Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit Connection, Entuit Gastrostomy BR Balloon Retention Feeding Tube with ENFit Connection
K171347 · Xeridiem Medical Devices · PIF · Jun 6, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K171347
Device Name
Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit Connection, Entuit Gastrostomy BR Balloon Retention Feeding Tube with ENFit Connection
Applicant
Xeridiem Medical Devices
Product Code
PIF · Gastroenterology, Urology
Decision Date
Jun 6, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector are indicated for placement in adult populations that require enteral feeding, medication administration or gastric decompression through an established gastrointestinal stoma tract. The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is intended to provide gastric access for enteral feeding, medication administration and decompression through an established gastrointestinal stoma tract. The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication administration or decompression through an established gastrointestinal stoma tract.
Device Story
Gastrostomy feeding tubes (G-tubes) provide gastric access for enteral nutrition, medication, and decompression. Devices consist of a silicone shaft, adjustable external bolster, and an inflatable retention balloon. Bi-Funnel models provide single-port access for feeding or medication; Tri-Funnel and Entuit BR models provide multi-port access for simultaneous delivery. The device is inserted percutaneously through an established stoma tract; the balloon is inflated with sterile/distilled water to secure the tube in the stomach. Clinicians (physicians/nurses) operate the device in clinical settings. Output is the delivery of nutrition or medication to the patient or decompression of the stomach. The modification involves changing the ENFit connector and cap material from ABS to Nylon to improve robustness.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is supported by design verification activities confirming the material change (Nylon vs. ABS) meets robustness requirements under torque and chemical exposure.
Technological Characteristics
Silicone shaft, bolster, and balloon; radiopaque barium RTV tip. ENFit connectors/caps made of Nylon Vydyne 21X1. Multi-lumen design (bi-lumen shaft). Balloon inflation via valve. Dimensions: 12-24 FR sizes. Non-powered, mechanical device.
Indications for Use
Indicated for adult and pediatric patients requiring enteral feeding, medication administration, or gastric decompression via an established gastrointestinal stoma tract.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Bi-Funnel Gastrostomy Feeding Tube and Tri-Funnel Gastrostomy Feeding Tube (K142075/A001)
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 6, 2017
Xeridiem Medical Devices Steve Murrav Regulatory Affairs Specialist 4700 S. Overland Dr. Tucson, Arizona 85714
Re: K171347
> Trade/Device Name: Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector, Entuit Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PIF Dated: May 5, 2017 Received: May 8, 2017
Dear Steve Murray:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### 4.0 Indications for Use Statement
### Bi-Funnel/Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector 4.1
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
**Indications for Use**
| 510(k) Number (if known) | K171347 |
|--------------------------|--------------------------------------------------------------------------|
| Device Name | Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector |
**Indications for Use (Describe)**
The Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector are indicated for placement in adult populations that require enteral feeding, medication administration or gastric decompression through an established gastrointestinal stoma tract.
| Type of Use (Select one or both, as applicable) | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/14)
ﻟﻠﻘﻀ
{3}------------------------------------------------
### Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® 4.2 Connection
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | Expiration Date: January 31, 2017 |
| Indications for Use | |
| 510(k) Number (if known) | K171347 |
| Device Name | Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection |
| Indications for Use (Describe) | The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is intended to provide gastric access for enteral feeding, medication administration and decompression through an established gastrointestinal stoma tract. |
| | The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication administration or decompression through an established gastrointestinal stoma tract. |
| Type of Use (Select one or both, as applicable) | <span> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </span> <span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1
(End of Section)
{4}------------------------------------------------
Image /page/4/Picture/3 description: The image shows the logo for Xeridiem Medical Devices. The logo consists of a circular graphic with four arrows pointing in a clockwise direction, each in a different color: green, blue, red, and purple. To the right of the graphic is the company name "xeridiem" in a bold, sans-serif font, with the words "MEDICAL DEVICES" in a smaller, lighter font underneath.
