← Product Code [MSD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD) · K063441

# 15.5FR ALTA GOLD, COATED HEMODIALYSIS CATHETER, MODEL ALC19SH24, ALC23SH28, ALC27SH32, ALC31SH36, ALC35SH40, ALC50SH55 (K063441)

_Spire Biomedical, Inc. · MSD · Feb 12, 2007 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K063441

## Device Facts

- **Applicant:** Spire Biomedical, Inc.
- **Product Code:** [MSD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD.md)
- **Decision Date:** Feb 12, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Spire Biomedical Inc's Alta Gold, Coated Twin Lumen Chronic Indications for Use: Hemodialysis Catheter with fixed tips is designed for chronic hemodialysis and apheresis. It is a radiopaque polyurethane with a heparin coating, designed for percutaneous insertion or insertion via cutdown. The ability of the Carmeda End point Bonded Heparin Coating to reduce clotting is supported by in-vitro testing. Catheters longer than 40cm are intended for femoral vein insertion.

## Device Story

15.5Fr twin lumen fixed tip chronic hemodialysis catheter; radiopaque polyurethane construction; processed with proprietary Carmeda BioActive Surface (CBAS) heparin coating. Coating anchors heparin to internal/external surfaces to mitigate thrombus formation and fibrin sheath propagation. Used in clinical settings for hemodialysis and apheresis; inserted percutaneously or via cutdown by healthcare professionals. Coating maintains bioactivity for 90 days; non-leaching; reduces total thrombus accumulation by 91% per in-vitro bovine blood loop studies. Benefits include improved thromboresistance and reduced clotting complications during long-term vascular access.

## Clinical Evidence

Bench testing only. In-vitro 90-day durability and stability testing confirmed heparin bioactivity. In-vitro 2-hour bovine blood loop study demonstrated 91% reduction in thrombus accumulation compared to uncoated catheters. Biocompatibility testing met ISO 10993 requirements for permanent contact devices.

## Technological Characteristics

Radiopaque polyurethane catheter; twin lumen; fixed tip. Carmeda BioActive Surface (CBAS) heparin coating (non-leaching, max 1mg heparin). 15.5Fr diameter. Permanent contact device. Mechanical properties: tensile strength (catheter-to-hub, extension-to-hub, luer-to-extension, cuff-to-catheter).

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Spire Biomedical, Inc.'s 15.5Fr Decathlon Coated Twin Lumen Chronic Hemodialysis Catheter with Separated Tips ([K060155](/device/K060155.md))
- Spire Biomedical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis Catheter ([K040509](/device/K040509.md))

## Reference Devices

- Medtronic Maxima cardiopulmonary bypass circuit ([K925626](/device/K925626.md) and [K933586](/device/K933586.md))
- Diametrics Paratrend intravascular blood gas sensor catheter ([K970906](/device/K970906.md))
- Cordis Bx Velocity coronary stent (P90043/S024)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K063441
page lot 3

Image /page/0/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in bold, sans-serif font, with the word "Biomedical" in a smaller, cursive font underneath. Above the word "spire" is a symbol that resembles a sun or starburst, with many rays emanating from a central circle.

Spire Biomedical, Inc., · One Patriots Park · Bedford, MA 01730-2396 TEL (781) 275-6000 • FAX 781 275 6010

# SECTION 5 - 510(k) SUMMARY

FEB 1 2 2007

15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter

Date: February 8, 2007

- Submitter: Spire Biomedical, Inc. One Patriots Park Bedford, MA 01730-2396 Phone: (781) 275-6000 Fax: (781) 275-7470
- Contact Person: Shekhar Nimkar Director of Product Development & Manufacturing Spire Biomedical, Inc. TEL: (781) 275-6001 x203 FAX: (781) 275-6010 email: snimkar@spirecorp.com

## Device Names:

| Trade Name:          | 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis<br>Catheter |
|----------------------|--------------------------------------------------------------------------------|
| Common Name:         | Catheter, Intravascular, Long-Term                                             |
| Classification Name: | Catheter, Hemodialysis, Implant (Long-Term)                                    |
| Classification:      | Class III                                                                      |
| Device Code:         | MSD                                                                            |

Legally Marketed Devices to Which Substantial Equivalence is Claimed:

- Spire Biomedical, Inc.'s 15.5Fr Decathlon Coated Twin Lumen Chronic Hemodialysis . Catheter with Separated Tips (K060155)
- Spire Biomedical, Inc.'s 15.5Fr Alta LR Twin Lumen Fixed Tip Chronic Hemodialysis . Catheter (K040509)
- Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K 933586) .
- Diametrics Paratrend intravascular blood gas sensor catheter (K970906) .
- . Cordis Bx Velocity coronary stent (P90043/S024)

{1}------------------------------------------------

K063441/page 2 of 3

Image /page/1/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with the word "Biomedical" in a smaller, italicized font underneath. Above the word "spire" is a circular graphic with a sun-like appearance, featuring a central dark circle surrounded by radiating lines.

Spire Biomedical, Inc., · One Patriots Park · Bedford, MA 01730-2396 TEL (781) 275-6000 • FAX 781 275 6010

### 510(k) Summary (Continued)

#### Device Description:

Spire Biomedical, Inc.'s 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter is processed with a proprietary Carmeda® BioActive Surface (CBAS®) coating technology that attaches a functionally active heparin to the surfaces of the device. The coating counteracts thrombus from forming on the catheter. Spire's 15.5Fr Alta Gold, Heparin Coated Catheter is fully coated with CBAS® on the internal surface and on the external surface of the catheter body (from 2cm distal to the ends of the distal tips; the cuff is not coated).

