URGENT PC NEUROMODULATION SYSTEM, MODELS UPC200-A AND UPC250-12
K061333 · Uroplasty, Inc. · NAM · Jul 3, 2006 · Gastroenterology, Urology
Device Facts
Record ID
K061333
Device Name
URGENT PC NEUROMODULATION SYSTEM, MODELS UPC200-A AND UPC250-12
Applicant
Uroplasty, Inc.
Product Code
NAM · Gastroenterology, Urology
Decision Date
Jul 3, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Urgent PC Neuromodulation System is intended to treat patients suffering from urinary urgency, urinary frequency, and urge incontinence.
Device Story
The Urgent PC Neuromodulation System is a minimally invasive device for Percutaneous Tibial Nerve Stimulation (PTNS). It consists of a battery-operated external pulse generator (Urgent PC Stimulator) and a single-use stimulation lead set with a needle electrode. The system delivers retrograde access to the sacral nerve via percutaneous electrical stimulation of the tibial nerve. The device is used in a clinical setting to treat overactive bladder symptoms. By stimulating the tibial nerve, the device modulates sacral nerve activity, potentially reducing urinary urgency, frequency, and urge incontinence. The healthcare provider operates the device to deliver therapy to the patient.
Clinical Evidence
No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to the predicate device.
Technological Characteristics
Percutaneous tibial nerve stimulator; consists of battery-operated external pulse generator and single-use needle electrode lead set. Delivers electrical stimulation to the tibial nerve. Class II device (21 CFR 876.5310).
Indications for Use
Indicated for patients suffering from urinary urgency, urinary frequency, and urge incontinence.
Regulatory Classification
Identification
A nonimplanted, peripheral electrical continence device is a device that consists of an electrode that is connected by an electrical cable to a battery-powered pulse source. The electrode is placed onto or inserted into the body at a peripheral location and used to stimulate the nerves associated with pelvic floor function to maintain urinary continence. When necessary, the electrode may be removed by the user.
Special Controls
*Classification.* Class II, subject to the following special controls:(1) That sale, distribution, and use of this device are restricted to prescription use in accordance with § 801.109 of this chapter.
(2) That the labeling must bear all information required for the safe and effective use of the device as outlined in § 801.109(c) of this chapter, including a detailed summary of the clinical information upon which the instructions are based.
In addition to the general controls of the act, the Percutaneous SANS Device is subject to the following special controls: (1) The sale, distribution and use of this device are restricted to prescription use in accordance with 21 CFR 801.109. (2) Clinical data are necessary. (3) The labeling must include (a) warnings indicating the need to evaluate the patient during treatment for pain or skin irritation/inflammation; (b) precautions indicating the need to read and understand all directions before using this device; (c) specific instructions to describe proper patients set-up prior to the start of treatment, including proper placement of the electrodes; (d) stimulation parameters and duration and frequency of treatments in accordance with the protocol used during clinical trial; and (e) a summary of the clinical data collected including a listing of adverse events reported.
K052025 — URGENT PC NEUROMODULATION SYSTEM · Uroplasty, Inc. · Oct 17, 2005
K101847 — URGENT PC STIMULATOR; URGENT PC LEAD SET · Uroplasty, Inc. · Oct 21, 2010
K071822 — URGENT PC NEUROMODULATION SYSTEM, MODELS UPC200 AND UPC250-12 · Uroplasty, Inc. · Aug 20, 2007
K132561 — NURO NEUROMODULATION SYSTEM · Advanced Uro-Solutions, Inc. · Nov 5, 2013
Submission Summary (Full Text)
{0}------------------------------------------------
Uroplasty, Inc. Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System
Page 27 of 28
JUL - 3 2006
Ko61333
## Section 10: 510(k) Summary
| Date Prepared | May 11, 2006 |
|------------------|--------------------------------------------------------------------------------------------------------------------------------|
| New Device Name | Urgent® PC Neuromodulation System |
| Predicate Device | Urgent® PC Neuromodulation System (K052025) |
| Contact | Uroplasty, Inc.<br>2718 Summer Street NE<br>Minneapolis, MN 55413-2820<br>Telephone: (612) 378-9399, Facsimile: (612) 378-2027 |
#### Intended Use
The Urgent PC Neuromodulation System is intended to treat patients suffering from urinary urgency, urinary frequency, and urge incontinence.
## Device Description
The Urgent PC Neuromodulation System is a minimally invasive neuromodulation system designed to deliver retrograde access to the sacral nerve through percutaneous electrical stimulation of the tibial nerve. The method of treatment is referred to as Percutaneous Tibial Nerve Stimulation (PTNS).
The Urgent PC Neuromodulation System is a combination of the Urgent PC Stimulator and the Urgent PC Stimulation Lead Set. The Urgent PC Stimulator is a battery-operated external pulse generator and is designed, constructed, and manufactured for multiple use, only in conjunction with the Urgent PC Stimulation Lead Set. The Urgent PC Stimulation Lead Set transfers the electrical current from the Urgent PC Stimulator to the tibial nerve via the Needle Electrode. The entire Stimulation Lead Set is intended for single use only and is not to be reused.
#### Indications Statement
The Urgent PC Neuromodulation System is intended to treat patients suffering from urinary urgency, urinary frequency, and urge incontinence by delivering retrograde access to the sacral nerve through percutaneous electrical stimulation of the tibial nerve.
### Technological Characteristics
The new and predicate devices are technologically the same; they are percutaneous tibial nerve stimulator devices with lead sets intended to deliver retrograde access to the sacral nerve for the overactive bladder symptoms of urinary urgency, urinary frequency, and urge incontinence. Both devices have similar intended uses and principles of action. In the few instances where the devices differ, no concerns about safety or effectiveness are raised.
#### Performance
The Urgent PC Neuromodulation System and Stimulation Lead Set allows for the successful performance of the product's intended use.
### Conclusion
The subject device of this special 510(k) submission is substantially equivalent to the previously cleared Urgent PC Neuromodulation System by Uroplasty, Inc. (K052025).
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines underneath them. The profiles and lines are black. The text "THE PARTNERSHIP FOR HEALTH AND HUMAN SERVICES, INC." is arranged in a circle around the image.
JUL - 3 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Lisa Gallatin, R.A.C. Regulatory Affairs Associate Uroplasty, Inc. 2718 Summer Street, N.E. MINNEAPOLIS MN 55413-2820
Re: K061333
Trade/Device Name: Urgent® PC Neuromodulation System Regulation Number: 21 CFR §876.5310 Regulation Name: Nonimplanted, peripheral electrical continence device Regulatory Class: II Product Code: NAM Dated: June 14, 2006 Received: June 15, 2006
Dear Ms. Gallatin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
Page 2 – Ms. Lisa Gallatin
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains a 70% isopropyl alcohol prep pad, which is subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
i. Panayi singelen.
Nancy C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Uroplasty, Inc. Special Premarket Notification [510(k)] Submission Urgent® PC Neuromodulation System Attachment 1
# Indications for Use
| 510(k) Number (if known): | K061333 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Urgent <sup>®</sup> PC Neuromodulation System |
| Indications for Use: | The Urgent PC Neuromodulation System is intended to treat patients suffering from urinary urgency, urinary frequency, and urge incontinence. |
Prescription Use_ X (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ylancy Inondon
(Division Sign Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number KC61333
Submission date: March 2006
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.