### 5.0 510(k) SUMMARY
### 5.1 General Information
| Date Prepared: | May 4, 2017 |
|-------------------|----------------------------------------------------------------------|
| 510(k) Submitter: | Xeridiem Medical Devices<br>4700 S. Overland Dr.<br>Tucson, AZ 85714 |
| Contact Person: | Steve Murray, Regulatory Affairs Specialist |
|-----------------|---------------------------------------------|
| | (520) 882-77694 ext. 178 |
| | smurray@xeridiem.com |
#### 5.2 Device Information
## 5.2.1 Bundled Device 1
| Trade Name: | Bi-Funnel Gastrostomy Feeding Tube with ENFit®<br>connector and Tri-Funnel Gastrostomy Feeding Tube<br>with ENFit® connector |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Gastrostomy Feeding Tube |
| Classification Name: | Gastrointestinal tube and accessories<br>(21 CFR 876.5980, Product Code PIF) |
| Classification Panel: | Gastroenterology/Urology |
# 5.2.2 Bundled Device 2
| Trade Name: | Entuit® Gastrostomy BR Balloon Retention Feeding<br>Tube with ENFit® Connection |
|-----------------------|---------------------------------------------------------------------------------|
| Common Name: | Gastrostomy Feeding Tube |
| Classification Name: | Gastrointestinal tube and accessories<br>(21 CFR 876.5980, Product Code PIF) |
| Classification Panel: | Gastroenterology/Urology |
#### 5.3 Predicate Device Information (Device Being Modified)
#### Bundled Device 1 5.3.1
| Trade Name: | Bi-Funnel Gastrostomy Feeding Tube and<br>Tri-Funnel Gastrostomy Feeding Tube |
|--------------|-------------------------------------------------------------------------------|
| Common Name: | Balloon Gastrostomy Feeding Tube |
{5}------------------------------------------------
| Classification Name: | Gastrointestinal tube and accessories<br>(21 CFR 876.5980, Product Code PIF) |
|-----------------------|------------------------------------------------------------------------------|
| Classification Panel: | Gastroenterology/Urology |
| 510(k) Number: | K142075/A001 |
# 5.3.2 Bundled Device 2
| Trade Name: | Entuit™ Thrive Balloon Retention Gastrostomy<br>Feeding Tube |
|-----------------------|------------------------------------------------------------------------------|
| Common Name: | Gastrostomy Tube |
| Classification Name: | Gastrointestinal tube and accessories<br>(21 CFR 876.5980, Product Code KNT) |
| Classification Panel: | Gastroenterology/Urology |
| 510(k) Number: | K130611/A001 |
- 5.4 Device Description
### Bi-Funnel Gastrostomy Feeding Tube with ENFit® connector and Tri-Funnel 5.4.1 Gastrostomy Feeding Tube with ENFit® connector
The Gastrostomy Feeding Tube (G-tube) consists of two product versions (models): a Bi-Funnel G-Tube and a Tri-Funnel G-Tube. Each G-Tube version is offered in varying shaft French (FR) sizes and balloon sizes. Both versions are offered in the same functional length. The overall difference between the Bi-Funnel and Tri-Funnel G-tubes is that the Tri-Funnel G-Tube allows for simultaneous delivery of feeding solutions and medications while the Bi-Funnel G-Tube allows for delivery of either feeding solutions or medications at a time. The table below summarizes the device offerings for the Bi-Funnel and Tri-Funnel G-Tubes.
| Model<br>P/N | Device Description | French Size<br>Offerings<br>(FR) | Balloon<br>Recommended<br>Fill Volume<br>(mL) | Functional<br>Length (in) |
|--------------|-----------------------|----------------------------------|-----------------------------------------------|---------------------------|
| 70-0047 | Bi-Funnel Gastrostomy | 12 | 5 | 10 |
| | Feeding Tube | 14 | | |
| | 16 | | | |
| | | 18 | 20 | |
| | 20 | | | |
| | 22 | | | |
| | | 24 | | |
| 70-0048 | Tri-Funnel | 12 | 5 | 10 |
| | Gastrostomy Feeding | 14 | | |
| | Tube | 16 | | |
# Table 5.4.1 - Device Offerings for Bi-Funnel/Tri-Funnel G-Tubes
{6}------------------------------------------------
| Model<br>P/N | Device Description | French Size<br>Offerings<br>(FR) | Balloon<br>Recommended<br>Fill Volume<br>(mL) | Functional<br>Length (in) |
|--------------|--------------------|----------------------------------|-----------------------------------------------|---------------------------|
| | | 18 | 20 | |
| | | 20 | | |
| | | 22 | | |
| | | 24 | | |
Both Bi-Funnel and Tri-Funnel G-Tubes consist of a silicone funnel, shaft, adjustable external bolster and balloon. The funnel for the Bi-Funnel G-Tube has two ports: one port houses the balloon inflation valve which allows for balloon inflation and deflation, and the other port houses the male ENFit connector which allows for delivery of feeding solutions and medication. The funnel for the Tri-Funnel G-Tube has three ports: one port houses the balloon inflation valve, the center port houses the male ENFit connector which allows for delivery of feeding solutions, and the last (side) port houses another male ENFit connector which allows for delivery of medications. Attached to the ENFit connectors are a tether and a cap. The cap is intended to cover the ENFit connector when the device is not in use and the tether is intended to secure the cap to the connector when the device is in use. The same ENFit connector, tether and cap components are used for both Bi-Funnel and Tri-Funnel G-Tubes. The remaining components (shaft, adjustable bolster and balloon) are commonly shared for the Bi-Funnel and Tri-Funnel G-Tubes and are identical in both versions.
# 5.4.2 Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection
The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is constructed primarily of silicone. The device consists of a tri-port funnel which is overmolded onto a previously extruded bi-lumen shaft. A retention bolster is molded separately and placed over the shaft. At the proximal end, a silicone balloon is bonded to the shaft using silicone RTV. The tip of the shaft is filled with radiopaque barium RTV. In use, the balloon end of the catheter is inserted through a gastrointestinal stoma tract and into the stomach lumen. The balloon is then inflated to the rated volume using sterile or distilled water and the bolster adjusted down to the skin to secure the device in place. The table below summarizes the device offerings for the Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection.