The coating is essentially non-leaching. Additionally, the maximum amount of heparin on the surface is only 1mg. Therefore; the effects of the entire coating on a patient's coagulation status would be insignificant.

#### Product Claims:

As demonstrated in an in-vitro study, the proprietary End-Point Bonded Heparin Coating attachment mechanism anchors heparin molecules to both internal and external surfaces of the catheter while maintaining heparin's bioactive properties for a minimum of 90 days. 1

Coating bioactivity was assessed in 90-day durability tests performed on coated catheters in saline. Surface-bound heparin bioactivity (in pmol/cm²) was assessed at each of seven regular intervals. Heparin bioactivity remained essentially constant throughout the test period, demonstrating that the coating's bioactive properties were maintained.

In-vitro studies have demonstrated that the coating reduces total thrombus accumulation by 91% compared to uncoated catheters. The coating was effective in mitigating both disturbed flow-mediated thrombosis (at the catheter tip) and fibrin sheath propagation (on the catheter shaft).1

The two-hour in-vitro thromboresistance study involved circulation of bovine blood through an outer loop, in which a coated or uncoated catheter was placed. Simultaneously, blood was circulated through the catheter at a constant flow rate. Three criteria were used to determine the effects of the coating in reducing thrombus formation: Pressure increase in the arterial lumen; visual evaluation of the catheters; and end point thrombus accumulation. Results show improved thromboresistance using each of these criteria for Alta Gold Heparin Coated catheters vs. uncoated catheters. End-point thrombus radiolabeled measurements showed an average of 91% reduction in thrombus accumulation for two hours for coated catheters.

<sup>1</sup> Data on file

{2}------------------------------------------------

K063441 page Sol 3

Image /page/2/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with the word "Biomedical" in a cursive font underneath. Above the word "spire" is a circular graphic with a sun-like appearance.

Spire Biomedical, Inc., · One Patriots Park · Bedford. MA 01730-2396 TEL (781) 275-6000 • FAX 781 275 6010

### Technological Characteristics Comparison to Predicate Devices:

The 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter is identical to the non-coated Alta catheter (K040509) in physical characteristics, has the same physical characteristics (except for the distal tip configuration), and exact same coating as the Decathlon coated catheter (K060155).

Carmeda® BioActive Surface (CBAS®) coating has been approved for the following legally marketed devices to which substantial equivalence is claimed:

- 1. Spire Biomedical Decathlon Gold coated catheter (K060155)
- 2. Medtronic Maxima cardiopulmonary bypass circuit (K925626 and K933586)
- 3. Diametrics Paratrend intravascular blood gas sensor catheter (K970906)
- 4. Cordis Bx Velocity coronary stent (P900043/S024)

#### Performance Data:

A series of physical, mechanical, and biological tests were conducted on the predicate, coated Decathion catheters (K060155) to demonstrate substantial equivalence to our nontreated Decathlon catheters. With the exception of the in-vitro blood loop test, these tests were not repeated due to the material, outside dimensions, and coating being identical for the Alta Gold catheter. Tests conducted on the Decathlon Gold predicate device are listed below:

- Catheter flows and mechanical properties (tensile strength) including: .
	- Catheter to hub .
	- Extension to hub .
	- Luer adapter to extensions .
	- Cuff to catheter �
	- Catheter extrusion .
	- . Catheter flow (bench testing)
- . Coating stability and performance claims:
	- Stability testing (leaching) for 90 days .
	- Durability testing (bioactivity) for 90 days .
	- Durability testing (Accelerated aging for up to 1 year) .
- . Biocompatibility testing:
	- Tests conducted on the 15.5Fr Decathlon coated catheter demonstrated that this . device met the requirements for a permanent contact device per ISO 10993.

The following test was performed on the coated Alta catheters:

- In-vitro blood loop testing 2 Hour study

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure with three wavy lines extending from the figure's head.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shekhar D. Nimkar Director of Product Development & Manufacturing Spire Biomedical, Inc. One Patriots Park BEDFORD MA 01730-2396

FEB 1 2 2007

Re: K063441

Trade/Device Name: 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: MSD Dated: November 2, 2006 Received: November 14, 2006

Dear Mr. Nimkar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Shekhar Nimkar

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

David W. Saylor

Nancy C. Brogdon 298 Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K063441 page 1 of 1

Image /page/5/Picture/1 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with a stylized sun or starburst symbol above the "i". Below "spire", in a smaller, italicized font, is the word "Biomedical".

Spire Biomedical, Inc., · One Patriots Park · Bedford, MA 01730-2396 TEL (781) 275-6000 • FAX 781 275 6010

### SECTION 4 - INDICATIONS FOR INTENDED USE

510(k) Number: KO63441

Device Name: 15.5Fr Alta Gold, Twin Lumen Fixed Tip Coated Chronic Hemodialysis Catheter.

- Spire Biomedical Inc's Alta Gold, Coated Twin Lumen Chronic Indications for Use: Hemodialysis Catheter with fixed tips is designed for chronic hemodialysis and apheresis. It is a radiopaque polyurethane with a heparin coating, designed for percutaneous insertion or insertion via cutdown. The ability of the Carmeda End point Bonded Heparin Coating to reduce clotting is supported by in-vitro testing. Catheters longer than 40cm are intended for femoral vein insertion.
Prescription Use ਵ and/or

Over-the-Counter Use

(Part 21 CFR 801 Subpart D)

(Part 21 CFR 801 Subpart C)

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE):

Daniel b. Legroon

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devic 510(k) Number

I ર

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K063441](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/MSD/K063441)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
