← Product Code [KPO](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPO) · K023985

# ACIDIFIED COMPONENT CONCENTRATES FOR BICARBONATE DIALYSIS CONTAINING DI-ACETATE (K023985)

_Rockwell Medical Technologies, Inc. · KPO · Feb 28, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPO/K023985

## Device Facts

- **Applicant:** Rockwell Medical Technologies, Inc.
- **Product Code:** [KPO](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPO.md)
- **Decision Date:** Feb 28, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Rockwell Medical Technologies, Inc., Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/ monitor and bicarbonate concentrate

## Device Story

Device consists of acidified component concentrates containing water, salt, dextrose, and non-sugar electrolytes. Used in three-stream hemodialysis machines; proportioned with purified water and bicarbonate concentrate to create final hemodialysis solution. Solution heated to body temperature; perfused through dialyzer fluid compartment. Separated from patient blood by semi-permeable membrane; allows removal of waste products (urea nitrogen, creatinine, potassium) via concentration gradient. Used in clinical settings by healthcare professionals. Enables effective toxin removal for patients with acute or end-stage renal failure.

## Clinical Evidence

No clinical data. Bench testing only performed to verify chemical composition and concentration ranges.

## Technological Characteristics

Liquid concentrate containing water, salt, dextrose, and non-sugar electrolytes. Formulated for specific dilution ratios (1:35.83, 1:44.00, or 1:34.00). Designed for use with three-stream hemodialysis machines. No software or electronic components.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Rockwell Medical Dri-Sate Acid Concentrate Solutions and Powders
- Fresenius USA, Inc. Granuflo Dialysate Concentrate

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023985

page 1 of 2

510K Notification Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate November 25th, 2002

# 510(k) SUMMARY

FFB 2 & 2003

SUBMITTER:

DATE PREPARED:

DEVICE NAME:

CLASSIFICATION NAMES:

PREDICATE DEVICE:

Rockwell Medical, Technologies, Inc. 30142 Wixom Road Wixom. MI 48393 Phone: 248-960-9009

November 25th, 2002

Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate

Concentrate Solutions for Hemodialysis Accessories to Hemodialysis

Rockwell Medical Dri-Sate Acid Concentrate Solutions and Powders

Fresenius USA, Inc. Granuflo Dialysate Concentrate

### Device Description:

The Rockwell Medical Technologies, Inc. Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acctate contain, water, salt, dextrose, and non-sugar electrolytes formulated and intended for use in hemodialysis when mixed or proportioned with the appropriate volume of purified water and bicarbonate concentrate solution. These concentrate solutions, when proportioned/ mixed in a threestream dialysis machine with pre-treated or purified water meeting AAMI Standards and a bicarbonate concentrate solution, may be used in conventional and commercially available hemodialysis machines or monitors as a hemodialysis solution. The Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate presented in this 510K Notification are intended to be used in three stream hemodialysis machines in which an acidified concentrate is proportioned into one stream, a bicarbonate, chloride and sodium concentrate solution is proportioned into the second stream of the hemodialysis machine, and water is proportioned into the third stream. These three streams are then mixed to prepare a final proportioned hemodialysis solution. These types of final hemodialysis solutions are commonly referred to as "Bicarbonate Hemodialysis Solutions." These proportioned hemodialysis solutions are then heated to body temperature and then perfused through the dialysis fluid compartment of artificial kidneys or hemodialyzers. These bicarbonate hemodialysis solutions are separated from the patient's blood by means of a semi-permeable cellulosic or non-cellulosic membrane which serves as a molecular weight selective barrier to the passage of molecules beyond a certain molecular weight. The molecular weigh cutoff of each type of membrane may vary depending on the membrane type, manufacturing process, etc. The semi-permeable membrane in a hemodialyzer permits the passage of smaller molecular weight (less than 5,000 Daltons for conventional cellulosic membranes), ionized molecules, waste products and toxins (i.e. blood urea nitrogen, creatinine, potassium, etc.) contained in the patient's blood circulating through the dialyzer, to pass through the semi-permeable membrane into the bicarbonate hemodialysis solutions, exit the hemodialyzer, enter the hemodialysis monitor and exit the monitor and are ultimately discarded. The ionic and molecular composition of the hemodialysis solution establishes the concentration gradient between the blood and the hemodialysis solution passing through the hemodialyzer which permits the effective removal of waste products and toxins found in excess in the patient's blood during acute and end-stage renal failure.

