- Resistance to separation from unscrewing ●
- . Resistance to overriding
- Disconnection by unscrewing ●
#### Predicate device
The Predicate Device selected for comparison with the Cedic ENFit Transition Connectors is:
| Predicate Device: | AMT Enteral Transition Adapters |
|----------------------|----------------------------------------|
| Sponsor: | Applied Medical Technology, Inc. |
| 510(k) Number: | K150034 |
| Clearance Date: | 9 June 2015 |
| FDA Product Code: | PIO |
| Classification Name: | Enteral specific transition connectors |
| Regulation No: | 876.5980 |
#### Substantial equivalence:
| Table 1: Predicate device comparison table | | | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Item | Predicate device | Subject device | Similarity |
| Device description | The AMT Enteral Transition Adapters come in 5 configurations:<br>• Male Enfit to Male Luer Adapter<br>• Female Enfit to Female Luer Adapter<br>• Female Enfit to Y-Port (Bolus/Luer) Adapter<br>• Female Enfit to Bolus Adapter<br>• Male Enfit to Christmas-tree Adapter | The Cedic Enteral ENFit Transition Connectors come in 2 configurations:<br>• Male ENFit to Female Luer Transition Connector with Cap<br>• Male Luer to Female ENFit Transition Connector with Cap | Very similar |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image features the word "Cedic" in a stylized, cursive font. The word is rendered in a light color, possibly white or a very light shade of blue, which contrasts with the background. The background appears to be a solid color, possibly a shade of blue or teal, providing a clean and simple backdrop for the text. The overall design is clean and professional.
| Table 1: Predicate device comparison table | | | |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Item | Predicate device | Subject device | Similarity |
| Indications for Use | The AMT Enteral Transition<br>Adapters are intended to<br>facilitate enteral specific<br>connections between<br>AAMI/CN3(PS) compliant<br>connectors and non ISO<br>80369-1 compliant legacy<br>enteral connectors. | The Cedic F 00126 ENFit to<br>Reverse Luer Transition<br>Connector is intended to<br>connect an enteral female<br>ENFit giving set to the "g" port<br>of the AbbVie PEG-J enteral<br>male Luer catheter.<br>The Cedic F 00128 Luer to<br>ENFit Transition Connector is<br>intended to connect the<br>AbbVie Duopa Medication<br>Cassette Reservoir delivery<br>system to an enteral male<br>ENFit catheter. | Substantially<br>equivalent |
| Geometry of ENFit<br>connecting portion | AAMI/CN3(PS) compliant | ISO 80369-3 compliant | The same |
| Geometry of Luer<br>connecting portion | ISO 594-2 compliant | ISO 594-2 compliant | The same |
| Materials | "DEHP and Latex free<br>materials" (from 510(k)<br>Summary), no further material<br>details provided | DEHP and Latex free<br>polypropylene | The same |
| Biocompatibility | Established in accordance<br>with ISO 10993 | Established in accordance<br>with ISO 10993 | The same |
| Sterility | Not sterile | Not sterile | The same |
| Usage | Single use | Single use | The same |
| Manufacturing method | Injection molding | Injection molding | The same |
| Shelf life | Unknown | 3 years | N/A |
The Subject Device and the Predicate Device have many identical or similar properties and features. The only differences that exist are related to the exact end connection geometry, for example the predicate has a Male ENFit to Male Luer combination, whereas the subject device has a Male ENFit to Female Luer combination. Similarly, where the predicate has a Female ENFit combination, the subject device has a Male Luer to Female ENFit combination. In addition, although different materials may be used for the predicate and subject devices, the modulus of elasticity in tension for all materials meets the requirements of the relevant FDA-recognised standard. These differences do not have any significant effect on the safety or effectiveness of the subject device.
# Conclusion:
Based on the information contained within this submission, it is concluded that the Cedic Enteral ENFit Transition Connectors are substantially equivalent to the identified predicate device already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.