The Vacurect vacuum component is intended to create an erection in men with erectile dysfunction by means of an applied vacuum to the penis. The constriction ring is intended to maintain penile rigidity in men with erectile dysfunction.
Device Story
Vacurect Vacuum Constriction Device is an over-the-counter external penile rigidity device. It consists of a vacuum tube (with pump sleeve, spacer ring, O-rings, valve stopper) and a flexible rubber constriction ring. Operation involves placing the constriction ring on the vacuum tube to form an airtight seal, positioning the tube over the penis, and manually moving the pump sleeve to create a vacuum. This vacuum draws the penis into the tube to induce an erection. Once rigidity is achieved, the pump is removed, and the constriction ring remains on the penis to restrict venous outflow for up to 30 minutes. The device is intended for patient self-use to assist in achieving and maintaining an erection.
Clinical Evidence
Bench testing only. Testing verified the pumping mechanism does not create a vacuum greater than 0.57 bar (17 inches).
Technological Characteristics
Device consists of a vacuum tube and flexible rubber constriction ring. Vacuum tube components include pump sleeve, spacer ring, O-rings, and valve stopper. Materials are smooth plastic. Operates via manual vacuum generation. No electronic components or software.
Indications for Use
Indicated for men with erectile dysfunction to create an erection and maintain penile rigidity.
Regulatory Classification
Identification
External penile rigidity devices are devices intended to create or maintain sufficient penile rigidity for sexual intercourse. External penile rigidity devices include vacuum pumps, constriction rings, and penile splints which are mechanical, powered, or pneumatic devices.
Special Controls
*Classification.* Class II (special controls). The devices are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: External Penile Rigidity Devices.” See § 876.1(e) for the availability of this guidance document.
Imagyn Medical Technologies, Inc. Impower System Vacuum Erection System (K980291)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary for Vacurect Vacuum Constriction Device
#### 1. SPONSOR
Vacurect Manufacturing (PTY) Ltd. 730 Tetra Avenue Moreletanark, Pretoria Republic of South Africa Postal Address: P.O. Box 39385 Moreletapark 0044 Republic of South Africa
## 2. DEVICE NAME
| Proprietary Name: | Vacurect Vacuum Constriction Device |
|----------------------|-------------------------------------|
| Common/Usual Name: | External Penile Rigidity Device |
| Classification Name: | Vacuum Pump and Constriction Rings |
| Submission Date: | September 4, 2003 |
### PREDICATE DEVICES 3.
Repro-Med Restore Vacuum Erection Device (K981506)
Imagyn Medical Technologies, Inc. Impower System Vacuum Erection System (K980291)
### DEVICE DESCRIPTION 4.
The Vacurect Vacuum Constriction Device is an over the counter device that consists of two main components: the vacuum tube and the constriction ring. The vacuum tube consists of the pump sleeve, spacer ring, dome/sleeve O-Rings, dome/valve stopper, and vacuum ring.
The Vacurect Vacuum Constriction Device works by creating a vacuum in the vacuum tube. This is accomplished by moving the pump sleeve up and down along the vacuum tube. The flexible rubber constriction ring when placed onto the vacuum tube (open end), forms an airtight seal with the vacuum tube. Once the opening of
{1}------------------------------------------------
1432776 PG. 2 F 2
the constriction ring is placed on the penis head, the vacuum chamber is sealed and operation of the pump sleeve will immediately draw the penis into the vacuum tube.
Once the required penis rigidity is reached. the vacuum pump mav be removed and the ring can remain on the penis for up to 30 minutes.
#### న్. INTENDED USE
The Vacurect vacuum component is intended to create an erection in men with erectile dysfunction by means of an applied vacuum to the penis. The constriction ring is intended to maintain penile rigidity in men with erectile dysfunction.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The Vacurect Vacuum Constriction Device and the predicate devices all use a vacuum tube and constriction ring to create and maintain an erection in men with erectile dysfunction. The Vacurect vacuum tube is similar to the predicate devices in that it is made of a smooth plastic material that provides an airtight seal with the penis during use. The constriction ring is also similar to the predicate devices in that they are all designed to restrict penis venous outflow after the patient has obtained an erection with the aid of a vacuum pump.
### 7. PERFORMANCE TESTING
Testing was performed to determine that the pumping mechanism does not create a vacuum greater than 0.57 bar = 17 inches.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure with three overlapping profiles, suggesting a sense of community and interconnectedness.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT - 3 2003
Vacurect Manufacturing (PTY) Ltd. c/o Mary McNamara-Cullinane, RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street NORTH ATTLEBORO MA 02760
Re: K032776
Trade/Device Name: Vacurect Vacuum Constriction Device Regulation Number: None Regulatory Class: Unclassified Product Code: 78 LK Y Dated: September 4, 2003 Received: September 12, 2003
Dear Ms. McNamara-Cullinane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Page 2
{4}------------------------------------------------
510(k) Number (if known): Ko32-776
Device Name: Vacurect Vacuum Constriction Device
Indications for Use:
;
. . . .
The Vacurect vacuum component is intended to create an erection in men with erectile dysfunction by means of an applied vacuum to the penis. The constriction ring is intended to maintain penile rigidity in men with erectile dysfunction.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal. and Radiological Devic 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Vacurect Manufacturing 510(k) Vacurect Vacuum Constriction Device September 4, 2002