← Product Code [PIF](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/PIF) · K170555

# U Deliver Bolink ENFit Enteral Feeding Sets (K170555)

_U Deliver Medical, LLC · PIF · Aug 17, 2017 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/PIF/K170555

## Device Facts

- **Applicant:** U Deliver Medical, LLC
- **Product Code:** [PIF](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/PIF.md)
- **Decision Date:** Aug 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The U Deliver Bolink ENFit Enteral Feeding Sets are intended to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).

## Device Story

Gravity-fed enteral feeding sets; connect nutritional reservoirs to patient enteral access devices. Components include PVC tubing, on/off clamp, and distal ENFit connector (ISO 80369-3). Two configurations: Small Cap with cross spike or Large Cap with 40mm thread. Used in clinical or home settings; operated by healthcare providers or patients/caregivers. Fluid flows from reservoir through tubing to feeding tube. Clamp controls flow. Non-sterile, single-patient use. Benefits include standardized, secure connection to enteral tubes, reducing risk of misconnection.

## Clinical Evidence

Bench testing only. Tests included fluid leakage, stress cracking, resistance to separation (axial load, unscrewing, overriding), dimensional analysis of ENFit connectors, biocompatibility (ISO 10993), flow rate testing, tensile testing, and packaging validation.

## Technological Characteristics

Medical-grade, biocompatible materials. Gravity-fed flow. Components: PVC tubing, on/off clamp, ENFit connector (ISO 80369-3), and reservoir connectors (ISO/DIS 18250-3). Non-sterile. Assembled in Class 8 clean room (ISO 14644).

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Metrixcare Enteral Feeding Sets ([K132424](/device/K132424.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2017

U Deliver Medical, LLC % Roger Grav VP. Quality And Regulatory Donawa Lifescience Consulting S.r.1. Piazza Albania, 10 Rome. 00153 Italy

Re: K170555 Trade/Device Name: U Deliver Bolink ENFit Enteral Feeding Sets Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: PIF Dated: July 11, 2017 Received: July 14, 2017

Dear Roger Gray:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K170555

Device Name

U Deliver Bolink ENFit Enteral Feeding Sets

Type of Use (Select one or both, as applicable)

Indications for Use (Describe)

The U Deliver Bolink ENFit Enteral Feeding Sets are intended to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).

| <div> <span>      ☒ Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> | <div> <span>      □ Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div> |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

 CONTINUE ON A SEPARATE PAGE IF NEEDED.

 This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

 The burden time for this collection of information is estimated to average 79 hours per response, including the
 time to review instructions, search existing data sources, gather and maintain the data needed and complete

eview instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Trade Name:                       | U Deliver Bolink ENFit Enteral Feeding Sets                                                                                                         |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Type of 510(k) submission:        | Traditional                                                                                                                                         |
| Date of Submission:               | 16 August 2017                                                                                                                                      |
| Manufacturer:                     | Cedic S.r.l.<br>Biomedical Division<br>Via Liberazione, 63/9<br>20068 Peschiera Borromeo (MI)<br>Italy                                              |
| 510(k) Owner and Submitter:       | U Deliver Medical, LLC<br>10150 City Walk Drive, Suite A<br>Woodbury, MN 55129<br>USA                                                               |
| Phone:                            | 612-381-7330                                                                                                                                        |
| FDA Registration Number:          | Not yet registered                                                                                                                                  |
| Owner/Operator Number:            | Not yet registered                                                                                                                                  |
| 510(k) Application Correspondent: | Mr Roger Gray<br>VP Quality and Regulatory<br>Donawa Lifescience Consulting<br>Piazza Albania 10<br>00153 Rome<br>Italy                             |
| Phone:<br>Fax:<br>Email:          | +39 06 578 2665<br>+39 06 574 3786<br>rgray@donawa.com                                                                                              |
| FDA Product Code:                 | PIF                                                                                                                                                 |
| FDA Regulation Number:            | 876.5980                                                                                                                                            |
| FDA Classification Name:          | Gastrointestinal tubes with enteral specific connectors                                                                                             |
| Classification Panel:             | Gastroenterology and Urology                                                                                                                        |
| Common Name:                      | Enteral Feeding Sets                                                                                                                                |
| FDA Classification:               | Class II                                                                                                                                            |
| FDA Identification:               | The connector is attached to an enteral tube and is intended to<br>facilitate the administration of hydration, medication, and feed<br>to a patient |

### Device Description:

The U Deliver Bolink ENFit Enteral Feeding Sets consist of two configurations:

- U Deliver Bolink Small Cap Gravity Feeding Set with Cross Spike and ENFit Connector (ref. . BK-1085)
- U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector (ref: BK-1405) .

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the text "U Deliver Medical LLC". The words "U Deliver" and "Medical" are in a darker shade of blue, while "LLC" is in a lighter shade of blue. The text is arranged in two lines, with "U Deliver" on the first line and "Medical LLC" on the second line.

The U Deliver Bolink Small Cap Gravity Feeding Set with Cross Spike and ENFit Connector uses the free flow of gravity to dispense enteral feeding solutions from a distal reservoir (not supplied) which connects via a safety screw connection meeting the requirements of clauses 4 "General requirement" and 5 "Dimensional requirements for enteral feed reservoir connectors" of ISO/DIS 18250-3. 'Connectors for reservoir delivery systems for healthcare applications - Part 3: Enteral applications', to a PVC tube with external clamp, to a distal ENFit connector, meeting the requirements of ISO 80369-3, 'Small-bore connectors for liquids and gases in healthcare applications - Part 3: Connectors for enteral applications'.

