MAHURKAR OPLUS CATHETER, MODEL 13.5 FR

K030209 · The Kendall Company · MPB · May 13, 2003 · Gastroenterology, Urology

Device Facts

Record IDK030209
Device NameMAHURKAR OPLUS CATHETER, MODEL 13.5 FR
ApplicantThe Kendall Company
Product CodeMPB · Gastroenterology, Urology
Decision DateMay 13, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

MAHURKAR® QPLUS® Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion.

Device Story

The MAHURKAR® QPLUS® Catheter is a 13.5Fr radiopaque polyurethane catheter designed for short-term central venous access. It features two longitudinal lumens (arterial and venous) and is available in straight or pre-curved shaft configurations with straight or curved extensions. The device is provided in a kit or tray containing necessary insertion accessories. It is intended for use by clinicians in a healthcare setting to facilitate hemodialysis, apheresis, or infusion procedures. The device functions as a conduit for blood or fluid transport between the patient's central venous system and external medical equipment.

Clinical Evidence

Bench testing only. Performance data compared the subject device to predicate devices to demonstrate substantial equivalence.

Technological Characteristics

13.5Fr radiopaque polyurethane shaft; dual-lumen design; straight or pre-curved shaft options; straight or curved extensions. Non-implanted blood access device. Kit includes accessory items for insertion.

Indications for Use

Indicated for patients requiring short-term central venous access for hemodialysis, apheresis, and infusion.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ KO30209 Page 1 of 2 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 · (508) 261-8000 #### 510(k) Premarket Notification MAHURKAR® QPLUS® Catheter MAY 1 3 2003 ## Section H - 510(K) Summary | Date Summary Was Prepared: | January 17, 2003 | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter's Information: | The Kendall Company<br>Division of Tyco Healthcare Group, LP<br>15 Hampshire Street<br>Mansfield, MA 02048<br>Phone: 508-261-8000<br>Fax: 508-261-8461 | | Contact: | Regina Yeh<br>Senior Specialist, Regulatory Affairs<br>The Kendall Company<br>Division of Tyco Healthcare Group, LP<br>Telephone: 508-261-8404<br>Fax: 508-261-8461 | | Device Trade Name: | MAHURKAR® QPLUS® Catheter for Hemodialysis, Apheresis, and Infusion | | Device Common Name: | Catheter, Hemodialysis, Non-implanted | Classification Panel: Gastroenterology Legally Marketed Devices To Which Substantial Equivalence Is Claimed To: The MAHURKAR® QPLUS® Catheter is substantially equivalent to MAHURKAR® Triple Lumen Catheter and MAHURKAR® Dual Lumen Catheter with respect to its design, materials, physical characteristics, and performance characteristics. Device Description: The MAHURKAR® QPLUS® Catheter has a 13.5Fr radiopaque polyurethane catheter shaft with two large lumens (arterial and venous) running longitudinally along the length of the catheter shaft. The catheter version with a straight shaft either has curved extensions or straight extensions. The catheter version with a pre-curved shaft has straight extensions. MAHURKAR® QPLUS® Catheter is available in kit or tray configuration which contains the sterile catheter as well as accessory items needed for insertion. {2}------------------------------------------------ KO30 209 Page 2 of 2 Image /page/1/Picture/2 description: The image shows the word "KENDALL" in bold, black letters. To the right of the word is the number "15" followed by the letters "HA". The text is all in uppercase. 15 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 • (508) 261-8000 510(k) Premarket Notification MAHURKAR® QPLUS® Catheter # Section H - 510(K) Summary Intended Use: MAHURKAR® QPLUS® Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion. Performance Data: Performance data for the MAHURKAR® QPLUS® Catheter were compared to that of the predicate devices identified in this 510(k) summary. The test results demonstrate that the proposed device is substantially equivalent to the legally marketed predicate devices. {3}------------------------------------------------ Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its head, representing the department's mission to promote health and well-being. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement. MAY 1 3 2003 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Regina Yeh Sr. Regulatory Affairs Specialist Tyco Healthcare Group, LP Division of the Kendall Co. 15 Hampshire Street MANSFIELD MA 02048 Re: K030209 Trade/Device Name: MAHURKAR® QPLUS® Catheter, 13.5 Fr. Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: 78 MPB Dated: April 9, 2003 Received: April 11, 2003 Dear Ms. Yeh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Regina Yeh Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. In addition, we have determined that your device kit contains Lidocaine which is subject to regulation as a drug. Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing this drug, we suggest you contact: Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html. Sincerely yours, Nancy C Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K030209 15 HAMPSHIRE STREET, MANSFIELD, MASSACHUSETTS 02048 · (508) 261-8000 510(k) Premarket Notification MAHURKAR® QPLUS® Catheter Appendix 1 # Indications for Use Statement Device Name: K MAHURKAR® QPLUS® Catheter Indications for Use: MAHURKAR® QPLUS® Catheter is intended for short-term central venous access for hemodialysis, apheresis, and infusion. #### Please Do Not Write Below This Line - Continue On Another Page If Needed Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use Nancy Brogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number. K030209
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...