VECTORFLOW; ANTEGRADE CHRONIC HEMODIALYSIS CATHETER, RETROGRADE CHRONIC HEMODIALYSIS CATHETER, RETROGRADE REPLACEMENT HU

K141051 · Arrow International, Inc. · MSD · Jul 7, 2014 · Gastroenterology, Urology

Device Facts

Record IDK141051
Device NameVECTORFLOW; ANTEGRADE CHRONIC HEMODIALYSIS CATHETER, RETROGRADE CHRONIC HEMODIALYSIS CATHETER, RETROGRADE REPLACEMENT HU
ApplicantArrow International, Inc.
Product CodeMSD · Gastroenterology, Urology
Decision DateJul 7, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is inserted percutaneously and is preferentially placed into the internal jugular (II) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is intended for use in adult patients. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is inserted perculaneously and is preferentially placed into the internal jugular (IJ) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow catheter is intended for use in adult patients. The Arrow VectorFlow Retrograde Replacement Hub Set is indicated for use in the replacement of a Arrow VectorFlow Retrograde hub connection assembly that has been damaged.

Device Story

VectorFlow catheters are long-term, single-use, two-lumen, 15 French symmetrical-tipped catheters for central venous access. VFA is a one-piece design for antegrade placement; VFR is a two-piece design for retrograde placement. Hub set allows repair of VFR luer connectors, clamps, and extension lines. Used in healthcare facilities by clinicians for hemodialysis and apheresis. Catheters provide vascular access; hub set restores device integrity. Benefits include reliable long-term blood access for patients with chronic renal failure.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including dialysis flow rate, mechanical hemolysis, and compliance with ISO 10555-1, 10555-3, and ISO 10993-1 standards.

Technological Characteristics

15 French, two-lumen, symmetrical-tipped catheters. Materials tested per ISO 10993-1. Mechanical performance tested per ISO 10555-1 and 10555-3. Sterile, single-use. No software or electronic components.