The U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector uses the free flow of gravity to dispense enteral feeding solutions from a distal reservoir (not supplied) which connects via a screw cap with a 40 mm thread, to a PVC tube with external clamp, to a distal ENFit connector, meeting the requirements of ISO 80369-3

Both sets are constructed from medical grade, biocompatible materials, and are supplied non-sterile for single patient use only.

# Indications for Use/Intended Use

The U Deliver Bolink ENFit Enteral Feeding Sets are intended to deliver liquid nutritional formulas or water to a patient's enteral access device (feeding tube).

# Principle of operation, mechanism of action, and interaction with the patient

The U Deliver Bolink ENFit Enteral Feeding Sets operate by providing a means of interconnecting enteral feeding solution reservoirs to a patient's enteral access device (feeding tube), so that patient enteral feeding can take place when 'new generation' enteral end fittings in accordance with ISO 80369-3 ('ENFit' connections) are present on the feeding tube. A clamp is provided on the tubes to allow feeding to be turned on or off. Although contact of components of the Enteral Feeding Sets may accidentally occur with patient's intact skin during use, the Enteral Feeding Sets have a central fluid path through which feeding solutions flow into the patient during the feeding process.

The U Deliver Bolink ENFit Enteral Feeding Sets are intended for prescription use only.

U Deliver Bolink ENFit Enteral Feeding Sets are not intended for a specific patient population: they can be used for adults, children, infants and neonates.

#### Device Specification

The ENFit connectors on the Enteral Feeding Sets are designed in compliance with FDA-recognized standard ISO 80369-3 "Small-bore connectors for liguids and gases in healthcare applications — Part 3: Connectors for enteral applications", and tested in accordance with FDA-recognized standard ISO 80369-20 "Small-bore connectors for liquids and gases in healthcare applications — Part 20: Common test methods".

The safety screw connector on the U Bolink ENFit Deliver Enteral Set with Safety Screw Connector is designed in accordance with ISO/DIS 18250-3.

The screw cap on the U Deliver Bolink Large Cap Gravity Feeding Set with ENFit Connector is designed with identical major dimensions to the identified predicate device, Metrixcare model 70040, Gravity Set with Feeding Cap (K132424).

#### Manufacture

The U Deliver Bolink ENFit Enteral Feeding Sets are assembled from components manufactured by subcontractors/suppliers to Cedic, the contract manufacturer, in environmentally controlled areas (clean room Class 8 according to ISO 14644), together with tubing and adhesive sourced 'off-the-shelf' by Cedic from other suppliers.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "U Deliver Medical LLC". The words "U Deliver" are in a larger, bolder font than "Medical LLC". The color of the text is blue, but the shade of blue is slightly different between the two lines of text.

# Performance Data

Bench tests have been carried out on samples of the U Deliver Bolink ENFit Enteral Feeding Sets, including:

- Fluid leakage ●
- Stress cracking ●
- . Resistance to separation from axial load
- Resistance to separation from unscrewing ●
- Resistance to overriding
- Disconnection by unscrewing
- Dimensional analysis of ENFit connectors
- . Biocompatibility testing according to ISO 10993
- Flow rate testing using water and enteral feeding formula ●
- . Tensile testing
- Packaging validation ●

#### Predicate device

The Predicate Device selected for comparison with the U Deliver Bolink ENFit Enteral Feeding Sets:

| Predicate Device:    | Metrixcare Enteral Feeding Sets        |
|----------------------|----------------------------------------|
| Sponsor:             | The Metrix Company, USA                |
| 510(k) Number:       | K132424                                |
| Clearance Date:      | 12 February 2014                       |
| FDA Product Code:    | KNT                                    |
| Classification Name: | Gastrointestinal tubes and accessories |
| Regulation No:       | 876.5980                               |

The Subject Device and the Predicate Device have many identical properties and features. The only differences that exist are:

The components included within the enteral feeding sets are different. The predicate includes a sight chamber, while the subject device does not, and the predicate includes a roller clamp, whereas the subject device includes an on/off clamp. These differences do not have any significant effect on the safety or effectiveness of the subject device.

The predicate device that uses the large screw cap at the proximal end has a minimum lumen diameter of 3.4 mm, whereas the equivalent subject device has a minimum lumen diameter of 4.0 mm. This difference allows a faster feeding rate for the subject device, but does not have any significant effect on the safety or effectiveness of the subject device.

The predicate device has a vented feeding cap, whereas the subject device cap is non-vented. The venting on the predicate device allows use of the predicate feeding set with both flexible and rigid feeding containers, while the subject device feeding set can be used only with flexible feeding containers. The subject device labeling includes a warning not to use with rigid containers.

#### Conclusion:

Based on the information contained within this submission, it is concluded that the U Deliver Bolink ENFit Enteral Feeding Sets are substantially equivalent to the identified sets within the predicate device that are already in interstate commerce within the USA.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/PIF/K170555](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/PIF/K170555)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
