← Product Code [MSD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD) · K101843

# DURAMAX CHRONIC HEMODIALYSIS CATHETER AND PROCEDURE KIT (K101843)

_AngioDynamics, Inc. · MSD · Oct 20, 2010 · Gastroenterology, Urology · SESK_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K101843

## Device Facts

- **Applicant:** AngioDynamics, Inc.
- **Product Code:** [MSD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD.md)
- **Decision Date:** Oct 20, 2010
- **Decision:** SESK
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The AngioDynamics, Inc. DuraMax™ Chronic Hemodialysis Catheter is intended for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. Alternative insertion sites include the subclavian vein as required. Catheters greater than 40cm are intended for femoral vein insertion. This catheter is indicated for > 30 days longterm placement.

## Device Story

DuraMax Chronic Hemodialysis Catheter is a multi-lumen catheter designed for long-term blood access during hemodialysis and apheresis. Device features distal lumens with side holes, a tapered/curved venous tip for vascular placement, and a guidewire lumen. Proximal end includes a tissue-ingrowth cuff for long-term stabilization, bifurcation hub, suture wing, and extension legs with color-coded clamps/luer connectors. Device is inserted percutaneously by clinicians in clinical settings. Catheter facilitates blood removal and return; tissue cuff promotes secure anchoring. Kit includes necessary insertion accessories. Performance relies on mechanical flow characteristics and biocompatibility for chronic indwelling use.

## Clinical Evidence

Bench testing only. Testing included shaft tensile strength, joint strength, leakage, flow rate, and kit component compatibility. No clinical data presented.

## Technological Characteristics

Multi-lumen catheter with integrated tissue-ingrowth cuff, bifurcation hub, and extension legs. Features distal side holes and tapered/curved venous tip. Materials and design are identical to predicate devices. Connectivity: N/A (mechanical device). Sterilization: Not specified.

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- AngioDynamics® DuraMax® Catheter Kits ([K080400](/device/K080400.md))
- AngioDynamics® EvenMore® Chronic Hemodialysis Catheter ([K040402](/device/K040402.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101843
. Page 1 of 2

AngioDynamics, Inc. DuraMax® Chronic Hemodialysis Catheter and Procedure Kit 510(k) Premarket Notification

OCT 2 0 2010

## 5. 510(K) SUMMARY AS REQUIRED BY 21 CFR 807.92

Submitter: AngioDynamics Inc. 14 Plaza Drive Latham, New York 12110 Tel. 518-795-1400 Fax 518-795-1402

Contact: David Greer AngioDynamics Inc. 14 Plaza Drive Latham, New York 12110 Tel. 518-795-1400 ext. 1676 Fax 518-795-1402

Date Summary was Prepared: 30-Jun-2010

DuraMax® Chronic Hemodialysis Catheter and Procedure Kit Trade Name: Common Name: Hemodialysis Catheter Classification Name: Catheter, Hemodialysis, Implanted 21 CFR 876.5540 Product Code MSD

Predicate Device(s): The DuraMax® Chronic Hemodialysis Catheter and Procedure Kit is substantially equivalent to the following devices:

| Device Name                                               | Manufacturer   | 510(k) No.                    |
|-----------------------------------------------------------|----------------|-------------------------------|
| AngioDynamics® DuraMax® Catheter Kits                     | AngioDynamics® | K080400, cleared<br>5/13/08   |
| AngioDynamics® EvenMore® Chronic<br>Hemodialysis Catheter | AngioDynamics® | K040402, cleared<br>4/19/2004 |

#### Device Description:

The DuraMax® Chronic Hemodialysis Catheter and Procedure Kit is used to remove and return blood during Hemodialysis and aphaeresis. The catheter lumens are open at the distal tip and have a total of 4 side holes (two at venous tip, two at arterial tip). The distal venous tip extends beyond the arterial lumen and is tapered and curved to facilitate placement into vasculature. The distal tip also includes a guidewire lumen to facilitate insertion.

{1}------------------------------------------------

#### AngioDynamics, Inc. DuraMax® Chronic Hemodialysis Catheter and Procedure Kit 510(k) Premarket Notification

The proximal section of the device contains a fixed cuff that allows for tissue in growth for long term placement, an integrated bifurcation hub, suture wing, and extension leg set with color-coded occlusion clamps and luer connectors (red and blue for the arterial and venous lumens respectively). The lumen priming volumes are printed on the clamps. The catheter is packaged individually or as a basic kit that includes the necessary accessories to correctly insert the catheter.

## Statement of Intended Use:

The Intended Use of the device described in this 510(k) is identical to the Intended Use of the predicate, DuraMax HemoDialysis Catheter and Procedure Kit, that was cleared via K080400. Specifically:

The AngioDynamics, Inc. DuraMax™ Chronic Hemodialysis Catheter is intended for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient. Alternative insertion sites include the subclavian vein as required. Catheters greater than 40cm are intended for femoral vein insertion. This catheter is indicated for > 30 days longterm placement.

## Substantial Equivalence:

The DuraMax® Chronic Hemodialysis Catheter and Procedure Kit has the same technological characteristics as the predicate devices that were cleared via K080400 and K040402 respectively.

## Discussion of Nonclinical Tests:

The safety and performance of the DuraMax® Chronic Hemodialysis Catheter and Procedure Kit have been substantiated through extensive non-clinical testing including shaft tensile strength, joint strength, leakage, flow rate, and kit component compatibility,

Results of testing show that the DuraMax® Chronic Hemodialysis Catheter and Procedure Kit can reliably perform as a conventional hemodialysis catheter for obtaining blood access for hemodialysis and apheresis. No new questions of safety or effectiveness have been raised.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, with a flowing, abstract design resembling a wing or ribbon extending from the head.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Mr. David A. Greer Sr. Regulatory Affairs Professional AngioDynamics, Inc. 14 Plaza Drive LATHAM NY 12110

OCT 2 0 2010

Re: K101843

Trade/Device Name: DuraMax® Chronic Hemodialysis Catheter and Procedure Kit Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: September 29, 2010 Received: September 30, 2010

Dear Mr. Greer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

#### Page 2 - Mr. David Greer

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Nicole L. Holanski

Herbert P. Lerner, M.D., Director (Acting) Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

K101843
Page 1 of 1

AngioDynamics, Inc. DuraMax® Chronic Hemodialysis Catheter and Procedure Kit 510(k) Premarket Notification

4. INDICATIONS FOR USE STATEMENT

# Indications for Use

510(k) Number (if known): New Application

OCT 2 0 2010

Device Name:

DuraMax® Chronic Hemodialysis Catheter and Procedure Kit

Indications for Use:

The AngioDynamics, Inc. DuraMax™ Chronic Hemodialysis Catheter is indicated for use in attaining long-term vascular access for Hemodialysis and apheresis.

It may be inserted percutaneously and is primarily placed in the internal jugular vein of an adult patient.

Alternative insertion sites include the subclavian vein as required.

· Catheters greater than 40cm are intended for femoral vein insertion.

This catheter is indicated for > 30 days long-term placement.

| Prescription Use            | ✓ |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) |   |

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CPRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _ K10184

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K101843](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K101843)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