{7}------------------------------------------------
| Table 5.4.2 - Device Offerings for Entuit® Gastrostomy BR Balloon Retention |
|-----------------------------------------------------------------------------|
| Feeding Tube with ENFit® Connection |
| Model P/N | French Size (FR) | Recommended<br>Balloon Fill<br>Volume (mL) | Functional Length<br>(in) |
|-----------|------------------|--------------------------------------------|---------------------------|
| 70-0060 | 12 | 5 - 7 | 10 |
| | 14 | | |
| | 16 | | |
| | 18 | 5 - 10 | |
| | 20 | 20 | |
| | 22 | | |
| | 24 | | |
The funnel has three ports. One port contains the inflation valve and used for inflation of the balloon. The center port houses the male ENFit connector which allows for the delivery of feeding solutions. The third (side) port houses another male ENFit connector which allows for delivery of medications. Attached to the ENFit connectors are a tether and a cap. The cap is intended to cover the ENFit connector when the device is not in use and the tether is intended to secure the cap to the connector when the device is in use. The funnel configuration and ENFit connectors/caps are identical to the Tri-Funnel Gastrostomy Feeding Tube with ENFit connector discussed in 5.4.1 above.
### 5.5 Intended Use/Indications for Use
### Bi-Funnel Gastrostomy Feeding Tube with ENFit® connector and Tri-Funnel 5.5.1 Gastrostomy Feeding Tube with ENFit® connector
The Bi-Funnel and Tri-Funnel Gastrostomy Feeding Tubes with ENFit® connector are indicated for placement in adult populations that require enteral feeding, medication administration or gastric decompression through an established gastrointestinal stoma tract.
# 5.5.2 Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection
The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is intended to provide gastric access for enteral feeding, medication administration and decompression through an established gastrointestinal stoma tract.
The Entuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection is indicated for use in percutaneous placement of an enteral feeding tube in adult and pediatric patients that require enteral feeding, medication
{8}------------------------------------------------
K171347 Page 5 of 6
administration or decompression through an established gastrointestinal stoma tract.
#### 5.6 Technological Characteristics - Limited Device Modification
The device modification for both devices bundled in this premarket notification is changing the ENFit connector and cap material from ABS LG HF-380 to Nylon Vydyne 21X1. Otherwise, each device and its predicate is identically the same. There is no other change in technological characteristics and no change in intended use or indications for use. The extent of the material change is summarized in the table below.
| Table 5.6 ENFit Connector Material Change | | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------|----------------------------------------------------------|
| Device | Material Description | |
| | Proposed | Predicate |
| Bi-Funnel and Tri-Funnel<br>Gastrostomy Feeding<br>Tubes with ENFit®<br>connector | Nylon Vydyne 21X1<br>(MEVOPUR Violet<br>AB4M665085 colorant) | ABS LG HF-380<br>(MEVOPUR Violet<br>SB4M664930 colorant) |
| Entuit® Gastrostomy BR<br>Balloon Retention<br>Feeding Tube with<br>ENFit® Connection | Nylon Vydyne 21X1 | ABS LG HF-380 |
### Table 5.6 - ENFit® Connector Material Change
### 5.6 Compliance with Design Controls
Substantial equivalence of the proposed devices with the predicate devices are by compliance to the required design control activities for the device modification. These activities are summarized in the table below.
| Table 5.6 – Design Control Activities | | |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| <b>Design Control<br/>Requirement</b> | <b>Design Control Activities</b> | <b>Status</b> |
| Design and<br>Development Planning | Design and Development Plan – outline scope of project and<br>major milestones, including risk management plan and<br>analysis | Completed |
| Design Input | Traceability Matrix update – add new design inputs for ENFit<br>connector robustness requirement under high torque and<br>certain chemical exposure conditions | Completed |
| Design Output | Traceability Matrix update - specify design outputs for the<br>added design inputs; update risk analysis (user and design<br>FMEAs) appropriately to address the change | Completed |
| Design Review | Design reviews conducted for key milestones such as design<br>inputs, design outputs, design verification completion, etc. | Completed |
| Design Verification | Documented in Verification and Validation Plan, summarized<br>in Table 12.2 | Completed |
### Table 5.6 - Design Control Activities
{9}------------------------------------------------
K171347
| Design Control<br>Requirement | Design Control Activities | Status |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Design Validation | Subject modification does not alter ENFit connector geometry<br>so there is no effect on validation previously conducted for<br>clearances K142075 and K130611 | Completed |
As summarized in the table above, required design controls activities were completed, including passing any required design verification activities.
#### 5.6 Substantial Equivalence Conclusion
Based on successful completion of the required design control activities for the device modification that is the subject of this Special 510(k) premarket notification, it may be concluded that the following modified devices are substantially equivalent to their respective predicate devices:
Bi-Funnel Gastrostomy Feeding Tube with ENFit® connector and Tri-Funnel Gastrostomy Feeding Tube with ENFit® connector
Enuit® Gastrostomy BR Balloon Retention Feeding Tube with ENFit® Connection
(End of Section)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.