K150034 · Applied Medical Technology, Inc. · PIO · Jun 9, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K150034
Device Name
AMT Enteral Transition Adapters
Applicant
Applied Medical Technology, Inc.
Product Code
PIO · Gastroenterology, Urology
Decision Date
Jun 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The AMT Enteral Transition Adapters are intended to facilitate enteral specific connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1 compliant legacy enteral connectors.
Device Story
AMT Enteral Transition Adapters facilitate connectivity between new ISO 80369-1 compliant (Enfit) enteral devices and legacy enteral feeding equipment (luer slip, catheter tip, bolus). Device consists of five adapter configurations: Male Enfit to Male Luer, Female Enfit to Female Luer, Female Enfit to Y-Port, Female Enfit to Bolus, and Male Enfit to Christmas-tree. Adapters are used by clinicians or patients in clinical or home settings to ensure compatibility during the industry-wide transition to Enfit connectors. By enabling use of existing legacy inventory with new feeding sets, the adapters prevent supply chain disruption and ensure continuity of enteral nutrition delivery. Adapters are provided non-sterile for single-user use.
Clinical Evidence
Bench testing only. Performance testing included tensile strength, fluid leakage, stress cracking, resistance to separation (axial load/unscrewing), resistance to overriding, disconnection from unscrewing, and flow testing. Biocompatibility data was leveraged from previous 510(k) K123716 for permanent (>30 days) and prolonged (<29 days) indirect mucosal contact.
Technological Characteristics
Non-sterile, single-user adapters. Materials are DEHP and Latex free. Five configurations: Male Enfit to Male Luer, Female Enfit to Female Luer, Female Enfit to Y-Port, Female Enfit to Bolus, and Male Enfit to Christmas-tree. Mechanical connection interface designed to bridge ISO 80369-1 compliant connectors and legacy luer/bolus connectors.
Indications for Use
Indicated for patients requiring enteral feeding who need to connect AAMI/CN3(PS) compliant enteral devices to legacy non-ISO 80369-1 compliant enteral connectors during the transition period to new connector standards.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Enteral Specific Transition Connectors (Cedic S.R.L.) (K140581)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2015
Applied Medical Technology, Inc. Joshua D. Meinke OA/Regulatory Supervisor 8006 Katherine Boulevard Brecksville, OH 44141
Re: K150034
> Trade/Device Name: AMT Enteral Transition Adapters Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIO Dated: March 16. 2015 Received: March 18, 2015
Dear Joshua D. Meinke,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/1/Picture/8 description: The image shows the text "Herbert P. Lerner -S" in a simple, sans-serif font. The text is black and appears to be the main focus of the image. The background is a light gray, and there is a faint watermark-like design behind the text.
Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150034
#### Device Name AMT Enteral Transition Adapters
#### Indications for Use (Describe)
The AMT Enteral Transition Adapters are intended to facilitate enteral specific connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1 compliant legacy enteral connectors.
|×| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION - 5
# 510(K) Summary
AMT Enteral Transition Adapters
| Date Prepared: | June 8, 2015 |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Joshua D. Meinke<br>QA/Regulatory Supervisor<br>Applied Medical Technology, Inc.<br>8006 Katherine Boulevard<br>Brecksville, OH 44141<br>Phone: 440-717-4252<br>Fax: 440-717-4200<br>Email: Josh.Meinke@appliedmedical.net<br>Contact Person: Joshua Meinke |
| Device<br>Information: | Trade Name: AMT Enteral Transition Adapters<br>Common Name: Enteral Transition Adapter<br>Classification Name: Gastroenterology and Urology (21 CFR 876.5980)<br>Regulatory Class: II<br>Product Code: PIO<br>FDA Identification: Facilitate enteral specific connections between<br>AAMI/CN3(PS) compliant connectors and non ISO 80369-1<br>compliant legacy enteral connectors. |
| Predicate<br>Device: | Enteral Specific Transition Connectors (Cedic S.R.L.)<br>(cleared under K140581) |
| Indications for<br>Use: | The AMT Enteral Transition Adapters are intended to facilitate enteral specific<br>connections between AAMI/CN3(PS) compliant connectors and non ISO 80369-1<br>compliant legacy enteral connectors. |