Indications for Use

Indicated for adult patients requiring long-term vascular access for hemodialysis and apheresis. Catheters are inserted percutaneously, preferentially into the internal jugular vein, with subclavian vein as an alternative. Catheters >40 cm are for femoral vein insertion. Hub set indicated for replacement of damaged hub connection assemblies.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Arrow International. The word "ARROW" is in large, white, sans-serif font on the top line. Below that, the word "INTERNATIONAL" is in a smaller, white, sans-serif font. The background is a solid dark blue. ### 510(k) SUMMARY SUMMARY OF SAFETY AND EFFECTIVENESS #### 1. Submitter Information | Name: | Arrow International, Inc. (subsidiary of Teleflex, Inc.) | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Address: | 2400 Bernville Road<br>Reading, PA 19605-9607 | | Telephone Number: | (610) 378-0131 | | Contact Person: | Frank Pelc<br>Regulatory Affairs/Quality Engineering Manager | | Telephone Number: | (610) 378-0131 Extension 603571 | | Fax Number: | (610) 478-3179 | | Email: | frank.pelc@teleflex.com | | Date Prepared: | April 22, 2014 | | Device Name | Device Trade Name: VectorFlow Antegrade Chronic Hemodialysis Catheter<br>VectorFlow Retrograde Chronic Hemodialysis Catheter<br>VectorFlow Retrograde Replacement Hub Set | | Common Name: | Chronic Hemodialysis Catheter<br>Catheter Repair Kit (Hub Set only) | | Classification Name: | Catheter, Hemodialysis, Implanted<br>Kit, Repair, Catheter, Hemodialysis (Hub Set only) | ### 3. Predicate Devices 2. Predicate 1: Arrow NextStep Antegrade Chronic Hemodialysis Catheter (K111900) Predicate 2: Arrow NextStep Retrograde Chronic Hemodialysis Catheter (K130192) Predicate 3: Arrow NextStep Retrograde Replacement Hub Set (K020430) #### 4. Device Description The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter (herein referred to as the proposed VFA catheter) and the Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter (herein referred to as the proposed VFR catheter) are long-term, single use catheters designed to provide access to the central venous system in a healthcare facility environment. The VFA catheter is a one piece catheter, two lumen, 15 French symmetrical-tipped catheter designed for antegrade placement. The VFR catheter is two-piece, two-lumen, 15 Fr, symmetrical tipped catheter designed for retrograde placement. The VFA and VFR catheters are available in multiple lengths. The Arrow VectorFlow Retrograde Replacement Hub Set (herein referred to as the proposed Hub {1}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 7, 2014 Arrow International, Inc. Frank Pelc Regulatory Affairs/Quality Engineering Manager 2400 Bernville Road Reading, PA 19605 Re: K141051 . > Trade/Device Name: Arrow VectorFlow™ Antegrade Chronic Hemodialysis Catheter Arrow VectorFlow™ Retrograde Chronic Hemodialysis Catheter Arrow VectorFlow™ Retrograde Replacement Hub Set Regulation Number: 21 CFR\$ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD, NFK Dated: June 5, 2014 Received: June 9, 2014 Dear Frank Pelc. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {2}------------------------------------------------ Page 2 - Frank Pelc found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ #### Indications for Use 510(k) Number (if known): _K141051 Device Name: Arrow VectorFlow™ Retrograde Chronic Hemodialysis Catheter Arrow VectorFlow™ Retrograde Replacement Hub Set #### Indications for Use: The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is inserted perculaneously and is preferentially placed into the internal jugular (IJ) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow catheter is intended for use in adult patients. The Arrow VectorFlow Retrograde Replacement Hub Set is indicated for use in the replacement of a Arrow VectorFlow Retrograde hub connection assembly that has been damaged. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH (Signature) Image /page/3/Picture/12 description: The image shows the name "Benjamin R. Fisher -S" in a bold, sans-serif font. Below the name, there is a date "2014.07.23-04'00'" also in a bold, sans-serif font. The text appears to be part of a document or label, with the name possibly indicating a person's name and the date potentially representing a timestamp or record. Page 1 of 1 {4}------------------------------------------------ ### Indications for Use K141051 510(k) Number (if known): Device Name: Anow VectorFlow™ Antegrade Chronic Hemodialysis Catheter #### Indications for Use: The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is inserted percutaneously and is preferentially placed into the internal jugular (J) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow catheter is intended for use in adult patients. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) #### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH (Signature) Benjamin R. Fisher -S 2014.07.22 22:30 -04'00' Page 1 of 1 {5}------------------------------------------------ Set) is sold separately to perform a repair of the VFR, replacing cracked or broken luer connectors, clamps, and/or damaged extension lines. The VFA and VFR catheters and Hub Set will be packaged sterile with various components to facilitate insertion. ## 5. Indications for Use The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is inserted percutaneously and is preferentially placed into the internal jugular (II) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is intended for use in adult patients. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is inserted percutaneously and is preferentially placed into the internal jugular (IJ) vein. Alternately, this catheter may be inserted into the subclavian vein although the jugular vein is the preferred site. Catheters greater than 40 cm are intended for femoral vein insertion. The Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is intended for use in adult patients. The Arrow VectorFlow Retrograde Replacement Hub Set is indicated for use in the replacement of a Arrow VectorFlow Retrograde hub connection assembly that has been damaged. ## 6. Technological Characteristics and Substantial Equivalence The Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter is substantially equivalent to the Arrow NextStep Antegrade Chronic Hemodialysis Catheter (K111900) and the Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter is substantially equivalent to the Arrow NextStep Retrograde Chronic Hemodialysis Catheter (K130192) in terms of overall design, manufacturing process, functional performance, and materials of construction. The proposed Arrow VectorFlow Retrograde Replacement Hub Set is substantially equivalent to the Arrow NextStep Retrograde Replacement Hub Set (K020430) in terms of overall design, manufacturing process, functional performance, and materials of construction. The indications for use for the proposed VFA and VFR catheters are identical to the Arrow NextStep Antegrade Chronic Hemodialysis Catheter and Arrow NextStep Retrograde Chronic Hemodialysis Catheter. The intended use of the product and the principle of operation of the devices are unchanged. The indications for use for the proposed Hub Set are identical to the Arrow NextStep Retrograde Replacement Hub Set. {6}------------------------------------------------ ## 7. Nonclinical Testing Bench testing was performed to demonstrate substantial equivalence between the predicate and proposed devices. Dialysis flow rate, mechanical hemolysis, and additional testing that is needed to comply with Draft Guidance for Industry and Food and Drug Administration Staff- Class II Special Controls Guidance Document: Implanted Blood Access Devices for Hemodialysis and FDA Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters was performed. Performance and material testing was completed on the proposed devices in accordance with ISO 10555-1, 10555-3, and ISO 10993-1. ### 8. Clinical Data Clinical performance data was not used to determine substantial equivalence. ### 9. Conclusions The results of the testing performed have demonstrated that the proposed Arrow VectorFlow Antegrade Chronic Hemodialysis Catheter, the Arrow VectorFlow Retrograde Chronic Hemodialysis Catheter and the Arrow VectorFlow Retrograde Replacement Hub Set are safe and perform as intended and therefore are considered substantially equivalent to the cited predicate devices. {7}------------------------------------------------ # Appendix G Class III Certification and Summary
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...