TUR,ARTHROSCOPIC,CYSTOSCOPY IRRIGATION SETS
Device Facts
| Record ID | K960787 |
|---|---|
| Device Name | TUR,ARTHROSCOPIC,CYSTOSCOPY IRRIGATION SETS |
| Applicant | Baxter Healthcare Corp |
| Product Code | LJH · Gastroenterology, Urology |
| Decision Date | May 14, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5130 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Baxter’s line of irrigation sets are intended for delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cystoscopy, transurethral resection (TUR) and arthroscopic procedures.
Device Story
Irrigation sets deliver sterile solutions from fluid containers to surgical sites; used during bladder irrigation, cystoscopy, TUR, and arthroscopy. Device consists of tubing segments connecting fluid source to endoscope. Modification replaces natural rubber latex distal tubing with styrene ethylene butadiene styrene (SEBS) to eliminate latex protein sensitivity risks. Operated by clinicians in surgical/procedural settings. Output is controlled fluid flow to irrigation site. Benefits include improved patient safety via allergen-free material.
Clinical Evidence
Bench testing only. Biological/chemical reactivity assessed per ISO 10993-1 and USP Physicochemical tests. Functional performance validated via kink, pull, and pressure testing. No clinical data presented.
Technological Characteristics
Materials: SEBS tubing (replacing natural rubber latex). Design: Identical to predicate; dimensions unchanged. Energy: Passive fluid delivery. Sterilization: Not specified. Connectivity: None. Software: None.
Indications for Use
Indicated for patients undergoing bladder irrigation or endoscopic procedures, including cystoscopy, transurethral resection (TUR), and arthroscopic procedures, requiring delivery of irrigating solutions.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
- Current Baxter Irrigation Sets
Related Devices
- K023808 — IRRIGATION SETS · International Medsurg Connection · Mar 10, 2003
- K032832 — IRRIGATION SETS · Vital Concepts, Inc. · Sep 26, 2003
- K062551 — NEXUS IRRIGATION TUBING SET · Nexus Medical, LLC · Dec 15, 2006
- K970946 — MEDLINE Y-TYPE TUR/BLADDER IRRIGATION SET · Medline Industries, Inc. · Jun 10, 1997
- K990909 — MTP UROPUMP DISPOSABLE TUBING SET · Mtp Medical Technical Promotion GmbH · Jun 9, 1999