| Device<br>Description: | New standard ISO-80369-1 is pushing feeding tube manufacturers to transition<br>current extension set designs to incorporate enteral only connectors. Most enteral<br>feeding devices currently on the market utilize luer slip or catheter tip connectors to<br>connect enteral giving sets and syringes to enteral receiving sets. Standard luer slip<br>and catheter tip connectors raise risk levels for patients as devices outside of the<br>enteral market use similar connections, including IVs. Therefore, the ISO/IEC 80369<br>series was developed to help reduce the possibility of misconnections across different<br>medical fields. |
| | The FDA recently cleared an updated AMT enteral extension set in 510(k) K142989.<br>This extension set remains identical to our previously marketed device, but includes<br>Enfit connectors. However, launching a enteral extension sets with Enfit connectors<br>will not be simultaneous across all manufacturers and distributors. Therefore,<br>transition adapters are necessary for a limited time. |
| | As the new Enfit connectors are not backwards compatible with current luer and<br>bolus connections currently used for enteral products, AMT has designed several<br>transition adapter configurations to help users make the switch to the new<br>connections. AMT estimates that there will be a time of at least a year and possibly<br>up to three years where users and distributors will still have access to enteral devices<br>with luer / bolus connectors. The transition adapters AMT has designed in this<br>510(k) will allow users to use their old giving/receiving sets until the market is<br>flushed of old inventory. |
| | The AMT Enteral Transition Adapters come in 5 configurations and are made up of 5<br>separate components. The five configurations include:<br>• Male Enfit to Male Luer Adapter<br>• Female Enfit to Female Luer Adapter<br>• Female Enfit to Y-Port (Bolus/Luer) Adapter<br>• Female Enfit to Bolus Adapter<br>• Male Enfit to Christmas-tree Adapter |
| | These 5 transition adapter configurations will allow AMT to offer users with options<br>for all of the common enteral connectors currently on the market. All components<br>and materials have been previously cleared in 510(k)s for similar use. It is planned to<br>include these transition adapters kitted with enteral devices previously cleared<br>through the 510(k) process, as well as selling the adapters as stand-alone items. |
| Technological<br>Characteristics: | The AMT Enteral Transition Adapters are provided non-sterile for single user use<br>only. They are made from DEHP and Latex free materials. The Enteral Transition<br>Adapters are provided in a number of configurations providing different connection<br>possibilities to enteral devices currently on the market. Description of the five<br>different transition adapters AMT will offer are listed below: |
| | Transition Adapter Types:<br>• Male Enfit to Male Luer Adapter (P/N: TRN101)<br>• Female Enfit to Female Luer Adapter (P/N: TRN201)<br>• Female Enfit to Y-Port (Bolus/Luer) Adapter ) (P/N: TRN203)<br>• Female Enfit to Bolus Adapter (P/N: TRN202)<br>• Male Enfit to Christmas-tree Adapter (P/N: TRN102) |
| Biocompatibility<br>Testing: | No additional biocompatibility testing was performed for this submission. The<br>device and materials in finished form for the AMT Enteral Transition Adapters were<br>previously tested under 510(k) K123716. All AMT Enteral Transition adapters have<br>been tested for permanent (greater than 30 days) indirect mucosal contact, except for<br>the Male Enfit to Christmas-tree Adapter (P/N: TRN 102) which has been tested to<br>prolonged (less than 29 days) indirect mucosal contact. |
{4}------------------------------------------------
Applied Medical Technology, Inc. -510(k) Submission
{5}------------------------------------------------
| Performance<br>Testing: | AMT conducted various performance tests on all components contained within AMT<br>Enteral Transition Adapters. The tests carried out included:<br>• Tensile strength<br>• Fluid leakage<br>• Stress cracking<br>• Resistance to separation from axial load<br>• Resistance to separation from unscrewing<br>• Resistance to overriding<br>• Disconnection from unscrewing<br>• Flow testing<br><br>Incompatibility testing was conducted on the Enfit connectors as recommended in the<br>ISO 80369-1 standard. Although the Enfit connectors will not misconnect to other<br>connectors in the 80369 series, misconnections with non ISO/IEC 80369 series<br>connectors may still occur. |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusions: | The AMT Enteral Transition Adapters are substantially equivalent to the predicate<br>device cleared under K140581 in intended use, patient population, design,<br>biocompatibility, testing criteria, and method of operation. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.