Since different patients have different requirements for the removal rates and quantities of various molecules and toxins (i.e. blood urea nitrogen, creatinine, potassium, phosphate, magnesium, chloride, sodium calcium water, etc.) in acute and chronic renal failure, it necessitates having a variety of different bicarbonate containing hemodialysis solutions to satisfy the needs of all acute and end-stage renal failure patients. In addition, there a number of different types of hemodialysis machines which have different proportioning rates. The concentrate mixes and solutions presented in this 510K Notification are designed

# 000123

{1}------------------------------------------------

page 207 x

510K Notification Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate November 25th, 2002

or formulated to be used with hemodialysis machines that proportion according to the following dilution ratios:

| TABLE I  |                                                      |                                                        |
|----------|------------------------------------------------------|--------------------------------------------------------|
| Stream 1 | Stream 2: Acidified Concentrate Proportioning Ratios | Stream 3: Bicarbonate Concentrate Proportioning Ratios |
| Water    | 1:35.83 or 1:44.00 or 1:34.00                        | 1:19.13 or 1:27.57 or 1:25.16                          |

It is for these reasons that a manufacturer of these hemodialysis concentrate solutions and must provide a number of different formulations to contain varying concentrations of the various molecular weight The concentrations of these various molecular components are varied in the final components. hemodialysis solution within physiological and non-physiological ranges to permit the efficient removal of lack thereof from the patient's blood during hemodialysis.

#### Predicate Devices:

The Rockwell Medical Technologies, Inc. Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate are substantially equivalent to the Rockwell Medical Acid concentrate solutions for bicarbonate hemodialysis as well as the Fresenius USA, Inc., Granuflo/Granulate Powder Dialysate Concentrates.

#### Intended Use:

#### Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate Indications:

The Rockwell Medical Technologies, Inc., Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/ monitor and bicarbonate concentrate

This indication statement is essentially the same as the indication statement for the predicate device.

#### Technological Characteristics:

Comparing the proposed device to the predicate devices utilize the same range of chemical compositions, packaging and formulations. There are no significant differences.

#### Summary of Non-Clinical Tests:

In vitro testing was performed to determine the chemical composition and range of composition.. The results of these tests confirmed that the proposed device is equivalent to the proposed device for these parameters.

#### Clinical Test Results:

Clinical testing was not performed

#### Conclusions:

Testing performed on the Rockwell Medical Technologies, Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate Inc. indicates that they are safe, effective, and performs as well as the predicate device, when used in accordance with the instructions for use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, with flowing lines beneath them, possibly representing water or movement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 8 2003

Mr. Gerry Fritz Director of Quality Assurance & Operations Rockwell Medical Technologies, Inc. 30142 Wixom Road WIXOM MI 48393

## Re: K023985

Trade/Device Name: Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 KPO Dated: November 25, 2002 Received: December 2, 2002

Dear Mr. Fritz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

**Page** / of /

## 510 (k) NUMBER (IF KNOWN):

K023985

DEVICE NAME: Rockwell Medical Technologies, Inc., Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate

INDICATIONS FOR USE:

Indications for Use Statement

The Rockwell Medical Technologies, Inc., Acidified Component Concentrates for Bicarbonate Dialysis Containing Di-Acetate are indicated for use in acute and chronic hemodialysis and to be used with the appropriate hemodialysis machine/ monitor and bicarbonate concentrate

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801.109)

OR

Over-The-Counter-Use (Optional Format 1-2-96)

David A. Lynn

(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number _

## 000128

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPO/K023985](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPO/K023985)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
